| Literature DB >> 30453842 |
G Carl1,2, D Reitz1,2, S Schönecker1, M Pazos1, P Freislederer1, M Reiner1, F Alongi3,4, M Niyazi1, U Ganswindt1,5, C Belka1, S Corradini1.
Abstract
PURPOSE/Entities:
Keywords: CBCT; catalyst; laser positioning; patient positioning; radiotherapy; surface scanner
Mesh:
Year: 2018 PMID: 30453842 PMCID: PMC6243634 DOI: 10.1177/1533033818806002
Source DB: PubMed Journal: Technol Cancer Res Treat ISSN: 1533-0338
Figure 1.The 3 Catalyst HD scanners in the treatment room. The scanners are arranged at an angle of 120° to each other in order to enable a continuous detection of the surface even if the gantry is rotating around the patient.
Figure 2.Positional deviations projected onto the patients’ surface (screenshot of Catalyst software c4D [C-Rad AB). Local deviations outside the tolerance limits compared to the reference scan are marked in red.
Figure 3.Daily workflow.
Figure 4.Histograms showing distributions of Catalyst and laser deviations along all 3 spatial directions.
Catalyst and Spatial Laser Deviations Sorted by Region.
| All Values in mm | Mean Value | 95% CI | SD | Range |
|---|---|---|---|---|
| Catalyst deviations sorted by therapeutic regions | ||||
|
| ||||
| Catalyst lat. | −0.21 | [−0.37; −0.05] | ±2.14 | −15.4; 11.8 |
| Laser lat. | −0.18 | [−0.31; −0.06] | ±1.71 | −5.3; 5.1 |
| Catalyst long. | 0.17 | [0.02; 0.33] | ±2.11 | −9.7; 9.1 |
| Laser long. | 0.29 | [0.14; 0.43] | ±1.89 | −29.6; 8.0 |
| Catalyst vert. | 0.13 | [−0.08; 0.35] | ±2.88 | −11.0; 15.3 |
| Laser vert. | −0.03 | [−0.20; 0.14] | ±2.27 | −10.1; 8.6 |
|
| ||||
| Catalyst lat. | −0.13 | [−0.55; 0.30] | ±4.67 | −18.1; 15.3 |
| Laser lat. | −0.21 | [−0.58; 0.16] | ±4.06 | −18.7; 15.1 |
| Catalyst long. | −1.14 | [−1.57; −0.71] | ±4.66 | −15.6; 16.8 |
| Laser long. | −0.65 | [−0.98; −0.31] | ±3.65 | −11.2; 12.6 |
| Catalyst vert. | −0.01 | [−0.44; 0.41] | ±4.64 | −18.9; 14.0 |
| Laser vert. | 0.22 | [−0.16; 0.61] | ±4.18 | −19.7; 14.1 |
|
| ||||
| Catalyst lat. | −0.46 | [−0.84; −0.08] | ±4.85 | −16.6; 12.9 |
| Laser lat. | −0.06 | [−0.37; 0.26] | ±4.03 | −13.8; 11.2 |
| Catalyst long. | 1.99 | [1.58; 2.40] | ±5.25 | −15.2; 23.0 |
| Laser long. | 1.56 | [1.23; 1.89] | ±4.18 | −29.1; 22.5 |
| Catalyst vert. | −0.59 | [−1.03; −0.16] | ±5.58 | −18.8; 19.8 |
| Laser vert. | 0.68 | [0.37; 0.98] | ±3.93 | −10.5; 13.8 |
|
| ||||
| Catalyst lat. | 1.71 | [0.55; 2.87] | ±6.50 | −17.1; 16.5 |
| Laser lat. | 1.18 | [0.18; 2.19] | ±5.63 | −12.5; 14.4 |
| Catalyst long. | 0.18 | [−0.96; 1.32] | ±6.38 | −17.4; 16.9 |
| Laser long. | 0.96 | [0.26; 1.66] | ±3.90 | −10.4; 11.2 |
| Catalyst vert. | −1.57 | [−2.52; −0.62] | ±5.34 | −16.6; 10.9 |
| Laser vert. | −1.20 | [−1.80; −0.59] | ±3.38 | −14.0; 6.0 |
Abbreviations: CI, confidence interval; SD, standard deviation; lat., lateral; long., longitudinal; vert., vertical.
Figure 5.Catalyst and laser deviations sorted by region along the 3 directions.
Two One-Sided t Tests for Equivalence Results (P Value) Sorted by Region and Spatial Axes.
| All Values in mm | Test Boundaries | 90% CI of the Differences of Mean Value |
|
|---|---|---|---|
| TOST (2 one-sided | |||
|
| |||
| Lateral | −0.5; 0.5 | [−0.140; 0.203] | <.001 |
| Longitudinal | −0.5; 0.5 | [−0.065; 0.290] | .000 |
| Vertical | −0.5; 0.5 | [−0.396; 0.064] | .008 |
|
| |||
| Lateral | −1.0; 1.0 | [−0.560; 0.391] | .001 |
| Longitudinal | −1.0; 1.0 | [0.036; 0.946] | .033 |
| Vertical | −1.0; 1.0 | [−0.246; 0.713] | .004 |
|
| |||
| Lateral | −1.5; 1.5 | [−0.009; 0.819] | <.001 |
| Longitudinal | −1.5; 1.5 | [−0.866; 0.014] | <.001 |
| Vertical | −1.5; 1.5 | [0.821; 1.716] | .198 |
|
| |||
| Lateral | −2.0; 2.0 | [−1.809; 0.752] | .029 |
| Longitudinal | −2.0; 2.0 | [−0.333; 1.893] | .036 |
| Vertical | −2.0; 2.0 | [−0.567; 1.315] | .002 |
Abbreviation: CI, confidence interval.