Supawitoo Sookpeng1, Colin J Martin2, Chitsanupong Butdee1.
Abstract
BACKGROUND: Computed tomography (CT) automatic tube current modulation (ATCM) systems and iterative reconstruction (IR) play an important role in CT radiation dose optimization. How the two can best be used together is one of the challenges faced by radiology professionals. AIM: To determine optimum settings of ATCM noise index (NI) together with adaptive statistic iterative reconstruction (ASIR) for a general electric (GE) scanner that aims to achieve similar image quality to the standard protocol used in the hospital (Smart mA technique with NI of 11.57 and 30% ASIR reconstruction) with a lower dose.Entities:
Keywords: Adaptive statistic iterative reconstruction; automatic tube current modulation; computed tomography optimization
Year: 2019 PMID: 31000942 PMCID: PMC6467048 DOI: 10.4103/ijri.IJRI_124_18
Source DB: PubMed Journal: Indian J Radiol Imaging ISSN: 0970-2016
CT protocols and dose data
| Protocol | Dose step | NI | ASIR level (%) | CTDIvol (mGy) | DLP (mGy cm) | Effective dose (mSv) |
|---|---|---|---|---|---|---|
| Standard | 0 | 11.57 | 30 | 5.2 | 211 | 3.0 |
| 1 | 50 | |||||
| 2 | 70 | |||||
| 3 | −3 | 13.3 | 30 | 3.9 | 160 | 2.2 |
| 4 | 50 | |||||
| 5 | 70 | |||||
| 6 | −6 | 15.04 | 30 | 3.1 | 125 | 1.8 |
| 7 | 50 | |||||
| 8 | 70 | |||||
| 9 | −9 | 16.78 | 30 | 2.5 | 100 | 1.4 |
| 10 | 50 | |||||
| 11 | 70 | |||||
| 12 | −12 | 18.52 | 30 | 2.0 | 82 | 1.2 |
| 13 | 50 | |||||
| 14 | 70 | |||||
| 15 | −15 | 20.26 | 30 | 1.7 | 69 | 1.0 |
| 16 | 50 | |||||
| 17 | 70 | |||||
| 18 | −18 | 22 | 30 | 1.4 | 59 | 0.8 |
| 19 | 50 | |||||
| 20 | 70 |
NI=Noise index, ASIR=Adaptive statistic iterative reconstruction, CTDIvol=Volume weighted computed tomography dose index, DLP=Dose length product
Figure 1CT images of the chest phantom at the middle of the heart obtained from standard protocol and ROI locations
Figure 2Tube current (mA) modulations for individual protocols
Quantitative image analysis for different CT protocols
| Protocol | NI | ASIR level (%) | SNR soft tissue | SNR lung | CNR |
|---|---|---|---|---|---|
| Standard | 11.57 | 30 | 3.3 | 90.7 | 137.1 |
| 1 | 50 | 3.8 | 99.1 | 158.1 | |
| 2 | 70 | 4.4 | 108.5 | 185.1 | |
| 3 | 13.3 | 30 | 2.8 | 82.7 | 114.3 |
| 4 | 50 | 3.2 | 92.7 | 130.9 | |
| 5 | 70 | 3.6 | 99.5 | 151.6 | |
| 6 | 15.04 | 30 | 2.6 | 77.0 | 108.8 |
| 7 | 50 | 3.0 | 84.9 | 126.2 | |
| 8 | 70 | 3.5 | 93.7 | 148.3 | |
| 9 | 16.78 | 30 | 2.3 | 70.3 | 96.5 |
| 10 | 50 | 2.7 | 77.4 | 111.4 | |
| 11 | 70 | 3.1 | 85.4 | 130.4 | |
| 12 | 18.52 | 30 | 2.1 | 67.2 | 88.9 |
| 13 | 50 | 2.5 | 74.6 | 103.3 | |
| 14 | 70 | 2.9 | 83.3 | 122.2 | |
| 15 | 20.26 | 30 | 2.0 | 63.7 | 80.1 |
| 16 | 50 | 2.3 | 70.5 | 93.4 | |
| 17 | 70 | 2.7 | 78.2 | 110.7 | |
| 18 | 22 | 30 | 1.8 | 58.1 | 75.6 |
| 19 | 50 | 2.1 | 64.5 | 87.9 | |
| 20 | 70 | 2.5 | 71.9 | 103.9 |
ASIR=Adaptive statistic iterative reconstruction, SNR=Signal-to-noise ratio, CNR=Contrast-to-noise ratio
Figure 3 (A-C)NR measured at the lungs (A) and soft tissue (B) and CNR (C) for individual protocols
Overall scores of subjective image quality analysis
| Protocol | Average qualitative image quality score ( | |||
|---|---|---|---|---|
| Image noise | Sharpness | Contrast | Image artifact | |
| Standard | 3.20±0.84 | 4.20±0.84 | 4.60±0.55 | 3.60±0.55 |
| 1 | 3.40±0.55 | 3.80±0.45 | 4.00±0.00 | 3.20±0.45 |
| 2 | 3.60±0.55 | 3.00±0.71 | 3.20±0.84 | 3.20±0.45 |
| 4 | 3.80±0.84 | 3.60±0.55 | 3.60±0.55 | 4.40±0.55 |
| 5 | 3.20±0.45 | 3.20±0.84 | 3.80±0.45 | 4.00±0.71 |
| 8 | 3.40±0.55 | 3.80±0.45 | 3.80±0.45 | 3.40±0.55 |
P>0.01
Figure 4 (A and B)CNR (A) and noise (B) for different reconstruction methods and dose levels
Figure 5Variation of spatial resolution with different reconstruction methods for the recommendation protocol of 3.1 mGy
Figure 6Variation of low-contrast detail detectability with dose levels and different reconstruction methods, for contrast levels of 1, 0.5, and 0.3%