| Literature DB >> 30867928 |
Jennifer L Wierman1, Olivier Paré-Labrosse2,3, Antoine Sarracini2, Jessica E Besaw2, Michael J Cook1, Saeed Oghbaey2, Hazem Daoud2, Pedram Mehrabi3, Irina Kriksunov1, Anling Kuo4, David J Schuller1, Scott Smith1, Oliver P Ernst4, Doletha M E Szebenyi1, Sol M Gruner1,5,6, R J Dwayne Miller2,3, Aaron D Finke1.
Abstract
A fixed-target approach to high-throughput room-temperature serial synchrotron crystallography with oscillation is described. Patterned silicon chips with microwells provide high crystal-loading density with an extremely high hit rate. The microfocus, undulator-fed beamline at CHESS, which has compound refractive optics and a fast-framing detector, was built and optimized for this experiment. The high-throughput oscillation method described here collects 1-5° of data per crystal at room temperature with fast (10° s-1) oscillation rates and translation times, giving a crystal-data collection rate of 2.5 Hz. Partial datasets collected by the oscillation method at a storage-ring source provide more complete data per crystal than still images, dramatically lowering the total number of crystals needed for a complete dataset suitable for structure solution and refinement - up to two orders of magnitude fewer being required. Thus, this method is particularly well suited to instances where crystal quantities are low. It is demonstrated, through comparison of first and last oscillation images of two systems, that dose and the effects of radiation damage can be minimized through fast rotation and low angular sweeps for each crystal.Entities:
Keywords: X-ray crystallography; fixed-target serial oscillation; oscillations; radiation damage; serial crystallography; storage rings; structural biology; structure determination
Year: 2019 PMID: 30867928 PMCID: PMC6400179 DOI: 10.1107/S2052252519001453
Source DB: PubMed Journal: IUCrJ ISSN: 2052-2525 Impact factor: 4.769
Figure 1Schematic of major beamline components. Shown is the arrangement with oscillation about the Z axis. The beam path is shown in yellow. A, beam-defining slit; B, CRL box with aperture; C, CRL positioning motors; D, air-bearing goniometer; E, three-axis piezo translation stage; F, chip holder; G, EIGER detector.
Figure 2Optical microscopy image of a fixed-target chip containing one grid of 40 × 40 microwells.
Data-collection parameters and refinement statistics
Statistics for the outermost resolution shell are in parentheses.
| FAcD, Chip 1 | FAcD, Chip 2 | Lysozyme | CO-SWMb | |
|---|---|---|---|---|
| Oscillation angle (°) | 1° | 3° | 3° | 5° |
| Oscillation velocity (° s−1) | 10 | 10 | 10 | 10 |
| Wavelength (Å) | 1.216 | 1.216 | 1.216 | 1.216 |
| Absorbed dose (kGy) | 68.2 | 147.9 | 169.3 | 303.1 |
| Dose per degree at 10° s−1 | 40.7 | 40.7 | 46.6 | 54.4 |
| Wells collected | 6400 | 13411 | 4040 | 6400 |
| Datasets with crystals (hit %) | 3293 (48.5%) | 2626 (19.6%) | 1003 (24.8%) | 2710 (42.3%) |
| Indexable datasets | 1554 | 2269 | 510 | 830 |
| Number of datasets used in refinement | 494 | 249 | 95 | 138 |
| Space group |
|
|
|
|
| Unit cell | 41.6, 79.1, 83.8, 90, 103, 90 | 41.6, 79.1, 83.8, 90, 103, 90 | 79.1, 79.1, 38.02, 90, 90, 90 | 37.03, 45.9, 82.91, 90, 90, 90 |
| Resolution range (Å) | 33.43–1.8 (1.864–1.8) | 39.55–1.8 (1.864–1.8) | 39.55–1.839 (1.905–1.839) | 40.16–1.97 (2.04–1.97) |
| Total reflections | 306210 (23803) | 613807 (45550) | 182350 (12878) | 236292 (21259) |
| Unique reflections | 48288 (4597) | 48608 (4680) | 10841 (981) | 10477 (1006) |
| Multiplicity | 6.3 (5.2) | 12.6 (9.7) | 16.8 (13.1) | 22.6 (21.1) |
| Completeness (%) | 98.42 (94.72) | 99.02 (96.37) | 98.87 (90.17) | 99.83 (99.41) |
| Mean | 9.33 (1.60) | 6.65 (0.53) | 10.01 (2.68) | 10.93 (0.73) |
| Wilson | 27.55 | 34.40 | 23.13 | 42.37 |
|
| 0.188 (3.651) | 0.2694 (8.233) | 0.3388 (7.666) | 0.2541 (16.48) |
|
| 0.07252 (1.554) | 0.07347 (2.599) | 0.08115 (2.037) | 0.05271 (3.53) |
| CC1/2 | 0.993 (0.176) | 0.992 (0.106) | 0.991 (0.352) | 0.994 (0.181) |
| CC* | 0.998 (0.548) | 0.998 (0.437) | 0.998 (0.722) | 0.999 (0.554) |
| Reflections used in refinement | 48270 (4592) | 48559 (4676) | 10840 (981) | 10472 (1006) |
| Reflections used for | 2413 (230) | 2422 (236) | 541 (49) | 524 (50) |
|
| 0.1518/0.1812 (0.3591/0.3796) | 0.1631/0.1984 (0.4598/0.5082) | 0.1703/0.2043 (0.2961/0.3054) | 0.1976/0.2731 (0.3943/0.3851) |
| Number of non-H atoms | 4976 | 4972 | 1058 | 1299 |
| Macromolecules | 4765 | 4754 | 1016 | 1225 |
| Ligands | 0 | 0 | 3 | 50 |
| Solvent | 211 | 217 | 39 | 24 |
| Protein residues | 596 | 595 | 129 | 155 |
| RMS (bonds) (Å) | 0.008 | 0.008 | 0.007 | 0.009 |
| RMS (angles) (°) | 1.30 | 1.29 | 1.15 | 1.24 |
| Ramachandran plot | ||||
| Favored (%) | 97.47 | 97.29 | 98.43 | 95.39 |
| Allowed (%) | 2.53 | 2.54 | 1.57 | 4.61 |
| Outliers (%) | 0.00 | 0.17 | 0 | 0.00 |
| Rotamer outliers (%) | 0.42 | 0.63 | 0 | 0.00 |
| Clashscore | 4.90 | 7.16 | 2.0 | 7.43 |
| Average | 34.42 | 40.26 | 26.16 | 50.67 |
| Macromolecules | 34.14 | 40.00 | 25.97 | 50.66 |
| Ligands | — | — | 29.87 | 48.61 |
| Solvent | 40.63 | 45.82 | 31.01 | 55.90 |
| Number of TLS groups | 15 | 17 | 9 | 4 |
| PDB codes | 6muh | 6muy | 6muz | 6mvo |
This value includes the initial 0.5° oscillation that precedes data collection.
‘Hit’ is defined as a dataset that had sufficient spots for XDS to attempt indexing.
Figure 3Composite OMIT maps (2mF o−DF c) of selected structures. Maps were calculated using PHENIX (Adams et al., 2010 ▸). Maps are contoured at 1σ. (a) FAcD from Chip 1, pictured here around the active site with the residues responsible for catalysis labeled and shown in magenta. (b) Carboxymyoglobin, pictured here around the heme moiety. (c) Lysozyme, around the active site.
Figure 4Superimposed structures of CO-SWMb data collected in the first 2° (green) and last 2° (blue), centered around the heme atom, detailing structural variations between the two models.