Clinical Validation of Automatable Gaussian Normalized CBV in Brain Tumor Analysis: Superior Reproducibility and Slightly Better Association with Survival than Current Standard Manual Normal Appearing White Matter Normalization.

Lei Qin1, Xiang Li2, Angie Li3, Suchun Cheng4, Jinrong Qu2, Katherine Reinshagen5, Jiani Hu4, Nathan Himes6, Gao Lu7, Xiaoyin Xu8, Geoffrey S Young9.   

Abstract

PURPOSE: To validate Gaussian normalized cerebral blood volume (GN-nCBV) by association with overall survival (OS) in newly diagnosed glioblastoma patients and compare this association with current standard white matter normalized cerebral blood volume (WN-nCBV).
METHODS: We retrieved spin-echo echo-planar dynamic susceptibility contrast MRI acquired after maximal resection and prior to radiation therapy between 2006 and 2011 in 51 adult patients (28 male, 23 female; age 23-87 years) with newly diagnosed glioblastoma. Software code was developed in house to perform Gaussian normalization of CBV to the standard deviation of the whole brain CBV. Three expert readers manually selected regions of interest in tumor and normal-appearing white matter on CBV maps. Receiver operating characteristics (ROC) curves associating nCBV with 15-month OS were calculated for both GN-nCBV and WN-nCBV. Reproducibility and interoperator variability were compared using within-subject coefficient of variation (wCV) and intraclass correlation coefficients (ICCs).
RESULTS: GN-nCBV ICC (≥0.82) and wCV (≤21%) were superior to WN-nCBV ICC (0.54-0.55) and wCV (≥46%). The area under the ROC curve analysis demonstrated both GN-nCBV and WN-nCBV to be good predictors of OS, but GN-nCBV was consistently superior, although the difference was not statistically significant.
CONCLUSION: GN-nCBV has a slightly better association with clinical gold standard OS than conventional WM-nCBV in our glioblastoma patient cohort. This equivalent or superior validity, combined with the advantages of higher reproducibility, lower interoperator variability, and easier automation, makes GN-nCBV superior to WM-nCBV for clinical and research use in glioma patients. We recommend widespread adoption and incorporation of GN-nCBV into commercial dynamic susceptibility contrast processing software.
Copyright © 2018 The Authors. Published by Elsevier Inc. All rights reserved.

Entities:  

Year:  2018        PMID: 30216765      PMCID: PMC6138997          DOI: 10.1016/j.tranon.2018.07.017

Source DB:  PubMed          Journal:  Transl Oncol        ISSN: 1936-5233            Impact factor:   4.243


Introduction

Dynamic susceptibility contrast (DSC) MRI estimates of brain tumor cerebral blood volume (CBV) reflect tumor vascularity and neoangiogenesis [1], are predictive of glioma grade and survival [2], [3], [4], [5], [6], [7], [8], [9], and aid in assessment of treatment response [10], [11] and differentiation of pseudoprogression from true tumor progression [12], [13]. DSC detects the transient decrease in signal intensity (ΔSI(t)) on continuously acquired echo-planar T2 or T2*-weighted images caused by passage of bolus gadolinium contrast through the brain capillaries. Integrating the area under the transverse relaxation rate (ΔR2 or ΔR2*) curve derived from the ΔSI curve yields the CBV for each voxel [14]. In addition to the number, size, and distribution of vessels within each voxel, CBV estimates vary with intravascular concentration, dispersion, delay, flow rate, choice of acquisition parameters [gradient vs spin-echo, repetition time (TR), echo time (TE), flip angle, contrast agent, contrast dose, leakage-reduction preload etc.] and choice of postprocessing algorithm that may or may not include gamma-variate fitting [15], baseline subtraction [16], and leakage correction [10], [17]. To compensate, white matter normalization (WN) is typically performed whereby tumor CBV is divided by the mean CBV in a region of interest (ROI) selected manually in the contralateral normal-appearing white matter (NAWM), yielding a unitless “normalized CBV” (nCBV) ratio [14], [17], [18], [19], [20]. The few published studies suggest substantial coefficient of variation (CV) in NAWM ROI CBV measurements including test-retest CV of 12%-14% in healthy volunteers [21] and an interscanner CV of 25%-30% in glioma patients [22]. The roughly 20% variation in NAWM measurements is likely responsible for a substantial part of the intra- and interobserver CV of 30%-41% reported in white matter normalized brain lesion nCBV [18]. Methods proposed to reduce this variation in CBV estimates include standardization [23], Z-score normalization [22], and Gaussian normalization (GN) [22]. In GN, the tumor ROI CBV is normalized to the standard deviation (SD) of CBV throughout the whole brain rather than ROI measurements of NAWM CBV. GN eliminates completely the subjectivity of NAWM ROI selection, reduces operator time, makes automation simpler, and has been reported to provide the lowest CV in NAWM and the highest tumor contrast for glioblastoma (GBM) [21]. Although GN decreases CV in NAWM, its effect on tumor nCBV estimates has not been studied in detail. Whether glioma GN-nCBV is as valid as current standard WM-nCBV remains to be established [24]. To address this, we compared GN-nCBV association with OS to the known strong OS association of WN-nCBV in a well-characterized cohort of newly diagnosed GBM.

