The effect of malalignment on proximal tibial strain in fixed-bearing unicompartmental knee arthroplasty: A comparison between metal-backed and all-polyethylene components using a validated finite element model.

I Danese1, P Pankaj1, C E H Scott2.   

Abstract

OBJECTIVES: Elevated proximal tibial bone strain may cause unexplained pain, an important cause of unicompartmental knee arthroplasty (UKA) revision. This study investigates the effect of tibial component alignment in metal-backed (MB) and all-polyethylene (AP) fixed-bearing medial UKAs on bone strain, using an experimentally validated finite element model (FEM).
METHODS: A previously experimentally validated FEM of a composite tibia implanted with a cemented fixed-bearing UKA (MB and AP) was used. Standard alignment (medial proximal tibial angle 90°, 6° posterior slope), coronal malalignment (3°, 5°, 10° varus; 3°, 5° valgus), and sagittal malalignment (0°, 3°, 6°, 9°, 12°) were analyzed. The primary outcome measure was the volume of compressively overstrained cancellous bone (VOCB) < -3000 µε. The secondary outcome measure was maximum von Mises stress in cortical bone (MSCB) over a medial region of interest.
RESULTS: Varus malalignment decreased VOCB but increased MSCB in both implants, more so in the AP implant. Varus malalignment of 10° reduced the VOCB by 10% and 3% in AP and MB implants but increased the MSCB by 14% and 13%, respectively. Valgus malalignment of 5° increased the VOCB by 8% and 4% in AP and MB implants, with reductions in MSCB of 7% and 10%, respectively. Sagittal malalignment displayed negligible effects. Well-aligned AP implants displayed greater VOCB than malaligned MB implants.
CONCLUSION: All-polyethylene implants are more sensitive to coronal plane malalignments than MB implants are; varus malalignment reduced cancellous bone strain but increased anteromedial cortical bone stress. Sagittal plane malalignment has a negligible effect on bone strain.Cite this article: I. Danese, P. Pankaj, C. E. H. Scott. The effect of malalignment on proximal tibial strain in fixed-bearing unicompartmental knee arthroplasty: A comparison between metal-backed and all-polyethylene components using a validated finite element model. Bone Joint Res 2019;8:55-64. DOI: 10.1302/2046-3758.82.BJR-2018-0186.R2.

Entities:  

Keywords:  Bone strain; Finite element analysis; Malalignment; Unicompartmental knee arthroplasty

Year:  2019        PMID: 30915211      PMCID: PMC6397332          DOI: 10.1302/2046-3758.82.BJR-2018-0186.R2

Source DB:  PubMed          Journal:  Bone Joint Res        ISSN: 2046-3758            Impact factor:   5.853


Finite element investigation of the effect of unicompartmental knee arthroplasty (UKA) tibial component alignment on cancellous and cortical bone strains and stresses in both metal-backed (MB) and all-polyethylene (AP) implants. Coronal plane malalignment affects bone strain more than sagittal malalignment, and this is more marked in AP UKA implants than in MB implants. Well-aligned AP implants display greater bone strains than malaligned MB implants. The strength of this finite element study is that it permits examination of bone strain for a wide range of malalignment scenarios for two different UKA implants. Limitations of this study include the use of composite tibias, the performance of a linearly elastic analysis, and the lack of kinematic analysis.

Introduction

The survival of unicompartmental knee arthroplasty (UKA) remains variable between implants and institutions, with joint registries reporting ten-year survival ranging from 80% to 96%.[1-3] The strong relationship between surgeon volume and UKA survivorship[4,5] implies that achieving well-aligned congruent UKAs is key to maximizing survival. Unexplained pain is a significant cause of UKA revision, with between 24% and 48% of revisions performed for this indication.[6,7] Elevated proximal tibial strain and microdamage may contribute to this pain.[8,9] Joint registries do not distinguish between metal-backed (MB) and all-polyethylene (AP) UKAs, but both experimental data[10,11] and finite element analysis (FEA)[12] have shown elevated proximal tibial bone strain in AP UKAs compared with MB medial UKAs. Finite element analysis has been used previously to investigate the influence of component alignment on cortical bone strain in MB implants.[13-15] Both coronal plane varus malalignment[13,14] and valgus malalignment[15] have been associated with elevated cortical bone strains in FEAs of a MB mobile-bearing UKA implant. In a MB fixed-bearing implant, mild valgus malalignment of between 2° and 4° has been associated with a reduction in cancellous bone strain.[13] Cortical rim support has also been identified as an important predictor of bone strain, with both overhang and underhang associated with elevated cortical bone strains again in a MB mobile-bearing UKA implant.[14] We have previously shown elevated cancellous bone strains in AP UKAs, compared with MB UKAs, in optimal alignment,[11,12] in addition to an increased revision rate for unexplained pain.[16] To our knowledge, no investigation of the influence of metal backing and malalignment has been published. The aims of this study were to investigate the effect of medial UKA tibial component alignment on bone strain in both MB and AP implants. We hypothesized that bone strains beneath AP components would be more reactive to malalignment than the bone beneath MB implants.

