Dosimetry of cone-defined stereotactic radiosurgery fields with a commercial synthetic diamond detector.

Johnny E Morales1, Scott B Crowe2, Robin Hill3, Nigel Freeman4, J V Trapp2.   

Abstract

PURPOSE: Small field x-ray beam dosimetry is difficult due to lack of lateral electronic equilibrium, source occlusion, high dose gradients, and detector volume averaging. Currently, there is no single definitive detector recommended for small field dosimetry. The objective of this work was to evaluate the performance of a new commercial synthetic diamond detector, namely, the PTW 60019 microDiamond, for the dosimetry of small x-ray fields as used in stereotactic radiosurgery (SRS).
METHODS: Small field sizes were defined by BrainLAB circular cones (4-30 mm diameter) on a Novalis Trilogy linear accelerator and using the 6 MV SRS x-ray beam mode for all measurements. Percentage depth doses (PDDs) were measured and compared to an IBA SFD and a PTW 60012 E diode. Cross profiles were measured and compared to an IBA SFD diode. Field factors, ΩQclin,Qmsr (fclin,fmsr) , were calculated by Monte Carlo methods using BEAMnrc and correction factors, kQclin,Qmsr (fclin,fmsr) , were derived for the PTW 60019 microDiamond detector.
RESULTS: For the small fields of 4-30 mm diameter, there were dose differences in the PDDs of up to 1.5% when compared to an IBA SFD and PTW 60012 E diode detector. For the cross profile measurements the penumbra values varied, depending upon the orientation of the detector. The field factors, ΩQclin,Qmsr (fclin,fmsr) , were calculated for these field diameters at a depth of 1.4 cm in water and they were within 2.7% of published values for a similar linear accelerator. The corrections factors, kQclin,Qmsr (fclin,fmsr) , were derived for the PTW 60019 microDiamond detector.
CONCLUSIONS: The authors conclude that the new PTW 60019 microDiamond detector is generally suitable for relative dosimetry in small 6 MV SRS beams for a Novalis Trilogy linear equipped with circular cones.

Entities:  

Mesh:

Substances:

Year:  2014        PMID: 25370616     DOI: 10.1118/1.4895827

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  6 in total

1.  Measurement of Total Scatter Factor for Stereotactic Cones with Plastic Scintillation Detector.

Authors:  Suresh H Chaudhari; Rishabh Dobhal; Rajesh A Kinhikar; Sudarshan S Kadam; Deepak D Deshpande
Journal:  J Med Phys       Date:  2017 Jan-Mar

2.  Detectors assessment for stereotactic radiosurgery with cones.

Authors:  Nicolas Garnier; Régis Amblard; Rémy Villeneuve; Rodolphe Haykal; Cécile Ortholan; Philippe Colin; Anaïs Gérard; Sarah Belhomme; Franck Mady; Mourad Benabdesselam; Benjamin Serrano
Journal:  J Appl Clin Med Phys       Date:  2018-09-14       Impact factor: 2.102

3.  "Characterization of ELEKTA SRS cone collimator using high spatial resolution monolithic silicon detector array".

Authors:  Khalsa Al Shukaili; Stéphanie Corde; Marco Petasecca; Vladimir Pereveratylo; Michael Lerch; Michael Jackson; Anatoly Rosenfeld
Journal:  J Appl Clin Med Phys       Date:  2018-05-22       Impact factor: 2.102

4.  The role of radiation-induced charge imbalance on the dose-response of a commercial synthetic diamond detector in small field dosimetry.

Authors:  Hui Khee Looe; Daniela Poppinga; Rafael Kranzer; Isabel Büsing; Tuba Tekin; Ann-Britt Ulrichs; Björn Delfs; Dennis Vogt; Jan Würfel; Björn Poppe
Journal:  Med Phys       Date:  2019-05-02       Impact factor: 4.071

5.  Evaluation of the PTW microDiamond in edge-on orientation for dosimetry in small fields.

Authors:  Owen J Brace; Sultan F Alhujaili; Jason R Paino; Duncan J Butler; Dean Wilkinson; Brad M Oborn; Anatoly B Rosenfeld; Michael L F Lerch; Marco Petasecca; Jeremy A Davis
Journal:  J Appl Clin Med Phys       Date:  2020-05-22       Impact factor: 2.102

6.  Characterization of a new commercial single crystal diamond detector for photon- and proton-beam dosimetry.

Authors:  Yuichi Akino; Archana Gautam; Len Coutinho; Jan Würfel; Indra J Das
Journal:  J Radiat Res       Date:  2015-08-12       Impact factor: 2.724

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.