| Literature DB >> 35557997 |
Itika Garg1,2, Raviv Katz1, Yifan Lu1,2, John B Miller1,2.
Abstract
Purpose: The COVID-19 pandemic has had a widespread impact on the health sector worldwide, both in the clinical and research subsector. This has necessitated for the invention of protocols compliant with social distancing policies to efficiently tackle the current needs. Our aim is to describe a novel remote imaging technique, which helped us continue our research with widefield swept-source optical coherence tomography angiography (WF SS-OCTA). Patients andEntities:
Keywords: COVID; COVID-19; optical coherence tomography; optical coherence tomography angiography; remote imaging; retinal imaging; teleimaging; telemedicine; teleophthalmology
Year: 2022 PMID: 35557997 PMCID: PMC9089192 DOI: 10.2147/OPTH.S352503
Source DB: PubMed Journal: Clin Ophthalmol ISSN: 1177-5467
Figure 1Workflow schema outlining the novel remote imaging approach used. (A) Both the on-site and off-site fellow communicate with each other as a part of the initial preparation stage. (B) The remote fellow then uses the institution approved remote computer to access an on-site workstation using the institutional user profile, (C) further connecting to and gaining full access to the advanced retinal imaging device. (D) Simultaneously, the on-site research fellow explains the study protocol and obtains a detailed written informed consent, (E) following which they accompany the patient to imaging room with the retinal imaging device and the on-site encrypted tablet computer device. (F) As the last step, the remote fellow uses the remote desktop to navigate the device for imaging and the remote encrypted tablet computer to communicate with the patient.
Figure 2Representative images of patients being remotely imaged using widefield swept-source optical coherence tomography angiography device by an experienced research fellow via an institute approved remote desktop. The relevant communication was done via video conference call on an encrypted tablet device with hospital approved video call application. (Written patient consent obtained for being photographed here for the purpose of publication).
Demographic Characteristics of Patients Imaged Remotely, Along with the Resultant Image Quality Metrics
| Patient Number | Age (Years) | Sex | Eye | Scan Protocol | Scan Length (mm) | Scan Depth (mm) | Signal Strength | Scan Duration (Seconds) |
|---|---|---|---|---|---|---|---|---|
| 1 | 69 | Female | OD | Angio (3 mm x 3 mm) | 3 | 3 | 9 | 8.1 |
| Angio (6 mm x 6 mm) | 6 | 3 | 10 | 11.6 | ||||
| Angio (12 mm x12 mm) | 12 | 3 | 9 | 18.4 | ||||
| OS | Angio (3 mm x 3 mm) | 3 | 3 | 9 | 11.5 | |||
| Angio (6 mm x 6 mm) | 6 | 3 | 10 | 14.3 | ||||
| Angio (12 mm x 12 mm) | 12 | 3 | 9 | 42.4 | ||||
| HD Angio (12 mm x 12mm) | 12 | 6 | 8 | 39.7 | ||||
| 2 | 67 | Male | OD | Angio (3 mm x 3 mm) | 3 | 3 | 8 | 7.2 |
| Angio (6 mm x 6 mm) | 6 | 3 | 9 | 9.4 | ||||
| Angio (12 mm x 12 mm) | 12 | 3 | 8 | 9.3 | ||||
| Angio (12 mm x 12 mm) | 12 | 3 | 8 | 12.2 | ||||
| Angio (6 mm x 6 mm) | 6 | 3 | 9 | 11.9 | ||||
| OS | Angio (3 mm x 3 mm) | 3 | 3 | 9 | 8.8 | |||
| Angio (6 mm x 6 mm) | 6 | 3 | 10 | 10.2 | ||||
| Angio (6 mm x 6 mm) | 6 | 3 | 10 | 11.5 | ||||
| Angio (12 mm x 12 mm) | 12 | 3 | 9 | 17.8 | ||||
| Angio (12 mm x 12 mm) | 12 | 3 | 9 | 22.2 | ||||
| 3 | 32 | Female | OD | Angio (3 mm x 3 mm) | 3 | 3 | 9 | 25.3 |
| Angio (6 mm x 6 mm) | 6 | 3 | 10 | 17.6 | ||||
| Angio (6 mm x 6 mm) | 6 | 3 | 10 | 12.6 | ||||
| Angio (12 mm x 12 mm) | 12 | 3 | 9 | 26.2 | ||||
| OS | Angio (3 mm x 3 mm) | 3 | 3 | 10 | 8.1 | |||
| Angio (6 mm x 6 mm) | 6 | 3 | 10 | 16.0 | ||||
| Angio (6 mm x 6 mm) | 6 | 3 | 10 | 13.8 | ||||
| Angio (12 mm x 12 mm) | 12 | 3 | 9 | 34.4 | ||||
| Angio (12 mm x 12 mm) | 12 | 3 | 9 | 56.3 | ||||
| 4 | 66 | Male | OD | Angio (3 mm x 3 mm) | 3 | 3 | 9 | 9.1 |
| Angio (6 mm x 6 mm) | 6 | 3 | 10 | 11.2 | ||||
| Angio (12 mm x 12 mm) | 12 | 3 | 10 | 17.1 | ||||
| OS | Angio (3 mm x 3 mm) | 3 | 3 | 9 | 8.5 | |||
| Angio (6 mm x 6 mm) | 6 | 3 | 10 | 51.0 | ||||
| Angio (12 mm x 12 mm) | 12 | 3 | 9 | 58.0 | ||||
| Mean | 9.25 | 19.7 | ||||||
| Median | 9 | 13.2 | ||||||
Figure 3Representative Angio 3 mm x 3 mm (A), Angio 6 mm x 6 mm (B), 12 mm x 12 mm scans (C), and the Cube 800×800 B-scan (D) from the right eye of patient number 1 in Table 1, obtained via the remote imaging technique outlined.