Materials and Methods

A finite element model (FEM) previously experimentally validated using digital image correlation (± 4.5% error) and acoustic emissions (± 12.5% error) was used.[12] This consisted of a third-generation left composite tibia with cortical and cancellous parts and fixed-bearing UKAs of 8 mm thickness (SIGMA Partial; DePuy Synthes, Raynham, Massachusetts) with 1.5 mm cement mantles (Fig. 1). Different component alignments were analyzed using Abaqus CAE Version 6.12 (Simulia, Dassault Systèmes Simulia Corp., Waltham, Massachusetts).

The finite element model: a) cortical bone part; b) cancellous bone part; c) metal-backed implant (polyethylene, metal tray) and cement meshes; and d) all-polyethylene implant and cement mesh. The area of distributed load is highlighted.

The finite element model: a) cortical bone part; b) cancellous bone part; c) metal-backed implant (polyethylene, metal tray) and cement meshes; and d) all-polyethylene implant and cement mesh. The area of distributed load is highlighted. Anatomical axes were defined in coronal and sagittal planes.[17] The tibia was cut distally 200 mm below the intercondylar eminence to reduce computational effort. Proximal medial bony resections of 6 mm were made in various alignments. The neutral alignment was 90° in the coronal plane with 6° of posterior slope in the sagittal plane. The following malalignments were investigated: 3°, 5°, and 10° varus; 3° and 5° valgus; and 0°, 3°, 9°, and 12° of posterior slope. All materials used in the FE model were assumed to exhibit isotropic, homogeneous, and linearly elastic material behaviour. Material properties are shown in Table I. Linear tetrahedral meshes with mean internodal distances of 1.5 mm to 2.0 mm were used. Mesh resolution was based on a 2% convergence criterion for the displacement magnitudes in the proximal tibia (Table I, Fig. 1). Mesh convergence studies showed a convergence of < 1.2% between element sizes of 1.5 mm and 2.5 mm; 2 mm was therefore chosen for cortical and cancellous bone (Table I). Tie constraints were used to bond cement to bone and tibial insert to metal baseplate.[18] The distal tibia was fully restrained. Proximally, the tibia was restrained against medial/lateral and anterior/posterior translations at a node representing the anterior cruciate ligament (ACL) footprint to prevent non-physiological bending.
Table I.

Material properties and elements for parts in the standard alignment models. Cortical and cancellous bone properties apply to loading in compression

PartElastic modulus, GPaPoisson’s ratioInternodal distance, mmElements, n
All-polyethylene (AP)
Cortical bone16.70.32108 542
Cancellous bone0.1550.3296 573
PMMA cement2.40.31.519 691
AP tibia0.690.46221 496
Metal-backed (MB)
Cortical bone16.70.32105 984
Cancellous bone0.1550.3299 485
PMMA cement2.40.31.525 630
MB tibial tray (CoCr)2100.31.519 851
Polyethylene insert0.690.461.525 630

