Abstract
AIMS: To perform a dosimetric evaluation of four different simultaneous integrated boost whole brain radiotherapy modalities with hippocampus and inner ear avoidance in the treatment of limited brain metastases.Entities:
Keywords: Brain metastases; Dosimetry; Hippocampus; Inner ear; Intensity-modulated radiation therapy; Tomotherapy; Volumetric-modulated arc therapy
Mesh:
Year: 2019 PMID: 30876444 PMCID: PMC6419811 DOI: 10.1186/s13014-019-1255-7
Source DB: PubMed Journal: Radiat Oncol ISSN: 1748-717X Impact factor: 3.481
Characteristics of 10 NSCLC patients with brain metastases
| Patient No. | Number of metastases | Total GTV volume (cc) | Bilateral hippocampus volume (cc) | Bilateral inner ear volume (cc) | Brain volume (cc) |
|---|---|---|---|---|---|
| 1 | 2 | 3.7 | 6.3 | 2.9 | 1189.6 |
| 2 | 1 | 12.9 | 5.7 | 2.0 | 1339.6 |
| 3 | 1 | 6.4 | 4.4 | 2.6 | 1656.0 |
| 4 | 4 | 9.0 | 5.8 | 2.0 | 1303.4 |
| 5 | 3 | 3.2 | 4.7 | 2.6 | 1239.4 |
| 6 | 1 | 7.3 | 5.0 | 2.3 | 1478.3 |
| 7 | 2 | 79.4 | 2.2(R) | 1.8 | 1335.7 |
| 8 | 3 | 1.5 | 6.3 | 2.0 | 1483.8 |
| 9 | 4 | 1.6 | 7.7 | 2.3 | 1387.2 |
| 10 | 5 | 3.3 | 2.7(L) | 2.3 | 1145.1 |
GTV Gross tumor volume, R right, L left
Average dose parameters for PTVmets according to the four planning modalities
| sIMRT | dIMRT | VMAT | Tomo | ||
|---|---|---|---|---|---|
| D2% (Gy) | 50.8 ± 2.1 | 50.9 ± 2.4 | 51.4 ± 2.7 | 50.2 ± 2.7 | > 0.05 |
| D98% (Gy) | 47.3 ± 2.5 | 47.6 ± 2.4 | 46.8 ± 2.2 | 46.9 ± 2.3 | > 0.05 |
| Dmedian (Gy) | 49.7 ± 2.5 | 49.7 ± 2.5 | 49.7 ± 2.5 | 49.3 ± 2.6 | > 0.05 |
| TC (%) | 99.1 ± 1.0 | 99.4 ± 0.7 | 98.9 ± 1.6 | 97.1 ± 1.7 | c,e,f |
| HI | 0.071 ± 0.021 | 0.066 ± 0.012 | 0.091 ± 0.024 | 0.065 ± 0.023 | b,d,f |
| CI | 0.74 ± 0.10 | 0.73 ± 0.10 | 0.73 ± 0.13 | 0.71 ± 0.10 | > 0.05 |
Each value was calculated from the data for all 10 patients. Values are expressed as mean ± SD. sIMRT step-and-shoot intensity-modulated radiotherapy, dIMRT dynamic intensity-modulated radiation therapy, VMAT volumetric-modulated arc therapy, Tomo helical tomotherapy. a: sIMRT vs dIMRT, b: sIMRT vs VMAT, c: sIMRT vs Tomo, d: dIMRT vs VMAT, e: dIMRT vs Tomo, f: VMAT vs Tomo. If c is in the “p value” column, it means that there is statistical significance between sIMRT and Tomo
Dose parameters for the PTVbrain according to the four planning modalities
| sIMRT | dIMRT | VMAT | Tomo | p value | |
|---|---|---|---|---|---|
| D2% (Gy) | 39.1 ± 1.2 | 39.4 ± 1.1 | 40.6 ± 2.4 | 37.5 ± 2.8 | e,f |
| D98% (Gy) | 14.7 ± 3.2 | 14.9 ± 3.3 | 21.7 ± 1.1 | 25.2 ± 2.0 | b,c,d,e,f |
| Dmedian (Gy) | 32.1 ± 0.7 | 32.1 ± 0.6 | 33.2 ± 1.1 | 31.6 ± 0.4 | b,d,f |
