Łukasz Waszczuk1, Maciej Guziński1, Jerzy Garcarek1, Marek Sąsiadek1.
Abstract
PURPOSE: The aim of the study was to test the hypothesis that unenhanced phase does not require as high image quality as subsequent phases acquired after contrast administration in triple-phase abdomen and pelvis computed tomography (CT), and to assess if attenuation value (AV) measurements may be obtained from unenhanced images acquired with three-fold reduced radiation dose.Entities:
Keywords: X-ray computed; abdomen; adult; radiation dosage; radiation protection; tomography
Year: 2018 PMID: 30627230 PMCID: PMC6323542 DOI: 10.5114/pjr.2018.75682
Source DB: PubMed Journal: Pol J Radiol ISSN: 1733-134X
Figure 1Study design: the planned radiation dose during modified triplephase abdomen computed tomography (CT) compared to the standard CT performed in our Institution. The bars represent dose length product (DLP, mGy–cm) in unenhanced, arterial, venous phase and the total dose
Scanning protocols in triple-phase abdomen and pelvis computed tomography. In the case group unenhanced phase performed with reduced dose. In the control group, all three phases performed with standard scanner settings
| Case | Control | |
|---|---|---|
| Tube potential | 120 kVp | |
| Automated exposure control | Enabled | |
| Noise index | 59.1/22.0/22.0 | 22.0/22.0/22.0 |
| Detector configuration | 64 × 0.625 mm | |
| Pitch | 0.984 : 1 | |
| Reconstructed slice thickness | 0.625 mm | |
| ASIR | Enabled | |
Noise index in unenhanced/arterial/venous phase
ASIR – Adaptive Statistical Iterative Reconstruction
The tissue attenuation values in unenhanced phase obtained with three-times reduced radiation dose compared to standard dose scan. Unenhanced phases are derived from triple-phase abdomen and pelvis computed tomography
| Reduced dose | Standard dose |
| |||
|---|---|---|---|---|---|
| Mean | SD | Mean | SD | ||
| Muscle | 51.0 | 6.0 | 50.0 | 7.5 | NS |
| Fat | –107.1 | 6.6 | –99.0 | 12.8 | NS |
| Air | –998.0 | 4.8 | –998.7 | 3.4 | NS |
| Aorta | 30.3 | 10.8 | 34.5 | 8.9 | NS |
| Liver | 50.4 | 10.0 | 43.5 | 20.5 | NS |
| Spleen | 43.5 | 7.5 | 45.1 | 5.3 | NS |
| Bladder | 6.0 | 9.2 | 10.9 | 11.0 | NS |
Figure 2The tissue attenuation value in unenhanced phase obtained with three-fold reduced radiation dose (Case) compared to standard dose scan (Control). The unenhanced phases are derived from triple-phase abdomen and pelvis computed tomography