Literature DB >> 30557331

Comparison of visual assessment and computer image analysis of intracoronary thrombus type by optical coherence tomography.

Timo P Kaivosoja1, Shengnan Liu2, Jouke Dijkstra2, Heini Huhtala3, Tej Sheth4, Olli A Kajander1.   

Abstract

BACKGROUND: Analysis of intracoronary thrombus type by optical coherence tomography (OCT) imaging is highly subjective. We aimed to compare a newly developed image analysis method to subjective visual classification of thrombus type identified by OCT.
METHODS: Thirty patients with acute ST elevation myocardial infarction were included. Thrombus type visually classified by two independent readers was compared with analysis using QCU-CMS software.
RESULTS: Repeatability of the computer-based measurements was good. By using a ROC, area under curve values for discrimination of white and red thrombi were 0.92 (95% confidence intervals (CI) 0.83-1.00) for median attenuation, 0.96 (95% CI 0.89-1.00) for mean backscatter and 0.96 (95% CI 0.89-1.00) for mean grayscale intensity. Median attenuation of 0.57 mm-1 (sensitivity 100%, specificity 71%), mean backscatter of 5.35 (sensitivity 92%, specificity 94%) and mean grayscale intensity of 120.1 (sensitivity 85%, specificity 100%) were identified as the best cut-off values to differentiate between red and white thrombi.
CONCLUSIONS: Attenuation, backscatter and grayscale intensity of thrombi in OCT images differentiated red and white thrombi with high sensitivity and specificity. Measurement of these continuous parameters can be used as a less user-dependent method to characterize in vivo thrombi. The clinical significance of these findings needs to be tested in further studies.

Entities:  

Mesh:

Year:  2018        PMID: 30557331      PMCID: PMC6296537          DOI: 10.1371/journal.pone.0209110

Source DB:  PubMed          Journal:  PLoS One        ISSN: 1932-6203            Impact factor:   3.240


Introduction

Intracoronary thrombus is a frequent finding in patients with acute coronary syndromes undergoing invasive angiography and percutaneous coronary intervention (PCI). Research on the clinical significance of thrombus type is ongoing, and with more detailed understanding of the thrombotic process, the goal is to facilitate the development of tailored therapies for patients with acute coronary syndromes. Optical coherence tomography (OCT) is a high-resolution intravascular imaging technique with excellent contrast between the vessel lumen and intravascular structures [1]. In a consensus paper, the evidence level for image interpretation of intracoronary thrombus using OCT has been considered to be high [2]. The ability of OCT to produce images allowing differentiation between thrombus types has been validated against histology in post-mortem samples of human coronary arteries [3]. It is widely accepted in the OCT community that thrombi in OCT images can be classified as high-backscattering with a signal-free shadowing (‘red’, erythrocyte-rich) or low-backscattering (‘white’, thrombocyte-rich) [2]. This encourages further validation in clinical patients. However, valid analysis of different thrombus types in vivo by OCT is hindered by the high degree of subjectivity of measurement and the lack of validated methodology. The aim of the present study is to develop a computer image analysis -based method for the assessment of thrombus morphology in OCT images that can be used for further in vivo validation. Computer algorithms determining backscatter, attenuation and intensity of the OCT signal are tested and compared to the present standard method, which is consensus classification of thrombus type by two independent analysts. Our hypothesis is that applying computer image analysis to assess thrombus type is feasible and can reduce observer variation in comparison to subjective evaluation.

Patients and methods

Patients

The study was approved by the Ethical board at Pirkanmaa Hospital District, approval number R10131. Written informed consent was obtained from all participants. The studied series included 30 patients enrolled in an OCT substudy of the TOTAL trial. The TOTAL trial was an international, multicentre, randomized trial of routine thrombectomy (using the Export catheter, Medtronic Cardiovascular, Santa Rosa, CA, USA) compared with PCI alone in STEMI patients treated with PCI (n = 10732) [4,5]. A prospective OCT sub-study (n = 214) of TOTAL trial evaluated thrombus burden in patients with symptoms of myocardial ischaemia lasting for ≥ 30 min and definite electrocardiographic changes indicating STEMI who were referred for primary PCI and randomized within 12 h of symptom onset [6]. Participation in the study required restoration of TIMI 2–3 flow after the first device. Patients were excluded from the OCT sub-study if they were in cardiogenic shock or had known renal failure. Informed consent was obtained from all individual participants included in the study. Of the patients in the OCT substudy who were randomized to thrombectomy, 72 had good quality OCT pullbacks available preceding balloon predilatation. Out of these, 30 patients with the largest maximal thrombus areas were selected for the present study (for detailed patient flow-chart, see S1 Fig). All procedures performed in studies involving human participants were in accordance with the ethical standards of the institutional and/or national research committee and with the 1964 Helsinki declaration and its later amendments or comparable ethical standards. This article does not contain any studies with animals performed by any of the authors.

OCT imaging

OCT imaging was performed using the Ilumien OCT system and C7 Dragonfly catheter (St Jude Medical, Minnesota, MN, USA). The radio-opaque distal marker of the OCT catheter was positioned 1–2 cm distal to the target lesion or stent and contrast was injected either manually or by automatic injector, according to local practice. Adequate image quality was checked and additional images were obtained, if the image quality was sub-optimal. OCT imaging raw data was exported in digital format for off-line analysis.

Optical coherence tomography image analysis

Data on thrombus cross-sectional areas at the culprit lesion was obtained from previous analysis (6). For each patient, three OCT cross-sections with the largest thrombus areas were analyzed.

Visual assessment of thrombus type

Visual thrombus characteristics were evaluated by two independent readers (T.P.K. and O.A.K.) to assess the inter-observer reproducibility. One observer (T.P.K.) repeated the analysis after > 4 week period to assess the intra-observer reproducibility. Thrombus was graded by a six-stage scale, where value 1 represents purely white thrombus and value 6 purely red thrombus, creating thrombus attenuation score (TAS). In accordance to published standards [1,2], thrombus was considered white or mostly white, when the abluminal border of the thrombus or the vessel wall behind it could be seen. Conversely, the more the thrombus caused shadowing and poor visibility of the structures behind it, the redder it was scored. In addition, we applied only two additional rules. If a cross-section contained multiple fragments with different visual appearances, TAS was defined for the largest fragment. In the case of multiple same-sized thrombi, TAS was defined for the most strongly attenuating part of the thrombus.