| TC (%) | 88.2 ± 4.2 | 90.6 ± 1.8 | 88.7 ± 2.6 | 92.6 ± 2.1 | |
| HI | 0.76 ± 0.11 | 0.76 ± 0.12 | 0.57 ± 0.08 | 0.39 ± 0.11 | b,c,d,e,f |
| CI | 0.75 ± 0.04 | 0.77 ± 0.03 | 0.76 ± 0.03 | 0.79 ± 0.06 | > 0.05 |
Each value was calculated from the data for all 10 patients. Values are expressed as mean ± SD. sIMRT step-and-shoot intensity-modulated radiotherapy, dIMRT dynamic intensity-modulated radiation therapy, VMAT volumetric-modulated arc therapy, Tomo helical tomotherapy. a: sIMRT vs dIMRT, b: sIMRT vs VMAT, c: sIMRT vs Tomo, d: dIMRT vs VMAT, e: dIMRT vs Tomo, f: VMAT vs Tomo. If c is in the “p value” column, it means that there is statistical significance between sIMRT and Tomo
Fig. 1Color wash comparison of dose distributions for four modalities in a representative patient. sIMRT: step-and-shoot intensity-modulated radiotherapy, dIMRT: dynamic intensity-modulated radiation therapy, VMAT: volumetric-modulated arc therapy, Tomo: helical tomotherapy
Fig. 2Dose–volume histograms for treatment planning with the four modalities for a representative patient. sIMRT: step-and-shoot intensity-modulated radiotherapy, dIMRT: dynamic intensity-modulated radiation therapy, VMAT: volumetric-modulated arc therapy, Tomo: helical tomotherapy
Doses to OARs
| sIMRT | dIMRT | VMAT | Tomo | p value | |
|---|---|---|---|---|---|
| Hippocampus Dmax (Gy) | 14.3 ± 1.1 | 14.5 ± 1.4 | 15.8 ± 0.7 | 15.5 ± 0.9 | b,c,d,e |
| Hippocampus Dmean (Gy) | 9.7 ± 0.2 | 10.1 ± 0.3 | 10.0 ± 0.6 | 10.0 ± 0.7 | > 0.05 |
| Hippocampus Dmin (Gy) | 8.9 ± 0.3 | 9.3 ± 0.2 | 8.3 ± 0.4 | 7.6 ± 0.7 | a,b,c,d,e,f |
| Left inner ear Dmean (Gy) | 13.7 ± 0.7 | 13.8 ± 0.8 | 13.4 ± 1.0 | 13.1 ± 1.0 | > 0.05 |
| Right inner ear Dmean (Gy) | 13.6 ± 0.8 | 13.8 ± 0.7 | 13.7 ± 1.6 | 13.3 ± 1.1 | > 0.05 |
| Eyes Dmax (Gy) | 32.5 ± 4.9 | 32.9 ± 4.8 | 26.7 ± 6.1 | 30.8 ± 4.7 | b,d,f |
| Lens Dmax (Gy) | 5.7 ± 0.6 | 5.8 ± 0.6 | 6.8 ± 1.3 | 5.5 ± 0.5 | b,d,f |
| Optic nerves Dmax (Gy) | 33.6 ± 1.7 | 33.8 ± 1.7 | 32.0 ± 2.0 | 31.3 ± 0.9 | c,e |
| Brainstem Dmax (Gy) | 35.4 ± 1.1 | 35.7 ± 0.9 | 35.9 ± 0.6 | 34.0 ± 0.2 | c,e,f |
Each value was calculated from the data of all 10 patients. Values are expressed as mean ± SD. sIMRT step-and-shoot intenseity-modulated radiotherapy, dIMRT dynamic-intensity modulated radiation therapy, VMAT volumetric-modulated arc therapy, Tomo helical tomotherapy. a: sIMRT vs dIMRT, b: sIMRT vs VMAT, c: sIMRT vs Tomo, d: dIMRT vs VMAT, e: dIMRT vs Tomo, f: VMAT vs Tomo. If c is in the “p value” column, it means that there is statistical significance between sIMRT and Tomo