Clinical evaluation of interfractional variations for whole breast radiotherapy using 3-dimensional surface imaging.

Amish P Shah1, Tomas Dvorak2, Michael S Curry2, Daniel J Buchholz2, Sanford L Meeks2.   

Abstract

PURPOSE: To evaluate the impact of 3-dimensional (3D) surface imaging on daily patient setup for breast radiotherapy.
MATERIALS AND METHODS: Fifty patients undergoing treatment for whole breast radiotherapy were setup daily using an AlignRT system (VisionRT, London, UK) for 3D surface-based alignment. Daily alignments were performed against a reference surface topogram and shifts from skin marks were recorded daily. This investigation evaluated the following: (1) the performance of the surface-based imaging system for daily breast alignment; (2) the absolute displacements between setup with skin marks and setup with the surface-based imaging system; and (3) the dosimetric effect of daily alignments with skin marks versus surface-based alignments.
RESULTS: Displacements from 1258 treatment fractions were analyzed. Sixty percent of those fractions (749) were reviewed against MV portal imaging in order to assess the performance of the AlignRT system. Daily setup errors were given as absolute displacements, comparing setup marks against shifts determined using the surface-based imaging system. Averaged over all patients, the mean displacements were 4.1 ± 2.6 mm, 2.7 ± 1.4 mm, and 2.6 ± 1.2 mm in the anteroposterior (AP), superoinferior (S/I), and left-right (L/R) directions, respectively. Furthermore, the standard deviation of the random error (σ) was 3.2 mm, 2.2 mm, and 2.2 mm in the A/P, S/I, and L/R directions, respectively.
CONCLUSIONS: Daily alignment with 3D surface imaging was found to be valuable for reducing setup errors when comparing with patient alignment from skin marks. The result of the surface-based alignments specifically showed that alignment with skin marks was noticeably poor in the anteroposterior directions. The overall dosimetric effect of the interfractional variations was small, but these variations showed a potential for increased dose deposition to both the heart and lung tissues. Although these interfractional variations would not negatively affect the quality of patient care for whole breast radiotherapy, it may require an increase in PTV margin, especially in cases of partial breast irradiation.
Copyright © 2013 American Society for Radiation Oncology. Published by Elsevier Inc. All rights reserved.

Entities:  

Year:  2012        PMID: 24674259     DOI: 10.1016/j.prro.2012.03.002

Source DB:  PubMed          Journal:  Pract Radiat Oncol        ISSN: 1879-8500


  37 in total

Review 1.  Optical and x-ray technology synergies enabling diagnostic and therapeutic applications in medicine.

Authors:  Brian W Pogue; Brian C Wilson
Journal:  J Biomed Opt       Date:  2018-10       Impact factor: 3.170

2.  Dosimetric Considerations in Respiratory-Gated Deep Inspiration Breath-Hold for Left Breast Irradiation.

Authors:  Steve Walston; Allison M Quick; Karla Kuhn; Yi Rong
Journal:  Technol Cancer Res Treat       Date:  2016-07-08

3.  A robust real-time surface reconstruction method on point clouds captured from a 3D surface photogrammetry system.

Authors:  Wenyang Liu; Yam Cheung; Amit Sawant; Dan Ruan
Journal:  Med Phys       Date:  2016-05       Impact factor: 4.071

4.  Setup accuracy for prone and supine whole breast irradiation.

Authors:  Thomas Mulliez; Akos Gulyban; Tom Vercauteren; Annick van Greveling; Bruno Speleers; Wilfried De Neve; Liv Veldeman
Journal:  Strahlenther Onkol       Date:  2016-02-10       Impact factor: 3.621

5.  Impact of surface-guided positioning on the use of portal imaging and initial set-up duration in breast cancer patients.

Authors:  Montserrat Pazos; Franziska Walter; Daniel Reitz; Stephan Schönecker; Dinah Konnerth; Annemarie Schäfer; Maya Rottler; Filippo Alongi; Philipp Freislederer; Maximilian Niyazi; Claus Belka; Stefanie Corradini
Journal:  Strahlenther Onkol       Date:  2019-07-22       Impact factor: 3.621

6.  Accuracy of surface registration compared to conventional volumetric registration in patient positioning for head-and-neck radiotherapy: a simulation study using patient data.

Authors:  Youngjun Kim; Ruijiang Li; Yong Hum Na; Rena Lee; Lei Xing
Journal:  Med Phys       Date:  2014-12       Impact factor: 4.071

7.  Comparison of three differently shaped ROIs in free breathing breast radiotherapy setup using surface guidance with AlignRT®.

Authors:  Marko Laaksomaa; Torsten Moser; Julia Kritz; Kiira Pynnönen; Maija Rossi
Journal:  Rep Pract Oncol Radiother       Date:  2021-08-12

8.  Automatic deformable surface registration for medical applications by radial basis function-based robust point-matching.

Authors:  Youngjun Kim; Yong Hum Na; Lei Xing; Rena Lee; Sehyung Park
Journal:  Comput Biol Med       Date:  2016-08-11       Impact factor: 4.589

9.  Prediction of high-dimensional states subject to respiratory motion: a manifold learning approach.

Authors:  Wenyang Liu; Amit Sawant; Dan Ruan
Journal:  Phys Med Biol       Date:  2016-06-14       Impact factor: 3.609

10.  Radiotherapy setup displacements in breast cancer patients: 3D surface imaging experience.

Authors:  Ana Cravo Sá; Ana Fermento; Dalila Neves; Sara Ferreira; Teresa Silva; Carina Marques Coelho; Aude Vaandering; Ana Roma; Sérgio Quaresma; Emmanuel Bonnarens
Journal:  Rep Pract Oncol Radiother       Date:  2018-01-16
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.