Literature DB >> 30728197

Separation of Plasma from Whole Blood by Use of the cobas Plasma Separation Card: a Compelling Alternative to Dried Blood Spots for Quantification of HIV-1 Viral Load.

Sergio Carmona1,2, Britta Seiverth3, Dieketseng Magubane2, Lucia Hans4,2, Matthias Hoppler3.   

Abstract

Plasma HIV viral load testing is the preferred means of monitoring antiretroviral treatment response. Dried blood spots (DBSs) hold considerable logistical advantages over EDTA samples, but they more frequently misclassify virological failure and have higher limits of detection (LoD). Plasma separation cards (PSCs) may overcome these limitations. Health workers collected EDTA whole blood by venipuncture and 140 μl of finger-prick blood by capillary tube from 53 HIV-infected adults. Capillary blood was immediately transferred to PSCs. Additionally, 432 EDTA samples from HIV-infected adults were spotted onto PSCs and analyzed together with the finger-prick samples. Specificity and sensitivity of PSC with paired EDTA-PSC samples tested on a cobas 6800/8800 system with the cobas HIV-1 test (cobas HIV) was determined. LoD (3rd HIV-1 WHO International Standard) and stability at a range of temperatures and storage durations was determined using cobas HIV and cobas AmpliPrep/cobas TaqMan HIV-1 test v2.0 (CAP/CTM). Of 132 specimens with quantitative values for paired EDTA-PSC samples, the mean log10 difference between samples was 0.05 copies/ml (95% confidence interval [CI], -0.01 to 0.11). The LoD for cobas HIV was 790.2 copies/ml and for CAP/CTM was 737.9 copies/ml. At 1,000 copies/ml, PSC sensitivity was 97.0% (128/132) and specificity was 97.2% (343/353). Results correlated well with those from EDTA samples (Deming R 2 = 0.90). PSC results were unaffected by temperature and storage conditions. PSC samples correlate well with plasma viral load and have adequate sensitivity and specificity. The improved performance may be as a result of a reduction in contribution from cell-associated viral nucleic acids. The card provides an alternative sample collection technology to DBSs.
Copyright © 2019 Carmona et al.

Entities:  

Keywords:  HIV; HIV patient monitoring; cobas; dried blood spots; filter paper cards; plasma; plasma separation card; viral load

Mesh:

Substances:

Year:  2019        PMID: 30728197      PMCID: PMC6440768          DOI: 10.1128/JCM.01336-18

Source DB:  PubMed          Journal:  J Clin Microbiol        ISSN: 0095-1137            Impact factor:   5.948