PMMA, poly(methyl methacrylate); CoCr, cobalt chromium

Material properties and elements for parts in the standard alignment models. Cortical and cancellous bone properties apply to loading in compression PMMA, poly(methyl methacrylate); CoCr, cobalt chromium Load was applied directly to the polyethylene articular surface medially and to the lateral tibial plateau, in order to replicate previous biomechanical testing, as distributed loads over a circle with a radius of 6 mm centred at the centre point of each polyethylene/plateau.[11] A 60:40 medial:lateral load division[19] was used, and a 2500 N medial load (4170 N maximum total load) was applied over five equal 500 N increments. This maximum load was approximately six times body weight (70 kg) and reflects physiological tibiofemoral loading.[20] Results are presented for 1000 N and 2500 N medial loads (1668 N and 4170 N total loads). The primary outcome measure was the volume of compressively overstrained cancellous bone (VOCB) experiencing minimum principal strain of less than -3000 µε, representing pathological overloading.[21] This limit has been used as a strain criterion previously.[12] In accordance with the commonly used convention in engineering, the negative sign denotes compression and positive tension. Secondary outcomes included maximum von Mises stress in cortical bone (MSCB). This was measured over a region of interest overlying the anteromedial tibia, as previously used by Simpson et al,[8] at the location frequently associated with patient reports of pain (Fig. 2).
Fig. 2

Region of interest for maximum von Mises stress in cortical bone.

Region of interest for maximum von Mises stress in cortical bone.

Results

Standard alignment

The AP tibial component generated a greater VOCB than the MB tibial component (Table II), with VOCB distributed over a larger part of the resected cancellous bone surface and extending deeper into it. For the MB implant, the VOCB was generated primarily at the implant peg and keel. Numerical data are displayed in Table II. Even when correctly aligned, there was a substantial difference in VOCB between the two tibial components at all loads (Table II).
Table II.

The volume of compressively overstrained cancellous bone (VOCB, minimum principal strain) in all-polyethylene (AP) and metal-backed (MB) tibial components in standard alignment

Medial load, NTotal load, NStrain value, µεAP VOCB, mm3MB VOCB, mm3% increase in VOCB for AP
10001668< -300097444310+ 126.1
25004170< -300028 44921 876+ 30.0
The volume of compressively overstrained cancellous bone (VOCB, minimum principal strain) in all-polyethylene (AP) and metal-backed (MB) tibial components in standard alignment

Coronal malalignment

Figures 3a and 3b show the VOCB and the MSCB in the anteromedial tibia for both implants at medial loads of 2500 N and 1000 N, respectively. The VOCB under the MB tibial component was lower than for the AP in all scenarios. In fact, even a poorly aligned MB tibial component yielded a smaller VOCB compared with a perfectly aligned AP component (Fig. 4). Overall, the MB implant was less sensitive to coronal plane malalignment in terms of VOCB than the AP implant, with 2% to 6% increases in VOCB compared with 5% to 10% increases in the AP tibial component (Table III).

Coronal plane malalignment. The volume of compressively overstrained cancellous bone (VOCB, < -3000 µε) and the anteromedial maximum von Mises stress in cortical bone (MSCB) for all-polyethylene (AP) and metal-backed (MB) implants at medial loads of a) 2500 N and b) 1000 N. Bars (left-hand y-axis) represent the VOCB and lines (right-hand y-axis) represent the MSCB.

Fig. 4

Mid-coronal plane contour maps showing for all-polyethylene and metal-backed implants in varus and valgus malalignment at a medial load of 2500 N.

Table III.

The volume of compressively overstrained cancellous bone (VOCB, < -3000 µε) by implant and alignment for a medial load of 2500 N

AlignmentAll-polyethylene implant
Metal-backed implant
VOCB, mm3% diff.MSCB, MPa% diff.VOCB, mm3% diff.MSCB, MPa% diff.
Coronal plane
5° valgus29 860+5.018.9-7.122 502+2.914.6-9.7
3° valgus29 300+3.019.5-4.422 325+2.115.2-6.1
0°[*]28 4490.020.40.021 8760.016.20.0
3° varus27 275-4.121.9+7.721 324-2.517.1+5.9
5° varus27 582-3.022.0+7.921 365-2.317.1+5.8
10° varus26 882-5.523.2+13.920 516-6.218.3+13.3
Sagittal plane
29 643+4.27.9-3.421 833-0.26.6+2.3
28 374-0.38.0-2.222 070+0.96.5+0.7
6°[*]28 4490.08.20.021 8760.06.50.0
28 287-0.68.3+2.421 772-0.56.3-1.8
12°28 473+0.18.5+4.321 707-0.86.50.0

