| Literature DB >> 30382341 |
F V Y Tjong1, B E Koop2.
Abstract
Cardiac implantable electronic devices have been successfully treating patients with brady- and tachyarrhythmias for decades. However, there are still significant complications related to this therapy modality, many related to the transvenous lead. Paradigm-shifting technologies, such as the subcutaneous implantable cardioverter-defibrillator (S-ICD) and leadless cardiac pacemakers (LCP), have emerged to address these complications. The novel modular cardiac rhythm management (mCRM) system, consisting of a communicating antitachycardia pacing-enabled LCP and S‑ICD, is the first system to integrate wireless intrabody communication between devices to allow for coordination of leadless pacing and defibrillator therapy delivery. In this review, the design and concept of the mCRM system are presented and available evidence is summarized.Entities:
Keywords: Antitachycardia pacing; Arrhythmias, cardiac; Leadless pacing; Subcutaneous implantable cardioverter–defibrillator; Tachycardia
Mesh:
Year: 2018 PMID: 30382341 PMCID: PMC6267407 DOI: 10.1007/s00399-018-0602-y
Source DB: PubMed Journal: Herzschrittmacherther Elektrophysiol ISSN: 0938-7412
Fig. 1The modular cardiac rhythm management (mCRM) system. a Depiction of implanted mCRM system. b Close-up images of EMPOWER and EMBLEM MRI devices (not to scale). S-ICD subcutaneous implantable cardioverter–defibrillator, LCP leadless cardiac pacemaker. © 2017 Boston Scientific Corporation or its affiliates. All rights reserved. Used with permission of Boston Scientific Corporation
Fig. 2Device–device communication of the mCRM system. Depiction of how communication messages are sent from S‑ICD to EMPOWER LCP via conducted signals. The EMPOWER LCP senses the conducted signals via its cathode and anode, which are also used for sensing intrinsic cardiac signals. mCRM modular cardiac rhythm management, S-ICD subcutaneous implantable cardioverter–defibrillator, LCP leadless cardiac pacemaker, ATP antitachycardia pacing
Acute and 3‑month EMPOWER LCP performance showing favorable pacing and sensing performance over the course of the study. Values are n (%) or mean ± standard deviation
| Acute Performance ( | 3-Month Performance ( | ||||||
|---|---|---|---|---|---|---|---|
| Baseline | Canine | ||||||
| Ovine ( | Swine ( | Canine ( | 7 Days | 28 Days | 90 Days | ||
| Implant Success | 8 (100) | 5 (100) | 26 (96) | – | – | – | – |
|
| |||||||
| RV Apex | 8 (100) | 4(80) | 12 (46) | – | – | – | – |
| RV Apical Septum | 0 | 0 | 14 (54) | – | – | – | – |
| RV Outflow Tract | 0 | 1 (20) | 0 | – | – | – | – |
|
| |||||||
| Pacing Threshold, V at 0.5 ms | 1.10 ± 0.81 | 0.53 ± 0.49 | 0.37 ± 0.19 | 0.56 ± 0.37b | 0.54 ± 0.30b | 0.72 ± 0.45b,c | <0.001 |
| R-Wave Amplitude, mV | 6.6 ± 1.4 | 28.3 ± 5.8 | 25.8 ± 5.1d | 26.3 ± 6.8d | 25.0 ± 9.4d | 23.3 ± 9.4 | <0.001 |
| Impedance, Ω | 665 ± 225 | 753 ± 118 | 826 ± 171d | 785 ± 129d | 827 ± 105d | 728 ± 141c | 0.04 |
|
| |||||||
| Pre- to Post Shock Change in Pacing Threshold, V at 0.5 ms | 0.0 ± 0.5 | 0.1 ± 0.1 | 0 | – | – | 0.0 ± 0.1 | – |
| Pre- to Post-Shock Change in Impedance, W | 18 ± 49 | 26 ± 40 | 5 | – | – | −48 ± 58 | – |
Table reprinted from Tjong et al. [23] with permission from the publisher
LCP leadless cardiac pacemaker, RV right ventricle
ap value between baseline and 90 days, for pacing threshold and impedance tested with the Student t test, for R wave calculated with a linear regression analysis
bPacing threshold data from 3 animals were excluded because the LCP prototypes did not have steroid-eluting electrode
cImpedance and pacing threshold data from 1 animal were excluded due to suspected device malfunction at day 90
dR-wave and impedance data from 7 animals at baseline, 9 animals at day 7, and 7 animals at day 28 were excluded due to programmer software malfunction
Fig. 3Device–device communication thresholds. Threshold for successful S‑ICD to LCP communication, where the threshold is based on the voltage amplitude setting in the S‑ICD used for sending the communication pulses. This is not equal to the voltage emitted by the S‑ICD, which is generally 30–75% lower. The communication voltage amplitude setting can range from 1 to 7 V, with a nominal setting of 4 V. S-ICD subcutaneous implantable cardioverter–defibrillator, LCP leadless cardiac pacemaker. Reprinted from Tjong et al. [23] with permission from the publisher
Fig. 4Device orientation in communication vector between S‑ICD and LCP. This fluoroscopy image shows the measurements of the orientation vector between S‑ICD lead and can, with the orientation of the LCP within this vector (angle α). S-ICD subcutaneous implantable cardioverter–defibrillator, LP leadless pacemaker. Reprinted from Tjong et al. [23] with permission from the publisher