Computer image analysis of thrombus

A special version of QCU-CMS version 4.69 software (Leiden University Medical Center, Leiden, The Netherlands) was used by two independent analysts. First, the thrombus areas were manually traced. The software then determined attenuation coefficient, backscatter term and grayscale intensity values for these areas and calculated several statistical parameters, including median, mean, standard deviation and 5th, 10th, 90th and 95th percentile values. Two patient cases with white and red thrombus showing the method are shown in Fig 1. To avoid the inclusion of noise pixels and dark areas (low attenuation) at the abluminal side of the thrombus, we experimented with different minimum threshold values for attenuation. For the present analyses, attenuation was analyzed using both 0.15 mm-1 and 0.5 mm-1 threshold values.
Fig 1

Examples of thrombus assessment by the QCU-CMS image analysis software in OCT image frames of two patients with white (A) and red thrombus (D).

After manual tracing of the thrombi (B,E), attenuation analysis was performed including regions with attenuation values above the designated threshold value (displayed in blue) (C,F). Accurate segmentation of the luminal border of the thrombus and the contrast-filled flow area of the vessel can be seen in C and F. *) OCT catheter, #) guidewire artefact, ¤) thrombus, §) vessel wall. OCT, optical coherence tomography.

Examples of thrombus assessment by the QCU-CMS image analysis software in OCT image frames of two patients with white (A) and red thrombus (D).

After manual tracing of the thrombi (B,E), attenuation analysis was performed including regions with attenuation values above the designated threshold value (displayed in blue) (C,F). Accurate segmentation of the luminal border of the thrombus and the contrast-filled flow area of the vessel can be seen in C and F. *) OCT catheter, #) guidewire artefact, ¤) thrombus, §) vessel wall. OCT, optical coherence tomography. The light attenuation coefficient and backscatter term have been determined based on a “depth-resolved” model [7]. The attenuation coefficient describes the total decreasing rate of light when it is traveling though the tissue. OCT images are generated by receiving the backscattered light. The backscatter coefficient describes the efficiency of tissue scattering light backwards. The backscatter term is estimated to be related to the backscatter coefficient. More information about the estimation can be referred to the technical paper [7]. The advantage of this method is its fast pixel-wise estimation, which requires no predefined delineation and is more flexible for post analysis. For each patient, a weighted average of each statistical parameter was calculated using the pixel count in each traced thrombus area. To study inter-observer reproducibility, two readers analyzed the same OCT cross-sections of 10 randomly selected patients independently. For intra-observer comparison, one observer repeated the analyses on these 10 patients after a 4-week period. For the repeated assessments, OCT raw data was re-imported into the QCU-CMS software. Then the thrombus areas were manually re-traced in the same transversal frames.

Statistical analysis

Bland–Altman (B-A) analysis was used to estimate the bias resulting from inter- and intra-observer variation [8]. In the analysis, the mean values of observer 1 and 2 were plotted against the difference of measurements between the observers. The 95% limits of agreement (equal to ±1.96 SD) were determined for the variables. Mean relative difference for each variable was defined as the mean absolute difference divided by mean of observers 1 and 2. Intraclass correlation coefficients (ICCs) and their 95% confidence intervals were calculated based on absolute-agreement, 2-way mixed effect model. Spearman correlations were calculated between the continuous variables and TAS. Inter- and intraobserver reliability of thrombus classification was assessed using the kappa statistic (κ). Receiver-operating characteristic (ROC) curve analyses were performed to determine the best cutoff values (using highest Youden Index) of QCU-CMS parameters for discriminating red and white thrombi and the areas under the curve (AUCs), as well as the sensitivities and specificities of the diagnostic test. In addition, binary logistic regression analysis was used to evaluate how the different continuous image analysis software parameters predicted the type of thrombus. Statistical analyses were performed with the Stata Statistical software: Release 13 (StataCorp, College Station, TX, USA) and the IBM SPSS 21 Statistics software (IBM, Armonk, NY, USA).

Results

Baseline patient characteristics are shown in Table 1.
Table 1

Baseline patient characteristics.

N = 30
Age, years (mean ± SD)57 ± 10.1
Male sex, N (%)22 (73.3)
Risk factor profile, N (%)
 Hypertension14 (46.6)
 Diabetes4 (13.3)
 Current smoking8 (26.7)
 Obesity (BMI ≥ 30)14 (46.7)
 Prior MI2 (6.7)
 Prior PCI1 (3.3)
Creatinine, μmol/L (mean ± SD)82.0 ± 16.6
Antithrombotic treatment, N (%)
 Upfront glycoprotein inhibitor IIb/IIIa use11 (36.7)
 UFH prior to procedure22 (73.3)
 Enoxaparin prior to procedure2 (6.7)
 Bivalirudin prior to procedure0 (0.0)
Culprit vessel, N (%)
 RCA16 (53.3)
 LAD12 (40.0)
 LCx2 (6.7)
TIMI flow*, N (%)
 015 (50.0)
 14 (13.3)
 23 (10.0)
 38 (26.7)
TIMI thrombus class*, N (%)
 < 34 (13.3)
 ≥ 326 (86.7)

* prior to thrombectomy.

BMI, body mass index; LAD, left anterior descending coronary artery; LCx, left circumflex coronary artery; MI, myocardial infarction; PCI, percutaneous coronary intervention; RCA, right coronary artery; SD, standard deviation; TIMI, thrombolysis in myocardial infarction; UFH, unfractionated heparin.

* prior to thrombectomy. BMI, body mass index; LAD, left anterior descending coronary artery; LCx, left circumflex coronary artery; MI, myocardial infarction; PCI, percutaneous coronary intervention; RCA, right coronary artery; SD, standard deviation; TIMI, thrombolysis in myocardial infarction; UFH, unfractionated heparin.

