| Literature DB >> 30210044 |
Adriano Contillo1, Anna Veronese2, Luca Brombal2, Sandro Donato2, Luigi Rigon2, Angelo Taibi1, Giuliana Tromba3, Renata Longo2, Fulvia Arfelli2.
Abstract
Background The SYRMA-3D collaboration is setting up the first clinical trial of phase-contrast breast CT with synchrotron radiation at the Elettra synchrotron facility in Trieste, Italy. In this communication, a quality control protocol for breast CT is proposed, and a first test of image quality measurements is performed by means of a custom-made radiographic phantom. Materials and methods A set of projections is acquired and used to perform a CT reconstruction of two selected portions of the phantom. Such portions contain a uniform layer of water and a set of radiographic inserts, respectively. Together, they allow to perform several image quality measurements, namely CT number linearity, reconstruction accuracy, uniformity, noise, and low contrast resolution. All measurements are repeated at different beam energies in the range of interest, and at two different dose values. Results Measurements show a good linearity in the soft tissue range, paired to a high accuracy of the CT number reconstruction. Uniformity and noise measurements show that reconstruction inhomogeneities are bound to a few percent of the average pixel values. However, low contrast detectability is limited to the higher portion of the explored energy range. Conclusions The results of the measurements are satisfactory in terms of their quality, feasibility and reproducibility. With minimal modifications, the phantom is promising to allow a set of image quality measurements to be used in the upcoming clinical trial.Entities:
Keywords: breast CT; quality control; radiographic phantom; synchrotron radiation
Mesh:
Year: 2018 PMID: 30210044 PMCID: PMC6137353 DOI: 10.2478/raon-2018-0015
Source DB: PubMed Journal: Radiol Oncol ISSN: 1318-2099 Impact factor: 2.991
Figure 1(A) QC phantom fixed at the breast position on the patient support. The upper portion is uniformly filled with water, while the radiographic details are visible in the lower portion. (B) Arrangement of the details at the bottom of the QC phantom. The rods are made of polyethylene (PE), nylon, polyoxymethylene (POM), polytetrafluoroethylene (PTFE), and BR12, a plastic material mimicking the attenuation properties of breast tissue. The structure of the low contrast PMMA insert, exhibiting the five holes of varying diameter, is clearly perceivable.
Figure 2Example of reconstructed water image (left panel) and detail image (right panel), acquired at 38 keV and high dose.
Parameters of the linear regressions between measured CT numbers and tabulated attenuation coefficients
| Energy (keV) | Dose | R2 | slope (cm–1) | intercept (cm–1) |
|---|---|---|---|---|
| 32 | high | 0.9996 | 175 ± 4 | 0.019 ± 0.009 |
| low | 0.9996 | 170 ± 4 | 0.025 ± 0.009 | |
| 35 | high | 0.9997 | 176 ± 4 | 0.018 ± 0.008 |
| low | 0.9996 | 173 ± 4 | 0.017 ± 0.009 | |
| 38 | high | 0.9998 | 177 ± 3 | 0.016 ± 0.005 |
| low | 0.9992 | 175 ± 3 | 0.013 ± 0.006 | |
| 40 | high | 0.9992 | 174 ± 7 | 0.021 ± 0.012 |
| low | 0.9991 | 172 ± 7 | 0.019 ± 0.012 |
Figure 3Examples of linear regressions, performed on high dose images acquired at 32, 35, 38 and 40 keV, from top left panel to bottom right panel. R2 coefficients, slopes and intercepts are given as well.
Accuracies of reconstructed attenuation coefficients for the plastic materials and water, acquired at 38 keV and high dose. Expected values, measured values and relative percentual differences are given for μ coefficients, absolute differences for the corresponding HU deviations
| Material | expected μ (cm–1) | measured μ (cm–1) | accuracyμ (%) | accuracyHU |
|---|---|---|---|---|
| PE | 0.216 | 0.221 ± 0.002 | 3.1 ± 1.3 | 24 ± 10 |
| BR12 | 0.260 | 0.249 ± 0.004 | 4.0 ± 1.6 | 37 ± 14 |
| Nylon | 0.271 | 0.273 ± 0.003 | 0.9 ± 1.0 | 8 ± 9 |
| PMMA | 0.290 | 0.291 ± 0.003 | 0.4 ± 1.2 | 4 ± 12 |
| POM | 0.359 | 0.360 ± 0.003 | 0.2 ± 0.8 | 3 ± 11 |
| PTFE | 0.622 | 0.633 ± 0.003 | 0.1 ± 0.5 | 3 ± 11 |
| Water | 0.282 | 0.278 ± 0.003 | 1.5 ± 0.9 | 14 ± 9 |
BR12 = breast-tissue equivalent material; PE = polyethylene; PMMA = polymethyl methacrylate; POM = polyoxymethylene; PTFE = polytetrafluoroethylene
Uniformity of reconstructed attenuation coefficient for water. Relative percentual values are given for μ coefficients, absolute values for the corresponding HU deviations
| Energy (keV) | Dose | uniformityμ (%) | uniformityHU |
|---|---|---|---|
| 32 | high | 1.79 | 17 |
| low | 3.19 | 32 | |
| 35 | high | 1.45 | 12 |
| low | 1.47 | 13 | |
| 38 | high | 2.58 | 26 |
| low | 1.35 | 13 | |
| 40 | high | 2.54 | 24 |
| low | 1.07 | 10 |
Noise measurement of the reconstructed attenuation coefficient for water. Relative percentual results are given for μ coefficients, absolute results for the corresponding HU values
| Energy (keV) | Dose | noiseμ (%) | noiseHU |
|---|---|---|---|
| 32 | high | 4.4 | 60 |
| low | 8.5 | 103 | |
| 35 | high | 3.8 | 83 |
| low | 7.3 | 114 | |
| 38 | high | 3.7 | 66 |
| low | 6.5 | 109 | |
| 40 | high | 3.5 | 66 |
| low | 6.2 | 111 |
Low contrast measurements. Relative percentual contrasts are given for μ coefficients, absolute contrasts for the corresponding Hounsfield unit (HU)
| Energy (keV) | Detail (mm) | μ contrast (%) | HU contrast | ||
|---|---|---|---|---|---|
| high dose | low dose | high dose | low dose | ||
| 32 | 10 | 0.81 | 0.71 | 7 | 5 |
| 8 | 0.96 | 0.87 | 9 | 7 | |
| 6 | 1.10 | 0.53 | 10 | 4 | |
| 4 | 1.08 | 1.12 | 10 | 10 | |
| 2 | 0.52 | 0.93 | 5 | 8 | |
| 35 | 10 | 1.80 | 1.85 | 15 | 15 |
| 8 | 1.68 | 1.73 | 14 | 14 | |
| 6 | 1.56 | 1.64 | 13 | 13 | |
| 4 | 1.21 | 0.87 | 10 | 8 | |
| 2 | 0.90 | 0.91 | 8 | 8 | |
| 38 | 10 | 4.14 | 3.86 | 41 | 39 |
| 8 | 3.99 | 3.41 | 40 | 34 | |
| 6 | 3.66 | 4.02 | 36 | 41 | |
| 4 | 3.91 | 3.45 | 39 | 35 | |
| 2 | 3.17 | 3.79 | 31 | 38 | |
| 40 | 10 | 5.35 | 5.49 | 54 | 55 |
| 8 | 5.09 | 5.43 | 52 | 55 | |
| 6 | 5.12 | 4.85 | 52 | 49 | |
| 4 | 5.31 | 3.41 | 54 | 34 | |
| 2 | 5.73 | 3.79 | 59 | 38 | |