Shigenori Ito1,2,3, Kanako Kinoshita1, Akiko Endo1, Masato Nakamura1,4.
Abstract
We evaluated intra- and interobserver variability of quantitative coronary angiography (QCA) due to cine frame selection for 9 coronary stenoses. The projection was selected in advance. Cine frames were selected by 2 blinded experts (blind frame QCA) followed by assignment by supervisor (pre-selected frame QCA). Each expert analyzed 18 frames twice with a 3-month interval. A total of 72 measurements by 2 experts were used for intra- and interobserver variability analysis in calibration factor (CF), minimal lumen diameter (MLD), percent diameter stenosis (%DS), interpolated reference diameter (Int R), and lesion length (LL). Accuracy, precision, and coefficient of variation (CV) were calculated based on 2 measurements. For interobserver variability, intraclass correlation coefficient (ICC) was evaluated. Regarding intraobserver variability, precision (CV) was 0.0026 (1.45), 0.220 (25.1), 0.282 (11.0), 7.626 (11.8), and 4.042 (28.7) for blind frame QCA and 0.0044 (2.46), 0.094 (11.2), 0.225 (8.6), 3.924 (5.9), and 1.941 (12.1) for pre-selected frame QCA and regarding interobserver variability, precision (CV) was 0.0037 (2.09), 0.271 (31.8), 0.307 (11.9), 10.10 (15.4), and 5.121 (39.5) for blind frame QCA and 0.0050 (2.82), 0.098 (11.4), 0.246 (9.5), 5.253 (8.0), and 2.857 (19.0) for pre-selected frame QCA in CF, MLD, Int R, %DS, and LL, respectively. Intraclass correlation coefficient of Int R was almost perfect in blind and pre-selected frame QCA. Intraclass correlation coefficient of MLD, %DS, and LL were substantial/lower by blind frame QCA and improved to almost perfect by pre-selected frame QCA. Blind cine film selection might affect intra- and interobserver variability, especially in MLD and LL. In the multiple linear regression analysis, blind frame QCA was selected as an explanatory factor of QCA variability in MLD, %DS, and LL. The error range due to frame selection must be taken into consideration in clinical use.Entities:
Keywords: cardiac cycle; core laboratory; coronary intervention; coronary stenosis; frame selection; minimal lumen diameter; quantitative coronary angiography
Year: 2019 PMID: 30967747 PMCID: PMC6444776 DOI: 10.1177/1179546819838232
Source DB: PubMed Journal: Clin Med Insights Cardiol ISSN: 1179-5468
Clinical and lesion characteristics.
| Case no. | Age | Sex | Diagnosis | Coronary risk factors | Lesion angle/skew | Location of stenosis | Calcification[ |
|---|---|---|---|---|---|---|---|
| 1 | 51 | Male | Angina pectoris | HT DL | LAD#6 | Proximal | None/None |
| 2 | 74 | Male | Angina pectoris | DM HT DL | LAD #6-7, 9 | Proximal | None/None |
| 3 | 87 | Female | Prior myocardial infarction | DL | LCx #11-13 | Proximal | None/None |
| 4 | 62 | Male | Angina pectoris | SAS | LCx#13 | Mid | None/None |
| 5 | 73 | Male | Angina pectoris | HT DL | LCx #15 | Mid | Moderate/None |
| 6 | 80 | Male | Prior myocardial infarction | HT CKD | LAD#7,9 | Mid | Moderate/None |
| 7 | 57 | Male | Angina pectoris | DM DL smoking | LAD #7 | Mid | Severe/None |
| 8 | 70 | Male | Prior myocardial infarction | CKD smoking | LAD #6-7, 9 | Proximal | Moderate/None |
| 9 | 66 | Male | Unstable angina pectoris | DL smoking | LAD#7 LAO/Cranial | Mid | None/None |
Abbreviations: AP, anterior posterior; CKD, chronic kidney disease; DL, dyslipidemia; DM, diabetes mellitus; HT, hypertension; LAD, left anterior descending coronary artery; LAO, left anterior oblique view; LCx, left circumflex artery; RAO, right anterior oblique view; SAS, sleep apnea syndrome.
Calcification includes none/mild, moderate (densities noted only during the cardiac cycle prior to contrast injection), and severe (radio-opacities noted without cardiac motion prior to contrast injection generally involving both sides of the artery wall).
Difference in cine frame number between 2 analysis experts.
| Case no. | Difference in cine frame number in blind frame QCA analysis (first analysis) | Difference in cine frame number in blind frame QCA analysis (second analysis) | Difference in cine frame number in pre-selected QCA analysis |
|---|---|---|---|
| 1 | 2 | 11 (different cardiac cycle) | 0 |
| 2 | 13 (different cardiac cycle)[ | 8 (different cardiac cycle)[ | 0 |
| 3 | 1 | 1 | 0 |
| 4 | 0 | 9 (different cardiac cycle) | 0 |
| 5 | 0 | 0 | 0 |
| 6 | 1 | 8 (different cardiac cycle)[ | 0 |
| 7 | 14 (different cardiac cycle) | 2 | 0 |
| 8 | 0 | 2 | 0 |
| 9 | 1 | 0 | 0 |
Abbreviation: QCA, quantitative coronary angiography.