Qualitative (visual) thrombus classification by two observers

For the purpose of statistical comparison of the thrombus assessment methods, a six-stage scale consensus TAS was used to form a binary classification, where scores 1–3 equaled white and 4–6 red thrombus. There were 13 cases classified as white and 17 as red thrombus (S2 Fig). Kappa values for binary classification of the thrombi were 0.74 and 0.53, respectively, for inter- and intraobserver comparisons.

Reproducibility of quantitative measurements using QCU-CMS software

Results presented in Tables 2 and 3 show inter- and intraobserver variability of measurements in thrombus regions. Mean differences for most statistical parameters of thrombus attenuation, backscatter and grayscale intensity were small, with narrow limits of agreement between observers. For all tested parameters, intraobserver variability was slightly smaller than interobserver variation. Limits of agreement for skewness and kurtosis of attenuation, backscatter and grayscale intensity were wider than for the other statistical parameters. B-A plot of median attenuation is shown in Fig 2. B-A plots of mean backscatter and mean grayscale intensity, respectively, are shown in S3 and S4 Figs.
Table 2

Interobserver variability of QCU-CMS measurements.

ParametersObserver 1 Mean (SD)Observer 2 Mean (SD)RangeMean absolute difference (CI)Mean relative difference (%)Limits of agreementICC (CI)
Attenuation (mm-1)
 Median0.61 (0.12)0.57 (0.13)0.35–0.770.03 (-0.02–0.08)5.8-0.11–0.170.91 (0.64–0.98)
 Mean0.72 (0.09)0.69 (0.11)0.49–0.830.03 (-0.02–0.08)4.3-0.10–0.160.88 (0.55–0.97)
 Standard deviation0.42 (0.09)0.42 (0.07)0.32–0.540.001 (-0.02–0.02)0.2-0.07–0.070.96 (0.84–0.99)
 Skewness1.36 (0.50)1.46 (0.40)0.89–2.18-0.10 (-0.26–0.06)-7.2-0.55–0.190.93 (0.73–0.98)
 Kurtosis3.01 (2.42)3.72 (2.45)0.55–7.21-0.72 (-2.21–0.78)-21.4-4.90–3.470.77 (0.14–0.94)
 5th percentile0.26 (0.06)0.23 (0.06)0.17–0.380.02 (-0.01–0.04)8.6-0.04–0.080.90 (0.52–0.98)
 10th percentile0.30 (0.07)0.28 (0.07)0.19–0.460.02 (-0.01–0.05)8.3-0.06–0.110.90 (0.58–0.97)
 90th percentile1.27 (0.16)1.24 (0.17)1.00–1.530.03 (-0.03–0.09)2.3-0.14–0.200.92 (0.71–0.98)
 95th percentile1.51 (0.20)1.48 (0.19)1.21–1.770.03 (-0.04–0.10)1.9-0.17–0.230.93 (0.75–0.98)
Backscatter
 Median5.18 (0.22)5.09 (0.28)4.77–5.490.10 (-0.01–0.20)1.9-0.20–0.400.87 (0.46–0.97)
 Mean5.19 (0.18)5.10 (0.23)4.79–5.460.10 (0.01–0.18)1.8-0.15–0.340.86 (0.36–0.97)
 Standard deviation0.61 (0.09)0.66 (0.07)0.47–0.72-0.04 (-0.07–-0.02)-6.9-0.10–0.010.89 (-0.09–0.98)
 Skewness-0.08 (0.29)-0.22 (0.44)-0.74–0.310.15 (-0.05–0.34)96.7-0.39–0.680.83 (0.37–0.96)
 Kurtosis-0.13 (0.49)0.44 (1.24)-0.70–2.37-0.57 (-1.16–0.02)-183-2.22–1.080.70 (-0.05–0.92)
 5th percentile4.22 (0.26)4.07 (0.26)3.77–4.700.15 (0.05–0.24)3.5-0.11–0.410.87 (0.05–0.97)
 10th percentile4.41 (0.25)4.28 (0.28)3.95–4.930.13 (0.04–0.21)2.9-0.12–0.370.90 (0.26–0.98)
 90th percentile5.99 (0.16)5.94 (0.18)5.65–6.260.05 (-0.02–0.11)0.8-0.14–0.230.91 (0.65–0.98)
 95th percentile6.17 (0.16)6.13 (0.17)5.85–6.410.04 (-0.02–0.10)0.7-0.14–0.210.92 (0.69–0.98)
Grayscale
 Median67.19 (32.99)63.86 (35.43)28.16–147.833.32 (-2.53–9.19)5.1-13.05–19.710.99 (0.94–1.00)
 Mean104.65 (26.87)100.04 (31.37)69.76–166.694.61 (-2.28–11.49)4.5-14.64–23.850.97 (0.88–0.99)
 Standard deviation106.03 (17.46)103.02 (15.95)80.03–134.783.02 (-3.78–9.81)2.9-15.98–22.010.91 (0.67–0.98)
 Skewness2.51 (0.88)2.65 (0.72)1.72–3.99-0.14 (-0.41–0.14)-5.3-0.91–0.630.94 (0.76–0.98)
 Kurtosis12.49 (11.67)14.62 (10.83)4.31–38.17-2.13 (-6.88–2.62)-15.7-15.40–11.140.91 (0.64–0.98)
 5th percentile16.59 (13.34)14.74 (11.59)6.66–48.921.85 (0.08–3.61)11.8-3.10–6.790.99 (0.91–1.00)
 10th percentile21.77 (16.88)20.14 (16.17)8.83–65.091.63 (-0.02–3.27)7.8-2.97–6.220.99 (0.96–1.00)
 90th percentile239.64 (40.18)230.74 (48.77)166.58–312.488.90 (-4.40–22.20)3.8-28.28–46.090.94 (0.80–0.99)
 95th percentile315.19 (44.98)305.05 (53.64)228.63–408.6510.15 (-6.64–26.94)3.3-36.78–57.080.94 (0.75–0.98)

N = 10. CI, confidence interval; SD, standard deviation; ICC, intra-class correlation coefficient.

