Assessment of effective radiation dose of an extremity CBCT, MSCT and conventional X ray for knee area using MOSFET dosemeters.

Juha Koivisto1, Timo Kiljunen, Jan Wolff, Mika Kortesniemi.   

Abstract

The objective of this study was to assess and compare the organ and effective doses in the knee area resulting from different commercially available multislice computed tomography devices (MSCT), one cone beam computed tomography device (CBCT) and one conventional X-ray radiography device using MOSFET dosemeters and an anthropomorphic RANDO knee phantom. Measurements of the MSCT devices resulted in effective doses ranging between 27 and 48 µSv. The CBCT measurements resulted in an effective dose of 12.6 µSv. The effective doses attained using the conventional radiography device were 1.8 µSv for lateral and 1.2 µSv for anterior-posterior projections. The effective dose resulting from conventional radiography was considerably lower than those recorded for the CBCT and MSCT devices. The MSCT effective dose results were two to four times higher than those measured on the CBCT device. This study demonstrates that CBCT can be regarded as a potential low-dose 3D imaging technique for knee examinations.

Mesh:

Year:  2013        PMID: 23825221     DOI: 10.1093/rpd/nct162

Source DB:  PubMed          Journal:  Radiat Prot Dosimetry        ISSN: 0144-8420            Impact factor:   0.972


  16 in total

1.  Radiographical measurements for distal intra-articular fractures of the radius using plain radiographs and cone beam computed tomography images.

Authors:  Nora Suojärvi; T Sillat; N Lindfors; S K Koskinen
Journal:  Skeletal Radiol       Date:  2015-08-14       Impact factor: 2.199

2.  Using cone-beam CT as a low-dose 3D imaging technique for the extremities: initial experience in 50 subjects.

Authors:  Ambrose J Huang; Connie Y Chang; Bijoy J Thomas; Peter J MacMahon; William E Palmer
Journal:  Skeletal Radiol       Date:  2015-02-05       Impact factor: 2.199

3.  [Design and optimization of a cone-beam CT system for extremity imaging].

Authors:  Kun Ma; Mingqiang Li; Xi Tao; Dong Zeng; Yongbo Wang; Zhaoying Bian; Ziquan Wei; Gaofeng Chen; Qianjin Feng; Jianhua Ma; Jing Huang
Journal:  Nan Fang Yi Ke Da Xue Xue Bao       Date:  2018-11-30

4.  The role of cone beam CT in the study of symptomatic total knee arthroplasty (TKA): a 20 cases report.

Authors:  Cosimo Nardi; Roberto Buzzi; Roberto Molteni; Carlo Cossi; Chiara Lorini; Linda Calistri; Stefano Colagrande
Journal:  Br J Radiol       Date:  2017-05-25       Impact factor: 3.039

Review 5.  Radiation dose in non-dental cone beam CT applications: a systematic review.

Authors:  Cosimo Nardi; Sergio Salerno; Roberto Molteni; Mariaelena Occhipinti; Giulia Grazzini; Niccolò Norberti; Cesare Cordopatri; Stefano Colagrande
Journal:  Radiol Med       Date:  2018-06-05       Impact factor: 3.469

6.  Reduction of radiation exposure by lead curtain shielding in dedicated extremity cone beam CT.

Authors:  C-H Lee; J H Ryu; Y-H Lee; K-H Yoon
Journal:  Br J Radiol       Date:  2015-03-26       Impact factor: 3.039

7.  Effective dose assessment in the maxillofacial region using thermoluminescent (TLD) and metal oxide semiconductor field-effect transistor (MOSFET) dosemeters: a comparative study.

Authors:  J Koivisto; D Schulze; J Wolff; D Rottke
Journal:  Dentomaxillofac Radiol       Date:  2014-08-21       Impact factor: 2.419

8.  Initial evaluation of image performance of a 3-D x-ray system: phantom-based comparison of 3-D tomography with conventional computed tomography.

Authors:  Robyn Melanie Benz; Meritxell Alzamora Garcia; Felix Amsler; Johannes Voigt; Andreas Fieselmann; Anna Lucja Falkowski; Bram Stieltjes; Anna Hirschmann
Journal:  J Med Imaging (Bellingham)       Date:  2018-03-06

9.  Cone-Beam CT in diagnosis of scaphoid fractures.

Authors:  Rolf Edlund; Mikael Skorpil; Gunilla Lapidus; Jenny Bäcklund
Journal:  Skeletal Radiol       Date:  2015-11-12       Impact factor: 2.199

Review 10.  [Digital volume tomography : Dedicated scanner and cone beam CT with C‑arm systems].

Authors:  M Fiebich; D Weber
Journal:  Radiologe       Date:  2018-03       Impact factor: 0.635

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.