Standard % difference compared with standard alignment

diff., difference; MSCB, maximum von Mises stress in cortical bone

Coronal plane malalignment. The volume of compressively overstrained cancellous bone (VOCB, < -3000 µε) and the anteromedial maximum von Mises stress in cortical bone (MSCB) for all-polyethylene (AP) and metal-backed (MB) implants at medial loads of a) 2500 N and b) 1000 N. Bars (left-hand y-axis) represent the VOCB and lines (right-hand y-axis) represent the MSCB. Mid-coronal plane contour maps showing for all-polyethylene and metal-backed implants in varus and valgus malalignment at a medial load of 2500 N. The volume of compressively overstrained cancellous bone (VOCB, < -3000 µε) by implant and alignment for a medial load of 2500 N Standard % difference compared with standard alignment diff., difference; MSCB, maximum von Mises stress in cortical bone The absolute VOCB was much lower for the 1000 N medial load case compared with the 2500 N medial load case (Fig. 3); for the lower load, the two tibial components only yielded 34% (AP) and 20% (MB) of the VOCB found for the higher load. Standard alignment was not associated with the lowest cancellous bone strains (VOCB) in either implant. Increasing varus alignment reduced VOCB, but this was balanced by a reciprocal increase in MSCB. This is highlighted in Figure 3 and by the contour maps in Figure 4. The same pattern existed for both compressive (minimum principal) and tensile (maximum principal) strains (Fig. 5). The MB tibial component distributes load more uniformly over the resected surface. Stress and strain concentrations are therefore less pronounced for the MB implant than for the AP implant. This explains why a higher MSCB was seen for the AP tibial component, even though the load-bearing support of the cortical bone was less for the AP implant than for the MB implant. Standard alignment appears to represent balance between cancellous bone strain and cortical bone stress.
Fig. 5

Anteromedial contour maps showing von Mises stress in cortical bone (MSCB) for all-polyethylene and metal-backed implants in varus and valgus malalignment at a medial load of 2500 N.

Anteromedial contour maps showing von Mises stress in cortical bone (MSCB) for all-polyethylene and metal-backed implants in varus and valgus malalignment at a medial load of 2500 N.

Sagittal malalignment

Changes in tibial slope had negligible effects on both VOCB and MSCB compared with coronal plane malalignments. A standard 6° posterior tibial slope produced the least VOCB. Both neutral slope (0°) and excessive slope (12°) increased the VOCB at medial loads of 1000 N and 2500 N by small amounts (Table III, Figs 6a and 6b). The MSCB increased as the slope increased in the AP implant but not in the MB implant. Considering that these differences are small, approaching values close to the mesh convergence criteria (< 1.2%), it can be concluded that sagittal plane malalignment is not detrimental to proximal tibial bone strain, neither for the AP nor for the MB tibial component in this model loaded without the femur.
Fig. 6

Sagittal plane malalignment: the volume of compressively overstrained cancellous bone (VOCB) (< -3000 µε) and the anteromedial maximum von Mises stress in cortical bone (MSCB) for all-polyethylene (AP) and metal-backed (MB) implants at medial loads of a) 2500 N and b) 1000 N. The VOCB is represented by bars (left-hand y-axis) and the MSCB by lines (right-hand y-axis).

Sagittal plane malalignment: the volume of compressively overstrained cancellous bone (VOCB) (< -3000 µε) and the anteromedial maximum von Mises stress in cortical bone (MSCB) for all-polyethylene (AP) and metal-backed (MB) implants at medial loads of a) 2500 N and b) 1000 N. The VOCB is represented by bars (left-hand y-axis) and the MSCB by lines (right-hand y-axis). Although the absolute cancellous strain and cortical stress differ little with sagittal alignment, contour maps show that the locations of stress/strain concentrations are affected by sagittal alignment (Fig. 7). Cancellous bone strain shifts posteriorly with increasing posterior slope, more so in the AP implant. Again, the MB tibial component yielded less VOCB, with concentrations primarily at the peg and keel. The small amount of variation in the MSCB with sagittal malalignment reflects constant cortical support across these malalignments (Figs 6a and 6b).
Fig. 7

a) Cancellous bone minimum principal strain and b) von Mises stress in cortical bone (MSCB) contours for all-polyethylene and metal-backed implants in different sagittal alignments.

a) Cancellous bone minimum principal strain and b) von Mises stress in cortical bone (MSCB) contours for all-polyethylene and metal-backed implants in different sagittal alignments.

