Francesca Giuffrida1, Sean Austin2, Denis Cuany2, Belén Sanchez-Bridge2, Karin Longet2, Emmanuelle Bertschy2, Julien Sauser2, Sagar K Thakkar2, Le Ye Lee3,4, Michael Affolter2.
Abstract
OBJECTIVE: The study aims at evaluating mid-infrared human milk analyzer (HMA) accuracy and precision, in human milk (HM). STUDYEntities:
Mesh:
Substances:
Year: 2018 PMID: 30552375 PMCID: PMC6514977 DOI: 10.1038/s41372-018-0291-8
Source DB: PubMed Journal: J Perinatol ISSN: 0743-8346 Impact factor: 2.521
Elution program for determination of lactose
| Time (min) | Flow rate (mL min−1) | Eluent ‘A’ (%) | Eluent ‘B’ (%) | Eluent ‘C’ (%) |
|---|---|---|---|---|
| 0.0 | 0.5 | 2.0 | 98.0 | 0.0 |
| 1.0 | 0.5 | 2.0 | 98.0 | 0.0 |
| 12.0 | 0.5 | 5.0 | 95.0 | 0.0 |
| 21.0 | 0.5 | 22.4 | 65.6 | 12.0 |
| 21.1 | 0.5 | 0.0 | 0.0 | 100.0 |
| 26.0 | 0.5 | 0.0 | 0.0 | 100.0 |
| 26.1 | 0.5 | 100.0 | 0.0 | 0.0 |
| 31.0 | 0.5 | 100.0 | 0.0 | 0.0 |
| 31.1 | 0.5 | 2.0 | 98.0 | 0.0 |
| 37.0 | 0.5 | 2.0 | 98.0 | 0.0 |
Fig. 1Total fat content in fully expressed HM samples collected at 30, 60 and 120 days postpartum, analyzed by MIRIS HMA. Error bars represent standard deviation
Fig. 2Total lactose content in fully expressed HM samples collected at 30, 60 and 120 days postpartum, analyzed by MIRIS HMA. Error bars represent standard deviation
Fig. 3Total protein content in fully expressed HM samples collected at 30, 60 and 120 days postpartum, analyzed by MIRIS HMA and BCA assay. Circles represent BCA results and squares HMA results. Error bars represent standard deviation
Fig. 4Correlation between total protein content in individual fully expressed HM samples collected at 30, 60 and 120 days postpartum, analyzed by MIRIS HMA and BCA assay