| Literature DB >> 35136911 |
Adolfo Vubil1, Carina Nhachigule1, Ana Flora Zicai1, Bindiya Meggi1, Paulino da Costa1, Nédio Mabunda1, Sofia Viegas1, Nádia Sitoe1, Ilesh Jani1.
Abstract
OBJECTIVES: Our study aimed to evaluate the stability of human immunodeficiency virus 1 (HIV-1) RNA on cobas plasma separation card (PSC) specimens for viral load (VL) testing after being exposed to varied temperatures and storage times.Entities:
Keywords: HIV viral load; Nucleic acids; Stability
Mesh:
Substances:
Year: 2022 PMID: 35136911 PMCID: PMC9247838 DOI: 10.1093/ajcp/aqac007
Source DB: PubMed Journal: Am J Clin Pathol ISSN: 0002-9173 Impact factor: 5.400
Mean Difference of Viral Load Results Over the 28-Day Study Period in Temperature Range of 25ºC to 42ºC
| Plasma VL Category | n | 24 Hours | 72 Hours | 7 Days | 21 Days | 28 Days |
|---|---|---|---|---|---|---|
| 2°C-8ºC PSC specimens | ||||||
| Target not detected, % | 3 | 100 | 100 | 100 | 100 | 100 |
| Target <20 cp/mL, % | 6 | 100 | 100 | 100 | 100 | 100 |
| Target quantifiable but <738 cp/mL, % | 7 | 100 | 100 | 100 | 100 | 100 |
| Target at 1,000 cp/mL, c/mL | 1 | 1,486 | TND | <738 | TND | 820 |
| >1,000 cp/mL (range, 3,000-800,000 cp/mL), mean (SD) difference | 13 | –0.31 (–0.09) | –0.30 (–0.15) | –0.29 (–0.04) | –0.22 (–0.09) | –0.25 (–0.12) |
| 25ºC PSC specimens | ||||||
| Target not detected, % | 3 | 100 | 100 | 100 | 100 | 100 |
| Target <20 cp/mL, % | 6 | 100 | 100 | 100 | 100 | 100 |
| Target quantifiable but <738 cp/mL, % | 7 | 100 | 100 | 100 | 100 | 100 |
| Target at 1,000 cp/mL, c/mL | 1 | 949 | 995 | <738 | 771 | <738 |
| >1,000 cp/mL (range, 3,000-800,000 cp/mL), mean (SD) difference | 13 | –0.31 (–0.17) | –0.30 (–0.09) | –0.29 (–0.13) | –0.22 (–0.17) | –0.25 (–0.10) |
| 37ºC PSC specimens | ||||||
| Target not detected, % | 3 | 100 | 100 | 100 | 100 | 100 |
| Target <20 cp/mL, % | 6 | 100 | 100 | 100 | 100 | 100 |
| Target quantifiable but <738 cp/mL, % | 7 | 100 | 100 | 100 | 86 | 86 |
| Target at 1,000 cp/mL, | 1 | 1,087 | TND | <738 | 749 | TND |
| >1,000 cp/mL (range, 3,000-800,000 cp/mL), mean (SD) difference | 13 | –0.28 (–0.10) | –0.25 (–0.11) | –0.22 (–0.23) | –0.05 (–0.28) | –0.07 (–0.19) |
| 42ºC PSC specimens | ||||||
| Target not detected, % | 3 | 100 | 100 | 100 | 100 | 100 |
| Target <20 cp/mL, % | 6 | 100 | 100 | 100 | 100 | 100 |
| Target quantifiable but <738 cp/mL, % | 7 | 100 | 100 | 100 | 100 | 100 |
| Target at 1,000 cp/mL, c/mL | 1 | <738 | TND | TND | 936 | 1,670 |
| >1,000 cp/mL (range, 3,000-800,000 cp/mL), mean (SD) difference | 13 | –0.28 (–0.06) | –0.28 (–0.10) | –0.28 (–0.06) | –0.20 (–0.07) | –0.05 (–0.17) |
cp/mL, copies per milliliter; PSC, plasma separation card; VL, viral load.
an = 1 at 1,354 cp/mL; n = 1 at 2,954 cp/mL.
FIGURE 1Bland-Altman analyses to evaluate agreement in human immunodeficiency virus 1 viral load quantification among plasma and cobas plasma separation card (PSC) specimens stored at temperatures ranging from 2ºC to 8ºC (A), 25ºC (B), 37ºC (C), and 42ºC (D) for a period of up to 28 days. LA, limit of agreement.