Discussion

Malalignment had less effect on the VOCB than implant selection; well-aligned AP implants displayed greater volumes of overstrained cancellous bone and greater anteromedial MSCB than poorly aligned MB implants at low and high loads. Coronal plane alignment had a greater influence on bone strain and stress than sagittal plane alignment, and AP implants were more sensitive to malalignment than MB implants. A balance appears to exist between cancellous bone strain and cortical bone stress; contour maps and numerical data suggest that increasing cancellous bone strain concentrations are offset by decreasing cortical bone stress concentrations, and vice versa for coronal plane malalignment. For both implants, valgus malalignment was associated with higher volumes of overstrained cancellous bone but lower cortical bone stresses. Varus malalignment led to lower volumes of overstrained cancellous bone, but greater maximum anteromedial cortical bone stresses. Our results suggest that although inaccurate alignment of the tibial component does affect cancellous bone strain, the effect of the selected tibial component material is much greater. Our findings of increasing cortical stresses in varus malalignment and increasing cancellous strains in valgus malalignment are similar to the findings of Simpson et al[14] and Sawatari et al.[13] Simpson et al[14] found that increasing varus alignment led to the largest increase in mean von Mises stress in cortical bone, with a 14% increase from the standard alignment to 2° varus, while a 2° valgus malalignment yielded only a 4% increase. Sawatari et al[13] reported a 77% increase in the yield area of the cancellous bone at the resected surface when a MB tibial component was malaligned from 0° to 6° varus, with a decrease of 30% with 4° of valgus. With increasing varus malalignment, the cancellous bone strain concentration shifts medially, resulting in more support from the cortical bone, thus lowering cancellous strains and increasing cortical stresses. It may be expected that stiffer MB implants, which shield the cancellous bone from compressive stresses and strains, convey more to the cortical bone. In our model, cortical bone stresses were consistently higher in the AP implants. Our use of tie constraints between bone and cement may have led to an overestimation of cortical stresses, as less stiff AP implants bend more under load. While we employed minimum principal strain for cancellous bone regions considered in this study (as the strain state is expected to be largely compressive), the cortical bone can experience both tension and compression due to possible bending (previous biomechanical testing indicated tensile and compressive strains over the proximal tibial cortex).[11] We chose von Mises stress to examine the stress state, since it is perhaps the most demonstrative scalar measure in stress state evaluation within a material (and is an important constituent of almost all yield criteria) and accounts for both tensile and compressive stresses. Also, von Mises stress enables comparison with much of the literature. Joint registries do not distinguish between AP and MB UKA implants. Poor survivorship of AP UKAs has been reported,[22-25] with early failures commonly experienced due to tibial loosening, subsidence, or pain. The increased sensitivity to malalignment demonstrated here in AP implants may well contribute to these modes of failure. It is uncertain whether greater cancellous bone strain translates into poorer clinical outcome, but greater strains with ongoing microfracture and remodelling may be the cause of unexplained pain, which is a larger problem in AP than in MB UKAs.[9] Our previous FEA of AP and MB UKAs demonstrated a significant effect of polyethylene thickness in AP but not in MB implants.[12] Malalignment was investigated here using 8 mm components only. It could be expected that the effects of coronal plane malalignment would be magnified with thinner implants. Although excessive or neutral tibial slope had little effect on bone strains, its influence is undoubtedly exerted kinematically with flexion instability or stiffness rather than bone strain concentrations. The advent of robotic UKA offers improved implant alignment with less variability.[26] This technology is based around fixed-bearing implants, and may well help to narrow the market gap between mobile-bearing (66% of the market in the United Kingdom)[1] and fixed-bearing UKAs (33% of the market in the United Kingdom).[1] Using MB implants, it remains unclear whether improved alignment of a robotic fixed-bearing UKA translates into superior clinical and patient-reported outcomes compared with a manually inserted mobile-bearing implant, although early results are promising.[27-30] Although AP UKAs are more sensitive than MB implants to malalignment, it also appears that associated bone strains for AP implants are higher than those for MB implants, even when perfectly aligned. Robotic-assisted surgery may not therefore provide a resurgence in AP implant use that would exploit the advantages of these implants in terms of ease of revision;[7] however, the clinical consequences of elevated strain are uncertain and may be resolved in a number of individuals by bone remodelling. A linear elastic FEM was used in this study. Although bone is viscoelastic with non-linear behaviour,[31,32] linear modelling can be used to reduce computing requirements when loading is not cyclical and not to failure.[19,33] It is argued that yielding and damage in bone is best described using strain rather than stress.[33] Strain-based criteria are numerically more efficient and accurate than stress-based criteria.[33] Previous UKA FEMs report von Mises stress and strains.[34-36] The threshold limit for cancellous bone strain of 3000 με was chosen to represent the strain at which cancellous bone is thought to be pathologically overloaded.[21] The limitations of this study include the use of composite tibias. These do not reflect the graduated trabecular structure of proximal tibial cancellous bone, but are applicable to the ‘average’ tibia.[37] Anisotropic, heterogeneous bone was modelled exhibiting isotropic and homogeneous material properties, and a linearly elastic analysis was performed. This is a common method and does not necessarily discredit the differences found between implants and alignments.[12] The use of strain-based criteria rather than stress-based criteria accommodates for material anisotropy and bone:volume ratio (or porosity). While yield stresses for bone are both anisotropy- and porosity-dependent, yield strains are largely uniform.[33,38] The threshold used for defining ‘overstrain’ (-3000 µε) is based on evidence[21,39] and has been used previously,[12] but it is unclear whether this cut-off accurately represents pathological overloading of bone and indeed this threshold may differ by patient. Gait was not modelled. As kinematic studies have shown the point of contact to change little throughout a range of movement in fixed-bearing UKAs, this was considered acceptable.[40] The soft tissues of the intact lateral compartment were not modelled and this will have affected lateral strain. The medial load was applied directly to the polyethylene, not through a femoral component. Although this is a recognized technique,[12,14] it may affect the results. In this study, all models were loaded in line with the mechanical axis. In reality, malaligning the implants, especially in extreme cases, may alter the loading axis relative to the native plateau. In conclusion, this FEM study has shown that UKA tibial component material has a greater effect than malalignment on proximal tibial bone strain. Sagittal plane malalignment has a negligible effect on bone strain for a given load. All-polyethylene implants are more sensitive than MB implants to coronal plane malalignments. Cancellous bone strain and anteromedial cortical bone stress appear to offset one another with varus malalignment, resulting in reduced cancellous bone strain but increased anteromedial cortical bone stress and vice versa in valgus alignment.
  35 in total

