Commissioning and dosimetric characteristics of TrueBeam system: composite data of three TrueBeam machines.

Zheng Chang1, Qiuwen Wu, Justus Adamson, Lei Ren, James Bowsher, Hui Yan, Andrew Thomas, Fang-Fang Yin.   

Abstract

PURPOSE: A TrueBeam linear accelerator (TB-LINAC) is designed to deliver traditionally flattened and flattening-filter-free (FFF) beams. Although it has been widely adopted in many clinics for patient treatment, limited information is available related to commissioning of this type of machine. In this work, commissioning data of three units were measured, and multiunit comparison was presented to provide valuable insights and reliable evaluations on the characteristics of the new treatment system.
METHODS: The TB-LINAC is equipped with newly designed waveguide, carousel assembly, monitoring control, and integrated imaging systems. Each machine in this study has 4, 6, 8, 10, 15 MV flattened photon beams, and 6 MV and 10 MV FFF photon beams as well as 6, 9, 12, 16, 20, and 22 MeV electron beams. Dosimetric characteristics of the three new TB-LINAC treatment units are systematically measured for commissioning. High-resolution diode detectors and ion chambers were used to measure dosimetric data for a range of field sizes from 10 × 10 to 400 × 400 mm(2). The composite dosimetric data of the three units are presented in this work. The commissioning of intensity modulated radiotherapy (IMRT), volumetric modulated arc therapy (VMAT), image-guided radiation therapy, and gating systems are also illustrated. Critical considerations of P(ion) of FFF photon beams and small field dosimetric measurements were investigated.
RESULTS: The authors found all PDDs and profiles matched well among the three machines. Beam data were quantitatively compared and combined through average to yield composite beam data. The discrepancies among the machines were quantified using standard deviation (SD). The mean SD of the PDDs among the three units is 0.12%, and the mean SD of the profiles is 0.40% for 10 MV FFF open fields. The variations of P(ion) of the chamber CC13 is 1.2 ± 0.1% under 6 MV FFF and 2.0 ± 0.5% under 10 MV FFF from dmax to the 18 cm-off-axis point at 35 cm depth under 40 × 40 cm(2). The mean penumbra of crossplane flattened photon beams at collimator angle of 0° is measured from 5.88 ± 0.09 to 5.99 ± 0.13 mm from 4 to 15 MV at 10 cm depth of 100 × 100 mm(2). The mean penumbra of crossplane beams at collimator angle of 0° is measured as 3.70 ± 0.21 and 4.83 ± 0.04 mm for 6 MV FFF and 10 MV FFF, respectively, at 10 cm depth with a field size of 5 × 5 cm(2). The end-to-end test procedures of both IMRT and VMAT were performed for various energy modes. The mean ion chamber measurements of three units showed less than 2% between measurement and calculation; the mean MultiCube ICA measurements demonstrated over 90% pixels passing gamma analysis (3%, 3 mm, 5% threshold). The imaging dosimetric data of KV planar imaging and CBCT demonstrated improved consistency with vendor specifications and dose reduction for certain imaging protocols. The gated output verification showed a discrepancy of 0.05% or less between gating radiation delivery and nongating radiation delivery.
CONCLUSIONS: The commissioning data indicated good consistency among the three TB-LINAC units. The commissioning data provided us valuable insights and reliable evaluations on the characteristics of the new treatment system. The systematically measured data might be useful for future reference.

Entities:  

Mesh:

Year:  2012        PMID: 23127092     DOI: 10.1118/1.4762682

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  43 in total

1.  Dosimetric accuracy of dynamic couch rotation during volumetric modulated arc therapy (DCR-VMAT) for primary brain tumours.

Authors:  Gregory Smyth; Philip M Evans; Jeffrey C Bamber; Henry C Mandeville; A Rollo Moore; Liam C Welsh; Frank H Saran; James L Bedford
Journal:  Phys Med Biol       Date:  2019-04-05       Impact factor: 3.609

2.  Technical Report: Evaluation of peripheral dose for flattening filter free photon beams.

Authors:  E L Covington; T A Ritter; J M Moran; A M Owrangi; J I Prisciandaro
Journal:  Med Phys       Date:  2016-08       Impact factor: 4.071

3.  Evaluation of beam modeling for small fields using a flattening filter-free beam.

Authors:  Daisuke Kawahara; Shuichi Ozawa; Takeo Nakashima; Masamichi Aita; Shintaro Tsuda; Yusuke Ochi; Takuro Okumura; Hirokazu Masuda; Yoshimi Ohno; Yuji Murakami; Yasushi Nagata
Journal:  Radiol Phys Technol       Date:  2016-06-21

4.  Technical Report: Reference photon dosimetry data for Varian accelerators based on IROC-Houston site visit data.

Authors:  James R Kerns; David S Followill; Jessica Lowenstein; Andrea Molineu; Paola Alvarez; Paige A Taylor; Francesco C Stingo; Stephen F Kry
Journal:  Med Phys       Date:  2016-05       Impact factor: 4.071

5.  Optimization of rotational arc station parameter optimized radiation therapy.

Authors:  P Dong; B Ungun; S Boyd; L Xing
Journal:  Med Phys       Date:  2016-09       Impact factor: 4.071

6.  Modification and validation of an analytical source model for external beam radiotherapy Monte Carlo dose calculations.

Authors:  Scott E Davidson; Jing Cui; Stephen Kry; Joseph O Deasy; Geoffrey S Ibbott; Milos Vicic; R Allen White; David S Followill
Journal:  Med Phys       Date:  2016-08       Impact factor: 4.071

7.  Beam data modeling of linear accelerators (linacs) through machine learning and its potential applications in fast and robust linac commissioning and quality assurance.

Authors:  Wei Zhao; Ishan Patil; Bin Han; Yong Yang; Lei Xing; Emil Schüler
Journal:  Radiother Oncol       Date:  2020-10-08       Impact factor: 6.280

8.  Validation of Dosimetric Leaf Gap (DLG) prior to its implementation in Treatment Planning System (TPS): TrueBeam™ millennium 120 leaf MLC.

Authors:  Ravindra Shende; Ganesh Patel
Journal:  Rep Pract Oncol Radiother       Date:  2017-10-21

9.  Evaluation of the Differences Between Measurements in Multiple Institutions and Calculation Modeled by Representative Beam Data in Prostate VMAT Plan.

Authors:  Hironao Goto; Hirokazu Mizuno; Yuichi Akino; Masaru Isono; Yoshihiro Tanaka; Norihisa Masai; Toshijiro Yamamoto; Masahiko Koizumi
Journal:  In Vivo       Date:  2020 May-Jun       Impact factor: 2.155

10.  Automatically gated image-guided breath-hold IMRT is a fast, precise, and dosimetrically robust treatment for lung cancer patients.

Authors:  Anna Simeonova-Chergou; Anika Jahnke; Kerstin Siebenlist; Florian Stieler; Sabine Mai; Judit Boda-Heggemann; Frederik Wenz; Frank Lohr; Lennart Jahnke
Journal:  Strahlenther Onkol       Date:  2016-01-15       Impact factor: 3.621

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.