Ouided Rouabhi1, Brandie Gross1, John Bayouth1,2, Junyi Xia1. 1. 1 Department of Radiation Oncology, University of Iowa, Iowa, IA, USA. 2. 2 Department of Human Oncology, University of Wisconsin-Madison, Madison, WI, USA.
Abstract
PURPOSE: To evaluate the dosimetric and temporal effects of high-dose-rate respiratory-gated radiation therapy in patients with lung cancer. METHODS: Treatment plans from 5 patients with lung cancer (3 nongated and 2 gated at 80EX-80IN) were retrospectively evaluated. Prescription dose for these patients varied from 8 to 18 Gy/fraction with 3 to 5 treatment fractions. Using the same treatment planning criteria, 4 new treatment plans, corresponding to 4 gating windows (20EX-20IN, 40EX-40IN, 60EX-60IN, and 80EX-80IN), were generated for each patient. Mean tumor dose, mean lung dose, and lung V20 were used to assess the dosimetric effects. A MATLAB algorithm was developed to compute treatment time. RESULTS: Mean lung dose and lung V20 were on average reduced between -16.1% to -6.0% and -20.0% to -7.2%, respectively, for gated plans when compared to the corresponding nongated plans, and between -5.8% to -4.2% and -7.0% to -5.4%, respectively, for plans with smaller gating windows when compared to the corresponding plans gated at 80EX-80IN. Treatment delivery times of gated plans using high-dose rate were reduced on average between -19.7% (-0.10 min/100 MU) and -27.2% (-0.13 min/100 MU) for original nongated plans and -15.6% (-0.15 min/100 MU) and -20.3% (-0.19 min/100 MU) for original 80EX-80IN-gated plans. CONCLUSION: Respiratory-gated radiation therapy in patients with lung cancer can reduce lung dose while maintaining tumor dose. Because treatment delivery during gated therapy is discontinuous, total treatment time may be prolonged. However, this increase in treatment time can be offset by increasing the dose delivery rate. Estimation of treatment time may be helpful in selecting patients for respiratory gating and choosing appropriate gating windows.
PURPOSE: To evaluate the dosimetric and temporal effects of high-dose-rate respiratory-gated radiation therapy in patients with lung cancer. METHODS: Treatment plans from 5 patients with lung cancer (3 nongated and 2 gated at 80EX-80IN) were retrospectively evaluated. Prescription dose for these patients varied from 8 to 18 Gy/fraction with 3 to 5 treatment fractions. Using the same treatment planning criteria, 4 new treatment plans, corresponding to 4 gating windows (20EX-20IN, 40EX-40IN, 60EX-60IN, and 80EX-80IN), were generated for each patient. Mean tumor dose, mean lung dose, and lung V20 were used to assess the dosimetric effects. A MATLAB algorithm was developed to compute treatment time. RESULTS: Mean lung dose and lung V20 were on average reduced between -16.1% to -6.0% and -20.0% to -7.2%, respectively, for gated plans when compared to the corresponding nongated plans, and between -5.8% to -4.2% and -7.0% to -5.4%, respectively, for plans with smaller gating windows when compared to the corresponding plans gated at 80EX-80IN. Treatment delivery times of gated plans using high-dose rate were reduced on average between -19.7% (-0.10 min/100 MU) and -27.2% (-0.13 min/100 MU) for original nongated plans and -15.6% (-0.15 min/100 MU) and -20.3% (-0.19 min/100 MU) for original 80EX-80IN-gated plans. CONCLUSION: Respiratory-gated radiation therapy in patients with lung cancer can reduce lung dose while maintaining tumor dose. Because treatment delivery during gated therapy is discontinuous, total treatment time may be prolonged. However, this increase in treatment time can be offset by increasing the dose delivery rate. Estimation of treatment time may be helpful in selecting patients for respiratory gating and choosing appropriate gating windows.
