Kai Joachim Borm1, Markus Oechsner1, Moritz Wiegandt1,2, Andreas Hofmeister1,2, Stephanie E Combs1,3, Marciana Nona Duma4,5.
Abstract
PURPOSE: Maximum (MIP) and average intensity projection (AIP) CTs allow rapid definition of internal target volumes in a 4D-CT. The purpose of this study was to assess the accuracy of these techniques in a large patient cohort in combination with simulations on a lung phantom.Entities:
Keywords: 4D-CT; AIP; MIP; Moving targets; SBRT
Mesh:
Year: 2018 PMID: 30041618 PMCID: PMC6056919 DOI: 10.1186/s12885-018-4647-4
Source DB: PubMed Journal: BMC Cancer ISSN: 1471-2407 Impact factor: 4.430
Fig. 1Illustration of the lung phantom (a) and delineation examples of the two targets (Ø 1 cm and Ø 2 cm) in one out of 10 phases of the 4DCT (b), in average intensity projection (c), in maximum intensity projection (d)
Fig. 2Lung tumors (*) adjoining different structures in the thorax: A) only lung tissue; B) the mediastinum; C) the diaphragm; D) the chest wall
Fig. 3Comparison of values measured in the phantom study (a,c) and in the clinical study (b,d). Relative size (c,d) and conformation numbers (a,b) of ITVs contoured in MIP and AIP in relation to the ITV10
Measured values in the lung phantom. Average absolute volumes±SD of the ITV10, ITVMIP, ITVAIP in cm3,(relative volumes±SD as percentage of the calculated values -ITVcalc- in %), conformation numbers (CN) of ITVMIP and ITVAIP with the ITV10. n = number of movement patterns
| Ø | n= | ITVcalc | ITV10 | ITVMIP | ITVAIP | CNMIP | CNAIP |
|---|---|---|---|---|---|---|---|
| 1.0 cm | 10 | 8.6 ± 1.3 cm3 | 9.3 ± 1.5 cm3 | 8.3 ± 10.9 cm3 | 7.3 ± 0.9 cm3 | 0.77 ± 0.2 | 0.77 ± 0.1 |
| 2.0 cm | 10 | 50.3 ± 4.8 cm3 | 54.8 ± 4.2 cm3 | 49.6 ± 3.8 cm3 | 45.7 ± 4.1 cm3 | 0.89 ± 0.0 | 0.89 ± 0.0 |
| Total | 20 | 29.4 ± 21.7 cm3 | 32.1 ± 23.5 cm3 | 28.9 ± 21.4 cm3 | 26.5 ± 19.9 cm3 | 0.79 ± 0.2 | 0.83 ± 0.1 |
Fig. 4Impact of the surrounding tissue (a,c) and motion amplitude (b,d) on the contoured target volumes in MIP and AIP. Relative size (a,b) in % and conformation numbers (c,d) of ITVs contoured in MIP and AIP in regard to the ITV10 CN: conformation number, PTT: peak-to-through motion
Measured values for ITVMIP and ITVAIP in dependence on the surrounding tissue and the tumor size in a 4D-CT. Relative size in % in regard to the ITV10 and overlapping volume (VOL) in cm3 as well as conformation numbers (CN) of IT VMIP and ITVAIP with the ITV10
| n= | MIP | AIP | ||||||
|---|---|---|---|---|---|---|---|---|
| % | VOL | CN | % | VOL | CN | |||
| Tumor surrouding | only lung tissue | 10 | 79.8 ± 11.6 | 77.5 ± 9.6 | 0.75 ± 0.1 | 68.4 ± 14.1 | 67.5 ± 13.3 | 0.67 ± 0.1 |
| chestwall | 20 | 76.4 ± 14.1 | 73.6 ± 11.3 | 0.71 ± 0.1 | 66.3 ± 11.8 | 65.2 ± 11.0 | 0.64 ± 0.1 | |
| mediastinum | 16 | 71.0 ± 15.4 | 68.9 ± 13.9 | 0.67 ± 0.1 | 61.8 ± 15.3 | 60.7 ± 14.5 | 0.59 ± 0.1 | |
| diaphragm | 4 | 46.1 ± 11.1 | 43.7 ± 11.1 | 0.41 ± 0.1 | 59.1 ± 11.9 | 58.0 ± 12.4 | 0.57 ± 0.1 | |
| Tumor size (Diameter) | ≤2.5 cm | 14 | 67.2 ± 19.4 | 64.8 ± 17.9 | 0.63 ± 0.2 | 55.2 ± 16.0 | 54.0 ± 14.7 | 0.53 ± 0.1 |
| 2.6–3.4 cm | 19 | 75.1 ± 13.9 | 72.6 ± 11.5 | 0.72 ± 0.1 | 66.0 ± 12.5 | 65.0 ± 11.8 | 0.64 ± 0.1 | |
| ≥3.5 cm | 17 | 76.7 ± 13.3 | 74.1 ± 11.7 | 0.71 ± 0.1 | 71.3 ± 6.9 | 70.1 ± 6.4 | 0.69 ± 0.1 | |
| PTT | < 0.5 cm | 21 | 77.1 ± 16.8 | 73.8 ± 15.0 | 0.71 ± 0.1 | 68.4 ± 10.4 | 67.2 ± 9.9 | 0.65 ± 0.1 |
| 0.5–1 cm | 22 | 72.5 ± 13.8 | 70.2 ± 12.1 | 0.68 ± 0.1 | 64.1 ± 12.8 | 63.1 ± 12.1 | 0.62 ± 0.1 | |
| > 1 cm | 7 | 65.5 ± 16.3 | 64.6 ± 15.3 | 0.63 ± 0.1 | 55.9 ± 20.2 | 55.0 ± 19.0 | 0.54 ± 0.1 | |
| Total | 50 | 73.4 ± 15.6 | 71.0 ± 13.9 | 0.69 ± 0.1 | 64.7 ± 13.5 | 63.7 ± 12.8 | 0.63 ± 0.1 | |