Clinical and hip simulator comparisons of ceramic-on-polyethylene and metal-on-polyethylene wear.

I C Clarke1, A Gustafson.   

Abstract

The benefit of reduced polyethylene wear with ceramic in hip replacements does not seem to have been universally appreciated. In this current study, wear predictions from laboratory and clinical studies were compared for ceramic-on-polyethylene and cobalt chrome-on-polyethylene combinations. Many laboratory studies included water-based lubrication and linear-tracking mechanisms. Now it is appreciated that these were inappropriate methods, because of a propensity for very low or virtually no polyethylene wear against ceramics in water. Thus, water-based studies predicting a 20- to 80-fold advantage for ceramic-on-polyethylene compared with metal-on-polyethylene clearly were in error. However, serum-based simulator studies with high protein-concentrations generally have shown greater wear with alumina-on-polyethylene than with metal-on-polyethylene. Controversy still remains over the use of such nonphysiologic protein levels. The simulator studies were just beginning to explore the role of serum protein concentrations and the influence on the various wear models. Polyethylene wear with zirconia systems was particularly affected by serum protein concentrations. In one simulator study, use of proteins in the physiologic range resulted in the alumina-on-polyethylene wear rate decreasing to approximately 50% of that of metal-on-polyethylene. In the literature, many hip design and polyethylene variations were reported which confounded the wear analysis. Overall, the clinical data supported the superior performance of ceramic-on-polyethylene systems by a factor of 1.5- to fourfold. However, the amount of supporting data was not large. This summary of laboratory and clinical data indicated that ceramic-on-polyethylene hip replacement systems offered on average a 50% wear reduction from metal-on-polyethylene systems.

Entities:  

Mesh:

Substances:

Year:  2000        PMID: 11039790     DOI: 10.1097/00003086-200010000-00006

Source DB:  PubMed          Journal:  Clin Orthop Relat Res        ISSN: 0009-921X            Impact factor:   4.176


  15 in total

1.  Wear performance of cobalt chromium, ceramic, and oxidized zirconium on highly crosslinked polyethylene at mid-term follow-up.

Authors:  Matthew G Teeter; Colin J MacLean; Lyndsay E Somerville; James L Howard; Richard W McCalden; Brent A Lanting; Edward M Vasarhelyi
Journal:  J Orthop       Date:  2018-05-07

Review 2.  [Implant wear and aseptic loosening. An overview].

Authors:  C Kaddick; I Catelas; P H Pennekamp; M A Wimmer
Journal:  Orthopade       Date:  2009-08       Impact factor: 1.087

3.  Modular sleeves with ceramic heads in isolated acetabular cup revision in younger patients-laboratory and experimental analysis of suitability and clinical outcomes.

Authors:  Peter Helwig; Lukas Konstantinidis; Anja Hirschmüller; Anke Bernstein; Oliver Hauschild; Norbert P Südkamp; Björn G Ochs
Journal:  Int Orthop       Date:  2012-12-07       Impact factor: 3.075

4.  Two Year Clinical Outcomes of Total Hip Arthroplasty Are Not Dependent on Femoral Head Composition.

Authors:  Michael P Ast; Stephen Lyman; Alexandra H Gorab; Natalie Parkes; John Boles; Yuo-Yu Lee; Geoffrey H Westrich
Journal:  HSS J       Date:  2015-03-20

5.  Reduced wear of cross-linked UHMWPE using magnesia-stabilized zirconia femoral heads in a hip simulator.

Authors:  Marcel E Roy; Leo A Whiteside; Mark E Magill; Brian J Katerberg
Journal:  Clin Orthop Relat Res       Date:  2011-02-11       Impact factor: 4.176

6.  Metal on metal hip resurfacing versus uncemented custom total hip replacement--early results.

Authors:  Nemandra A Sandiford; Sarah K Muirhead-Allwood; John A Skinner; Jia Hua
Journal:  J Orthop Surg Res       Date:  2010-02-18       Impact factor: 2.359

7.  Alumina-on-Polyethylene Bearing Surfaces in Total Hip Arthroplasty.

Authors:  Yup Lee Jung; Shin-Yoon Kim
Journal:  Open Orthop J       Date:  2010-02-11

8.  Risk of complication and revision total hip arthroplasty among Medicare patients with different bearing surfaces.

Authors:  Kevin J Bozic; Kevin Ong; Edmund Lau; Steven M Kurtz; Thomas P Vail; Harry E Rubash; Daniel J Berry
Journal:  Clin Orthop Relat Res       Date:  2010-09       Impact factor: 4.176

9.  THA with Delta ceramic on ceramic: results of a multicenter investigational device exemption trial.

Authors:  William G Hamilton; James P McAuley; Douglas A Dennis; Jeffrey A Murphy; Thomas J Blumenfeld; Joel Politi
Journal:  Clin Orthop Relat Res       Date:  2009-09-19       Impact factor: 4.176

10.  Clinical comparison of polyethylene wear with zirconia or cobalt-chromium femoral heads.

Authors:  Maiken Stilling; Kjeld Anton Nielsen; Kjeld Søballe; Ole Rahbek
Journal:  Clin Orthop Relat Res       Date:  2009-03-27       Impact factor: 4.176

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.