Methods

Human Subjects

The study was in compliance with the Health Insurance Portability and Accountability Act and approved by our institutional review board. Informed consent was waived for this retrospective study. For this study, 51 patients were retrieved from a database of adult glioma patients newly diagnosed between 2006 and 2011 (28 men, 23 women; age mean: 56.6 years and range: 23-87 years). All patients had WHO grade IV glioblastoma, known survival, and SE-EPI DSC acquired after maximal surgical resection and before radiation therapy.

Data Acquisition

Thirty-five patients underwent 1.5-T MRI and 16 patients underwent 3-T MRI on whole body MRI scanners (GE Medical Systems, Milwaukee, WI). Axial DSC was performed utilizing a series of SE-EPI images (scan parameters: 1900-2000 milliseconds TR/80 milliseconds TE, 128*128 matrix size; 10-mm slice thickness; 40 time points) acquired 10 seconds prior to, during, and after intravenous administration of gadopentetate dimeglumine (Magnevist, Bayer Healthcare) with a power injector at a rate of 4 ml/s, followed by 20-ml saline flush. Double-dose (0.2 mmol/kg) contrast was used if acquisition was performed with 1.5 T, while one and half dose (0.15 mmol/kg) was used for 3 T, with a maximum 30 ml.

Image Analysis

The dynamic source images were visually inspected for to exclude datasets with substantial patient motion. No substantially motion-degraded scans were detected. DSC analysis was performed using the Functool software package on Advantage Window workstation (GE Medical Systems). A lower threshold was manually adjusted to remove the background noise. All pixels with lower intensity than the threshold were removed. The remaining pixels define the brain volume used to calculate whole brain SD for GN. The beginning and end of the bolus passage were defined on the time-intensity curve to set the integration range for calculation of CBV maps. Three readers (clinical radiologists with 15, 5, and 5 years of experience, respectively) independently selected tumor and NAWM ROIs directly on the resulting CBV maps. Each reader picked three tumor ROIs on the high CBV spots and another three ROIs on the contralateral NAWM. The radius of these ROIs was 1-2 image pixels (2-4 mm). Because it has been previously shown that the maximum or mean CBV of several tumor ROIs generates better intra- and interobserver reproducibility than a single ROI and that the maximum is slightly better than the mean [18], we calculated both the maximum and the mean of the three tumor ROIs. For NAWM, the mean of the three ROIs was calculated and used for normalization of the measured tumor CBV values. To assess the influence of variation in ROI size on NAWM estimates, one of the three readers (reader 2) picked one large ROI (radius = 10 mm) on NAWM for each patient, and the mean CBV inside the ROI was recorded for normalization. This second evaluation was performed more than 2 months later than the first evaluation to avoid recall bias. A fourth reader (with clinical experience of 14 years) picked a single large ROI in the NAWM using the same radius independently. GN was performed using Matlab (Mathworks, Natick, MA) code produced in-house [22]. The SD of the whole brain CBV map, σ, was calculated and used for normalization: Conventional NAWM ROIs are selected to exclude tumor, but whole brain SD used in GN includes tumor. To examine whether the presence of variable amounts of tumor introduces significant variation between patients in whole brain SD, we selected 10 patients with the largest volume of enhancing tumor in this cohort. One investigator manually contoured the ROIs slice by slice on both contrast-enhanced pGd-T1WI and fluid-attenuated inversion recovery T2WI (FLAIR-T2). The volume of enhancing and nonenhancing tumor in these ROIs was calculated as a percentage of whole brain volume. After the two sets of ROIs were copied to the aligned CBV maps, the whole brain SD was calculated after first excluding contrast-enhancing tumor (σ) and then after excluding both enhancing and nonenhancing tumor (σ). The percentage change in SD resulting from exclusion of the tumor ROI was calculated as: Or Because the ROI size subanalysis demonstrated that the small NAWM ROI mean CBV was similar to the large NAWM ROI CBV (see results section), only small ROI NAWM estimates were used for nCBV normalization. Twelve nCBVs in total were generated: ; ; and, , where i = reader 1,2,3.