Table 3

Intraobserver variability of QCU-CMS measurements.

ParametersObserver 1 first time Mean (SD)Observer 1 second time Mean (SD)RangeMean absolute difference (CI)Mean relative difference (%)Limits of agreementICC (CI)
Attenuation (mm-1)
 Median0.61 (0.12)0.59 (0.13)0.37–0.750.02 (-0.03–0.06)3.3-0.11–0.140.93 (0.74–0.98)
 Mean0.72 (0.09)0.70 (0.11)0.51–0.840.02 (-0.02–0.06)2.8-0.10–0.130.92 (0.69–0.98)
 Standard deviation0.42 (0.09)0.41 (0.08)0.31–0.540.001 (-0.01–0.02)0.2-0.04–0.050.98 (0.94–1.00)
 Skewness1.36 (0.50)1.41 (0.43)0.75–2.14-0.05 (-0.24–0.14)-3.6-0.58–0.480.92 (0.68–0.98)
 Kurtosis3.01 (2.42)3.64 (2.81)0.52–6.99-0.64 (-2.48–1.21)-19.2-5.79–4.520.69 (-0.24–0.92)
 5th percentile0.26 (0.06)0.25 (0.06)0.18–0.390.01 (-0.01–0.02)3.9-0.03–0.050.97 (0.90–0.99)
 10th percentile0.30 (0.07)0.29 (0.07)0.20–0.450.01 (-0.01–0.03)3.4-0.05–0.060.97 (0.88–0.99)
 90th percentile1.27 (0.16)1.25 (0.17)1.02–1.540.02 (-0.03–0.07)1.6-0.12–0.160.95 (0.82–0.99)
 95th percentile1.51 (0.20)1.48 (0.21)1.23–1.790.03 (-0.03–0.08)2.0-0.13–0.180.96 (0.86–0.99)
Backscatter
 Median5.18 (0.22)5.13 (0.29)4.78–5.470.05 (-0.06–0.16)1.0-0.25–0.350.91 (0.65–0.98)
 Mean5.19 (0.18)5.14 (0.25)4.80–5.460.04 (0.05–0.13)0.8-0.20–0.290.92 (0.69–0.98)
 Standard deviation0.61 (0.09)0.62 (0.09)0.43–0.69-0.01 (-0.03–-0.01)-1.6-0.06–0.040.98 (0.92–1.00)
 Skewness-0.08 (0.29)-0.10 (0.31)-0.47–0.290.02 (-0.10–0.14)-22.2-0.32–0.360.92 (0.68–0.98)
 Kurtosis-0.13 (0.49)0.02 (0.56)-0.76–0.86-0.15 (-0.44–0.13)-136-0.94–0.640.83 (0.37–0.96)
 5th percentile4.22 (0.26)4.16 (0.32)3.78–4.770.05 (-0.04–0.14)1.2-0.19–0.290.95 (0.82–0.99)
 10th percentile4.41 (0.25)4.36 (0.31)3.96–4.940.05 (-0.05–0.14)1.1-0.21–0.300.95 (0.80–0.99)
 90th percentile5.99 (0.16)5.96 (0.19)5.66–6.270.03 (-0.03–0.09)0.5-0.13–0.190.94 (0.79–0.99)
 95th percentile6.17 (0.16)6.14 (0.18)5.86–6.430.03 (-0.02–0.08)0.5-0.12–0.320.95 (0.81–0.99)
Grayscale
 Median67.19 (32.99)70.05 (36.06)30.67–151.52-2.86 (-8.00–2.28)-4.2-17.23–11.510.99 (0.96–1.00)
 Mean104.65 (26.87)107.50 (29.72)73.22–170.61-2.85 (-9.16–3.45)-1.3-20.48–14.770.98 (0.91–0.99)
 Standard deviation106.03 (17.46)107.19 (13.59)91.86–136.54-1.16 (-7.88–5.56)-1.1-19.95–17.640.91 (0.63–0.98)
 Skewness2.51 (0.88)2.52 (0.70)1.63–3.99-0.003 (-0.33–0.32)-0.1-0.91–0.900.92 (0.70–0.98)
 Kurtosis12.49 (11.67)12.84 (10.82)3.81–38.65-0.35 (-5.00–4.30)-2.8-13.35–12.650.92 (0.66–0.98)
 5th percentile16.59 (13.34)17.19 (14.43)7.17–54.29-0.60 (-1.97–0.76)-3.6-4.41–3.211.00 (0.98–1.00)
 10th percentile21.77 (16.88)22.96 (18.30)9.50–69.53-1.20 (-2.97–0.58)-5.4-6.15–3.760.99 (0.98–1.00)
 90th percentile239.64 (40.18)244.76 (38.86)195.65–319.26-5.12 (-19.99–9.75)-2.1-46.69–36.460.93 (0.73–0.98)
 95th percentile315.19 (44.98)320.95 (41.95)273.50–414.96-5.76 (-24.41–12.90)-1.8-57.91–46.390.91 (0.63–0.98)

N = 10. CI, confidence interval; SD, standard deviation; ICC, intra-class correlation coefficient.

Fig 2

Intra- and interobserver variability of measurement of thrombus attenuation in OCT images using image analysis software.

Scatterplot (left) and Bland-Altman plot (right) of intraobserver (A) and interobserver (B) comparison for median attenuation. OCT, optical coherence tomography; SD, standard deviation.

Intra- and interobserver variability of measurement of thrombus attenuation in OCT images using image analysis software.

Scatterplot (left) and Bland-Altman plot (right) of intraobserver (A) and interobserver (B) comparison for median attenuation. OCT, optical coherence tomography; SD, standard deviation. N = 10. CI, confidence interval; SD, standard deviation; ICC, intra-class correlation coefficient. N = 10. CI, confidence interval; SD, standard deviation; ICC, intra-class correlation coefficient.

