| Literature DB >> 29850228 |
M K Badawy1,2, T Clark3, D Carrion1,4, P Deb2, O Farouque5,6.
Abstract
Radiological interventions play an increasingly relevant role in cardiology. Due to the inherent risks of ionizing radiation, proper care must be taken with monitoring and optimizing the dose delivered in angiograms to pose as low risk as possible to staff and patients. Dose optimization is particularly pertinent in teaching hospitals, where longer procedure times are at times necessary to accommodate the teaching needs of junior staff, and thus impart a more significant radiation dose. This study aims to analyze the effects of different protocol settings in routine coronary angiograms, from the perspective of a large tertiary center implementing a rapid dose reduction program. Routine coronary angiograms were chosen to compare baseline levels of radiation, and the dose imparted before and after dose optimization techniques was measured. Such methods included lowering dose per pulse, fluoroscopic pulse rates, and cine acquisition frame rates. The results showed up to 63% reduction in radiation dose without adverse impact on clinical or teaching outcomes. A 10 fps/low and 5 pps/low setting was found to achieve maximum dose optimization, with the caveat that settings require incremental changes to accommodate for patient complexities.Entities:
Year: 2018 PMID: 29850228 PMCID: PMC5925206 DOI: 10.1155/2018/6912841
Source DB: PubMed Journal: Cardiol Res Pract ISSN: 2090-0597 Impact factor: 1.866
Selected frame rate and pulse rate settings during the different phases of this study.
| Phase | Frame rate (fps)/dose setting | Pulse rate (pps)/dose setting |
|---|---|---|
| 1 | 15/normal | 7.5/normal |
| 2 | 15/normal | 7.5/low |
| 3 | 15/normal | 5/low |
| 4 | 10/normal | 5/low |
| 5 | 10/low | 5/low |
Patient characteristics of phase-specific populations.
| Characteristics | Phase 1 | Phase 2 | Phase 3 | Phase 4 | Phase 5 |
|
|---|---|---|---|---|---|---|
|
| 129 | 79 | 156 | 76 | 51 | |
| Male, no./total (%) | 81/129 (63) | 52/79 (66) | 98/156 (63) | 43/76 (57) | 34/51 (67) | 0.8 |
| Age, mean (SD) (years) | 64 (11) | 64.1 (12.8) | 65.2 (10.28) | 65.9 (11.9) | 63.4 (10.7) | 0.6 |
| Body mass index, mean (SD) (kg·m−2) | 29.5 (5.65) | 29.3 (6.2) | 30.3 (6.6) | 29.6 (6.2) | 29.0 (4.5) | 0.6 |
The CA procedural characteristics of phase-specific populations.
| Characteristics | Phase 1 | Phase 2 | Phase 3 | Phase 4 | Phase 5 |
|
|---|---|---|---|---|---|---|
| LV% | 87 | 86 | 86 | 79 | 80 | 0.8 |
| Radial % | 57 | 72 | 66 | 61 | 59 | 0.2 |
| No. of acquisitions | 8.4 (1.6) | 8.3 (1.7) | 8.4 (1.5) | 8.6 (1.5) | 8.5 (1.7) | 0.8 |
| Air kerma (mGy) | 0.68 (0.47–0.95) | 0.63 (0.4–0.84) | 0.59 (0.42–0.80) | 0.46 (0.29–0.57) | 0.3 (0.2–0.38) | <0.0001 |
| DAP (Gy·cm2) | 60.2 (43–84.6) | 53.0 (36.5–69.7) | 49.2 (36.5–68.4) | 36.4 (25.2–46.9) | 28.0 (17.6–35.2) | <0.0001 |
| Total study time | 4.5 (3.4–6.4) | 4.8 (3.6–7.6) | 5 (2.8–7.0) | 5 (2.6–7.6) | 4.2 (3.2–6.4) | 0.2 |
The phase-specific significant difference in cases where a significant difference is found (p < 0.001 in all instances where there was a statistically significant difference).
| Phase | 1 | 2 | 3 | 4 | 5 |
|---|---|---|---|---|---|
| 1 | — | NSD | NSD | SD | SD |
| 2 | NSD | — | NSD | SD | SD |
| 3 | NSD | NSD | — | SD | SD |
| 4 | SD | SD | SD | — | SD |
| 5 | SD | SD | SD | SD | — |
Quantitative image quality results using the CIRS model 901 20 cm phantom for standard NEMA measurements.
