Azadeh Hojreh1, Peter Homolka2, Jutta Gamper3, Sylvia Unterhumer1, Daniela Kienzl-Palma1, Csilla Balassy1, Thomas Wrba4, Helmut Prosch1.
Abstract
BACKGROUND: Modern CT scanners provide automatic dose adjustment systems, which are promising options for reducing radiation dose in pediatric CT scans. Their impact on patient dose, however, has not been investigated sufficiently thus far.Entities:
Mesh:
Substances:
Year: 2018 PMID: 30281614 PMCID: PMC6169939 DOI: 10.1371/journal.pone.0204794
Source DB: PubMed Journal: PLoS One ISSN: 1932-6203 Impact factor: 3.240
Scan parameters of the ATCM group and of the ATVS +ATCM group.
| Scan Parameters | ATCM group | ATVS + ATCM group |
|---|---|---|
| 2x64x0.6mm | 2x64x0.6mm | |
| on | on | |
| 3 | 3 | |
| average | average | |
| Not implemented | On (DSO 3) | |
| Fixed Setting | Reference kV: 100 | |
| - < 45 kg | 80 kV | |
| - > 45 kg | 100 kV | |
| 50mAs | 80mAs @ 100kV |
ATCM, automated tube current modulation; ATVS, automated tube voltage selection; DSO 3, dose saving optimized level 3 for non-contrast, as suggested by the manufacturer for non-contrast CT scans
Fig 1Boxplots of selected kV at effective diameter, in the ATCM group and in the ATVS + ATCM group.
Mean value of effective diameter (EDM), kV, mAs, SDNR, and CTDIvol of chest CTs from the study age groups (32 cm PMMA phantom), with ATCM and with ATVS + ATCM.
| Age group | Patients’ number | EDM (cm) | Mean kV | Mean mAs | Mean SDNR | CTDIvol | CTDIvol with ATVS and ATCM vs. CTDIvol with ATCM | ||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| ATCM | ATVS | ATCM | ATVS | ATCM | ATVS | ATCM | ATVS | ATCM | ATVS | ATCM | ATVS | ||
| 0 | 0 | - | - | - | - | - | - | - | - | - | - | - | |
| 6 | 6 | 13.40 | 11.81 | 80 | 100 | 72.67 | 17.83 | 11.05 | 8.64 | 1.16 | 0.63 | - % 46 | |
| 22 | 16 | 15.93 | 15.54 | 80 | 100 | 80.05 | 32.13 | 11.37 | 10.74 | 1.27 | 1.11 | - % 13 | |
| 16 | 12 | 18.74 | 18.89 | 80 | 100 | 86 | 36.83 | 10.99 | 10.62 | 1.37 | 1.25 | - % 9 | |
| 31 | 19 | 21.56 | 22.29 | 81.94 | 101.05 | 95.19 | 56.11 | 10.91 | 10.29 | 1.71 | 1.96 | + %15 | |
| 17 | 15 | 23.57 | 23.26 | 88.24 | 104 | 101.35 | 49.33 | 10.03 | 9.04 | 2.42 | 1.91 | - %21 | |
EDM, Effective diameter; SDNR, Signal-difference-to-noise ratio; PMMA, polymethyl methacrylate; ATCM, automated tube current modulation; ATVS, automated tube voltage selection.
Fig 2CTDIvol (CTDI) as a function of effective diameter in the ATCM group and in the ATVS + ATCM group.
Fig 3Boxplots of mean SDNR measurements in the ATCM group and in the ATVS + ATCM group.
Fig 4Distribution of image quality for each reader, both in the ATCM and in the ATVS + ATCM group.
Weighted kappa coefficients (95% CI) assessing agreement between the readers.
| 0.1 (0.03–0.16) | 0.27 (0.09–0.45) | |
| 0.1 (0.06–0.14) | 0.12 (0.05–0.2) | |
| 0.22 (0.11–0.34) | 0.36 (0.24–0.49) |
ATCM, automated tube current modulation; ATVS, automated tube voltage selection; PR1, pediatric radiologist with seven years experience; PR2, pediatric radiologist with eight years experience; GR, general radiologist with seven years experience.
Fig 5Two Examples of image quality for the chest scan a) with ATCM only and b) with ATVS + ATCM.