Comparison of visual TAS and measurements by QCU-CMS software

There was an inverse relationship between several QCU-CMS parameters of attenuation (median values, r = -0.79, p<0.001); 10th percentile values r = -0.71, p<0.001), backscatter (mean values r = -0.78, p<0.001; 10th percentile value r = -0.79, p<0.001) and grayscale intensity (mean values r = -0.80, p<0.001; 10th percentile values r = -0-80, p<0.001)) and the visual TAS. Distribution of selected statistical parameters of attenuation, backscatter and grayscale intensity in the six-stage TAS is shown (Fig 3). For the distribution of the parameters using two-stage TAS, see S5 Fig. In these analyses, a minimum threshold 0.15 mm-1 was used for attenuation.
Fig 3

Relationship of thrombus attenuation score and parameters measured by image analysis software in thrombus areas in OCT images.

Scatterplots for median attenuation (A), 10th percentile of attenuation (B), mean backscatter (C), 10th percentile of backscatter (D), mean grayscale intensity (E) and 10th percentile of grayscale intensity (F). OCT, optical coherence tomography.

Relationship of thrombus attenuation score and parameters measured by image analysis software in thrombus areas in OCT images.

Scatterplots for median attenuation (A), 10th percentile of attenuation (B), mean backscatter (C), 10th percentile of backscatter (D), mean grayscale intensity (E) and 10th percentile of grayscale intensity (F). OCT, optical coherence tomography. The same thrombus cross-sectional areas were analyzed using a higher minimum threshold (0.5 mm-1) for attenuation. This resulted in exclusion of an average of 36% (minimum 5.3%, maximum 69.8%) of the pixels of each patient previously included with the 0.15 mm-1 threshold. In this analysis, only low percentiles of attenuation values (for 10th percentile values, r = 0.59, p<0.001) had an inverse relationship with visual TAS. Median (r = -0.05, p = 0.78) or 95th percentile values (r = 0.34, p = 0.07) of attenuation values were not statistically significantly related to TAS (S6 Fig). In addition, using ratio of attenuation and intensity measurements, a new ‘normalized’ variable was constructed. The ratio of 95th percentile of attenuation and median intensity had a direct relationship with TAS (r = 0.86, p<0.001) (S7 Fig).

ROC analysis

A ROC curve analysis was applied to assess the best parameters to distinguish between different types of thrombus. The AUCs represented the ability of QCU-CMS software parameters to differentiate red and white thrombi (Table 4). The best cutoff values to diagnose red thrombus in each class of variables were <5.35 for mean backscatter (AUC = 0.959, sensitivity 92%, specificity 94%), <120.1 for mean grayscale intensity (AUC = 0.959, sensitivity 85%, specificity 100%), <0.57 mm-1 for median attenuation (AUC = 0.919, sensitivity 100%, specificity 71%) and >0.022 mm-1 for 95th percentile of attenuation/median grayscale intensity (AUC = 0.950, sensitivity 88%, specificity 85%) (Fig 4). In an additional analysis, using binary logistic regression, best predictors of thrombus type were unchanged.
Table 4

Results of receiver operating characteristic (ROC) curve analysis of values determined by QCU-CMS software in prediction of thrombus type (white vs. red).

VariableAUC95% CISens (%)Spec (%)CP
Attenuation (mm-1)
 Median0.9190.83–1.00100710.568
 Mean0.8960.78–1.0085820.753
 5th percentile0.8960.78–1.0085880.246
 10th percentile0.8910.77–1.0077940.294
 90th percentile0.6790.48–0.8746881.384
 95th percentile0.5840.38–0.79100241.222
 Standard deviation0.6240.42–0.8385180.322
 Skewness0.1860.02–0.3585120.778
 Kurtosis0.3120.11–0.5285120.849
Backscatter
 Median0.9460.87–1.0092885.33
 Mean0.9590.89–1.0092945.35
 5th percentile0.9460.87–1.0092824.19
 10th percentile0.9550.89–1.00100774.36
 90th percentile0.8010.63–0.97621006.20
 95th percentile0.7100.51–0.9162886.33
 Standard deviation0.8370.67–1.0015770.694
 Skewness0.0720.00–0.17626-0.417
 Kurtosis0.8820.76–1.00691000.110
Grayscale intensity
 Median0.9410.86–1.001007757.11
 Mean0.9590.89–1.0085100120.1
 5th percentile0.9280.84–1.00928212.03
 10th percentile0.9370.85–1.00928215.95
 90th percentile0.9410.84–1.009294261.6
 95th percentile0.9000.78–1.009282337.6
 Standard deviation0.7560.58–0.946977106.7
 Skewness0.0810.00–0.188532.72
 Kurtosis0.1000.00–0.2388214.86
95th percentile of attenuation/median intensity (mm-1)0.9500.88–1.0088850.022

N = 30. AUC, area under curve; CI, confidence interval; CP, cut-off point defined by the largest value of J (Youden) index; Sens, sensitivity; Spec, specificity.

Fig 4

Receiver operating characteristic (ROC) curves of parameters determined by QCU-CMS software in prediction of thrombus type (white vs. red).

Different statistical parameters of attenuation (A), backscatter (B), grayscale intensity (C) and the ratio of 95th percentile of attenuation and median grayscale intensity (D). Perc, percentile.

Receiver operating characteristic (ROC) curves of parameters determined by QCU-CMS software in prediction of thrombus type (white vs. red).

Different statistical parameters of attenuation (A), backscatter (B), grayscale intensity (C) and the ratio of 95th percentile of attenuation and median grayscale intensity (D). Perc, percentile. N = 30. AUC, area under curve; CI, confidence interval; CP, cut-off point defined by the largest value of J (Youden) index; Sens, sensitivity; Spec, specificity. AUCs for different statistical variables of attenuation measurements from the second analysis using a higher 0.5 mm-1 threshold were low (in range of 0.50–0.60) except for 5th percentile and 10th percentile of attenuation (AUCs 0.878 and 0.851, respectively).