INTRODUCTION

Presently, of the estimated 36.7 million people infected with HIV worldwide, about 20 million are receiving antiretroviral treatment (ART) (1). Those who are not yet receiving ART are eligible for treatment regardless of their CD4 cell count or clinical status, and concerted efforts are under way to ensure these patients are identified and commence treatment. Once treatment begins, patients require careful monitoring of their responses. Viral load is the most reliable monitoring tool and is preferred over clinical assessment or immunological monitoring through CD4 enumeration (2). The WHO recommends viral load testing biannually in the first year of treatment and annually thereafter (3). Although viral load testing is increasingly becoming available, only about half of patients worldwide who are receiving ART have access to these tests (1). The timely detection of treatment failure allows health workers to provide additional adherence support for patients or to make changes in treatment regimen (4, 5). In the absence of viral load monitoring, treatment failure is often missed, resulting in HIV disease progression, the accumulation of HIV drug resistant strains, and higher risks of further HIV transmission (6–8). In addition, inadequate access to viral load testing makes it difficult to monitor the overall impact of ART programs, including measurement of progress toward the “third 90”—the goal that 90% of people on ART achieve viral suppression (9). While plasma is the preferred sample type, viral load measurement from plasma is currently challenging in many settings. Challenges are especially marked in rural areas due to the logistics of transporting plasma samples to centralized testing facilities and the need for sophisticated laboratory infrastructure, qualified personnel, and stringent storage requirements to maintain sample integrity (10). One potential alternative to plasma samples has been the spotting of whole blood from a finger prick onto filter paper that is allowed to dry (11–13). This “dried blood spot” (DBS) facilitates the decentralization of specimen collection while maintaining high throughput at centralized laboratories where blood is eluted off the card for testing (3). The DBS is easy to collect, does not require phlebotomy, minimizes infection risks, and can stabilize biological material at ambient temperature during transport and storage. There are a range of applications of DBSs within the HIV field, including testing for viral resistance and antiretroviral drug levels (14–16), diagnosis of pediatric HIV infection, self-collection of samples for HIV self-testing (17), and population surveillance, including for HIV incidence and for measuring the impact of HIV treatment programs (18–20). Outside of the HIV field, the DBS sample type is well known for enabling the measurement of glucose (21), screening newborns for inherited metabolic disorders, and recent large population-based surveys of infectious diseases, such as hepatitis B and C virus (22). Although DBSs present many compelling advantages and are included in WHO recommendations for measuring HIV-1 viral loads (3), they have several limitations (3). Most notably, DBS samples can yield higher HIV viral load measurements than plasma-based methods because proviral DNA and intracellular RNA found in whole blood contribute to the HIV molecular signal (23). Viral load testing using DBS may thus misclassify patients as having failed therapy, necessitating additional laboratory testing and even unnecessary switching of treatment regimens. At the same time, there is a concern that DBS samples containing only 50 to 100 μl of whole blood may be suboptimal for adequate representation of the viral population, with inadequate lower limits of viral detection and false-negative results (5, 24). Innovations in sample collection matrices that overcome the limitations of DBSs could help address the large need for accurate viral load measures (25). This study examined the performance of a novel plasma collection device, the cobas plasma separation card (PSC), which shares many features of the DBS but which collects plasma rather than whole blood. Using paired EDTA plasma and PSC samples, we examined the card’s bias and misclassification at a viral load threshold of 1,000 copies/ml, as well as its usability, limits of detection (LoD), and stability. A successful PSC approach would retain the sample collection and transport advantages of the DBS but have improved sensitivity, specificity, and reliability.

MATERIALS AND METHODS

The sensitivity component of the study was performed in the Development Laboratory at Roche Molecular Diagnostics International, Rotkreuz, Switzerland, from July to August 2017. All other study components were done in Johannesburg, South Africa, from May to September 2017. The Human Research Ethics Committee (Medical) of the University of the Witwatersrand (no. M160617) and the District Research Committee of Johannesburg approved the study. The health workers and laboratory technicians in the usability assessment gave written informed consent, as did patients who provided blood samples. Test results were not used for patient care. All analyses were done using SAS JMP v12.

Description of the plasma separation card.

The cobas plasma separation Card is a sampling device that is spotted with whole blood samples obtained from finger pricks or venous blood (140 μl blood per spot). A porous membrane allows only plasma to pass and to be collected on an underlying polyester fleece. The fleece is impregnated with an RNA-stabilizing reagent, which stabilizes the samples over a range of temperatures, humidity levels, and storage conditions. A defined area of the fleece (“the spot”) is then removed and subjected to an elution step before the obtained sample is loaded on either the cobas 6800/8800 system using the cobas HIV-1 test (cobas HIV) or the cobas AmpliPrep/cobas TaqMan HIV-1 test, v2.0 (CAP/CTM).

Usability assessment.

Participants were recruited from three settings for the usability assessment, conducted per international guidelines (26, 27). First, six phlebotomists collected PSC finger-prick samples at an outpatient clinic of Charlotte Maxeke Johannesburg Academic Hospital (CMJAH), Johannesburg, South Africa. Thereafter, the PSC and its workflow were presented to twelve nurses and one clinic manager drawn from four primary care clinics. Finally, five technologists and medical scientists at the National Health Laboratory Service (NHLS) at CMJAH assessed the preparation and processing of the PSC prior to testing with cobas HIV. In all assessments, participants completed a Likert-type questionnaire covering usability statements for the PSC and its associated workflows.

PSC performance assessment: sample collection and preparation.

