| Literature DB >> 30302910 |
Wei Yang1, Xiaochen Yu1, Dan Liu1, Xiuru Guan1.
Abstract
BACKGROUND: The UF-1000i has been widely used in screening urinary sediments. However, the interference factor of the UF-1000i in the screening urinary sediments of pregnant women has not been reported. The aim of the study was to demonstrate that epithelial cells (ECs) cause a high false positive rate of white blood cells (WBCs) by the UF-1000i in pregnant women.Entities:
Keywords: UF-1000i; epithelia; misclassified; pregnant women; white blood cell
Mesh:
Year: 2018 PMID: 30302910 PMCID: PMC6899535 DOI: 10.1002/cyto.b.21727
Source DB: PubMed Journal: Cytometry B Clin Cytom ISSN: 1552-4949 Impact factor: 3.058
Figure 1Receiver operating characteristic (ROC) curve of WBC and EC counts on the UF‐1000i. [Color figure can be viewed at http://wieyonlinelibrary.com]
Comparison Between UF‐1000i and the Microscopic Method in Urinary WBC and EC of Pregnant Women
| Positive samples by UF‐1000i | Positive samples by microscopic method | |||
|---|---|---|---|---|
|
| % |
| % | |
| WBC | 152 | 73.43 | 40 | 19.32 |
| EC | 77 | 37.20 | 151 | 72.95 |
WBC, white blood cell; EC, epithelial cell.
P < 0.01;
P < 0.01
Results Tested by Two Methods in Urinary WBC of Pregnant Women
| UF‐1000 i | Microscopic method | ||
|---|---|---|---|
| + ( | − ( | Total ( | |
| + | 39 | 113 | 152 |
| − | 1 | 54 | 55 |
| Total | 40 | 167 | 207 |
Results Tested by Two Methods in Urinary EC of Pregnant Women
| UF‐1000i | Microscopic method | ||
|---|---|---|---|
| +( | −( | Total ( | |
| + | 74 | 3 | 77 |
| − | 77 | 53 | 130 |
| Total | 151 | 56 | 207 |
The Performance of UF‐1000i and Microscopic Method in Screening Urinary WBC and EC of Pregnant Women
| Positive results for both UF1000i and microscopic ( | Negative results for both UF1000i and microscopic ( | UF1000i positive/microscopic negative ( | UF1000i negative/microscopic positive ( | Positive predictive value (%) | False positive rate (%) | Negative predictive value (%) | False negative rate (%) | |
|---|---|---|---|---|---|---|---|---|
| WBC | 39 | 54 | 113 | 1 | 25.66 | 74.34 | 98.18 | 1.82 |
| EC | 74 | 53 | 3 | 77 | 96.1 | 3.9 | 40.77 | 59.23 |
Figure 2Correlations between the UF‐1000i WBC false positives and the respective microscopic EC counts (R = 0.503, P < 0.01)