Discussion

As a first step in order to resolve the clinical significance and predictive potential of OCT-based thrombus analysis in clinical patients, a robust method to translate the traditional visual classification into reproducibly measured variables is needed. Validated methodology to assess different thrombus types by OCT imaging has not been available, and previously published data has relied on subjective analysis methods based on observers’ visual perception of the thrombus type [9,10]. In this methodological study, we measured attenuation, backscatter and grayscale intensity of in vivo thrombi using newly developed image analysis software features. For methodological validation, we assessed repeatability of the measurements and the correlation between traditional visual thrombus classification and the computer software-based analysis. We were able to demonstrate for the first time the use of image analysis software to assess in vivo thrombi in OCT images. Measurements of attenuation, backscatter and grayscale intensity of the thrombi using the software were highly repeatable and there was good agreement between two observer consensus visual evaluation of thrombus type and several parameters measured using the software on the OCT images. Visual assessment of thrombus type in OCT images is per se highly observer-dependent. There is not much previous data on the repeatability of the operator-based assessment of thrombus type and there is also a lack of standardized definitions and cut-off values differentiating red and white thrombus. We observed that the visual classification, as expected, was only moderately reproducible. Of note, the present analysis was performed off-line by experienced analysts and reproducibility outside the OCT core lab setting is most probably even lower. In the present study, we showed that computer-based image analysis of in vivo thrombus can be performed with good repeatability. The OCT-derived mean and 10th percentile backscatter and mean grayscale intensity values and the ratio of 95th percentile of attenuation to median grayscale intensity had excellent performance (AUC >0.95) in discriminating red from white thrombus. In addition, median values of attenuation, backscatter and grayscale intensity, low percentile values of the same variables and high percentile values of grayscale intensity all had good performance (AUC>0.90). Earlier histopathological studies of aspirated coronary thrombi have provided important insight to the composition of the thrombus [11,12]. Specimens retrieved by aspiration devices, however, do not necessarily represent the whole thrombotic burden in a patient with STEMI. Some of the thrombotic material stays attached to the vessel wall and may differ from the aspirate. Further, the structure of thrombus is easily distorted and the material fragmented during aspiration. A thorough analysis of aspirated thrombus fragments would be highly laborious and as such subject to a variety of sources of error. The three-dimensional structure of the in vivo thrombus is lost and the co-localization of thrombus and the culprit lesion is not possible. The ability of intravascular OCT to provide high-resolution images of thrombus in vivo during angiography and PCI procedures [13] makes it a robust tool to study thrombus. A further advantage of OCT imaging in comparison to histological analysis of thrombus aspirate is that the images are available for intraoperative decision-making. Considering the frequent occurrence of thrombus in OCT imaging of clinical patients, especially in case of acute coronary syndromes, the data on thrombus image analysis is surprisingly scarce. In a single cadaver series, OCT light attenuation characteristics have been validated against histology to classify thrombi as red or white. When the authors applied a cut-off value defined in that study, OCT had good sensitivity and specificity for differentiating red and white thrombi. [3] In addition to a case report [14] the method has not been used successively. The goal of analyzing thrombus type in vivo and gaining insight to the pathogenesis of acute myocardial infarction is to facilitate the development of novel therapies and to tailor the intervention for each patient. There is some evidence that aspirates consisting mainly of white thrombus might be predictive of lower mortality after the infarction [15]. Presence of older thrombus in even 40% of thrombus aspirates of acute STEMI patients in a large series [16] supports the concept that the acute closure of the culprit artery is preceded by days or weeks of thrombotic process. Interestingly, histological signs of older thrombus have been associated with higher mortality after the myocardial infarction [16]. In addition, red thrombus in the aspirate has been suggested to associate with more frequent angiographically visible distal embolization [17]. Development of reliable image analysis methodology is essential to enable investigation of thrombus types in vivo. In the future, development of smaller size imaging catheters may reduce the need for routine thrombus aspiration prior to OCT imaging of highly thrombotic lesions, such as those in STEMI patients. In a recent in vitro study, the correlation between red blood cell concentration and attenuation or backscatter of the OCT signal was shown to have a limit, above which the correlation was not seen [18]. In our series of in vivo OCT imaging, we did not observe any saturation threshold level for these variables. A feasible explanation for this might be that the red blood cell concentration achieved in an experimental model is higher than in thrombus in vivo, which contains other components as well. Unexpectedly, in the present study we were not able to demonstrate a higher actual attenuation in red than white thrombus using the software algorithm Instead, there was an inverse relationship between attenuation values and red thrombus, including high percentiles of attenuation. Even after setting a higher threshold for minimum attenuation, which removed all ‘dark’ areas in the abluminal side of the thrombus when strong signal attenuation was present, a positive correlation was not observed. This seems to be due to high heterogeneity of analyzed thrombus regions. In red thrombi, there are highly attenuating parts, but these are also observed in the white thrombus. We tested different parameters measured by the software from the thrombus images. Attenuation is a parameter, which describes the relative decrease in the light intensity per distance travelled in the tissue. This is different from backscatter and grayscale intensity, which both describe the absolute values of light properties. In the present series, we observed that grayscale intensity performed as well as attenuation in differentiating thrombus types. There were also no differences in observer variation between attenuation and grayscale intensity parameters. An explanation for this finding may be the high technical quality of the patient subset selected for the analysis. In presence of more blood speckle i.e. residual artefactual blood in the image frames, attenuation is expected to be a more stable parameter. On the other hand, intraluminal thrombus is in most occasions located very proximally to the imaging catheter. This may render thrombus analysis less vulnerable to artefacts, such as small quantities of artefactual blood in the OCT recording. The tested image analysis methodology relies on automatic measurement of attenuation and backscatter along the A-lines of the OCT image. The analysis is not automatic, and user-dependent interaction is needed, because the software does not integrate global information of the cross-section, considering artefacts and thrombus-like structures. Although it would be possible to use image analysis algorithms to determine most of the borders of thrombus cross-sections, in most cases some manual correction would be needed, especially for the red thrombus where no abluminal border is visible. For that reason and to simplify the testing of observer variation, in this study we decided to use manual tracing of the thrombus cross-sections. In addition, we applied a minimum threshold value for attenuation, which excludes noise pixels and dark regions accidentally included in the regions of interest. By using this threshold, we also aimed at avoiding inclusion of non-informative pixels in the abluminal side of the thrombus, where OCT signal has attenuated to a minimum. For further development of the methodology, it can be possible to aim for a semi-automatic analysis method, starting from delineation of intraluminal space followed by automatic detection of luminal border of the thrombus with necessary manual corrections.