A study nurse collected finger-prick blood samples using a safety lancet (MiniCollect safety lancet, with a penetration depth of 2.0 mm; Greiner Bio-One) from HIV-infected adults attending the HIV clinic at CMJAH (n = 53). The nurse collected blood using an EDTA-coated capillary tube positioned directly beneath the puncture site, and 140 μl of blood was transferred onto each of three delineated areas of the card. Venous whole blood samples were then collected by venipuncture using EDTA tubes. We also prepared PSCs using leftover whole blood samples from HIV-infected patients who had recently had a CD4 cell count (n = 185) or viral load test (n = 241). At the NHLS laboratory, the PSCs were placed in a single-use drying rack for a minimum of 4 h at room temperature. PSCs were then packed in separate gas-impermeable zip lock bags with a desiccant and stored at room temperature. In a laminar flow hood, the top layer of the PSC (the “spotting layer”), together with the membrane, was manually removed by pulling on a removable flap on the PSC and then discarded (Fig. 1). A single spot was removed with tweezers and transferred into a cryogenic tube (Cryo.sTM, 5 ml, PP, round bottom; Greiner Bio-One), to which 1,300 μl of cobas specimen preextraction reagent was added. Tubes were placed in a preheated thermomixer (Eppendorf 5355 Thermomixer R with Thermoblock for 24 cryotubes) with 1.5 ml Thermoblock and incubated for 10 min at 56°C and at 1,000 rpm. Thereafter, the tubes were uncapped and loaded into the system for testing. Whole blood in EDTA tubes was centrifuged to generate plasma.
FIG 1

cobas plasma separation card showing the three spots (left) and how plasma is separated from whole blood in the spot (right).

cobas plasma separation card showing the three spots (left) and how plasma is separated from whole blood in the spot (right). Plasma samples (EDTA-plasma and PSC) were analyzed in single determinations using cobas HIV. Handling of systems, specimens, controls, and reagents was carried out according to procedures described in the cobas 8800 Systems User Guide, Publication version 3.0, software version 1.2. Execution of the study was performed using one cobas 8800 system. A single lot of PSC and a single lot of cobas specimen preextraction reagent were used. Assessments of the limits of detection (LoD) and sample stability were tested with both the cobas HIV and CAP/CTM assays. The cobas HIV LoD is 13.2 copies/ml, and the lower limit of quantification (LLOQ) is 20 copies/ml in plasma, using 0.5 ml of sample input volume. The CAP/CTM has an LoD of 16.5 copies/ml and an LLOQ of 20 copies/ml, using 0.85 ml of sample input volume. PSC results were not adjusted for hematocrit level.

PSC performance assessment: equivalency, sensitivity and specificity.

To determine matrix (plasma type) equivalency, results from EDTA and PSC samples were compared and tested by cobas HIV. Samples with large differences between the viral load from EDTA and by PSC were also tested with CAP/CTM. We calculated the log10 viral load difference and the mean deviation between matched samples. All valid results with a quantitative value of >0 copies/ml in both matrices (EDTA and PSC) were included in the equivalency calculation. A linear regression analysis using Deming regression was also conducted. Test results generated for the matrix equivalency testing were analyzed to determine the sensitivity and specificity at a viral load threshold of 1,000 copies/ml (WHO recommended threshold for treatment failure) (3).

PSC performance assessment: limits of detection and stability assessments.

Three independent dilution series of five concentration levels for HIV-1 (plus a blank control) were prepared by diluting positive spike-in panel members (HIV-1 group M subtype B from MVP899 cell culture supernatant) 1:100 in HIV-negative EDTA whole blood. The study panels were pipetted onto three different PSCs and multiple replicates tested using cobas HIV and CAP/CTM. A part of the first panel member was centrifuged and titer assigned by the calibrator bracketing method using cobas HIV with the 3rd HIV-1 International Standard (WHO). Probit values were determined at the 95% hit rate, together with the lowest panel member with at least a 95% reactive rate. To assess stability, PSCs were prepared using HIV-negative whole blood that had been spiked with HIV-1 at approximately five times the LoD. Viral load was measured in control PSC, with the remainder evaluated after being stored for various time periods at temperatures ranging from −10°C or lower to 45°C and at a humidity of 85%.

RESULTS

Usability.