Entities:
Keywords:
4D dose; 4DCT; gated radiation therapy; gating; treatment time
When treating lung tumors with radiation therapy, respiration-induced tumor motion can
yield large uncertainties in target delineation and localization.[1-6] In an effort to characterize and reduce the impact of this motion, various techniques
have been presented, including respiratory-gated techniques, breath-hold techniques, forced
shallow breathing methods, and real-time tumor tracking methods.[7-9]Respiratory-gated radiation therapy (RGRT) involves monitoring the respiratory signal and
only delivering radiation when the patient is within a specific window of the respiratory
cycle, thereby reducing the size of the target volume and increasing the amount of normal
tissue spared. Several recent studies using 4-dimensional computed tomography (4DCT) have
shown that the use of patient-specific treatment margins and respiratory gating, rather than
standard population-based treatment margins, can reduce normal tissue toxicity.[8-12] However, some have pointed out that the additional dosimetric benefits from
respiratory gating may be modest and not justify the challenges of RGRT, including the
increased time for treatment delivery and, consequently, the increased potential for
uncertainty introduced from patient movement.[11,13] While increased treatment delivery time may not be of great concern for cases which
only require a few minutes to complete, as may be true for some treatment plans using
volumetric modulated arc therapy (VMAT), cases requiring longer treatment times, such as
those using step-and-shoot intensity-modulated radiotherapy (IMRT), may benefit from the
ability to estimate treatment time when assessing potential gating windows. Currently, no
established guidelines concerning the use of respiratory gating have been defined.[8 , 13]Treatment delivery during gated therapy is not continuous, and as a result, total treatment
time is prolonged due to the decreased duty cycle (ie, percentage of the beam-on time).
Although various gating methods have been suggested, a limited number of publications were
found to report on treatment times for gated radiation therapy.[14-16] One common method for reducing gated treatment times involves the use of breath hold
during treatment delivery. Berson et al
[14] compared gated therapy using breath-hold coaching versus free breathing without
coaching and found that treatment times decreased from 4.0 minutes/100 MU to 2.0 minutes/100
MU. Increasing the delivery dose rate has also been suggested for reducing gated treatment
times. Preliminary results from Linthout et al
[15] showed that increasing dose rate from 480 to 800 MU/min decreased treatment delivery
time from 0.9 min/100 MU to 0.4 min/100 MU. Likewise, Willoughby et al
[16] found that increasing dose rate from 480 to 800 MU/min reduced gated treatment times
by 40% when using 20% and 40% gating windows while maintaining an output consistency within
0.5%.In this study, we investigated the dosimetric and temporal effects of RGRT by evaluating
the dosimetric impacts of 4 gating windows and developing a novel algorithm for estimating
the corresponding treatment delivery times when using dose rates of 500 and 1500 MU/min.
Methods
Following institutional review board approval (#200905703), 5 patients with lung cancer
treated at our institution were included in this retrospective study. All of the patients
had undergone 4DCT scans. A summary of patient characteristics is listed in Table 1. Among the patients, maximum
tumor motion varied from 6 to 12 mm. Two patients were treated with respiratory gating using
an 80EX-80IN gating window (Table
2), while 3 patients were treated without respiratory gating. Of the patients, 1
was treated with IMRT, while the remaining 4 were treated using stereotactic body radiation
therapy (SBRT). Table 3
summarizes the planning methods and prescriptions for the SBRT patients. The prescription
dose varied from 8 to 18 Gy/fraction with treatment fractions varied from 3 to 5 fractions.
Both static beams and IMRT were used in the SBRT planning.
Summary of Planning Methods and Prescriptions for SBRT Patients.
Patient
Prescription
Static Beam/IMRT
Couch Angles
# of Beams
P1
10 Gy × 5
Static
0°
6
P6
10 Gy × 4
IMRT
0°
11
P7
18 Gy × 3
Static
0°, 15°, 345°
16
P9
8 Gy × 5
Static
0°, 15°, 345°
16
Abbreviations: IMRT, intensity-modulated radiation therapy; SBRT, stereotactic body
radiation therapy.
Summary of Patient Characteristics.Abbreviations: IMRT, intensity-modulated radiation therapy; PTV, planning target
volume; SBRT, stereotactic body radiation therapy.Gating Windows.Summary of Planning Methods and Prescriptions for SBRT Patients.Abbreviations: IMRT, intensity-modulated radiation therapy; SBRT, stereotactic body
radiation therapy.
