Detector dose response in megavoltage small photon beams. I. Theoretical concepts.

Hugo Bouchard1, Jan Seuntjens2, Simon Duane1, Yuji Kamio3, Hugo Palmans4.   

Abstract

PURPOSE: To explain the reasons for significant quality correction factors in megavoltage small photon fields and clarify the underlying concepts relevant to dosimetry under such conditions.
METHODS: The validity of cavity theory and the requirement of charged particle equilibrium (CPE) are addressed from a theoretical point of view in the context of nonstandard beams. Perturbation effects are described into four main subeffects, explaining their nature and pointing out their relative importance in small photon fields.
RESULTS: It is demonstrated that the failure to meet classical cavity theory requirements, such as CPE, is not the reason for significant quality correction factors. On the contrary, it is shown that the lack of CPE alone cannot explain these corrections and that what matters most, apart from volume averaging effects, is the relationship between the lack of CPE in the small field itself and the density of the detector cavity. The density perturbation effect is explained based on Fano's theorem, describing the compensating effect of two main contributions to cavity absorbed dose. Using the same approach, perturbation effects arising from the difference in atomic properties of the cavity medium and the presence of extracameral components are explained. Volume averaging effects are also discussed in detail.
CONCLUSIONS: Quality correction factors of small megavoltage photon fields are mainly due to differences in electron density between water and the detector medium and to volume averaging over the detector cavity. Other effects, such as the presence of extracameral components and differences in atomic properties of the detection medium with respect to water, can also play an accentuated role in small photon fields compared to standard beams.

Entities:  

Mesh:

Year:  2015        PMID: 26429279     DOI: 10.1118/1.4930053

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  12 in total

1.  DNA double-strand breaks as a method of radiation measurements for therapeutic beams.

Authors:  Mohammad Obeidat; Kristen A McConnell; Xiaolei Li; Brian Bui; Sotirios Stathakis; Niko Papanikolaou; Karl Rasmussen; Chul Soo Ha; Sang Eun Lee; Eun Yong Shim; Neil Kirby
Journal:  Med Phys       Date:  2018-05-23       Impact factor: 4.071

2.  Survey of 5 mm small-field output factor measurements in Australia.

Authors:  Christopher P Oliver; Duncan J Butler; Viliami Takau; Ivan Williams
Journal:  J Appl Clin Med Phys       Date:  2018-01-25       Impact factor: 2.102

3.  AAPM-RSS Medical Physics Practice Guideline 9.a. for SRS-SBRT.

Authors:  Per H Halvorsen; Eileen Cirino; Indra J Das; Jeffrey A Garrett; Jun Yang; Fang-Fang Yin; Lynne A Fairobent
Journal:  J Appl Clin Med Phys       Date:  2017-08-08       Impact factor: 2.102

4.  Measurement of Total Scatter Factor for Stereotactic Cones with Plastic Scintillation Detector.

Authors:  Suresh H Chaudhari; Rishabh Dobhal; Rajesh A Kinhikar; Sudarshan S Kadam; Deepak D Deshpande
Journal:  J Med Phys       Date:  2017 Jan-Mar

5.  Assessment of Imprecise Small Photon Beam Modeling by Two Treatment Planning System Algorithms.

Authors:  Hadi Keivan; Daryoush Shahbazi-Gahrouei; Ahmad Shanei; Alireza Amouheidari
Journal:  J Med Signals Sens       Date:  2018 Jan-Mar

6.  "Characterization of ELEKTA SRS cone collimator using high spatial resolution monolithic silicon detector array".

Authors:  Khalsa Al Shukaili; Stéphanie Corde; Marco Petasecca; Vladimir Pereveratylo; Michael Lerch; Michael Jackson; Anatoly Rosenfeld
Journal:  J Appl Clin Med Phys       Date:  2018-05-22       Impact factor: 2.102

7.  The role of radiation-induced charge imbalance on the dose-response of a commercial synthetic diamond detector in small field dosimetry.

Authors:  Hui Khee Looe; Daniela Poppinga; Rafael Kranzer; Isabel Büsing; Tuba Tekin; Ann-Britt Ulrichs; Björn Delfs; Dennis Vogt; Jan Würfel; Björn Poppe
Journal:  Med Phys       Date:  2019-05-02       Impact factor: 4.071

8.  Three-dimensional characterization of the active volumes of PTW microDiamond, microSilicon, and Diode E dosimetry detectors using a proton microbeam.

Authors:  Daniela Poppinga; Rafael Kranzer; Ann-Britt Ulrichs; Björn Delfs; Ulrich Giesen; Frank Langner; Björn Poppe; Hui Khee Looe
Journal:  Med Phys       Date:  2019-07-22       Impact factor: 4.071

9.  Technical Note: Characterization of the new microSilicon diode detector.

Authors:  Ann-Britt Schönfeld; Daniela Poppinga; Rafael Kranzer; Rudy Leon De Wilde; Kay Willborn; Björn Poppe; Hui Khee Looe
Journal:  Med Phys       Date:  2019-07-31       Impact factor: 4.071

10.  Commissioning of a PTW 34070 large-area plane-parallel ionization chamber for small field megavoltage photon dosimetry.

Authors:  Tom Kupfer; Joerg Lehmann; Duncan J Butler; Ganesan Ramanathan; Tracy E Bailey; Rick D Franich
Journal:  J Appl Clin Med Phys       Date:  2017-10-04       Impact factor: 2.102

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.