Statistical Analysis

Tumor CBVimean, Tumor CBVimax, NAWM CBVimeaninside the selected ROIs, nCBVimean and nCBVimax, nCBVGimean and nCBVGimax measurements were summarized by means and SDs for i = reader 1,2,3. Pairwise comparisons among readers in these outcome measurements were summarized by mean differences and corresponding standard errors (SEs) and assessed by paired t tests, separately. For reader 2 who selected both NAWM large ROIs and small ROIs, a paired t test was also used to evaluate the effect of ROI size. Mixed-effects models were built for each outcome variable with readers as fixed effects to evaluate differences among the three readers and subjects as random effects to account for correlation among measurements from the three readers on the same subject. Only the small ROI measurements were used in building mixed-effects models. The between-subject standard deviation (bSD), within-subject standard deviation (wSD), repeatability coefficient (RC), within-subject coefficient of variation (wCV), intraclass correlation coefficient (ICC), and concordance correlation coefficient (CCC) were calculated based on variance components of the aforementioned mixed-effects models [25] to compare repeatability across readers.. Bland-Altman plots provided an intuitive methodology using the concept of limits of agreement for assessing agreement between two readers. Lastly, we calculated the receiver operating characteristics (ROC) curves to associate nCBV measurements with 15-month overall survival (OS) based on the reported 12-15 month median post-operative OS in GBM patients [26], [27]. We evaluated the strength of association by the estimated area under the ROC curve (AUC). A P ≤ .05 was considered to be statistically significant. Statistics were computed using Stata (Stata v14, StataCorp., College Station, TX), with the exception that CCC was calculated in R-package [28].

Results

Summary statistics of reader-specific Tumor CBVmean, Tumor CBVmax, and NAWM CBVmean measurements are listed in Table 1 and plotted in Figure 1. All three measured parameters using small ROIs showed statistically significant differences among readers. The differences came largely from reader 1 in Tumor CBVmean and Tumor CBVmax, and between readers 1 and 3 in NAWM CBVmean. The NAWM measurements using large ROI also showed a significant difference between readers 2 and 4. The variations in NAWM mean measurements were similar regardless of ROI size (Table 1). This is illustrated by the data from reader 2 who performed measurements using both small and large ROIs while achieving comparable means and SDs (P = .80). This suggests that a larger ROI radius does not offer better precision. Therefore, in order to reduce the overall number of comparisons, when calculating nCBV for testing the effects of normalization, we only used the NAWM CBVmean measured from the small ROI measurements.
Table 1

Summary Statistics of Tumor CBVmean, Tumor CBVmax, and NAWM CBVmean

ReaderROI SizeTumor CBVmeanTumor CBVmaxNAWM CBVmean
1*Small (2-4 mm)125.19 ± 85.04P = .002146.61 ± 102.22P < .00131.41 ± 11.38P = .04
2*135.95 ± 83.92160.39 ± 100.3930.93 ± 12.62
3*140.87 ± 86.48172.00 ± 100.8527.95 ± 14.95
(1-2)−10.76 ± 3.59§P = .004−13.78 ± 4.28§P = .0020.48 ± 1.12P = .67
(1-3)−15.68 ± 4.88§P = .002−25.39 ± 6.44§P < .0013.45 ± 1.67§P = .044
(2-3)−4.91 ± 4.84P = .31−11.61 ± 6.86P = .102.98 ± 1.64P = .075
2*,Large (10 mm)N/AN/A31.26 ± 12.14
4*N/AN/A36.37 ± 13.70
(2-4)N/AN/A−5.11 ± 1.23§P < .001

Data are represented as mean ± SD.

