Peter Maloca1,2,3, Pascal W Hasler1,2, Daniel Barthelmes4,5, Patrik Arnold6, Mooser Matthias6, Hendrik P N Scholl7,2,8, Heinrich Gerding9,10, Justus Garweg11, Tjebo Heeren12, Konstantinos Balaskas13,3, J Emanuel Ramos de Carvalho3, Catherine Egan3, Adnan Tufail3, Sandrine A Zweifel4.
Abstract
PURPOSE: To study a novel and fast optical coherence tomography (OCT) device for home-based monitoring in age-related macular degeneration (AMD) in a small sample yielding sparse OCT (spOCT) data and to compare the device to a commercially available reference device.Entities:
Keywords: age-related macular degeneration; monitoring; optical coherence tomography; retina
Year: 2018 PMID: 30050725 PMCID: PMC6058910 DOI: 10.1167/tvst.7.4.8
Source DB: PubMed Journal: Transl Vis Sci Technol ISSN: 2164-2591 Impact factor: 3.283
Figure 1Comparison of standard Heidelberg Spectralis desktop OCT to novel investigational spOCT. (A) Scanning laser ophthalmoscopy (SLO) of AMD. (B) Corresponding cross-section scan of the same eye as in (A) with illustration of manual central retinal thickness (CRT) measurement of standard Spectralis OCT (B) and spOCT (C). Note that in the spOCT system the scan resolution, thus the redundancy of image information, is reduced to allow rapid and repeated measurement of OCT volumes. However, despite the low sampling, CRT measurement was achieved in spOCT images comparable to the reference device.
Figure 2Illustration and performance of the novel sparseOCT device (spOCT, MIMO_02) compared to standard Heidelberg Spectralis desktop OCT. (A) Illustration of the complete prototype spOCT (black box) shows compact design and novel headrest position to reduce moving artifacts. (B) The direction of the light beam can be adjusted to the position of the subject to provide a comfortable user experience. High sample recording of the single line scan of a healthy macula obtained by spOCT (C) compared to Heidelberg desktop OCT (D) of the same eye with comparable scan settings.
Figure 3Bland-Altman plot of difference against the mean of spOCT scanner and Heidelberg Spectralis (reference) measurements of CRT (μm). The lines represent mean difference (solid) and mean ±1.96 standard deviations (dashed) as the 95% limits of agreement. Sixty-nine percent of the differences are in the range of ±20 μm (dotted lines).
Summary of Patient Feedback on the Examination Performed With the Investigational, spOCT Device and Reference OCT System