An Evaluation of the IOLMaster 700.

Mark A Bullimore1, Stephen Slade, Paul Yoo, Todd Otani.   

Abstract

OBJECTIVES: To evaluate the repeatability and reproducibility of the Carl Zeiss Meditec IOLMaster 700, which uses swept-source optical coherence tomography technology, along with its agreement with the IOLMaster 500 and Lenstar LS 900.
METHODS: In a clinical practice, complete measurements were taken on one eye of 100 subjects: 51 with cataracts and 49 with clear lenses. Three sets of measurements were taken by 3 operators with three different IOLMaster 700 units, and one operator took 3 measurements with the IOLMaster 500 and the Lenstar. A random-effects model of analysis of variance was used to estimate the repeatability and reproducibility. The 95% limits of agreement (95% LoA) were calculated for all comparisons.
RESULTS: Comparing the IOLMaster 700 and IOLMaster 500 in cataract patients, 95% LoA were -0.01 to +0.06 mm for axial length, -0.44 to +0.27 D for corneal power, and -0.18 to +0.17 mm for anterior chamber depth. Comparing the IOLMaster 700 and Lenstar for corneal thickness and lens thickness yielded 95% LoA of -4 to +13 μm and -0.26 to +0.41 mm, respectively. The repeatability and reproducibility limits for the IOLMaster 700 were ±0.014 and ±0.023 mm for axial length; ±0.26 and ±0.27 D for corneal power; ±7 and ±11 μm for corneal thickness; ±0.02 and ±0.02 mm for anterior chamber depth; and ±0.02 and ±0.05 mm for lens thickness, respectively.
CONCLUSIONS: There was good agreement between the IOLMaster 700 and comparator instruments and superior or equivalent precision. Some differences were noted, but not considered clinically meaningful.

Entities:  

Mesh:

Year:  2019        PMID: 30325761     DOI: 10.1097/ICL.0000000000000552

Source DB:  PubMed          Journal:  Eye Contact Lens        ISSN: 1542-2321            Impact factor:   2.018


  11 in total

1.  Comparison of IOL-Master 700 and IOL-Master 500 biometers in ocular biological parameters of adolescents.

Authors:  Qiang Shi; Guang-Yan Wang; Yu-Hong Cheng; Cheng Pei
Journal:  Int J Ophthalmol       Date:  2021-07-18       Impact factor: 1.779

2.  Accuracy of segmented measurement of axial length in ultra-high myopia filled with silicone oil using immersion B-scan ultrasonography.

Authors:  Qing-Hua Yang; Hong-Tao Zhang; Xiao-Qi Li; Bing Chen; Zhao-Hui Li; Yi-Fei Huang; Xin Jin; Ying Zhang; Li-Qiang Wang
Journal:  Int J Ophthalmol       Date:  2022-05-18       Impact factor: 1.645

3.  Comparison of Vault Measurements Using a Swept-Source OCT-Based Optical Biometer and Anterior Segment OCT.

Authors:  Yali Du; Chuang Jin; Shengjie Yin; Geng Wang; Qian Ma; Yuancun Li; Binyao Chen; Hongxi Wang; Kunliang Qiu; Mingzhi Zhang
Journal:  Front Med (Lausanne)       Date:  2022-06-23

4.  Measurement of crystalline lens tilt in high myopic eyes before cataract surgery using swept-source optical coherence tomography.

Authors:  Qiang Lu; Wenwen He; Dongjin Qian; Yi Lu; Xiangjia Zhu
Journal:  Eye Vis (Lond)       Date:  2020-03-06

5.  Anterior segment optical coherence tomography imaging and ocular biometry in cataract patients with open angle glaucoma comorbidity.

Authors:  Giedre Pakuliene; Kirilas Zimarinas; Irena Nedzelskiene; Brent Siesky; Loreta Kuzmiene; Alon Harris; Ingrida Januleviciene
Journal:  BMC Ophthalmol       Date:  2021-03-08       Impact factor: 2.209

6.  Effect of Implantable Collamer Lens on Anterior Segment Measurement and Intraocular Lens Power Calculation Based on IOLMaster 700 and Sirius.

Authors:  Xiaoyong Chen; Di Zhang; Ziyuan Liu; Yinan Liu; Hongyuan Cai; Qianru Wu; Yu Zhang
Journal:  J Ophthalmol       Date:  2021-12-21       Impact factor: 1.909

7.  Interocular biometric parameters comparison measured with swept-source technology.

Authors:  César Albarrán-Diego; Francisco Poyales; Esther López-Artero; Nuria Garzón; María García-Montero
Journal:  Int Ophthalmol       Date:  2021-08-21       Impact factor: 2.031

8.  Agreement of Anterior Segment Parameter Measurements With CASIA 2 and IOLMaster 700.

Authors:  Xiaoting Ruan; GuangYao Yang; Zhaoxia Xia; Jiaqing Zhang; Xiaoxun Gu; Yuan Tan; Zhenzhen Liu; Lixia Luo
Journal:  Front Med (Lausanne)       Date:  2022-02-11

9.  Accuracy of OCT-derived net corneal astigmatism measurement.

Authors:  Clara Llorens-Quintana; Elias Pavlatos; Omkar Thaware; Seema Gupta; Daniel Gradin; Denzil Romfh; Yan Li; David Huang
Journal:  J Cataract Refract Surg       Date:  2022-03-01       Impact factor: 3.528

10.  [Optometric eye screening in schools : First epidemiological data for children and adolescents in grades 5-7].

Authors:  Hakan Kaymak; Kai Neller; Birte Graff; Kristina Körgesaar; Achim Langenbucher; Berthold Seitz; Hartmut Schwahn
Journal:  Ophthalmologe       Date:  2021-06-10       Impact factor: 1.059

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.