Accuracy of electrical impedance cardiography for measuring cardiac output in children with congenital heart defects.

D S Miles1, R W Gotshall, J C Golden, D T Tuuri, R H Beekman, T Dillon.   

Abstract

This study determined whether noninvasive electrical impedance cardiography accurately measures systemic blood flow (cardiac output) in children with congenital heart defects. A total of 37 patients ranging in age from 2 to 171 months underwent complete right- and left-sided heart catheterizations that included simultaneous Fick and impedance measurement of cardiac output. Based on the diagnosis, 4 groups were formed consisting of a control group (n = 11) with no shunts, a group with intracardiac left-to-right shunting and an atrial septal defect (n = 7), another with a ventricular septal defect (n = 12) and an extracardiac left-to-right shunting with patent ductus arteriosus group (n = 7). Impedance values for systemic blood flow were compared with systemic and pulmonary blood flow obtained by the direct Fick method with measured oxygen consumption. The difference between impedance and Fick systemic blood flow was less than or equal to 5% in each of the 4 groups. The highest correlation between impedance and Fick systemic blood flow was with the atrial septal defect group (r = 0.89) and lowest with the ventricular septal defect and control (r = 0.69) groups. Fick pulmonary blood flow was significantly greater than impedance or Fick systemic flow in all 3 shunt groups. Impedance cardiography accurately measured systemic blood flow in children without shunts or valvular insufficiency. Likewise, systemic blood flow was accurately measured by impedance in the presence of intracardiac left-to-right shunts (atrial and ventricular septal defects) and extracardiac left-to-right shunts (patent ductus arteriosus).(ABSTRACT TRUNCATED AT 250 WORDS)

Entities:  

Mesh:

Year:  1988        PMID: 3344687     DOI: 10.1016/0002-9149(88)90775-8

Source DB:  PubMed          Journal:  Am J Cardiol        ISSN: 0002-9149            Impact factor:   2.778


  4 in total

1.  Comparison of two impedance cardiographic techniques for measuring cardiac output.

Authors:  R W Gotshall; V C Wood; D S Miles
Journal:  Ann Biomed Eng       Date:  1989       Impact factor: 3.934

2.  Validation of intracardiac shunt using thoracic bioimpedance and inert gas rebreathing in adults before and after percutaneous closure of atrial septal defect in a cardiology research unit: study protocol.

Authors:  Laura Filaire; Aurelie Chalard; Hélène Perrault; Romain Trésorier; Jean-René Lusson; Bruno Pereira; Frederic Costes; Claire Dauphin; Ruddy Richard
Journal:  BMJ Open       Date:  2019-05-27       Impact factor: 2.692

3.  Feasibility, reproducibility and accuracy of electrical velocimetry for cardiac output assessment in congenital heart disease.

Authors:  Alexander C Egbe; Muhammad Wajih Ullah; Arslan Afzal; Keerthana Banala; Rahul Vojjini; Maria Najam; Karim Osman; Sahith Thotamgari; Donald J Hagler
Journal:  Int J Cardiol Heart Vasc       Date:  2020-01-09

4.  Ginsenoside Rg2 Ameliorates Brain Injury After Intracerebral Hemorrhage in a Rat Model of Preeclampsia.

Authors:  Liying Cai; Feifei Hu; Wenwen Fu; Xiaofeng Yu; Weijie Zhong; Fangcong Liu; Tian Wang; Dayuan Sui
Journal:  Reprod Sci       Date:  2021-07-16       Impact factor: 3.060

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.