GEANT4 for breast dosimetry: parameters optimization study.

C Fedon1, F Longo, G Mettivier, R Longo.   

Abstract

Mean glandular dose (MGD) is the main dosimetric quantity in mammography. MGD evaluation is obtained by multiplying the entrance skin air kerma (ESAK) by normalized glandular dose (DgN) coefficients. While ESAK is an empirical quantity, DgN coefficients can only be estimated with Monte Carlo (MC) methods. Thus, a MC parameters benchmark is needed for effectively evaluating DgN coefficients. GEANT4 is a MC toolkit suitable for medical purposes that offers to the users several computational choices. In this work we investigate the GEANT4 performances testing the main PhysicsLists for medical applications. Four electromagnetic PhysicsLists were implemented: the linear attenuation coefficients were calculated for breast glandularity 0%, 50%, 100% in the energetic range 8-50 keV and DgN coefficients were evaluated. The results were compared with published data. Fit equations for the estimation of the G-factor parameter, introduced by the literature for converting the dose delivered in the heterogeneous medium to that in the glandular tissue, are proposed and the application of this parameter interaction-by-interaction or retrospectively is discussed. G4EmLivermorePhysicsList shows the best agreement for the linear attenuation coefficients both with theoretical values and published data. Moreover, excellent correlation factor (r2>0.99) is found for the DgN coefficients with the literature. The final goal of this study is to identify, for the first time, a benchmark of parameters that could be useful for future breast dosimetry studies with GEANT4.

Mesh:

Year:  2015        PMID: 26267405     DOI: 10.1088/0031-9155/60/16/N311

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  13 in total

1.  Monochromatic breast computed tomography with synchrotron radiation: phase-contrast and phase-retrieved image comparison and full-volume reconstruction.

Authors:  Luca Brombal; Bruno Golosio; Fulvia Arfelli; Deborah Bonazza; Adriano Contillo; Pasquale Delogu; Sandro Donato; Giovanni Mettivier; Piernicola Oliva; Luigi Rigon; Angelo Taibi; Giuliana Tromba; Fabrizio Zanconati; Renata Longo
Journal:  J Med Imaging (Bellingham)       Date:  2018-11-27

2.  A Monte Carlo model for mean glandular dose evaluation in spot compression mammography.

Authors:  Antonio Sarno; David R Dance; Ruben E van Engen; Kenneth C Young; Paolo Russo; Francesca Di Lillo; Giovanni Mettivier; Kristina Bliznakova; Baowei Fei; Ioannis Sechopoulos
Journal:  Med Phys       Date:  2017-06-13       Impact factor: 4.071

3.  Comparison of propagation-based CT using synchrotron radiation and conventional cone-beam CT for breast imaging.

Authors:  Seyedamir Tavakoli Taba; Patrycja Baran; Yakov I Nesterets; Serena Pacile; Susanne Wienbeck; Christian Dullin; Konstantin Pavlov; Anton Maksimenko; Darren Lockie; Sheridan C Mayo; Harry M Quiney; Diego Dreossi; Fulvia Arfelli; Giuliana Tromba; Sarah Lewis; Timur E Gureyev; Patrick C Brennan
Journal:  Eur Radiol       Date:  2020-01-23       Impact factor: 5.315

4.  Optimization of the energy for Breast monochromatic absorption X-ray Computed Tomography.

Authors:  Pasquale Delogu; Vittorio Di Trapani; Luca Brombal; Giovanni Mettivier; Angelo Taibi; Piernicola Oliva
Journal:  Sci Rep       Date:  2019-09-11       Impact factor: 4.379

5.  Monte Carlo study on optimal breast voxel resolution for dosimetry estimates in digital breast tomosynthesis.

Authors:  Christian Fedon; Carolina Rabin; Marco Caballo; Oliver Diaz; Eloy García; Alejandro Rodríguez-Ruiz; Gabriel A González-Sprinberg; Ioannis Sechopoulos
Journal:  Phys Med Biol       Date:  2018-12-19       Impact factor: 3.609

6.  Report on G4-Med, a Geant4 benchmarking system for medical physics applications developed by the Geant4 Medical Simulation Benchmarking Group.

Authors:  P Arce; D Bolst; M-C Bordage; J M C Brown; P Cirrone; M A Cortés-Giraldo; D Cutajar; G Cuttone; L Desorgher; P Dondero; A Dotti; B Faddegon; C Fedon; S Guatelli; S Incerti; V Ivanchenko; D Konstantinov; I Kyriakou; G Latyshev; A Le; C Mancini-Terracciano; M Maire; A Mantero; M Novak; C Omachi; L Pandola; A Perales; Y Perrot; G Petringa; J M Quesada; J Ramos-Méndez; F Romano; A B Rosenfeld; L G Sarmiento; D Sakata; T Sasaki; I Sechopoulos; E C Simpson; T Toshito; D H Wright
Journal:  Med Phys       Date:  2020-12-12       Impact factor: 4.071

7.  Effect of thyroid shielding during mammography: measurements on phantom and patient as well as estimation with Monte Carlo simulation.

Authors:  Miriam Pyka; Patrik Eschle; Christian Sommer; Mathias S Weyland; Rahel Kubik; Stephan Scheidegger
Journal:  Eur Radiol Exp       Date:  2018-06-28

8.  A proposal for a quality control protocol in breast CT with synchrotron radiation.

Authors:  Adriano Contillo; Anna Veronese; Luca Brombal; Sandro Donato; Luigi Rigon; Angelo Taibi; Giuliana Tromba; Renata Longo; Fulvia Arfelli
Journal:  Radiol Oncol       Date:  2018-03-24       Impact factor: 2.991

9.  Internal breast dosimetry in mammography: Monte Carlo validation in homogeneous and anthropomorphic breast phantoms with a clinical mammography system.

Authors:  Christian Fedon; Marco Caballo; Ioannis Sechopoulos
Journal:  Med Phys       Date:  2018-06-29       Impact factor: 4.071

10.  Motion artifacts assessment and correction using optical tracking in synchrotron radiation breast CT.

Authors:  Luca Brombal; Lucia Mariel Arana Peña; Fulvia Arfelli; Renata Longo; Francesco Brun; Adriano Contillo; Francesca Di Lillo; Giuliana Tromba; Vittorio Di Trapani; Sandro Donato; Ralf Hendrik Menk; Luigi Rigon
Journal:  Med Phys       Date:  2021-07-29       Impact factor: 4.506

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.