Data are represented as mean ± SE.

Data from reader 2 in big ROIs.

The measurements from the two readers differed statistically (paired t tests).

Differences among three readers were assessed by mixed-effects models.

Figure 1

Box plots of NAWM mean CBV measurements among different readers, varying ROI radius size, and the SDs of the brains used for normalization in GN.

Summary Statistics of Tumor CBVmean, Tumor CBVmax, and NAWM CBVmean Data are represented as mean ± SD. Data are represented as mean ± SE. Data from reader 2 in big ROIs. The measurements from the two readers differed statistically (paired t tests). Differences among three readers were assessed by mixed-effects models. Box plots of NAWM mean CBV measurements among different readers, varying ROI radius size, and the SDs of the brains used for normalization in GN. The substantially smaller variation among patients resulting from σ used in GN compared to manually selected NAWM CBVmeanused in WN is illustrated in Figure 1. Figure 2 shows an example of ROIs selected on T1 and FLAIR images, and the same ROIs copied onto the CBV map. These two ROIs were excluded to calculate SD change (%) using Eq. (2A or B). Figure 3 shows the SD change when enhancing and nonenhancing lesions were excluded in the 10 selected patients. The x-axis of Figure 3 is the tumor volume percentage in the whole brain. The SD variation was less than 1% for all but 1 of the 10 patients analyzed. In the outlying patient with tumor occupying 12% of the brain, the SD changed by only 2.5%-3%. This analysis shows that variation in tumor volume has minimal effect on SD even in the worst case where tumor involves a very large fraction of brain.
Figure 2

Conventional images and perfusion imaging CBV map demonstrating ROIs manually contoured on contrast-enhanced T1WI and FLAIR-T2WI, respectively, and copied to CBV map.

Figure 3

Change (%) in SD of the whole brain CBV as a function of the excluded volume of contrast-enhancing tumor (A) and enhancing and nonenhancing tumor (B).

Conventional images and perfusion imaging CBV map demonstrating ROIs manually contoured on contrast-enhanced T1WI and FLAIR-T2WI, respectively, and copied to CBV map. Change (%) in SD of the whole brain CBV as a function of the excluded volume of contrast-enhancing tumor (A) and enhancing and nonenhancing tumor (B). Table 2 summarizes the statistical analysis of the GN-nCBV and WM-nCBV estimates. Both nCBVmean and nCBVmax measurements obtained from WN differed significantly among the three independent readers (P < .001). The difference is largely attributable to reader 3 whose measurements were significantly higher than readers 1 and 2. GN of nCBV measurements did not resolve the significant differences among the three readers for either nCBVmean or nCBVmax measurements (P = .004 and P < .001, respectively) but did substantially decrease the SEs of each pairwise comparison.
Table 2

Comparison of nCBV Measurements Derived from NAWM Normalization (WN) Versus GN

ReadernCBVmeannCBVmax
WN 1*4.04 ± 2.26P < .001§4.69 ± 2.62P < .001§
WN 2*4.70 ± 3.095.58 ± 3.55
WN 3*6.53 ± 4.957.85 ± 5.67
WN (1-2)−0.66 ± 0.33P = .051−0.88 ± 0.37P = .02
WN (1-3)−2.50 ± 0.54P < .001−3.16 ± 0.62P < .001
WN (2-3)−1.83 ± 0.56P = .002−2.27 ± 0.63P < .001
GN 1*3.65 ± 1.93P = .004§4.26 ± 2.27P < .001§
GN 2*3.99 ± 1.944.73 ± 2.31
GN 3*4.12 ± 1.925.06 ± 2.33
GN (1-2)−0.34 ± 0.12P = .005−0.46 ± 0.15P = .003
GN (1-3)−0.47 ± 0.15P = .004−0.80 ± 0.21P < .001
GN (2-3)−0.13 ± 0.16P = .40−0.34 ± 0.22P = .14

Data are represented as mean ± SD.