The phlebotomists rated the sample collection positively, scoring most subtasks above 4 out of 5 (“agree”) (Figs 2A and B). The lancet, capillary tube, and workflow were considered user friendly. However, the phlebotomists—who had not been trained—had some difficulties obtaining sufficient samples, determining when the capillary was adequately filled with blood, obtaining samples without air bubbles, and applying the correct amount of blood to the PSC. Of the 51 capillaries collected, 5 clotting events occurred, due to air bubbles in the tube or delays in spotting the blood on the PSC. Once these phlebotomists became familiar with the use of capillary tubes, their performance improved markedly. Health care workers participating in the study believed that the PSC would be very useful in their setting. A few health workers believed that the sample collection workflow did not appear user friendly, mainly due to the use of a capillary and the volume of blood to be collected.
FIG 2

Usability assessment of sample collection workflow.

Usability assessment of sample collection workflow. Laboratory staff rated all the subtasks in the preanalytical workflow assessment as above 4 (“agree”). A few errors occurred on isolation of the spot and application, in which the filtration membrane was transferred rather than the spot. Overall, however, across all components of the usability assessment, no critical product risks were identified.

Equivalency, sensitivity, and specificity.

A total of 535 PSCs were processed, of which 23 yielded invalid results (4.3%), 9 were duplicates (1.7%), 6 had no paired plasma viral load results (1.1%) and 8 had no HIV-positive status confirmed (1.5%). Of the 521 EDTA-plasma samples processed, 19 were invalid (3.6%). A further 13 samples were excluded; 1 was a duplicate, 4 had no paired PSC result, and 8 had no HIV-positive status confirmed. Additionally, 4 tests on both PSC and EDTA-plasma were excluded due to operator errors. A total of 485 samples had valid paired EDTA and PSC results; of these, 132 samples had quantifiable results for both matrices. The overall mean log10 difference in titer between EDTA-plasma and PSC-plasma was 0.05 (95% confidence interval [CI], −0.01 to 0.11; Fig. 3). The Deming regression analysis line slope was 0.92, the y intercept was 0.42, and R2 was 0.90 (Fig. 4).
FIG 3

Bland-Altman graph for matrix (plasma type) equivalency with the cobas HIV-1 test, showing the difference in matched pairs in log10 titer PSC-log10 titer plasma (mean difference, 0.05; standard error, 0.03; upper 95%, 0.11; lower 95%, −0.01; n = 132. A total of 353 samples did not have a quantifiable result on both matrices; 333 were target not detected, 13 were below the limit of detection, and 7 were above the limit of detection).

FIG 4

Deming regression for matrix (plasma type) equivalency with the cobas HIV-1 test.

Bland-Altman graph for matrix (plasma type) equivalency with the cobas HIV-1 test, showing the difference in matched pairs in log10 titer PSC-log10 titer plasma (mean difference, 0.05; standard error, 0.03; upper 95%, 0.11; lower 95%, −0.01; n = 132. A total of 353 samples did not have a quantifiable result on both matrices; 333 were target not detected, 13 were below the limit of detection, and 7 were above the limit of detection). Deming regression for matrix (plasma type) equivalency with the cobas HIV-1 test. Analysis of all 485 paired EDTA and PSC samples at a 1,000 copies/ml threshold using the cobas HIV yielded a sensitivity of 97.0% (128/132; 95% CI, 92.4 to 99.2%) and specificity of 97.2% (343/353; 95% CI, 94.9 to 98.6%). In the 14 specimens with discordant EDTA and PSC results, almost all had a viral load of around 1,000 copies/ml, with a median log10 titer of 3.13 (interquartile range [IQR] = 3.00 to 3.35). The median log10 difference between the PSC and plasma viral load for the misclassified specimens was 0.60 (IQR = −0.14 to 0.80). No differences in findings were noted by sample type.

Limits of detection and stability.

