| Literature DB >> 30970039 |
Velicia Bachtiar1, Matthew D Kelly1, Henry R Wilman1,2, Jaco Jacobs1, Rexford Newbould1, Catherine J Kelly1, Michael L Gyngell1, Katherine E Groves1, Andy McKay1, Amy H Herlihy1, Carolina C Fernandes1, Mark Halberstadt1, Marion Maguire1, Naomi Jayaratne1, Sophia Linden1, Stefan Neubauer1,3, Rajarshi Banerjee1.
Abstract
As the burden of liver disease reaches epidemic levels, there is a high unmet medical need to develop robust, accurate and reproducible non-invasive methods to quantify liver tissue characteristics for use in clinical development and ultimately in clinical practice. This prospective cross-sectional study systematically examines the repeatability and reproducibility of iron-corrected T1 (cT1), T2*, and hepatic proton density fat fraction (PDFF) quantification with multiparametric MRI across different field strengths, scanner manufacturers and models. 61 adult participants with mixed liver disease aetiology and those without any history of liver disease underwent multiparametric MRI on combinations of 5 scanner models from two manufacturers (Siemens and Philips) at different field strengths (1.5T and 3T). We report high repeatability and reproducibility across different field strengths, manufacturers, and scanner models in standardized cT1 (repeatability CoV: 1.7%, bias -7.5ms, 95% LoA of -53.6 ms to 38.5 ms; reproducibility CoV 3.3%, bias 6.5 ms, 95% LoA of -76.3 to 89.2 ms) and T2* (repeatability CoV: 5.5%, bias -0.18 ms, 95% LoA -5.41 to 5.05 ms; reproducibility CoV 6.6%, bias -1.7 ms, 95% LoA -6.61 to 3.15 ms) in human measurements. PDFF repeatability (0.8%) and reproducibility (0.75%) coefficients showed high precision of this metric. Similar precision was observed in phantom measurements. Inspection of the ICC model indicated that most of the variance in cT1 could be accounted for by study participants (ICC = 0.91), with minimal contribution from technical differences. We demonstrate that multiparametric MRI is a non-invasive, repeatable and reproducible method for quantifying liver tissue characteristics across manufacturers (Philips and Siemens) and field strengths (1.5T and 3T).Entities:
Mesh:
Year: 2019 PMID: 30970039 PMCID: PMC6457552 DOI: 10.1371/journal.pone.0214921
Source DB: PubMed Journal: PLoS One ISSN: 1932-6203 Impact factor: 3.240
Fig 1Study design.
Two manufacturers (Siemens and Philips) and a range of scanner models were used to systematically test the repeatability and reproducibility of multiparametric-MRI derived measurements in human participants and phantoms.
Fig 2Phantom T1 Standardisation.
Bland-Altman plots demonstrating T1 measurements in phantoms before and after standardisation at (a) 1.5T and (b) 3T.
Fig 3Repeatability and reproducibility of phantom measurements.
Bland-Altman plots from phantom measurements across manufacturer and field strength for (a) T1, (b) T2*, and (c) PDFF.
Fig 4Repeatability and reproducibility of human multiparametric MRI measurements.
Bland-Altman plots from human measurements across manufacturer and field strength for (a) cT1, (b) T2*, and (c) PDFF.