Data are represented as mean ± SE.

nCBV measurements between two readers were significantly different (paired t tests).

Differences among three readers were assessed by mixed-effects models.

Comparison of nCBV Measurements Derived from NAWM Normalization (WN) Versus GN Data are represented as mean ± SD. Data are represented as mean ± SE. nCBV measurements between two readers were significantly different (paired t tests). Differences among three readers were assessed by mixed-effects models. Reproducibility metrics obtained for Tumor CBVmean, Tumor CBVmax, and NAWM CBVmean and nCBV measurements among the three readers are listed in Table 3. Tumor CBVmean, Tumor CBVmax, and nCBV measurements derived from GN all had an excellent ICC value (≥0.82) and CCC (≥0.79). NAWM CBVmean measurements had good ICC and CCC values both of 0.66. On the other hand, nCBV measurements derived from WN had a fair ICC value between 0.54 and 0.55 and CCC of 0.48. The RC of the absolute measured values on CBV maps was relatively high, with Tumor CBVmaxRC of 83.43 and NAWM CBVmean RC of 20.94. But because RC measures the within-subject variance, it is expected to be relatively high for high mean values. The RC of normalized nCBV was smaller using GN compared to WN, demonstrating a better reproducibility of using GN. The wCV, already scaled by the mean, can be used to compare across all measurement types. The Tumor CBVmean, Tumor CBVmax, and wCV were less than 20%, but NAWM CBVmean wCV was higher than 25%, suggesting that readers were more consistent in placing hot spot ROI in tumor than in NAWM. Using WN, the wCV of nCBV was >46%. GN reduced the wCV to 18.5% and 21.1% for nCBVGmeanand nCBVGmax, respectively.
Table 3

ICC, CCC, wCV, bSD, wSD, and RC for Tumor CBVmean, Tumor CBVmax, and NAWM CBVmean and nCBV Measurements Among Three Independent Readers

ICCCCCRCwCVbSDwSD
Tumor CBVmean0.930.9262.6016.87%82.1022.60
Tumor CBVmax0.910.9083.4318.86%96.5730.12
NAWM CBVmean0.660.6620.9425.11%10.667.56
nCBVmean0.540.486.8248.39%2.642.46
nCBVmax0.550.487.7446.28%3.062.80
nCBVGmean0.860.851.9918.49%1.790.72
nCBVGmax0.820.792.7421.10%2.080.99

A value <0.40 was considered poor, values between 0.40 and 0.59 were considered fair, values between 0.60 and 0.74 were considered good, and values >0.75 was considered excellent.

ICC, CCC, wCV, bSD, wSD, and RC for Tumor CBVmean, Tumor CBVmax, and NAWM CBVmean and nCBV Measurements Among Three Independent Readers A value <0.40 was considered poor, values between 0.40 and 0.59 were considered fair, values between 0.60 and 0.74 were considered good, and values >0.75 was considered excellent. Figure 4 shows Bland-Altman scatter plots of Tumor CBVmean and NAWM CBVmean measurements. No trend was observed in the difference of Tumor CBVmean (A-C) and NAWM CBVmean (D-F) between readers. Figure 5 reveals that, for all pairwise comparisons, the nCBV difference between readers increases with increasing WN-nCBV measurements (A-C). On the contrary, no systematic trend is observed in GN-nCBV measurements between readers (D-F). In addition, the y-axis scale is the same across the plots, clearly illustrating that the GN-nCBV has a much smaller variation between readers than the WN-nCBV. Note that the x-axis scale representing the mean of two readers nCBV is smaller for the GN-nCBV (0-8) than WN-nCBV (0-16) plot.
Figure 4

Bland-Altman scatter plots of tumor and NAWM mean CBV. The dotted lines show the 95% confidence interval. Y-axis is the difference between the two readers' measurements.

Figure 5

Bland-Altman scatter plots of manual ROI normalized and Gaussian normalized nCBV between readers. The dotted lines show the 95% confidence interval. Y-axis is the difference between the two readers' measurements.