Limitations

The current study has a number of important limitations. First, the thrombus regions were manually delineated. To reduce user-dependence, we applied a minimal threshold value to attenuation, which effectively delineated the endoluminal border of the thrombus. The abluminal border of the thrombus area, especially in the case of red thrombi was, however, partly dependent on manual tracing. Second, since this was an in vivo study including clinical patients, no histological golden standard could be applied to the ROC curve analyses. This prevents us from establishing a direct correlation between the software-derived parameters and biological thrombus contents. The basic principle of increased light attenuation by increased red blood cell content of the thrombus has, however, been validated earlier [3]. Third, the current study was not designed to assess the real-world reproducibility, but was limited to a relatively small sample of selected cases. We focused on improving the reproducibility of thrombus type analysis by the use of computer-based methodology. More data is needed in order to evaluate the true absolute reproducibility.

Conclusion

Subjective thrombus type assessment in OCT images can be complemented by less user-dependent computer-based image analysis methodology. Computer-based measurement of attenuation, backscatter and grayscale intensity of thrombi is repeatable and can be used to differentiate red and white thrombi with high sensitivity and specificity. This study stimulates further studies with larger sample size to confirm the results and their clinical value.

Patient flow chart of the study.

OCT, optical coherence tomography. (TIF) Click here for additional data file.

Qualitative thrombus classification.

Consensus classification by two observers is shown. Thrombus attenuation score (TAS) from 1 to 3 equals white thrombus and from 4 to 6 red thrombus. (TIF) Click here for additional data file.

Intra- and interobserver variability of measurement of thrombus backscatter in OCT images using image analysis software.

Scatterplot (left) and Bland-Altman plot (right) of intraobserver (A) and interobserver (B) comparison for mean backscatter. OCT, optical coherence tomography; SD, standard deviation. (TIF) Click here for additional data file.

Intra- and interobserver variability of measurement of thrombus grayscale intensity in OCT images using image analysis software.

Scatterplot and Bland-Altman plot of intraobserver (A) and interobserver (B) comparison for mean grayscale intensity. OCT, optical coherence tomography; SD, standard deviation. (TIF) Click here for additional data file.

Relationship of binary thrombus type and parameters measured by image analysis software in thrombus areas in OCT images.

Scatterplots for median attenuation (A), 10th percentile of attenuation (B), mean backscatter (C), 10th percentile of backscatter (D), mean grayscale intensity (E) and 10th percentile of grayscale intensity (F). OCT, optical coherence tomography. (TIF) Click here for additional data file.

Relationship of thrombus type and parameters measured by image analysis software in thrombus areas in OCT images using a higher, 0.5 mm-1 threshold for minimal attenuation.

Scatterplots for median attenuation (A,D), 10th percentile of attenuation (B,E) and 95th percentile of attenuation (C,F). Thrombus attenuation score in the upper panel and bivariate thrombus type in the lower panel. OCT, optical coherence tomography. (TIF) Click here for additional data file.

Relationship of thrombus type and the ratio of 95th percentile of attenuation and median grayscale intensity.

Thrombus attenuation score (A) and binary thrombus type (B). (TIF) Click here for additional data file.
  18 in total

1.  Consensus standards for acquisition, measurement, and reporting of intravascular optical coherence tomography studies: a report from the International Working Group for Intravascular Optical Coherence Tomography Standardization and Validation.

Authors:  Guillermo J Tearney; Evelyn Regar; Takashi Akasaka; Tom Adriaenssens; Peter Barlis; Hiram G Bezerra; Brett Bouma; Nico Bruining; Jin-man Cho; Saqib Chowdhary; Marco A Costa; Ranil de Silva; Jouke Dijkstra; Carlo Di Mario; Darius Dudek; Darius Dudeck; Erling Falk; Erlin Falk; Marc D Feldman; Peter Fitzgerald; Hector M Garcia-Garcia; Hector Garcia; Nieves Gonzalo; Juan F Granada; Giulio Guagliumi; Niels R Holm; Yasuhiro Honda; Fumiaki Ikeno; Masanori Kawasaki; Janusz Kochman; Lukasz Koltowski; Takashi Kubo; Teruyoshi Kume; Hiroyuki Kyono; Cheung Chi Simon Lam; Guy Lamouche; David P Lee; Martin B Leon; Akiko Maehara; Olivia Manfrini; Gary S Mintz; Kyiouchi Mizuno; Marie-angéle Morel; Seemantini Nadkarni; Hiroyuki Okura; Hiromasa Otake; Arkadiusz Pietrasik; Francesco Prati; Lorenz Räber; Maria D Radu; Johannes Rieber; Maria Riga; Andrew Rollins; Mireille Rosenberg; Vasile Sirbu; Patrick W J C Serruys; Kenei Shimada; Toshiro Shinke; Junya Shite; Eliot Siegel; Shinjo Sonoda; Shinjo Sonada; Melissa Suter; Shigeho Takarada; Atsushi Tanaka; Mitsuyasu Terashima; Troels Thim; Thim Troels; Shiro Uemura; Giovanni J Ughi; Heleen M M van Beusekom; Antonius F W van der Steen; Gerrit-Anne van Es; Gerrit-Ann van Es; Gijs van Soest; Renu Virmani; Sergio Waxman; Neil J Weissman; Giora Weisz
Journal:  J Am Coll Cardiol       Date:  2012-03-20       Impact factor: 24.094

