| Literature DB >> 30972756 |
Hui Khee Looe1, Daniela Poppinga2, Rafael Kranzer2, Isabel Büsing1, Tuba Tekin1, Ann-Britt Ulrichs1, Björn Delfs1, Dennis Vogt2, Jan Würfel2, Björn Poppe1.
Abstract
PURPOSE: Discrepancy between experimental and Monte Carlo simulated dose-response of the microDiamond (mD) detector (type 60019, PTW Freiburg, Germany) at small field sizes has been reported. In this work, the radiation-induced charge imbalance in the structural components of the detector has been investigated as the possible cause of this discrepancy.Entities:
Keywords: diamond detector; polarity effect; radiation-induced charge imbalance; small fields; volume effect
Mesh:
Substances:
Year: 2019 PMID: 30972756 PMCID: PMC6849526 DOI: 10.1002/mp.13542
Source DB: PubMed Journal: Med Phys ISSN: 0094-2405 Impact factor: 4.071
Figure 1Left insert: x‐ray image of a standard microDiamond (mD) detector. The diamond chip with the sensitive volume is shown as a green line. The contacts of the detector are marked in blue and red. The type of charge carries collected depends on which contact is connected to the electrometer. Right insert: schematic drawings of the four versions of mD detector used in this work. [Color figure can be viewed at http://www.wileyonlinelibrary.com]
Raw measured signals, in nC, of the mD detectors and W1 scintillator integrated over 60 s. The dose–response (nC/Gy) of each mD detector, obtained using a 10 cm × 10 cm 6 MV photon beam, is provided. |Q EDEP| have been computed from Eqs. (2) and (3) by considering the signals of mD_noChip as Q net in Eq. (1); and from Eq. (4) by taking the mean of the absolute signals of the standard and mD_reversed* detectors
|
| 6.4 | 8.2 | 10.2 | 16 | 20 | 30 | 40 |
|---|---|---|---|---|---|---|---|
| W1 | 0.066 | 0.076 | 0.083 | 0.093 | 0.096 | 0.101 | 0.105 |
| mD_standard (1.0 nC/Gy) | 2.672 | 3.084 | 3.330 | 3.668 | 3.770 | 3.941 | 4.093 |
| mD_reversed (1.3 nC/Gy) | −3.340 | −3.904 | −4.245 | −4.718 | −4.865 | −5.095 | −5.288 |
| mD_reversed* (1.0 nC/Gy) | −2.569 | −3.003 | −3.265 | −3.629 | −3.742 | −3.919 | −4.068 |
| mD_shortened (1.0 nC/Gy) | 2.674 | 3.094 | 3.351 | 3.706 | 3.815 | 3.988 | 4.136 |
| mD_noChip | 0.0451 | 0.0335 | 0.0237 | 0.0106 | 0.0080 | 0.0073 | 0.0084 |
| | | |||||||
| Eq. | 2.627 | 3.050 | 3.306 | 3.657 | 3.762 | 3.934 | 4.085 |
| Eq. | 2.614 | 3.037 | 3.289 | 3.640 | 3.750 | 3.926 | 4.076 |
| Eq. | 2.621 | 3.043 | 3.298 | 3.649 | 3.756 | 3.930 | 4.080 |
Figure 2(a) Measured output ratio (OR) using the standard microDiamond (mD_standard) and two modified mD detectors (mD_reversed* and mD_shortened); (b) measured small field output correction factor, k meas, obtained using the OR in (a); (c) Corrected OR according to Eqs. (2), (3), and (4) and the OR measured using the modified detector with shortened contacts (mD_shortened); (d) small field output correction factor, k meas, obtained using the corrected OR in (c) and the OR measured with the mD_shortened detector. [Color figure can be viewed at http://www.wileyonlinelibrary.com]
Figure 3Simulated output correction factors, k sim, and the volume‐averaging correction factors, P vol, for the microDiamond (mD) detector obtained using the IAEA phase space files of Varian Clinac iX 6 MV photon beam. The ratios k sim/P vol represent the correction for the remaining perturbation effects except for the volume‐averaging effect. The measured output correction factors, k meas, computed from the corrected output ratio (OR) according to Eq. (4) show good agreement with the simulated values. [Color figure can be viewed at http://www.wileyonlinelibrary.com]
Figure 4The field size dependent ratio computed from measurements according to Eqs. (8), (9) and using Monte Carlo simulations.