The viral load of 63 replicates per concentration level were measured in the LoD assessment (Table 1). Based on probit analysis for all PSC lots combined, the LoD was 790.2 copies/ml with cobas HIV (95% CI, 658.9 to 1,003.6 copies/ml) and 737.9 copies/ml with CAP/CTM (95% CI, 614.3 to 938.5 copies/ml). For each of the storage conditions listed in Table 2, there was a 100% detection rate for HIV. The mean difference between the stored specimens and the controls was <0.2 log10 copies/ml for all the stability conditions assessed.
TABLE 1

Limits of detection of the plasma separation card for cobas HIV

Concn (copies/ml)No. of valid samplesa No. of reactive samplesDetection rate (%)
1,971.16262100.0
1,925.36363100.0
1,358.16262100.0
985.5636298.4
962.6636298.4
679635993.7
657635993.7
641.8635790.5
452.7635282.5
328.5634774.6
320.9634673.0
226.3624979.0
164.3633657.1
160.4633555.6
113.2633860.3
06300.0
06300.0
06300.0

Some sample sizes were <63 due to an insufficient number of spots. Data for limit of detection analysis for CAP/CTM are not shown.

Limit of detection by probit analysis (95% hit rate), 790.2 copies/ml. 95% confidence interval (CI), 658.9 to 1,003.6 copies/ml. Lowest panel member with ≥95% reactive rate, 962.6 copies/ml 98.4% (62/63); 95% CI = 91.5 to 100.0%; lower one-sided 95% CI = 92.7%.

TABLE 2

Clinical specimen stability results

Time pointStorage conditions (no. of days, temperature [°C], relative humidity [%])a Layer separation after 21 days (transport stability)b Valid replicatesHIV-1 reactive rate (%)Mean log10 titerMean log10 titer difference from that at T0
1T0NA12/12c 1003.450.00
221, 45, 85NA15/151003.42−0.03
321, 45, 85 + 28, 18–30No15/151003.480.03
21, 45, 85 + 28, 2–8No15/151003.38−0.07
21,45, 85 + 28, ≤10No14/141003.460.01
421, 45, 85 + 35, 18–30Yes14/141003.35−0.10
21, 45, 85 + 35, 2-8Yes15/151003.43−0.02
21, 45, 85 + 35, ≤10Yes15/151003.460.01
521, 45, 85 + 56, 18–30No15/151003.35−0.09
21, 45, 85 + 56, 2–8No15/151003.42−0.02
21, 45, 85 + 56, ≤10No15/151003.37−0.08
21, 45, 85 + 56, 18–30Yes14/141003.480.03
21, 45, 85 + 56, 2–8Yes15/151003.450.00
21, 45, 85 + 56, ≤10Yes15/151003.43−0.02

T0, time of initial measurement.

n/a, not applicable.

Could not repeat as no additional T0 samples were available.

Limits of detection of the plasma separation card for cobas HIV Some sample sizes were <63 due to an insufficient number of spots. Data for limit of detection analysis for CAP/CTM are not shown. Limit of detection by probit analysis (95% hit rate), 790.2 copies/ml. 95% confidence interval (CI), 658.9 to 1,003.6 copies/ml. Lowest panel member with ≥95% reactive rate, 962.6 copies/ml 98.4% (62/63); 95% CI = 91.5 to 100.0%; lower one-sided 95% CI = 92.7%. Clinical specimen stability results T0, time of initial measurement. n/a, not applicable. Could not repeat as no additional T0 samples were available.

