Long-term performance of a pacing lead family: A single-center experience.

Mikhael F El-Chami1, Birju Rao2, Anand D Shah2, Carolyn Wood2, Michael Sayegh2, Patrick Zakka2, Krista Ginn2, Lauren Pallotta2, Bryanna Evans2, Michael H Hoskins2, David Delurgio2, Michael Lloyd2, Jonathan Langberg2, Angel R Leon2, Faisal M Merchant2.   

Abstract

BACKGROUND: The performance of Abbott/St. Jude Medical (Sylmar CA) Tendril pacing leads has not been well characterized.
OBJECTIVE: We sought to assess the performance of Tendril leads as compared with that of different pacing leads.
METHODS: We retrospectively identified patients implanted with the following leads: Tendril leads 1888 TC, 2088 TC, and 1688 TC, Medtronic (Fridely, MN) 4076 CapSureFix Novus, and Boston Scientific (Natick, MA) FINELINE II Sterox Pacing EZ leads (models 4469, 4470, and 4471). The primary end point was the incidence of lead malfunction assessed by Kaplan-Meier analysis.
RESULTS: During the study period, 9782 leads were implanted, including 8512 Tendril leads, 731 Medtronic 4076 CapSureFix Novus leads, and 539 FINELINE II leads. A total of 540 leads (5.5%) malfunctioned during a mean follow-up of 3.6 ± 2.9 years. Lead malfunction manifested predominantly as noise and/or low impedance (95%). Lead malfunction rates were significantly higher at 5 years for Tendril vs non-Tendril leads (7.0% vs 2.1%; P < .001). The highest rate of failure at 5 years was seen in the Tendril 1888 TC leads (9.9%), followed by Tendril 1688 (5.7%) and Tendril 2088 (5.2%) leads. In contrast, malfunction rates were significantly lower for the Medtronic 4076 (2.6%) and FINELINE II (1.7%) leads. During follow-up to 10 years, the incidence of lead malfunction for Optim-insulated Tendril leads (models 1888 TC and 2088 TC) was significantly higher than that for the non-Optim-insulated Tendril 1688 TC lead (24.5% vs 7.1%) (P = .008).
CONCLUSION: Tendril leads appear to have a higher rate of malfunction than do comparator leads. Optim insulation may partly explain the higher failure rate.
Copyright © 2018 Heart Rhythm Society. Published by Elsevier Inc. All rights reserved.

Entities:  

Keywords:  Insulation abrasion; Lead failure; Lead fracture; Pacing leads; Tendril leads

Mesh:

Year:  2018        PMID: 30366161     DOI: 10.1016/j.hrthm.2018.10.024

Source DB:  PubMed          Journal:  Heart Rhythm        ISSN: 1547-5271            Impact factor:   6.343