Image Acquisition
Patient computed tomography (CT) images were acquired using the Siemens Biograph positron
emission tomography CT scanner (Siemens Medical System, Knoxville, Tennessee). For each
patient, a breath-hold CT scan at the end of exhale was first taken, followed by a
free-breathing 4DCT scan, during which the patient’s respiratory motion was recorded using
a commercially available strain gauge pressure sensing system (Anzai medical Co Ltd,
Tokyo, Japan) fixed to the upper abdominal region using an elastic belt. Retrospective
sorting of the 4DCT projections was performed using the CT console. Amplitude-based
binning was used for image reconstruction. Each reconstructed CT image corresponded to one
of 10 respiratory amplitudes: 0EX, 20EX, 40EX, 60EX, 80EX, 100IN, 80IN, 60IN, 40IN, and
20IN. The name of each respiratory phase identifies the percentage of full inhalation
reached, as well as whether the phase occurs during inhalation or exhalation (eg, the 20IN
phase occurs when the patient is inhaling and reaches 20% of full inhalation). A sample
respiratory cycle is depicted in Figure
1.
Figure 1.
Sample respiratory cycle.
Sample respiratory cycle.
Deriving Target Volumes
For each patient, the breath-hold CT scan was used to generate the gross tumor volume
(GTV). Contouring was performed using the Pinnacle3 treatment planning system (TPS;
Philips Radiation Oncology Systems, Milpitas, California). This GTV was automatically
copied to each 4DCT data set using Velocity AI deformable image registration software
(Velocity Medical Systems, Atlanta, Georgia). The GTVs for each phase CT were verified to
ensure conformity with the corresponding 4DCT image. Four internal target volumes (ITVs),
corresponding to 4 gating windows (Table 2), were then created by performing a union of the GTVs for the phases
included within the gating window.Planning target volumes (PTVs) were derived from ITVs by adding 5 mm margins, accounting
for uncertainty in patient setup. This resulted in 4 gated treatment plans (20EX-20IN,
40EX-40IN, 60EX-60IN, and 80EX-80IN) for each patient, generated to satisfy the same
planning criteria as the original patient treatment plan.
Dosimetric Evaluation
Mean tumor dose (MTD), mean lung dose (MLD), and lung V20 (percentage of the lung volume
receiving at least a 20 Gy dose) were used for dosimetric evaluation. The Pinnacle3 TPS
was used to generate and analyze the dose–volume histogram. Percentage and absolute dose
differences were computed between each gated treatment and the original treatment
plan.
Calculation of Treatment Time
A MATLAB (MathWorks, Natick, Massachusetts) algorithm was developed to compute the
treatment time for each plan, including time for gantry rotation, time for collimator
leaves and jaws motion, time to deliver dose, and time for communication overhead. For
gated treatment plans, dose delivery time was scaled relative to the time spent within the
gating window based on the patient’s respiratory trace. The treatment time calculation
algorithm was validated using the actual patient treatment time. Treatment times were
first compared using dose rates of 500 MU/min for both gated and nongated treatment plans
and then compared again using an increased dose rate of 1500 MU/min for gated plans.
Treatment time calculation
The MATLAB algorithm computed treatment delivery time by utilizing the treatment plan
exported from the Pinnacle3 TPS (see Figure 2). Equation 1 describes the calculation for total
treatment time (t
total), where beam is the total number of beams, t
overhead is the time of communication overhead, t
gantry is the gantry rotation time, CP is the total number
of control points, t
mech is the mechanical time for the collimator leaves and jaws, and
t
dose is the time to deliver dose.
Figure 2.
Workflow for treatment time calculation. The MATLAB algorithm reads a “.Trial” file
exported from the Pinnacle3 TPS and calculates the treatment delivery time
(t
total), including time for gantry rotation (t
gantry), time to position collimator leaves and jaws (t
mech), time to deliver dose (t
dose), and time for communication overhead (t
overhead).