1.  Mechanical validation of whole bone composite tibia models.

Authors:  L Cristofolini; M Viceconti
Journal:  J Biomech       Date:  2000-03       Impact factor: 2.712

2.  The routine of surgical management reduces failure after unicompartmental knee arthroplasty.

Authors:  O Robertsson; K Knutson; S Lewold; L Lidgren
Journal:  J Bone Joint Surg Br       Date:  2001-01

3.  In vivo determination of knee kinematics for subjects implanted with a unicompartmental arthroplasty.

Authors:  Jean-Noël A Argenson; Richard D Komistek; Jean-Manuel Aubaniac; Douglas A Dennis; Eric J Northcut; Dylan T Anderson; Serge Agostini
Journal:  J Arthroplasty       Date:  2002-12       Impact factor: 4.757

4.  Three-dimensional finite element analysis of unicompartmental knee arthroplasty--the influence of tibial component inclination.

Authors:  T Sawatari; H Tsumura; K Iesaka; Y Furushiro; T Torisu
Journal:  J Orthop Res       Date:  2005-03-23       Impact factor: 3.494

5.  Incidence and reasons for reoperation after minimally invasive unicompartmental knee arthroplasty.

Authors:  William G Hamilton; Matthew B Collier; Eshan Tarabee; James P McAuley; C Anderson Engh; Gerard A Engh
Journal:  J Arthroplasty       Date:  2006-09       Impact factor: 4.757

6.  Failure mechanisms after unicompartmental and tricompartmental primary knee replacement with cement.

Authors:  O Furnes; B Espehaug; S A Lie; S E Vollset; L B Engesaeter; L I Havelin
Journal:  J Bone Joint Surg Am       Date:  2007-03       Impact factor: 5.284