One analysis expert selected non-diastolic frame.
Variability of CF, MLD, Int R, %DS, and LL between QCAs for the blind and pre-selected frames.
| CF | Blind | Pre-selected | |
|---|---|---|---|
| Mean ± SD (range) | 0.17783 ± 0.00672 (0.0248) | 0.17847 ± 0.00493 (0.0166) | .827 |
| Accuracy (pre-selected-blind) | –0.00026 | ||
| Precision (pre-selected-blind) | 0.00489 | ||
| Coefficient of variation (%) | 2.74 | ||
| 0.6864 | .0017 | ||
| MLD (mm) | Blind | Pre-selected | |
| Mean ± SD (range) | 0.877 ± 0.270 (0.880) | 0.838 ± 0.244 (0.730) | .4461 |
| Accuracy (pre-selected-blind) | –0.039 | ||
| Precision (pre-selected-blind) | 0.215 | ||
| Coefficient of variation (%) | 25.7 | ||
| 0.654 | .003 | ||
| Int R (mm) | Blind | Pre-selected | |
| Mean ± SD (range) | 2.564 ± 0.674 (2.260) | 2.607 ± 0.816 (2.710) | .542 |
| Accuracy (pre-selected-blind) | 0.043 | ||
| Precision (pre-selected-blind) | 0.292 | ||
| Coefficient of variation (%) | 11.2 | ||
| 0.961 | <.0001 | ||
| %DS (%) | Blind | Pre-selected | |
| Mean ± SD (range) | 64.610 ± 11.419 (35.700) | 66.383 ± 10.159 (36.300) | .391 |
| Accuracy (pre-selected-blind) | 1.773 | ||
| Precision (pre-selected-blind) | 8.542 | ||
| Coefficient of variation (%) | 12.9 | ||
| 0.585 | .011 | ||
| LL (mm) | Blind | Pre-selected | |
| Mean ± SD (range) | 14.079 ± 6.600 (21.210) | 16.084 ± 6.681 (22.640) | .166 |
| Accuracy (pre-selected-blind) | 2.004 | ||
| Precision (pre-selected-blind) | 5.868 | ||
| Coefficient of variation (%) | 36.5 | ||
| 0.674 | .002 | ||
Abbreviations: CF, calibration factor; %DS, percent diameter stenosis; Int R, interpolated reference diameter; LL, lesion length; MLD, minimal lumen diameter; QCA, quantitative coronary angiography.
Intraobserver variability in 2 analysis experts for blind and pre-selected frame QCA.
| CF | MLD | Int R | %DS | LL | |
|---|---|---|---|---|---|
| Blind frame QCA | |||||
| A and B first | 0.1787 ± 0.0067 | 0.877 ± 0.270 | 2.564 ± 0.674 | 64.6 ± 11.4 | 14.08 ± 6.60 |
| A and B second | 0.1779 ± 0.0066 | 0.854 ± 0.221 | 2.499 ± 0.626 | 64.4 ± 11.8 | 12.31 ± 5.95 |
| | .1303 | .659 | .827 | .755 | .084 |
| Accuracy | 0.0008 | 0.023 | 0.064 | 0.238 | 1.771 |
| Precision | 0.0026 | 0.220 | 0.282 | 7.626 | 4.042 |
| Coefficient of variation (%) | 1.45 | 25.1 | 11.0 | 11.8 | 28.7 |
| Pre-selected frame QCA | |||||
| A and B first | 0.1785 ± 0.0049 | 0.838 ± 0.244 | 2.607 ± 0.829 | 66.4 ± 10.2 | 16.08 ± 6.68 |
| A and B second | 0.1773 ± 0.0043 | 0.872 ± 0.230 | 2.574 ± 0.738 | 65.2 ± 8.90 | 15.65 ± 6.23 |
| | .0625 | .140 | .551 | .206 | .356 |
| Accuracy | 0.0012 | −0.034 | 0.032 | 1.217 | 0.434 |
| Precision | 0.0044 | 0.094 | 0.225 | 3.924 | 1.941 |
| Coefficient of variation (%) | 2.46 | 11.2 | 8.6 | 5.9 | 12.1 |
Abbreviations: CF, calibration factor; %DS, percent diameter stenosis; Int R, interpolated reference diameter; LL, lesion length; MLD, minimal lumen diameter; QCA, quantitative coronary angiography.
Accuracy, precision, and coefficient of variation (%): same as in Table 3.
A and B indicate 2 analysis experts.