Workflow for treatment time calculation. The MATLAB algorithm reads a “.Trial” file
exported from the Pinnacle3 TPS and calculates the treatment delivery time
(t
total), including time for gantry rotation (t
gantry), time to position collimator leaves and jaws (t
mech), time to deliver dose (t
dose), and time for communication overhead (t
overhead).Time of communication overhead (t
overhead), accounting for machine communication time, was 3 seconds for each
beam based on observation. Gantry rotation time (t
gantry), as seen in Equation 2, was computed by calculating the time to
travel between the current gantry angle (θ) and previous (θ) gantry angle, and the gantry rotation speed (s) was set to 3°/sec based on observation of the gantry rotation.Mechanical time (t
mech), calculated Equation 3, accounts for the time to move the jaws
and collimator leaves at each control point. First, an optimal sequence for the control
points, which provided the shortest time, was determined by finding the sequence
requiring the least movement between control points. Next, for each control point, the
distance moved by each leaf (y) and jaw (j), based on
current (i) and previous (i − 1) position, was
computed and divided by constants sleaf and sjaw for the leaves
and jaws, respectively. For each control point, the maximum amount of time taken to move
all leaves and jaws was defined as the mechanical time. In order to determine
sleaf and sjaw, the treatment time algorithm calculated the
treatment time for each patient using varying leaf and jaw speed combinations ranging
from 0.1 up to 4.0 cm/s and compared to the actual treatment time obtained from the
treatment record. Based on our calculation, sleaf and sjaw were
set to 2.0 and 1.0 cm/s accordingly.As shown in Equation 4, time to deliver dose (t
dose) was also calculated for each control point of the beam. Number of
monitor units for each control point (MUCP) was first calculated by
multiplying the total prescribed monitor units (MUtotal) by the beam
weighting (w
beam) and the control point weighting (w
cp). Time to deliver dose was then calculated by dividing
MU by the dose rate (DR) and gating factor (GF). Gating factor was derived from
the patient’s respiratory trace and defined as the fraction of the respiratory signal
contained within the gating window (RespGW) versus the total respiratory
signal (Resptotal).
Validation of treatment time algorithm
To validate the treatment time algorithm, actual treatment times for 5 enrolled
patients were obtained from the MOSAIQ treatment record and verify system (Elekta,
Stockholm, Sweden) and compared against computed treatment time. Because patients
received multiple fractions, treatment time for each fraction could vary depending on
the consistency of breathing, and whether the machine encountered problems during
delivery, the actual treatment time was defined as the average treatment time across all
fractions. Any obvious outliers were excluded during calculation of the average
treatment time.
Results
Results were divided into 2 groups: (1) patients whose original treatment plan was nongated
and (2) patients whose original treatment plan was gated using an 80EX-80IN gating window
(refer to Table 2). In the first
group, the evaluated 20EX-20IN to 80EX-80IN treatment plans were compared against the
original nongated plan. In the second group, the evaluated 20EX-20IN to 60EX-60IN treatment
plans were compared against the original 80EX-80IN gated plan.
Dosimetric Evaluation for Nongated Patients
The average reduction in PTV volume was −26.9% ± 4.4%, −21.8% ± 4.2%, −15.4% ± 2.6%, and
−9.4% ± 6.0%, respectively, for the 20EX-20IN, 40EX-40IN, 60EX-60IN, and 80EX-80IN gated
plans when compared to the nongated plan (Figure 3). The PTV coverage for all gated plans was kept within 1% of the
original PTV coverage, and the MTD was kept within <1Gy of the original MTD. The
percentage of difference in MTD was <2% across plans for all patients (Figure 4). As shown in Figures 5 and 6, MLD and lung V20 were found to decrease as we
reduced the gating window. On average, relative percentage differences in MLD and lung V20
were reduced by −16.1% ± 1.0% and −20.0% ± 2.3% for 20EX-20IN-gated plans, −12.5% ± 1.0%
and −15.6% ± 2.3% for 40EX-40IN-gated plans, −8.7% ± 3.5% and −11.1% ± 4.2% for
60EX-60IN-gated plans, and −6.0% ± 4.7% and −7.2% ± 5.7% for 80EX-80IN−gated plans when
compared to the nongated plan.
Figure 3.
Percentage difference in planning target volume (PTV) volume.
Figure 4.
Percentage difference in mean tumor dose (MTD).
Figure 5.
Percentage difference in mean lung dose (MLD).
Figure 6.
Percentage difference in lung V20.