7.  Experimental validation of a finite element model of a composite tibia.

Authors:  H A Gray; A B Zavatsky; F Taddei; L Cristofolini; H S Gill
Journal:  Proc Inst Mech Eng H       Date:  2007-04       Impact factor: 1.617

8.  Some ABC's of skeletal pathophysiology. 5. Microdamage physiology.

Authors:  H M Frost
Journal:  Calcif Tissue Int       Date:  1991-10       Impact factor: 4.333

9.  Experimental validation of a finite element model of a human cadaveric tibia.

Authors:  Hans A Gray; Fulvia Taddei; Amy B Zavatsky; Luca Cristofolini; Harinderjit S Gill
Journal:  J Biomech Eng       Date:  2008-06       Impact factor: 2.097

10.  Early failure of unicompartmental knee arthroplasty.

Authors:  E Marc Mariani; Michael H Bourne; Richard T Jackson; Scott T Jackson; Portia Jones
Journal:  J Arthroplasty       Date:  2007-09       Impact factor: 4.757

View more
  8 in total

1.  Cement augmentation of calcar screws may provide the greatest reduction in predicted screw cut-out risk for proximal humerus plating based on validated parametric computational modelling: Augmenting proximal humerus fracture plating.

Authors:  Peter Varga; Jason A Inzana; James W A Fletcher; Ladina Hofmann-Fliri; Armin Runer; Norbert P Südkamp; Markus Windolf
Journal:  Bone Joint Res       Date:  2020-09-03       Impact factor: 5.853

2.  Biomechanical effect of tibial slope on the stability of medial unicompartmental knee arthroplasty in posterior cruciate ligament-deficient knees.

Authors:  Jin-Ah Lee; Yong-Gon Koh; Paul Shinil Kim; Ki Won Kang; Yoon Hae Kwak; Kyoung-Tak Kang
Journal:  Bone Joint Res       Date:  2020-09-23       Impact factor: 5.853

Review 3.  [Research progress on finite element analysis of unicompartmental knee arthroplasty in medial knee compartmental osteoarthritis].

Authors:  Huazhang Xiong; Yi Zeng; Haibo Si; Yuangang Wu; Bin Shen
Journal:  Zhongguo Xiu Fu Chong Jian Wai Ke Za Zhi       Date:  2021-06-15

4.  Robotic-assisted unicompartmental knee arthroplasty has a greater early functional outcome when compared to manual total knee arthroplasty for isolated medial compartment arthritis.

Authors:  N D Clement; A Bell; P Simpson; G Macpherson; J T Patton; D F Hamilton
Journal:  Bone Joint Res       Date:  2020-05-16       Impact factor: 5.853

5.  Metaphyseal cones in revision total knee arthroplasty: The role of stems.

Authors:  Shuqiao Xie; Noel Conlisk; David Hamilton; Chloe Scott; Richard Burnett; Pankaj Pankaj
Journal:  Bone Joint Res       Date:  2020-05-16       Impact factor: 5.853

6.  The risk of tibial eminence avulsion fracture with bi-unicondylar knee arthroplasty : a finite element analysis.

Authors:  Jennifer C Stoddart; Amy Garner; Mahmut Tuncer; Justin P Cobb; Richard J van Arkel
Journal:  Bone Joint Res       Date:  2022-08       Impact factor: 4.410

7.  Computational analysis of tibial slope adjustment with fixed-bearing medial unicompartmental knee arthroplasty in ACL- and PCL-deficient models.

Authors:  Hyuck M Kwon; Jin-Ah Lee; Yong-Gon Koh; Kwan K Park; Kyoung-Tak Kang
Journal:  Bone Joint Res       Date:  2022-07       Impact factor: 4.410

8.  Exposure to hypergravity during zebrafish development alters cartilage material properties and strain distribution.

Authors:  Elizabeth Anna Lawrence; Jessye Aggleton; Jack van Loon; Josepha Godivier; Robert Harniman; Jiaxin Pei; Niamh Nowlan; Chrissy Hammond
Journal:  Bone Joint Res       Date:  2021-02       Impact factor: 5.853

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.