Daddy Mata-Mbemba1, Aissam Labani2, Soraya El Ghannudi2, Mi-Young Jeung2, Patrick Ohlmann3, Catherine Roy2, Mickaël Ohana2,4.
Abstract
OBJECTIVE: To reduce the iodine load required for CT Transcatheter Aortic Valve Replacement (TAVR) planning on a 320-row scanner by acquiring the two CT TAVR steps (ECG-gated aortic root CTA and non-gated aorto-ilio-femoral CTA) within a single contrast media bolus injection.Entities:
Mesh:
Substances:
Year: 2018 PMID: 30212567 PMCID: PMC6136810 DOI: 10.1371/journal.pone.0204145
Source DB: PubMed Journal: PLoS One ISSN: 1932-6203 Impact factor: 3.240
Description of the five-point likert scale used in qualitative analysis of CTA.
| Score | Diagnostic value | Description | Clinical value |
|---|---|---|---|
| 1 | Uninterpretable examination | Absent opacification, extremely severe artifacts | Non-diagnostic |
| 2 | Poor image quality | Insufficient opacification, major artifacts | |
| 3 | Acceptable image quality | Sufficient opacification, minor artifacts | Diagnostic |
| 4 | Good image quality | Excellent arterial opacification, negligible artifacts | |
| 5 | Excellent image quality | Excellent arterial opacification, no artifacts |
General characteristics of patients.
| Parameters | Study Group | Control Group | |
|---|---|---|---|
| 82.6 (±6.9) | 84.3 (±4.8) | 0.21 | |
| 0.85 | |||
| Male | 22 (44%) | 10 (42%) | |
| Female | 28 (56%) | 14 (58%) | |
| 0.89 | |||
| Mean | 26.2 (±5.5) | 26.1 (±5.2) | |
| Range | 17.6–40.8 | 16.7–34.8 | |
| 11 (22%) | 6 (25%) | 0.78 | |
| 32 (64%) | 17 (71%) | 0.56 | |
| 14 (28%) | 8 (33%) | 0.65 | |
| 28 (54%) | 11 (45%) | 0.42 | |
| 0.63 | |||
| Mean | 81.2 (±18.4) | 79.3 (±21.1) | |
| Range | 51–130 | 55–121 |
Fig 1Comparative example of image quality.
Up is the aortic root CTA, depicting the aortic annulus in axial and frontal planes. Down is the aortoiliac CTA in 3D and curved mutliplanar reconstructions. Study group images are from a 76 years-old man from the study group, whose BMI was 32.5, and were obtained after a single bolus injection of 70mL of Iohexol 350mgI/ml. Total radiation dose was 935mGy.cm. Control group images are from a 86 years-old man from the control group, with a BMI of 30.1, and were obtained after a 2 steps acquisition protocol using a total amount of 120mL of Iohexol 350mgI/ml. Total radiation dose was 1653 mGy.cm.
Detailed HU attenuation within the aortic root and the aortoiliac CTA.
| Measures | Study Group | Control Group | |
|---|---|---|---|
| HU attenuation | 480 ±139 | 372 ±114 | <0.01 |
| SD | 36 ±6 | 38 ±8 | 0.21 |
| SNR | 14 ±5.3 | 10.3 ±4.2 | <0.01 |
| CNR | 10.4 ±4.5 | 6.8 ±3.3 | <0.01 |
| HU attenuation | 364 ±170 | 400 ±95 | 0.16 |
| SD | 23 ±10 | 28 ±4 | <0.01 |
| SNR | 16.5 ±8 | 14.7 ±5.5 | 0.42 |
| CNR | 14.1 ±7.9 | 12.5 ±5 | 0.66 |
Numbers are given in a mean ±SD (minimum–maximum) fashion.