| Phase 1 | Phase 2 | Phase 3 | Phase 4 | Phase 5 | |||||||
|---|---|---|---|---|---|---|---|---|---|---|---|
| Fluoroscopy | Acquisitions | Fluoroscopy | Acquisitions | Fluoroscopy | Acquisitions | Fluoroscopy | Acquisitions | Fluoroscopy | Acquisitions | ||
| X-ray parameters | kVp | 81 | 70 | 78 | 70 | 76 | 70 | 76 | 70 | 76 | 71 |
| Tube current (mA) | 6.5 | 342 | 5 | 342 | 3 | 342 | 3 | 373 | 3 | 400 | |
| Dose setting | Normal | Normal | Low | Normal | Low | Normal | Low | Normal | Low | Low | |
| Frame rate | 7.5 | 15 | 7.5 | 15 | 5 | 15 | 5 | 10 | 5 | 10 | |
| Added filteration | Al 2.0 + Cu 0.1 | Al 1.0 + Au 0.01 | Al 1.5 + Cu 0.6 | Al 1.0 + Au 0.01 | Al 1.5 + Cu 0.6 | Al 1.0 + Au 0.01 | Al 1.5 + Cu 0.6 | Al 1.0 + Au 0.01 | Al 1.5 + Cu 0.6 | Al 2.0 + Cu 0.1 | |
|
| |||||||||||
| Static image quality | Line pairs | 1.8 | 2.2 | 1.8 | 2.2 | 1.4 | 2.2 | 1.4 | 2 | 1.4 | 2 |
| Iodine group 1 | 6 | 8 | 6 | 8 | 6 | 8 | 6 | 6 | 6 | 8 | |
| Iodine group 2 | 4 | 6 | 0 | 6 | 0 | 6 | 0 | 6 | 0 | 6 | |
| Iodine group 3 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | |
| Iodine group 4 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | |
| Air cylinders | 4 | 4 | 4 | 4 | 4 | 4 | 4 | 4 | 4 | 4 | |
| Aluminum cylinders | 4 | 4 | 4 | 4 | 4 | 4 | 4 | 3 | 4 | 3 | |
|
| |||||||||||
| X-ray parameters | kVp | 75 | 71 | 77 | 71 | 77 | 71 | 77 | 72 | 77 | 73 |
| Tube current (mA) | 5 | 432 | 5.7 | 432 | 3 | 432 | 3 | 400 | 3 | 399 | |
| Dose setting | Normal | Normal | Low | Normal | Low | Normal | Low | Normal | Low | Low | |
| Frame rate | 7.5 | 15 | 7.5 | 15 | 5 | 15 | 5 | 10 | 5 | 10 | |
| Added filteration | Al 2.0 + Cu 0.1 | Al 1.0 + Au 0.01 | Al 1.5 + Cu 0.6 | Al 1.0 + Au 0.01 | Al 1.5 + Cu 0.6 | Al 1.0 + Au 0.01 | Al 1.5 + Cu 0.6 | Al 1.0 + Au 0.01 | Al 1.5 + Cu 0.6 | Al 2.0 + Cu 0.1 | |
|
| |||||||||||
| Temporal image quality | Moving wires | 3 | 4 | 3 | 4 | 3 | 4 | 3 | 4 | 3 | 4 |
Quantitative image quality results using the CIRS model 901 30 cm phantom for standard NEMA measurements.
| Phase 1 | Phase 2 | Phase 3 | Phase 4 | Phase 5 | |||||||
|---|---|---|---|---|---|---|---|---|---|---|---|
| Fluoroscopy | Acquisitions | Fluoroscopy | Acquisitions | Fluoroscopy | Acquisitions | Fluoroscopy | Acquisitions | Fluoroscopy | Acquisitions | ||
| X-ray parameters | kVp | 118 | 90 | 113 | 90 | 110 | 90 | 110 | 102 | 110 | 102 |
| Tube current (mA) | 11.4 | 746 | 11 | 746 | 7.1 | 746 | 7.1 | 341 | 7.1 | 340 | |
| Dose setting | Normal | Normal | Low | Normal | Low | Normal | Low | Normal | Low | Low | |
| Frame rate | 7.5 | 15 | 7.5 | 15 | 5 | 15 | 5 | 10 | 5 | 10 | |
| Added filteration | Al 2.0 + Cu 0.1 | Al 1.0 + Au 0.01 | Al 1.5 + Cu 0.6 | Al 1.0 + Au 0.01 | Al 1.5 + Cu 0.6 | Al 1.0 + Au 0.01 | Al 1.5 + Cu 0.6 | Al 1.0 + Au 0.01 | Al 1.5 + Cu 0.6 | Al 2.0 + Cu 0.1 | |
|
| |||||||||||
| Static image quality | Line pairs | 1.6 | 1.8 | 1.4 | 1.8 | 1.2 | 1.8 | 1.2 | 1.6 | 1.2 | 1.6 |
| Iodine group 1 | 4 | 6 | 4 | 6 | 2 | 6 | 2 | 6 | 2 | 6 | |
| Iodine group 2 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | |
| Iodine group 3 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | |
| Iodine group 4 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | |
| Air cylinders | 4 | 4 | 4 | 4 | 4 | 4 | 4 | 4 | 4 | 4 | |
| Aluminum cylinders | 4 | 4 | 4 | 4 | 4 | 4 | 4 | 4 | 4 | 4 | |
|
| |||||||||||
| X-ray parameters | kVp | 106 | 90 | 102 | 90 | 102 | 90 | 102 | 103 | 102 | 103 |
| Tube current (mA) | 10 | 744 | 9 | 744 | 6.6 | 744 | 6.6 | 339 | 6.6 | 339 | |
| Dose setting | Normal | Normal | Low | Normal | Low | Normal | Low | Normal | Low | Low | |
| Frame rate | 7.5 | 15 | 7.5 | 15 | 5 | 15 | 5 | 10 | 5 | 10 | |
| Added filteration | Al 2.0 + Cu 0.1 | Al 1.0 + Au 0.01 | Al 1.5 + Cu 0.6 | Al 1.0 + Au 0.01 | Al 1.5 + Cu 0.6 | Al 1.0 + Au 0.01 | Al 1.5 + Cu 0.6 | Al 1.0 + Au 0.01 | Al 1.5 + Cu 0.6 | Al 2.0 + Cu 0.1 | |
|
| |||||||||||
| Temporal image quality | Moving wires | 2 | 3 | 1 | 3 | 1 | 3 | 1 | 3 | 1 | 3 |
Figure 1Median reference air kerma values during the different phases of the study. The error bars represent first and third quartile values.
Figure 2Median dose-area product values during the different phases of the study. The error bars represent first and third quartile values.
Figure 3Median total fluoroscopic time values during the different phases of the study. The error bars represent first and third quartile values.