Percentage difference in planning target volume (PTV) volume.Percentage difference in mean tumor dose (MTD).Percentage difference in mean lung dose (MLD).Percentage difference in lung V20.
Dosimetric Evaluation for 80EX-80IN-Gated Patients
For patients originally gated at 80EX-80IN, the average reduction in PTV volume was
−29.1% ± 7.7% for the 20EX-20IN-gated plans, −26.9% ± 7.8% for the 40EX-40IN-gated plans,
and −21.3% ± 5.2% for the 60EX-60IN-gated plans when compared to the 80EX-80IN-gated plan
(Figure 3). The PTV coverage for
all gated plans was kept within 1% of the original PTV coverage, and the MTD was kept
within <1 Gy of the original MTD. The percentage of difference in MTD was ≤1% across
plans for all patients (Figure 4).
Figures 5 and 6 show the relative percentage of
differences in MLD and lung V20 across plans for each patient. Both MLD and lung V20 were
on average found to be reduced −5.8% ± 1.4% and −7.0% ± 4.3% for the 20EX-20IN-gated
plans, −4.7% ± 0.0% and −6.0% ± 2.9% for the 40EX-40IN-gated plans, 4.2% ± 0.3% and −5.4%
± 2.1% for the 60EX-60IN-gated plans.
Validation of Treatment Time Algorithm
Table 4 compares the results
of the treatment time calculation algorithm with the actual treatment time obtained from
the treatment records. On average, the algorithm was able to estimate treatment time
within −0.8 minutes (−7.0%) of the actual treatment time. In 4 of 5 patients, the
algorithm was able to predict treatment time within <1 minute of the actual time.
Table 4.
Validation of Treatment Time Algorithm.
Patient
Actual Treatment Time, minutes
Computed Treatment Time, minutes
Time Difference, minutes
Percentage Difference
1
7.6
8.0
0.4
5.3
6
5.5
5.3
−0.2
−3.6
7
11.7
8.0
−3.7
−31.6
8
5.0
5.2
0.2
4.0
9
9.0
8.2
−0.8
−8.9
Average
−0.8
−7.0
Standard deviation
1.7
14.9
Validation of Treatment Time Algorithm.
Treatment Time Comparison for Nongated Patients
The left-hand side of Figure 7
depicts the percentage difference in treatment times between the nongated plans and their
corresponding 20EX-20IN- to 80EX-80IN-gated plans. When using a dose rate of 500 MU/min
for both the nongated and gated plans, treatment delivery times were on average found to
increase 29.0% ± 21.3% (0.13 ± 0.07 min/100 MU), 18.5% ± 15.0% (0.08 ± 0.05 min/100 MU),
10.2% ± 11.0% (0.04 ± 0.04 min/100 MU), and 4.9% ± 7.1% (0.02 ± 0.03 min/100 MU),
respectively.
Figure 7.
Percentage treatment time increase for gated delivery when using a dose rate of 500
MU/min for original and gated plans.
Percentage treatment time increase for gated delivery when using a dose rate of 500
MU/min for original and gated plans.Treatment times were next compared using an increased dose rate for the gated plans. dose
rate for the nongated plans was kept at 500 MU/min, while dose rate for the 20EX-20IN- to
80EX-80IN-gated plans was raised to 1500 MU/min. Figure 8 outlines these results. By increasing the
dose rate for the gated plans, treatment times were found to be reduced compared to the
original nongated plans, with larger gating windows requiring less time than smaller
windows. On average, treatment delivery times were found to decrease by −19.7% ± 7.0%
(−0.10 ± 0.03 min/100 MU), −22.7% ± 5.5% (−0.11 ± 0.02 min/100 MU), −25.5% ± 6.0% (−0.13 ±
0.02 min/100 MU), and −27.2% ± 6.6% (−0.13 ± 0.02 min/100 MU), respectively.
Figure 8.
Percentage difference in time for gated delivery when using a dose rate of 500 MU/min
for original plans and 1500 MU/min for gated plans.
Percentage difference in time for gated delivery when using a dose rate of 500 MU/min
for original plans and 1500 MU/min for gated plans.
