Accuracy of a 3D laser∕camera surface imaging system for setup verification of the pelvic and thoracic regions in radiotherapy treatments.

Stefania Pallotta1, Gabriele Simontacchi, Livia Marrazzo, Marco Ceroti, Fabiola Paiar, Giampaolo Biti, Marta Bucciolini.   

Abstract

PURPOSE: The aim of this paper is to investigate the accuracy of a laser∕camera surface imaging system (Sentinel) when used on patients treated in the thorax or pelvic regions and to evaluate system employment conditions and patient setup procedures that provide more accurate results.
METHODS: The system was tested on two groups of patients for whom different patient setup procedures and Sentinel employment conditions were considered. For the first group of 33 patients (FG) no changes to the usual setup procedures were made and a surface extension limited to the treated region was considered. For the second group of 14 patients (SG) the reproducibility of external body surfaces, including body parts not in close proximity to the treatment site, was optimized and a wider surface was captured. In all cases the system accuracy was evaluated comparing registration results from concurrent Sentinel and cone beam CT (CBCT) acquisitions for a total of 192 occasions. External body surfaces, extracted from planning CT studies, were used as reference in both cases, but for SG also surface data captured by Sentinel system at the first treatment were employed.
RESULTS: In the 90th percentile of the distributions reporting CBCT and Sentinel registration parameters, absolute differences for FG were less than 6.4 mm and 3.8°. Better performances were observed for SG (≤5.7 mm and 2°). Mean absolute differences between three translation and three rotation parameters of CBCT and Sentinel were: less than 3.5 mm and 2.1° and 3.7 mm and 1.3° in FG for thorax and pelvis, respectively, and less than 2.8 mm and 1° and 2.7 mm and 0.9° for pelvis and thorax, respectively, in SG. No advantage in considering surface data captured by Sentinel as a reference instead of the surface extracted from the planning CT was observed.
CONCLUSIONS: The accuracy of Sentinel system in detecting errors is influenced by the extension and reliability of the surface used. When the reproducibility of external body surfaces was optimized differences between CBCT and Sentinel registration parameters resulted less than 5.7 mm and 2° in the 90% of the pelvis and thorax considered cases. No advantage in considering a Sentinel acquisition as reference was observed.

Entities:  

Mesh:

Year:  2013        PMID: 23298080     DOI: 10.1118/1.4769428

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  8 in total

1.  Surface imaging, portal imaging, and skin marker set-up vs. CBCT for radiotherapy of the thorax and pelvis.

Authors:  Stefania Pallotta; Eleonora Vanzi; Gabriele Simontacchi; Livia Marrazzo; Marco Ceroti; Fabiola Paiar; Lorenzo Livi; Marta Bucciolini
Journal:  Strahlenther Onkol       Date:  2015-06-19       Impact factor: 3.621

2.  Accuracy of surface registration compared to conventional volumetric registration in patient positioning for head-and-neck radiotherapy: a simulation study using patient data.

Authors:  Youngjun Kim; Ruijiang Li; Yong Hum Na; Rena Lee; Lei Xing
Journal:  Med Phys       Date:  2014-12       Impact factor: 4.071

3.  A novel surface imaging system for patient positioning and surveillance during radiotherapy. A phantom study and clinical evaluation.

Authors:  F Stieler; F Wenz; M Shi; F Lohr
Journal:  Strahlenther Onkol       Date:  2013-09-27       Impact factor: 3.621

4.  Development and clinical evaluation of a simple optical method to detect and measure patient external motion.

Authors:  Benigno Barbés; Juan Diego Azcona; Elena Prieto; José Manuel de Foronda; Marina García; Javier Burguete
Journal:  J Appl Clin Med Phys       Date:  2015-09-08       Impact factor: 2.102

Review 5.  AAPM task group report 302: Surface-guided radiotherapy.

Authors:  Hania A Al-Hallaq; Laura Cerviño; Alonso N Gutierrez; Amanda Havnen-Smith; Susan A Higgins; Malin Kügele; Laura Padilla; Todd Pawlicki; Nicholas Remmes; Koren Smith; Xiaoli Tang; Wolfgang A Tomé
Journal:  Med Phys       Date:  2022-03-15       Impact factor: 4.506

6.  Setup error assessment based on "Sphere-Mask" Optical Positioning System: Results from a multicenter study.

Authors:  Yan Zhang; Han Zhou; Kaiyue Chu; Chuanfeng Wu; Yun Ge; Guoping Shan; Jundong Zhou; Jing Cai; Jianhua Jin; Weiyu Sun; Ying Chen; Xiaolin Huang
Journal:  Front Oncol       Date:  2022-10-04       Impact factor: 5.738

7.  Delivery efficiency of an Elekta linac under gated operation.

Authors:  Guoqiang Cui; David J Housley; Fan Chen; Vivek K Mehta; David M Shepard
Journal:  J Appl Clin Med Phys       Date:  2014-09-08       Impact factor: 2.102

8.  Optical Surface Scanning for Patient Positioning in Radiation Therapy: A Prospective Analysis of 1902 Fractions.

Authors:  G Carl; D Reitz; S Schönecker; M Pazos; P Freislederer; M Reiner; F Alongi; M Niyazi; U Ganswindt; C Belka; S Corradini
Journal:  Technol Cancer Res Treat       Date:  2018-01-01
  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.