Acute and 3-Month Performance of a Communicating Leadless Antitachycardia Pacemaker and Subcutaneous Implantable Defibrillator.

Fleur V Y Tjong1, Tom F Brouwer2, Brendan Koop3, Brian Soltis3, Allan Shuros3, Brian Schmidt3, Bryan Swackhamer3, Anne-Floor E B Quast2, Arthur A M Wilde2, Martin C Burke4, Reinoud E Knops2.   

Abstract

OBJECTIVES: The primary objective was to assess the acute and 3-month performance of the modular antitachycardia pacing (ATP)-enabled leadless pacemaker (LP) and subcutaneous implantable cardioverter-defibrillator (S-ICD) system, particularly device-device communication and ATP delivery.
BACKGROUND: Transvenous pacemakers and implantable cardioverter-defibrillators (ICDs) have considerable rates of lead complications. We examined the next step in multicomponent leadless cardiac rhythm management: feasibility of pacing (including ATP) by a LP, commanded by an implanted S-ICD through wireless, intrabody, device-device communication.
METHODS: The combined modular cardiac rhythm management therapy system of the LP and S-ICD prototypes was evaluated in 3 animal models (ovine, porcine, and canine) both in acute and chronic (90 days) experiments. LP performance, S-ICD to LP communication, S-ICD and LP rhythm discrimination, and ATP delivery triggered by the S-ICD were tested.
RESULTS: The LP and S-ICD were successfully implanted in 98% of the animals (39 of 40). Of the 39 animals, 23 were followed up for 90 days post-implant. LP performance was adequate and exhibited appropriate VVI behavior during the 90 days of follow-up in all tested animals. Unidirectional communication between the S-ICD and LP was successful in 99% (398 of 401) of attempts, resulting in 100% ATP delivery by the LP (10 beats at 81% of the coupling interval). Adequate S-ICD sensing was observed during normal sinus rhythm, LP pacing, and ventricular tachycardia/ventricular fibrillation.
CONCLUSIONS: This study presents the preclinical acute and chronic performance of the combined function of an ATP-enabled LP and S-ICD. Appropriate VVI functionality, successful wireless device-device communication, and ATP delivery were demonstrated by the LP. Clinical studies on safety and performance are needed.
Copyright © 2017 The Authors. Published by Elsevier Inc. All rights reserved.

Entities:  

Keywords:  ATP; S-ICD; leadless pacemaker; modular therapy; wireless communication

Mesh:

Year:  2017        PMID: 29759829     DOI: 10.1016/j.jacep.2017.04.002

Source DB:  PubMed          Journal:  JACC Clin Electrophysiol        ISSN: 2405-500X


  9 in total

Review 1.  Current Review of Implantable Cardioverter Defibrillator Use in Patients With Left Ventricular Assist Device.

Authors:  Jacinthe Boulet; Emmanuelle Massie; Blandine Mondésert; Yoan Lamarche; Michel Carrier; Anique Ducharme
Journal:  Curr Heart Fail Rep       Date:  2019-12

Review 2.  [Leadless pacemakers and subcutaneously implantable cardioverter defibrillators].

Authors:  C Stellbrink; B Hansky; D Meyer Zu Vilsendorf
Journal:  Internist (Berl)       Date:  2018-10       Impact factor: 0.743

3.  Is ventricular sensing always right, when it is left?

Authors:  Mauro Biffi; Giulia de Zan; Giulia Massaro; Andrea Angeletti; Cristian Martignani; Giuseppe Boriani; Igor Diemberger; Matteo Ziacchi
Journal:  Clin Cardiol       Date:  2018-09-21       Impact factor: 2.882

Review 4.  Nanostim-leadless pacemaker.

Authors:  Johannes Sperzel; Christian Hamm; Andreas Hain
Journal:  Herzschrittmacherther Elektrophysiol       Date:  2018-10-19

Review 5.  Innovations in Cardiac Implantable Electronic Devices.

Authors:  Khurrum Khan; Jitae A Kim; Andra Gurgu; Muzamil Khawaja; Dragos Cozma; Mihail G Chelu
Journal:  Cardiovasc Drugs Ther       Date:  2021-03-02       Impact factor: 3.947

Review 6.  Innovations in Clinical Cardiac Electrophysiology: Challenges and Upcoming Solutions in 2018 and Beyond.

Authors:  Vaibhav R Vaidya; Alan Sugure; Samuel J Asirvatham
Journal:  J Innov Card Rhythm Manag       Date:  2017-12-15

7.  Highlights from Heart Rhythm 2019: Cardiac Ablation, Pacing, and Monitoring.

Authors:  Claudio Tondo
Journal:  J Innov Card Rhythm Manag       Date:  2019-08-15

Review 8.  Non-traditional implantable cardioverter-defibrillator configurations and insertion techniques: a review of contemporary options.

Authors:  Johanna B Tonko; Christopher A Rinaldi
Journal:  Europace       Date:  2022-02-02       Impact factor: 5.214

Review 9.  The modular cardiac rhythm management system: the EMPOWER leadless pacemaker and the EMBLEM subcutaneous ICD.

Authors:  F V Y Tjong; B E Koop
Journal:  Herzschrittmacherther Elektrophysiol       Date:  2018-10-31
  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.