Treatment Time Comparison for 80EX-80IN-Gated Patients
The right-hand side of Figure 7
depicts the percentage difference in treatment time between treatment plans originally
gated 80EX-80IN and their corresponding 20EX-20IN- to 60EX-60IN-gated plans. When using a
dose rate of 500 MU/min for both the original and new gated plans, treatment delivery
times were on average found to increase by 16.9% ± 23.4% (0.08 ± 0.10 min/100 MU), 4.9% ±
12.8% (−0.01 ± 0.10 min/100 MU), and 4.6% ± 0.9% (0.04 ± 0.03 min/100 MU), respectively.
Patient 8 did not exhibit the same pattern in treatment time as seen in the remaining
patients (smaller gating windows required more time) but rather had relatively similar
treatment times across the 4 gated plans.Treatment times were next compared using an increased dose rate of 1500 MU/min for the
20EX-20IN- to 60EX-60IN-gated plans while keeping the original 80EX-80IN-gated plan at 500
MU/min. Figure 8 outlines these
results. By increasing the dose rate for the gated plans, treatment times were on average
found to be reduced by −15.6% ± 5.2% (−0.15% ± 0.09 min/100 MU), −20.3% ± 7.9% (−0.19 ±
0.11 min/100 MU), and −17.7% ± 15.5% (−0.12 ± 0.01 min/100 MU), respectively. In the case
of patient 8, the 60EX-60IN-gated plan took longer to deliver than the 20EX-20IN- and
40EX-40IN-gated plans.
Discussion
As has been demonstrated in previous RGRT studies,[8,9,11,12] the PTV for all patients could be reduced, indicating a smaller total volume received
the target dose. The PTV coverage and MTD were able to be maintained, suggesting that the
target was just as effectively able to be treated. Furthermore, MLD and lung V20 were shown
to decrease when smaller gating windows were selected. For the smallest gating window,
20EX-20IN, MLD was on average reduced by −16.1% ± 1.0% in originally nongated plans and by
−5.8% ± 1.4% in plans originally gated 80EX-80IN, an absolute difference of −87.9 ± 19.9 cGy
and −46.7 ± 48.8 cGy, respectively. Likewise, lung V20 was on average reduced by −20.0% ±
2.3% in originally nongated plans and by −6.9% ± 4.3% in plans originally gated 80EX-80IN
and an absolute difference of −1.5% ± 0.1% and −1.0% ± 1.2%, respectively. Some authors[11,13] have pointed out that the direct links between the dosimetric benefits of respiratory
gating and successful patient outcome are unknown. However, MLD and lung V20 have been used
clinically as predictive factors for radiation pneumonitis.[17,18] Although no studies have directly evaluated the effects of respiratory gating on
radiation pneumonitis, results of these dosimetric studies imply that respiratory gating may
reduce the risk of radiation pneumonitis. When selecting patients for respiratory gating and
choosing appropriate gating windows, the dosimetric benefits of gated treatment should be
weighed along with the clinical risk for radiation pneumonitis.As shown in Figures 7 and 8, smaller gating windows required
longer delivery times when dose rate was unchanged. This was expected due to the fact that
dose delivery for gated treatment is intermittent rather than continuous. However, time
increase was not found to be linearly proportional to the number of respiratory phases
within the gating window. This is likely because patients spent more time in the exhalation
phase of respiration. For example, the 20EX-20IN gating window includes 3 respiratory phases
(20EX, 0EX, and 20IN), while the 80EX-80IN gating window includes 9 respiratory phases
(80EX, 60EX, 40EX, 20EX, 0EX, 20IN, 40IN, 60IN, and 80IN), although, as we see Figure 9, patients did not spend 3 times
longer in the 80EX-80IN window as they did in the 20EX-20IN window but rather spent more
than half of the breathing cycle within the 20EX-20IN window. The 20EX-20IN window was
estimated to have the greatest increase in delivery time: 29.0% ± 21.3% on average,
corresponding to an absolute difference of 0.13 ± 0.07 min/100 MU, while the 80EX-80IN
gating window was on average only found to increase time by 4.9% ±7.1%, an absolute
difference of 0.02 ± 0.03 min/100 MU. However, treatment delivery times for gated treatment
plans may be reduced by increasing the dose rate. As can be seen in Figure 8, raising the dose rate from 500 to 1500 MU/min
was shown to decrease delivery time to even less than the time taken for nongated treatment
at 500 MU/min. Patient 8 did not exhibit the same patterns in treatment time as seen in the
remaining patients (smaller gating windows required more time) and instead had relatively
similar treatment times across the 4 gated plans. This is likely due to this patient
spending a majority of the breathing cycle in the full exhalation phase (0EX) of
respiration, thereby giving similar gating factor for the 20EX-20IN (60%), 40EX-40IN (70%),
60EX-60IN (78%), and 80EX-80IN (89%) gating windows. Further investigation also indicates
that for this patient, the mechanical time took longer for the 60EX-60IN (87 seconds) and
80EX-80IN (82 seconds) gating windows than for the 20EX-20IN (69 seconds) and 40EX-40IN (63
seconds) gating windows. The combination of similar gating factor and shorter mechanical
times for the smaller gating windows allowed for comparable treatment times across the
20EX-20IN- to 80EX-80IN-gated plans.