Bland-Altman scatter plots of tumor and NAWM mean CBV. The dotted lines show the 95% confidence interval. Y-axis is the difference between the two readers' measurements. Bland-Altman scatter plots of manual ROI normalized and Gaussian normalized nCBV between readers. The dotted lines show the 95% confidence interval. Y-axis is the difference between the two readers' measurements. Table 4 shows the AUC estimate to evaluate the association between nCBV and 15-month OS in 51 subjects, where 46 deaths were observed and the other 5 were still alive at the cutoff for this analysis. Twenty of the 46 deaths occurred within 15 months of follow-up and the other 26 after 15 months. We compared GN-nCBV and WN-nCBV measurements among the three readers. As reflected in Table 4, all nCBV measurements, except for reader 1 WN-nCBV, were significantly different from 0.5, suggesting that both WN and GN are good predictors of OS. The mean GN-nCBV among the three readers (i.e.,; ), was slightly better than the mean WN-nCBV maps. This held true for both tumor mean nCBV (AUC of nCBV 0.68 vs nCBV 0.73, P = .33) and tumor max nCBV (AUC of nCBV 0.70 vs nCBV 0.73, P = .54) measurements.
Table 4

AUC Correlating nCBV with 15-Month OS

ReaderAUC (SE)
nCBVmean
nCBVmax
WNGNWNGN
10.60 (0.08)P = .250.67 (0.08)P = .030.62 (0.08)P = .160.69 (0.08)P = .01
20.67 (0.08)P = .030.74 (0.08)P = .0020.67 (0.08)P = .020.71 (0.07)P = .004
30.70 (0.07)P = .0080.71 (0.07)P = .0030.72 (0.07)P = .0030.75 (0.07)P = .001
Mean0.68 (0.08)P = .020.73 (0.07)P = .0020.70 (0.07)P = .0070.73 (0.07)P = .001

P values evaluate the differences of the estimated AUC to 0.5.

AUC Correlating nCBV with 15-Month OS P values evaluate the differences of the estimated AUC to 0.5.

Discussion

Translation of novel quantitative, reproducible, validated imaging biomarkers into clinical practice is an urgent need and the mission of the Quantitative Imaging Network supported by the National Cancer Institute [29]. For early detection of tumor response to treatment, the algorithm that generates the smallest measurement noise will perform best [30]. DSC is widely available, but interoperator variation and time needed for ROI measurement have limited clinical translation of published semiquantitative nCBV analysis techniques. GN, a promising automatable statistical method for producing nCBV maps, has been demonstrated to be superior to other calculation-based methods [22] but has not been validated in comparison to clinical gold standard OS or compared to current standard manual NAWM ROI normalization.

Validation

In our sample of newly diagnosed GBM patients, GN-nCBV had a slightly stronger association with OS than expert reader WN-nCBV (Table 4). WN and GN nCBVmax and nCBVmean were comparable, with nCBVmax offering slightly higher AUC for prediction of survival. GN-nCBV, the average of all readers’ WN-nCBV, and individual reader 2 and reader 3 WN-nCBV were good predictors of OS (Table 4), but reader 1 WN-nCBV did not correlate with OS, and reader 2 and reader 3 nCBVmean differed significantly (Table 2). These results both validate GN-nCBV and illustrate the importance of such automatable methods to reduce interoperator variation in clinical practice and research.

Reproducibility

Interobserver wCVs for tumor nCBVmax (46%) and nCBVmean (48%) in our sample were similar to previous literature when using comparable statistical methods [18]. Applying the formula CV = 100*SD/mean to each lesion in our sample individually before averaging across all patients, per Wetzel et al. [18], yields “average” CV of 32% for both nCBVmax and nCBVmean, nearly identical to their published “average” CV of 30% for nCBVmax and 35% for nCBVmean. For NAWM CBV, our “average” CV of 23% accords well with the 20% in that report. As an operator-independent statistical method, GN is completely reproducible within individual patients. The σ used for GN is very robust and reliable (Figure 3). It has substantially smaller interpatient variation than NAWM CBV regardless of reader or NAWM ROI size (Figure 1), which may be an important advantage for use in clinical trial grouped analyses.