DISCUSSION

The PSC performed well in measures of equivalency, LoD, and sample stability. Most importantly, the mean deviation of the log10 viral load between PSC and matched EDTA-plasma specimens was 0.05 log10, suggesting that the card is adequate for use in the monitoring of HIV viral load. The rate of upward misclassification of results, where specimens are incorrectly categorized as being above 1,000 copies/ml or virological failure, was 3%. This is considerably lower than that in many reports of DBS analyses, where upward misclassification rates can reach 10 to 15% or higher (25, 28). The LoD of around 750 copies/ml with cobas HIV and with CAP/CTM is considerably higher than that with EDTA specimens on those systems (around 20 copies/ml). Even with the introduction of point-of-care viral load devices, it is expected that DBS and related technologies, such as the PSC, will remain relevant tools for expanding access to viral load monitoring. This is especially true in rural settings, but also even in some of the urban clinics visited in this study. There is clearly a need, however, for optimizing the performance of DBS technologies. The impact of interference of cell-associated viral nucleic acids in viral load measures is highest when plasma viral load is low, as in patients successfully treated with ART. Indeed, most DBSs perform poorly at thresholds of around 1,000 copies/ml (25, 29–31). Several DBS protocols have been designed to minimize the contribution of cell-associated viral nucleic acid by selectively purifying RNA during sample preparation or by preferentially eluting plasma-associated virus from a DBS, with some success (32, 33). The PSC described in this paper presents a novel, alternative strategy for overcoming this vexing problem. The stability of the PSC is reassuring, indicating that a specimen collected with the PSC can be transported for at least 28 days and then stored for 56 days at a wide range of temperatures without compromising the HIV viral load result. Although DBS specimens generally have high levels of stability (34–36), some evidence suggests that caution is still required in this regard (37). Overall, the sample collection and the preanalytical workflow of the PSC appear to be of low complexity and user friendly, especially once staff become familiar with the procedures. The usability findings, however, highlighted areas that need to be foregrounded in the training and documentation that accompany the card. These mostly concern the correct use of capillary tubes to avoid blood clots and to ensure correct spotting of the PSC. Rates of invalid samples linked to the PSC specimens were around 4% and would likely reduce as health workers become more proficient in sample collection. Health workers are increasingly familiar with collecting DBS samples, and this expertise would likely be transferable to PSC use. Having multiple spots on the PSC is a strength of the card, allowing for repeat testing or for additional tests to be done. Collection of three samples of relatively high volume, however, may raise the likelihood of blood clots and pose challenges for health workers, especially with pediatric patients. The study had several limitations. Most notably, it was conducted in a single urban center, distinct from many of the settings where the card would be used. Direct comparison between the PSC and DBS cards under field conditions would assist countries in selecting the optimum approach for their circumstances. To date, many of the studies that measure the performance of DBS and related technologies in field conditions are of poor quality (38). Also, it is necessary to confirm the performance of PSCs in other assays, such as those for HIV viral resistance, drug level monitoring (e.g., ART), and for diagnosing hepatitis B and C viral infections. In conclusion, a technology that allows for the collection of plasma using finger-prick specimens would retain the sample collection and transport advantages of a DBS but produce plasma as the testing matrix; this is a major development for the field. The findings presented here on the performance and usability of the PSC are promising. This technology supports the efforts to scale up viral load testing, which is essential for ART monitoring, and indeed, most importantly, for the achieving of the “third 90.” Success of the ongoing global rollout of antiretroviral treatment depends in no small measure on the levels of access to and quality of tests for monitoring of viral loads.
  31 in total

Review 1.  The use of the dried blood spot sample in epidemiological studies.

Authors:  S P Parker; W D Cubitt
Journal:  J Clin Pathol       Date:  1999-09       Impact factor: 3.411

2.  Limited utility of dried-blood- and plasma spot-based screening for antiretroviral treatment failure with Cobas Ampliprep/TaqMan HIV-1 version 2.0.

Authors:  Souleymane Sawadogo; Andreas Shiningavamwe; Joy Chang; Andrew D Maher; Guoqing Zhang; Chunfu Yang; Esegiel Gaeb; Harold Kaura; Dennis Ellenberger; David W Lowrance
Journal:  J Clin Microbiol       Date:  2014-08-20       Impact factor: 5.948

3.  Field evaluation of dried blood spots for routine HIV-1 viral load and drug resistance monitoring in patients receiving antiretroviral therapy in Africa and Asia.

Authors:  Marjorie Monleau; Avelin F Aghokeng; Sabrina Eymard-Duvernay; Anoumou Dagnra; Dramane Kania; Nicole Ngo-Giang-Huong; Coumba Touré-Kane; Lien X T Truong; Marie-Laure Chaix; Eric Delaporte; Ahidjo Ayouba; Martine Peeters
Journal:  J Clin Microbiol       Date:  2013-12-11       Impact factor: 5.948

4.  Estimation of HIV-1 DNA Level Interfering with Reliability of HIV-1 RNA Quantification Performed on Dried Blood Spots Collected from Successfully Treated Patients.