Figure 9.
Temporal probability for the 20EX-20IN to 80EX-80IN gating windows. These gating
factors were used to scale delivery time for the gated treatments.
Temporal probability for the 20EX-20IN to 80EX-80IN gating windows. These gating
factors were used to scale delivery time for the gated treatments.Treatment delivery time is an important consideration when selecting an optimal treatment
plan. Prolonged treatment delivery times can affect patient comfort and compliance. Advances
in radiotherapy delivery techniques, such as VMAT, have led to the potential for reduction
in treatment delivery times. Arc therapy allows for continuous delivery of radiation,
thereby increasing the delivery efficiency. In contrast, techniques such as “step-and-shoot”
IMRT require increased time for therapy because delivery is not continuous. Particularly for
cases that require longer treatment delivery times, having the ability to estimate delivery
time through this algorithm may aid in selecting patients for respiratory gating and
choosing appropriate gating windows. When presented with 2 treatment plans with similar dose
distributions, treatment time information may be useful in selecting the optimal plan.
Additionally, the temporal effects of factors such as dose rate may also be assessed and
used to aid in planning. The biggest limitation to this study was the small patient sample
size (n = 5), which was made smaller by separating patients who had been treated using
80EX-80IN-gated treatment (n = 2) from those who had undergone nongated treatment (n = 3).
The next step in this research will be to repeat this study using a larger patient sample
size and perform a statistical analysis. An increased sample size will allow us to further
validate the treatment time calculation algorithm as well as determine factors that have the
largest impact on delivery time. Additionally, we will be able to investigate which
parameters, such as maximum tumor motion or size of GTV, are most useful in predicting the
potential benefits of respiratory gating.
Conclusion
Respiratory-gated radiation therapy in patients with lung cancer can reduce lung dose while
maintaining tumor dose. Because treatment delivery during gated therapy is discontinuous,
total treatment time may be prolonged. However, this increase in treatment time can be
offset by increasing the dose delivery rate. Estimation of treatment time may be helpful in
selecting patients for respiratory gating and choosing appropriate gating windows.
Authors: R Bryan Barriger; Jeffrey A Forquer; Jeffrey G Brabham; David L Andolino; Ronald H Shapiro; Mark A Henderson; Peter A S Johnstone; Achilles J Fakiris Journal: Int J Radiat Oncol Biol Phys Date: 2010-10-29 Impact factor: 7.038
Authors: Eike Rietzel; Arthur K Liu; Karen P Doppke; John A Wolfgang; Aileen B Chen; George T Y Chen; Noah C Choi Journal: Int J Radiat Oncol Biol Phys Date: 2006-09-01 Impact factor: 7.038
Authors: René W M Underberg; Frank J Lagerwaard; Ben J Slotman; Johan P Cuijpers; Suresh Senan Journal: Int J Radiat Oncol Biol Phys Date: 2005-06-01 Impact factor: 7.038
Authors: Jaehyeon Park; Ji Woon Yea; Se An Oh; Jongmoo Park; Jae Won Park; Jeong Eun Lee Journal: Cancers (Basel) Date: 2022-09-02 Impact factor: 6.575