Sources of Variation

Tumor CBV measured by reader 1was significantly different from reader 2 (P = .004) and reader 3 (P = .002) (Table 1). While neither WN nor GN eliminated interreader variation completely (Table 2), SD is much lower (1.9 vs 2.3-5.0) in the GN-nCBV data, suggesting that the variation in NAWM ROI selection is responsible for a substantial degree of interoperator variation in nCBV. Analysis of ICC between readers supports this, revealing excellent Tumor CBVmean (0.93) and Tumor CBVmax (0.91) ICC (Table 3) but relatively poor NAWM CBVmean ICC (0.66.). WN markedly decreases tumor CBV ICC from 0.93 to 0.54 for nCBVmean and 0.91 to 0.55 for nCBVmax. Normalization inevitably decreases ICC somewhat by combining measurement error from the denominator (NAWM CBVmeanor σ) with that of the numerator (Tumor CBVmean or Tumor CBVmax), but this marked decrease produced by WN contrasts with the relatively slight ICC decrease (GN-nCBVmean 0.86 and GN-nCBVmax 0.82), providing further evidence that NAWM ROI variability is the primary source of nCBV variation (Table 3). Possible explanations for the observed lower reproducibility of NAWM ROI measurements compared to tumor ROI measurements include larger degree of freedom in selection of NAWM ROI compared to tumor ROI and greater impact of intravoxel noise and partial volume averaged blood vessels and gray matter on lower NAWM CBV estimates compared with higher tumor CBV estimates. We varied the size of NAWM ROI to test the effect of differences in partial volume averaging and voxel SNR but detected no significant difference in NAWM mean or SE between large and small ROI (Table 1).

Weaknesses

The major weakness of our analysis is that the nCBV ROC AUC in our cohort is lower than previously reported AUC of 0.86 and 1.0 for all gliomas and pure astrocytic tumors respectively [1]. In part, this difference likely reflects the larger sample size and more heterogeneous mix of tumor histologies and treatments in our patient group. As such, 0.7 may be a more realistic estimate of the diagnostic value of nCBV in a typical mixed clinical practice. Also, we chose not to use leakage correction methods for this analysis in order to avoid introducing additional computational model complexity and/or sources of variation. This may have contributed to the lower AUCs observed. Genetic differences between tumors were not considered in this study. Many known genetics tumor markers reflect differences in tumor biology that affect the patient's OS, including IDH-1 mutation and MGMT methylation, among others. In addition, although all subjects received standard of care consisting of maximal resection followed by Stupp protocol temozolamide chemoradiation, many subsequently underwent different experimental treatments on trial. None of these experimental treatments has been proven to improve OS, but it is possible that variations in therapy may have affected survival in some patients. This genetic and treatment heterogeneity in our dataset may have lowered the overall AUC for nCBV association with OS in our dataset, but the observed association with OS nevertheless remains substantial. Since our analysis was designed to validate the newer more reproducible and automatable GN method of CBV analysis and compare it to the existing standard white matter normalization method, rather than to investigate the absolute strength of nCBV association with OS per se, this should not affect our conclusions. In other words, because the same test population with the identical perfusion and survival data was used for both the GN and conventional WN analyses, the heterogeneity should not introduce any bias into our comparison between these normalization methods. From a clinical translation point of view, this heterogeneous population is advantageous. Since it closely simulates a typical clinical population of patients with different tumor genetics undergoing different treatments, our results suggest that GN-nCBV should perform robustly in the clinic.

Conclusion

Both brain tumor nCBV maps produced by GN and by current standard manual NAWM ROI normalization correlate strongly with 15-month OS in our newly diagnosed GBM patients, but the GN-nCBV had a consistently stronger association and far lower interoperator and intersubject variability. This slightly better validity and superior reproducibility, combined with computational simplicity and potential for full automation, argue for implementation of fully automated GN in DSC processing software and for clinical and research use of the GN-nCBV maps produced. Implementation of GN by MRI and postprocessing software vendors is needed to allow widespread use of this technique and can be expected to improve patient care.
  30 in total

Review 1.  Intracranial mass lesions: dynamic contrast-enhanced susceptibility-weighted echo-planar perfusion MR imaging.

