J Rueckel1, M Stockmar1, F Pfeiffer1, J Herzen2.
Abstract
We report on an experimental characterization of the spatial resolution of a commercial X-ray micro-computed tomography scanner. We have measured the full modulation transfer function (MTF) to assess the spatial resolution. The MTF and those spatial frequencies corresponding to a contrast loss of 50% were determined as a function of different applied X-ray tube parameters and magnification-dependent pixel sizes. A significant influence of the focal spot enlargement on the achievable spatial resolution could be shown. Our results allow for the designation of optimal X-ray tube parameters for a specific application requirement.Keywords: General electric; MTF; Micro-CT; Spatial resolution; X-ray imaging
Mesh:
Year: 2014 PMID: 25233529 DOI: 10.1016/j.apradiso.2014.08.014