2.  Evaluation of the characterization of thrombi in vitro by optical coherence tomography.

Authors:  Guang Yang; Xiaoyuan Zhang; Qingyang Meng; Dan Ma; Huai Yu; Xuefeng Ren; Bo Lv; Shaosong Zhang; Shuang Yang; Bo Yu
Journal:  Int J Cardiol       Date:  2016-06-15       Impact factor: 4.164

3.  Statistical methods for assessing agreement between two methods of clinical measurement.

Authors:  J M Bland; D G Altman
Journal:  Lancet       Date:  1986-02-08       Impact factor: 79.321

4.  Randomized trial of primary PCI with or without routine manual thrombectomy.

Authors:  Sanjit S Jolly; John A Cairns; Salim Yusuf; Brandi Meeks; Janice Pogue; Michael J Rokoss; Sasko Kedev; Lehana Thabane; Goran Stankovic; Raul Moreno; Anthony Gershlick; Saqib Chowdhary; Shahar Lavi; Kari Niemelä; Philippe Gabriel Steg; Ivo Bernat; Yawei Xu; Warren J Cantor; Christopher B Overgaard; Christoph K Naber; Asim N Cheema; Robert C Welsh; Olivier F Bertrand; Alvaro Avezum; Ravinay Bhindi; Samir Pancholy; Sunil V Rao; Madhu K Natarajan; Jurriën M ten Berg; Olga Shestakovska; Peggy Gao; Petr Widimsky; Vladimír Džavík
Journal:  N Engl J Med       Date:  2015-03-16       Impact factor: 91.245

5.  Tissue characterization with depth-resolved attenuation coefficient and backscatter term in intravascular optical coherence tomography images.

Authors:  Shengnan Liu; Yohei Sotomi; Jeroen Eggermont; Gaku Nakazawa; Sho Torii; Takeshi Ijichi; Yoshinobu Onuma; Patrick W Serruys; Boudewijn P F Lelieveldt; Jouke Dijkstra
Journal:  J Biomed Opt       Date:  2017-09       Impact factor: 3.170

6.  Eroded Versus Ruptured Plaques at the Culprit Site of STEMI: In Vivo Pathophysiological Features and Response to Primary PCI.

Authors:  Francesco Saia; Kenichi Komukai; Davide Capodanno; Vasile Sirbu; Giuseppe Musumeci; Giacomo Boccuzzi; Giuseppe Tarantini; Massimo Fineschi; Gabriele Tumminello; Chiara Bernelli; Giampaolo Niccoli; Micol Coccato; Barbara Bordoni; Hiram Bezerra; Giuseppe Biondi-Zoccai; Renu Virmani; Giulio Guagliumi
Journal:  JACC Cardiovasc Imaging       Date:  2015-04-15

Review 7.  Thrombus aspiration therapy and coronary thrombus components in patients with acute ST-elevation myocardial infarction.

Authors:  Kei Yunoki; Takahiko Naruko; Kenichi Sugioka; Mayumi Inaba; Akira Itoh; Kazuo Haze; Minoru Yoshiyama; Makiko Ueda
Journal:  J Atheroscler Thromb       Date:  2013-04-30       Impact factor: 4.928

8.  Red versus white thrombi in patients with ST-elevation myocardial infarction undergoing primary percutaneous coronary intervention: clinical and angiographic outcomes.

Authors:  Alexandre S Quadros; Eduardo Cambruzzi; Juliana Sebben; Renato B David; Anibal Abelin; Dulce Welter; Rogério Sarmento-Leite; Rajendra H Mehta; Carlos A Gottschall; Renato D Lopes
Journal:  Am Heart J       Date:  2012-10       Impact factor: 4.749

9.  Design and rationale of the TOTAL trial: a randomized trial of routine aspiration ThrOmbecTomy with percutaneous coronary intervention (PCI) versus PCI ALone in patients with ST-elevation myocardial infarction undergoing primary PCI.

Authors:  Sanjit S Jolly; John Cairns; Salim Yusuf; Brandi Meeks; Olga Shestakovska; Lehana Thabane; Kari Niemelä; Philippe Gabriel Steg; Olivier F Bertrand; Sunil V Rao; Alvaro Avezum; Warren J Cantor; Samir B Pancholy; Raul Moreno; Anthony Gershlick; Ravinay Bhindi; Robert C Welsh; Asim N Cheema; Shahar Lavi; Michael Rokoss; Vladimír Džavík
Journal:  Am Heart J       Date:  2013-12-14       Impact factor: 4.749

Review 10.  Expert review document on methodology, terminology, and clinical applications of optical coherence tomography: physical principles, methodology of image acquisition, and clinical application for assessment of coronary arteries and atherosclerosis.

Authors:  Francesco Prati; Evelyn Regar; Gary S Mintz; Eloisa Arbustini; Carlo Di Mario; Ik-Kyung Jang; Takashi Akasaka; Marco Costa; Giulio Guagliumi; Eberhard Grube; Yukio Ozaki; Fausto Pinto; Patrick W J Serruys
Journal:  Eur Heart J       Date:  2009-11-04       Impact factor: 29.983

View more
  3 in total

1.  MRI, CT and high resolution macro-anatomical images with cryosectioning of a Beagle brain: Creating the base of a multimodal imaging atlas.

Authors:  Kálmán Czeibert; Gábor Baksa; András Grimm; Szilvia Anett Nagy; Enikő Kubinyi; Örs Petneházy
Journal:  PLoS One       Date:  2019-03-07       Impact factor: 3.240

Review 2.  Advantages and prospects of optical coherence tomography in interventional therapy of coronary heart disease (Review).

Authors:  Jie Wang; Shuo Yuan; Jingjing Qi; Qinggao Zhang; Zheng Ji
Journal:  Exp Ther Med       Date:  2022-02-02       Impact factor: 2.447

Review 3.  Molecular Imaging and Non-molecular Imaging of Atherosclerotic Plaque Thrombosis.

Authors:  Bingchen Guo; Zhaoyue Li; Peiyang Tu; Hao Tang; Yingfeng Tu
Journal:  Front Cardiovasc Med       Date:  2021-07-05
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.