Authors:  Soonmee Cha; Edmond A Knopp; Glyn Johnson; Stephan G Wetzel; Andrew W Litt; David Zagzag
Journal:  Radiology       Date:  2002-04       Impact factor: 11.105

2.  Test-retest reliability of DSC MRI CBV mapping in healthy volunteers.

Authors:  M E Henry; M J Kaufman; N Lange; M E Schmidt; S Purcell; J Cote; D M Perron-Henry; E Stoddard; B M Cohen; P F Renshaw
Journal:  Neuroreport       Date:  2001-06-13       Impact factor: 1.837

3.  Comparison between intensity normalization techniques for dynamic susceptibility contrast (DSC)-MRI estimates of cerebral blood volume (CBV) in human gliomas.

Authors:  Benjamin M Ellingson; Taryar Zaw; Timothy F Cloughesy; Kourosh M Naeini; Shadi Lalezari; Sandy Mong; Albert Lai; Phioanh L Nghiemphu; Whitney B Pope
Journal:  J Magn Reson Imaging       Date:  2012-01-26       Impact factor: 4.813

4.  Relative cerebral blood volume values to differentiate high-grade glioma recurrence from posttreatment radiation effect: direct correlation between image-guided tissue histopathology and localized dynamic susceptibility-weighted contrast-enhanced perfusion MR imaging measurements.

Authors:  L S Hu; L C Baxter; K A Smith; B G Feuerstein; J P Karis; J M Eschbacher; S W Coons; P Nakaji; R F Yeh; J Debbins; J E Heiserman
Journal:  AJNR Am J Neuroradiol       Date:  2008-12-04       Impact factor: 3.825

5.  Signal-to-noise analysis of cerebral blood volume maps from dynamic NMR imaging studies.

Authors:  J L Boxerman; B R Rosen; R M Weisskoff
Journal:  J Magn Reson Imaging       Date:  1997 May-Jun       Impact factor: 4.813

6.  Correlation between cerebral blood volume measurements by perfusion-weighted magnetic resonance imaging and two-year progression-free survival in gliomas.

Authors:  M V Spampinato; C Schiarelli; A Cianfoni; P Giglio; C T Welsh; S Bisdas; Z Rumboldt
Journal:  Neuroradiol J       Date:  2013-08-27

7.  Spin-echo echo-planar perfusion prior to chemoradiation is a strong independent predictor of progression-free and overall survival in newly diagnosed glioblastoma.

Authors:  Ayca Akgoz; Rifaquat Rahman; Hui You; Jinrong Qu; Alhafidz Hamdan; Ravi T Seethamraju; Patrick Y Wen; Geoffrey S Young
Journal:  J Neurooncol       Date:  2014-05-04       Impact factor: 4.130

8.  Standardization of relative cerebral blood volume (rCBV) image maps for ease of both inter- and intrapatient comparisons.

Authors:  Devyani Bedekar; Todd Jensen; Kathleen M Schmainda
Journal:  Magn Reson Med       Date:  2010-09       Impact factor: 4.668

9.  Survival analysis in patients with glioblastoma multiforme: predictive value of choline-to-N-acetylaspartate index, apparent diffusion coefficient, and relative cerebral blood volume.

Authors:  Joonmi Oh; Roland G Henry; Andrea Pirzkall; Ying Lu; Xiaojuan Li; Isabelle Catalaa; Susan Chang; William P Dillon; Sarah J Nelson
Journal:  J Magn Reson Imaging       Date:  2004-05       Impact factor: 4.813

10.  Repeatability of Cerebral Perfusion Using Dynamic Susceptibility Contrast MRI in Glioblastoma Patients.

Authors:  Kourosh Jafari-Khouzani; Kyrre E Emblem; Jayashree Kalpathy-Cramer; Atle Bjørnerud; Mark G Vangel; Elizabeth R Gerstner; Kathleen M Schmainda; Kamran Paynabar; Ona Wu; Patrick Y Wen; Tracy Batchelor; Bruce Rosen; Steven M Stufflebeam
Journal:  Transl Oncol       Date:  2015-06       Impact factor: 4.243

View more
  1 in total

Review 1.  CDC7 as a novel biomarker and druggable target in cancer.

Authors:  Runze Liu; Yong Huang
Journal:  Clin Transl Oncol       Date:  2022-06-03       Impact factor: 3.340

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.