| Literature DB >> 29937782 |
Jan Gilden1, Kristen Lewis2, Ginger Grant3, Jillian Crosson3.
Abstract
OBJECTIVE: To demonstrate the performance benefit of the Automatic Scene Classifier (SCAN) algorithm available in the Nucleus® 6 (CP900 series) sound processor over the default processing algorithms of the previous generation Nucleus 5 (CP810) and Freedom® Hybrid™ sound processors.Entities:
Keywords: Automatic scene classifier (SCAN); Cochlear implant; Nucleus 6; SmartSound iQ; Speech perception
Year: 2015 PMID: 29937782 PMCID: PMC6002557 DOI: 10.1016/j.joto.2015.09.001
Source DB: PubMed Journal: J Otol ISSN: 1672-2930
Subject demographics by group.
| Nucleus 5 Group | Hybrid group |
|---|---|
| 4 sites | 9 sites |
| 40 traditional CI recipients | 42 Hybrid CI recipients |
| Mean age = 47 yrs | Mean age = 67 yrs |
| CI24RE, CI512, CI422 implants | Hybrid L24 implants |
| Nucleus 5 sound processor | Freedom Hybrid sound processor |
Study schedule and test conditions by group.
| Nucleus 5 group | Hybrid group |
|---|---|
Test Nucleus 5 (ASC + ADRO) Test Nucleus 6 Custom (ASC + ADRO) Test Nucleus 6 Default (SCAN) | Test Freedom Hybrid (ASC + ADRO) Test Nucleus 6 Custom (ASC + ADRO) Test Nucleus 6 Default (SCAN) |
Test Nucleus 6 Default (SCAN) | Test Nucleus 6 Default (SCAN) |
Fig. 1AzBio scores by group at Visit 1 and Visit 2.
Fig. 2Results of the Speech, Spatial and Qualities of Hearing Scale (Comparative) by subtests for Nucleus 5 group.