Authors:  Sylvie Zida; Edouard Tuaillon; Makoura Barro; Arnaud Kwimatouo Lekpa Franchard; Thérèse Kagoné; Boubacar Nacro; Abdoul Salam Ouedraogo; Karine Bolloré; Armen Sanosyan; Jean-Christophe Plantier; Nicolas Meda; Lassana Sangaré; Christine Rouzioux; François Rouet; Dramane Kania
Journal:  J Clin Microbiol       Date:  2016-03-23       Impact factor: 5.948

5.  Sustainable HIV treatment in Africa through viral-load-informed differentiated care.

Authors:  Andrew Phillips; Amir Shroufi; Lara Vojnov; Jennifer Cohn; Teri Roberts; Tom Ellman; Kimberly Bonner; Christine Rousseau; Geoff Garnett; Valentina Cambiano; Fumiyo Nakagawa; Deborah Ford; Loveleen Bansi-Matharu; Alec Miners; Jens D Lundgren; Jeffrey W Eaton; Rosalind Parkes-Ratanshi; Zachary Katz; David Maman; Nathan Ford; Marco Vitoria; Meg Doherty; David Dowdy; Brooke Nichols; Maurine Murtagh; Meghan Wareham; Kara M Palamountain; Christine Chakanyuka Musanhu; Wendy Stevens; David Katzenstein; Andrea Ciaranello; Ruanne Barnabas; R Scott Braithwaite; Eran Bendavid; Kusum J Nathoo; David van de Vijver; David P Wilson; Charles Holmes; Anna Bershteyn; Simon Walker; Elliot Raizes; Ilesh Jani; Lisa J Nelson; Rosanna Peeling; Fern Terris-Prestholt; Joseph Murungu; Tsitsi Mutasa-Apollo; Timothy B Hallett; Paul Revill
Journal:  Nature       Date:  2015-12-03       Impact factor: 49.962

6.  Alternative Sample Types for HIV-1 Antiretroviral Drug Resistance Testing.

Authors:  Dolly Singh; Adit Dhummakupt; Lilly Siems; Deborah Persaud
Journal:  J Infect Dis       Date:  2017-12-01       Impact factor: 5.226

Review 7.  Measurement of HIV-1 viral load for drug resistance surveillance using dried blood spots: literature review and modeling of contribution of DNA and RNA.

Authors:  Neil T Parkin
Journal:  AIDS Rev       Date:  2014 Jul-Sep       Impact factor: 2.500

8.  Dried blood spots for viral load monitoring in Malawi: feasible and effective.

Authors:  Sarah E Rutstein; Mina C Hosseinipour; Deborah Kamwendo; Alice Soko; Memory Mkandawire; Andrea K Biddle; William C Miller; Morris Weinberger; Stephanie B Wheeler; Abdoulaye Sarr; Sundeep Gupta; Frank Chimbwandira; Reuben Mwenda; Steve Kamiza; Irving Hoffman; Ronald Mataya
Journal:  PLoS One       Date:  2015-04-21       Impact factor: 3.240

9.  Comparison of cross-sectional HIV incidence assay results from dried blood spots and plasma.

Authors:  Katherine E Schlusser; Christopher Pilcher; Esper G Kallas; Breno R Santos; Steven G Deeks; Shelley Facente; Sheila M Keating; Michael P Busch; Gary Murphy; Alex Welte; Thomas Quinn; Susan H Eshleman; Oliver Laeyendecker
Journal:  PLoS One       Date:  2017-02-23       Impact factor: 3.240

10.  Screening for HIV, hepatitis B and syphilis on dried blood spots: A promising method to better reach hidden high-risk populations with self-collected sampling.

