Local organ dose conversion coefficients for angiographic examinations of coronary arteries.

H Schlattl1, M Zankl, J Hausleiter, C Hoeschen.   

Abstract

New organ dose conversion coefficients for coronary angiographic interventions are presented, as well as dose distributions and resulting maximal local dose conversion coefficients in the relevant organs. For the Monte Carlo based simulations, voxel models of the human anatomy were employed which represent the average Caucasian adult man and woman as defined by the International Commission on Radiological Protection. In the 21 investigated projections, the mean organ dose conversion coefficients vary from a few 0.01 to 2 mGy(Gy cm(2))(-1), depending on the projections. However, especially in portions of the lungs and the active bone marrow, the conversion coefficients can locally amount up to 10 mGy(Gy cm(2))(-1), which is half the average conversion coefficient of the skin at the field entrance. In addition to the dose conversion coefficients, the dependence of the patient dose on the projection has been estimated. It could be shown that the patient doses are highest for left anterior oblique views with strong caudal or cranial orientation. Nevertheless, for a large range of image-intensifier positions no significant dose differences could be found.

Entities:  

Mesh:

Year:  2007        PMID: 17634640     DOI: 10.1088/0031-9155/52/15/003

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  4 in total

1.  DNA double-strand breaks as potential indicators for the biological effects of ionising radiation exposure from cardiac CT and conventional coronary angiography: a randomised, controlled study.

Authors:  Dominik Geisel; Elke Zimmermann; Matthias Rief; Johannes Greupner; Michael Laule; Fabian Knebel; Bernd Hamm; Marc Dewey
Journal:  Eur Radiol       Date:  2012-04-17       Impact factor: 5.315

Review 2.  Task-based measures of image quality and their relation to radiation dose and patient risk.

Authors:  Harrison H Barrett; Kyle J Myers; Christoph Hoeschen; Matthew A Kupinski; Mark P Little
Journal:  Phys Med Biol       Date:  2015-01-07       Impact factor: 3.609

3.  Organ doses in pediatric patients undergoing cardiac-centered fluoroscopically guided interventions: Comparison of three methods for computational phantom alignment.

Authors:  Emily L Marshall; David Borrego; James C Fudge; Dhanashree Rajderkar; Wesley E Bolch
Journal:  Med Phys       Date:  2018-06-13       Impact factor: 4.071

4.  Computed tomography dosimetry with high-resolution detectors commonly used in radiotherapy - an energy dependence study.

Authors:  Mario Liebmann; Bjoern Poppe; Heiner von Boetticher
Journal:  J Appl Clin Med Phys       Date:  2015-09-08       Impact factor: 2.102

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.