Interobserver variability between 2 analysis experts for blind and pre-selected frame QCA.
| CF | MLD | Int R | %DS | LL | |
|---|---|---|---|---|---|
| Blind frame QCA | |||||
| A first and second | 0.1772 ± 0.0058 | 0.852 ± 0.258 | 2.577 ± 0.605 | 65.7 ± 12.2 | 12.97 ± 6.573 |
| B first and second | 0.1795 ± 0.0072 | 0.879 ± 0.235 | 2.487 ± 0.691 | 63.3 ± 10.9 | 13.42 ± 6.113 |
| | .0209 | .682 | .231 | .333 | .717 |
| Accuracy | −0.0022 | −0.027 | 0.090 | 2.373 | −0.446 |
| Precision | 0.0037 | 0.271 | 0.307 | 10.10 | 5.121 |
| Coefficient of variance | 2.09 | 31.8 | 11.9 | 15.4 | 39.5 |
| Pre-selected frame QCA | |||||
| A first and second | 0.1775 ± 0.0044 | 0.863 ± 0.246 | 2.587 ± 0.727 | 65.5 ± 10.5 | 15.00 ± 6.18 |
| B first and second | 0.1793 ± 0.0063 | 0.847 ± 0.229 | 2.594 ± 0.838 | 66.0 ± 8.5 | 16.74 ± 6.62 |
| | .1498 | .482 | .910 | .682 | .019 |
| Accuracy | −0.0018 | 0.017 | −0.007 | −0.517 | −1.741 |
| Precision | 0.0050 | 0.098 | 0.246 | 5.253 | 2.857 |
| Coefficient of variance | 2.82 | 11.4 | 9.5 | 8.0 | 19.0 |
Abbreviations: CF, calibration factor; %DS, percent diameter stenosis; Int R, interpolated reference diameter; LL, lesion length; MLD, minimal lumen diameter; QCA, quantitative coronary angiography.
Accuracy, precision, and coefficient of variation (%): same as in Table 3.
A and B same as in Table 4.
Interobserver reliability between 2 analysis experts in the average of first and second for blind cine film analysis (N = 18).
| Variables | ICC | Method | Intraclass correlation | 95% CI | |
|---|---|---|---|---|---|
| MLD blind A:B | ICC (2, 1) | Single | 0.316 | −0.361 to 0.787 | .175 |
| ICC (2, 2) | Average | 0.480 | −1.132 to 0.881 | .175 | |
| Int R blind A:B | ICC (2, 1) | Single | 0.941 | 0.778 to 0.986 | <.001 |
| ICC (2, 2) | Average | 0.970 | 0.875 to0.993 | <.001 | |
| %DS blind A:B | ICC (2, 1) | Single | 0.492 | −0.162 to 0.855 | .065 |
| ICC (2, 2) | Average | 0.660 | −0.397 to 0.922 | .065 | |
| LL blind A:B | ICC (2, 1) | Single | 0.643 | 0.057 to 0.905 | .018 |
| ICC (2, 2) | Average | 0.783 | 0.109 to 0.950 | .018 |
Abbreviations: CI, confidence interval; %DS, percent diameter stenosis; ICC, intraclass correlation; Int R, interpolated reference diameter; LL, lesion length; MLD, minimal lumen diameter.
A and B: same as in Table 4.
Interobserver reliability between 2 analysis experts in the average of first and second for selected cine film analysis (N = 18).
| Variables | ICC | Method | Intraclass Correlation | 95% CI | |
|---|---|---|---|---|---|
| MLD check A:B | ICC (2, 1) | Single | 0.888 | 0.608 to 0.973 | <.001 |
| ICC (2, 2) | Average | 0.941 | 0.756 to 0.986 | <.001 | |
| Int R check A:B | ICC (2, 1) | Single | 0.941 | 0.777 to 0.986 | <.001 |
| ICC (2, 2) | Average | 0.969 | 0.874 to 0.993 | <.001 | |
| %DS check A:B | ICC (2, 1) | Single | 0.835 | 0.462 to 0.960 | .001 |
| ICC (2, 2) | Average | 0.910 | 0.632 to 0.979 | .001 | |
| LL check A:B | ICC (2, 1) | Single | 0.842 | 0.479 to 0.961 | .001 |
| ICC (2, 2) | Average | 0.914 | 0.647 to 0.980 | .001 |
Abbreviations: CI, confidence interval; %DS, percent diameter stenosis; ICC, intraclass correlation; Int R, interpolated reference diameter; LL, lesion length; MLD, minimal lumen diameter.
A and B: same as in Table 4.
Figure 1.Coronary angiograms for quantitative coronary angiography with a blindly selected and pre-selected cine frames (left and right, respectively).
Figure 3.Pictures after final quantitative coronary angiography analysis with functional lines and measured parameters. There were some differences in the values between the blind and pre-selected cine frames. %DS indicates percent diameter stenosis; Int R, interpolated reference diameter; LL, lesion length; MLD, minimal lumen diameter.
Figure 2.Pictures after borderline determination by the edge detection algorithm.