Authors:  Inge H M van Loo; Nicole H T M Dukers-Muijrers; Rosalie Heuts; Marianne A B van der Sande; Christian J P A Hoebe
Journal:  PLoS One       Date:  2017-10-20       Impact factor: 3.240

View more
  9 in total

1.  Stability of HIV-1 Nucleic Acids in cobas Plasma Separation Card for Viral Load Measurement.

Authors:  Adolfo Vubil; Carina Nhachigule; Ana Flora Zicai; Bindiya Meggi; Paulino da Costa; Nédio Mabunda; Sofia Viegas; Nádia Sitoe; Ilesh Jani
Journal:  Am J Clin Pathol       Date:  2022-07-01       Impact factor: 5.400

2.  Field Suitability and Diagnostic Accuracy of the Biocentric Open Real-Time PCR Platform for Dried Blood Spot-Based HIV Viral Load Quantification in Eswatini.

Authors:  Bernhard Kerschberger; Nombuso Ntshalintshali; Qhubekani Mpala; Paola Andrea Díaz Uribe; Gugu Maphalala; Sydney Kalombola; Addissu Bekele Telila; Tiwonge Chawinga; Mukelo Maphalala; Aditi Jani; Nomcebo Phugwayo; Roberto de la Tour; Nomxolise Nyoni; Javier Goiri; Sindisiwe Dlamini; Iza Ciglenecki; Emmanuel Fajardo
Journal:  J Acquir Immune Defic Syndr       Date:  2019-09-01       Impact factor: 3.731

3.  Assessment of a dried blood spot C-reactive protein method to identify disease flares in rheumatoid arthritis patients.

Authors:  Leon G D'Cruz; Kevin G McEleney; Chris Cochrane; Kyle B C Tan; Priyank Shukla; Philip V Gardiner; Dawn Small; Shu-Dong Zhang; David S Gibson
Journal:  Sci Rep       Date:  2020-12-03       Impact factor: 4.379

4.  Diagnostic Accuracy of Dried Plasma Spot Specimens for HIV-1 Viral Load Testing: A Systematic Review and Meta-analysis.

Authors:  Youyi Fong; Jessica Markby; Mauro Andreotti; Ingrid Beck; Thomas Bourlet; Don Brambilla; Lisa Frenkel; Rosalia Lira; Julie A E Nelson; Georgios Pollakis; Sandrine Reigadas; Douglas Richman; Souleymane Sawadogo; Laura Waters; Chunfu Yang; Clement Zeh; Meg Doherty; Lara Vojnov
Journal:  J Acquir Immune Defic Syndr       Date:  2022-03-01       Impact factor: 3.771

5.  Non-Lethal Blood Sampling of Fish in the lab and Field With Methods for Dried Blood Plasma Spot Omic Analyses.

Authors:  S Pollard; J C Anderson; F Bah; M Mateus; M Sidhu; Dbd Simmons
Journal:  Front Genet       Date:  2022-03-24       Impact factor: 4.599

6.  903 Protein Saver cards: the best alternative for dried blood spot storage at room temperature for HCV RNA.

Authors:  Ricardo Madrid; Verónica Briz; Sonia Arca-Lafuente; Cristina Casanueva-Benítez; Celia Crespo-Bermejo; Violeta Lara-Aguilar; Luz Martín-Carbonero; Ignacio de Los Santos
Journal:  Sci Rep       Date:  2022-06-16       Impact factor: 4.996

7.  Viral load monitoring for people living with HIV in the era of test and treat: progress made and challenges ahead - a systematic review.

Authors:  Minh D Pham; Huy V Nguyen; David Anderson; Suzanne Crowe; Stanley Luchters
Journal:  BMC Public Health       Date:  2022-06-16       Impact factor: 4.135

Review 8.  Pooled testing: A tool to increase efficiency of infant HIV diagnosis and virological monitoring.

Authors:  Wolfgang Preiser; Gert U van Zyl
Journal:  Afr J Lab Med       Date:  2020-08-11

9.  Utility of the Cobas® Plasma Separation Card as a Sample Collection Device for Serological and Virological Diagnosis of Hepatitis C Virus Infection.

Authors:  Fernando Velásquez-Orozco; Ariadna Rando-Segura; Joan Martínez-Camprecios; Paula Salmeron; Adrián Najarro-Centeno; Àngels Esteban; Josep Quer; María Buti; Tomás Pumarola-Suñe; Francisco Rodríguez-Frías
Journal:  Diagnostics (Basel)       Date:  2021-03-08
  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.