| Literature DB >> 30671668 |
Erik Leijte1, Elke Arts2, Bart Witteman3, Jack Jakimowicz4,5, Ivo De Blaauw2, Sanne Botden2.
Abstract
BACKGROUND: The purpose of this study was to validate the eoSim, an affordable and mobile inanimate laparoscopic simulator with instrument tracking capabilities, regarding face, content and construct validity on complex suturing tasks.Entities:
Keywords: Augmented reality simulator; Construct validity; Content validity; Face validity; Laparoscopy training; Simulation
Mesh:
Year: 2019 PMID: 30671668 PMCID: PMC6795635 DOI: 10.1007/s00464-018-06652-3
Source DB: PubMed Journal: Surg Endosc ISSN: 0930-2794 Impact factor: 4.584
Fig. 1Set up of the eoSim as used in this study
Fig. 4Screenshot of the anastomosis needle transfer (Task 3)
Fig. 2Screenshot of the vertical defect suturing (Task 1)
Fig. 3Screenshot of the up side down needle transfer (Task 2)
Demographics of participants
| Novice ( | Target ( | Expert ( | Total ( | |
|---|---|---|---|---|
| Mean age (SD) | 25 (3.2) | 34 (4.7) | 43 (11.2) | 30 (7.9) |
| Dexterity | ||||
| Right handed | 88.1% | 87.1% | 92.3% | 88.5% |
| Left handed | 11.9% | 9.7% | 0.0% | 9.6% |
| Ambidextrous | 0.0% | 3.2% | 7.7% | 1.9% |
| Education | ||||
| Intern | 50 | 0 | 0 | 50 |
| Postgraduate year (years) | ||||
| 1 | 9 | 2 | 0 | 11 |
| 2 | 1 | 1 | 0 | 2 |
| 3 | 0 | 4 | 0 | 4 |
| 4 | 0 | 12 | 0 | 12 |
| 5 | 0 | 9 | 0 | 9 |
| 6 | 0 | 0 | 1 | 1 |
| Specialized surgeon | 0 | 3 | 13 | 16 |
SD standard deviation
Mean opinion scores (SD) on realism, didactic value and usability of the three performed tasks on the eoSim
| Intracorporeal suturing | Expert ( | Target ( | Novice ( | Total ( | |
|---|---|---|---|---|---|
| Realism | 3.9 (0.8) | 3.9 (0.5) | 3.9 (0.6) | 3.9 (0.6) | 0.953 |
| Didactic value | 4.0 (0.8) | 4.0 (0.5) | 3.9 (0.7) | 4.0 (0.6) | 0.573 |
| Usability | 4.0 (1.0) | 4.2 (0.7) | 4.2 (0.6) | 4.2 (0.7) | 0.618 |
Differences in opinion scores were calculated with the ANOVA test. A p-value of < 0.05 was considered a significant difference
SD standard deviation
*p-value after Welch correction for unequal variances
Construct validity of the eoSim
| Intracorporeal suturing | Expert ( | Target ( | Novice ( | |
|---|---|---|---|---|
| Time (s) | 339 (114) | 607 (229) | 1124 (578) | < |
| Distance (m) | 8.1 (3.5) | 15.6 (6.3) | 21.7 (15.3) | < |
| Off-screen (%) | 6.5 (4.8) | 9.9 (7.4) | 8.2 (7.0) | 0.302 |
| Speed (mm/s) | 2.67 (0.69) | 4.42 (2.62) | 7.52 (29.31) | 0.698 |
| Acceleration (mm/s2) | 1.27 (0.49) | 2.15 (1.08) | 1.67 (1.23) |
|
| Smoothness (mm/s3) | 0.008 (0.009) | 0.024 (0.039) | 0.011 (0.018) | 0.088* |
| Working area (cm) | 5.72 (0.96) | 5.14 (0.94) | 5.41 (1.17) | 0.242 |
Data in this table represent the mean value (standard deviation). Statistical differences were calculated with the ANOVA test. A p-value < 0.05 (displayed in italic) was considered significant
*p-value after Welch correction for unequal variances
**p-value after Chi-square test
Post hoc ANOVA analysis of significant results
| Parameter | Expert versus novice | Expert versus intermediate | Intermediate versus novice | ||
|---|---|---|---|---|---|
| Task 1: Intracorporeal suturing | Time | < | < | < | |
| Distance | < | < |
| ||
| Acceleration | 0.769 | 0.059 | 0.188 | ||
| Smoothness | 0.558 | 0.080 | 0.213 | ||
| Off screen | 1.000 | 0.435 | 0.827 | ||
| Working area | 0.770 | 0.183 | 0.830 | ||
| Task 2: Upside down needle transfer | Time | < | 0.885 | < | |
| Distance | 0.267 | 0.429 | < | ||
| Off screen |
| < | 0.736 | ||
| Working area |
| < | 0.079 | ||
| Task 3: Anastomosis needle transfer | Time | 0.128 | 0.671 | 0.347 | |
| Number of stitches | < | 0.057 |
| ||
| Time per stitch | < | 0.126 | < | ||
| Distance per stitch | < | 0.054 |
| ||
| Completed 16 stitches* | < | 0.392 |
| ||
| Completed 8 stitches* | < | 0.078 | 0.138 | ||
After testing for homogeneity of variance either Bonferroni correction (equal variances) or the Games-Howell correction (unequal variance) was used
*p-value after Chi-square test. A p-value of < 0.05 (displayed in italic) was considered significant