Dosimetric study of GZP6 60 Co high dose rate brachytherapy source.

Qin Lei1, Anjian Xu2, Chengjun Gou1, Yumei Wen1, Donglin He1, Junxiang Wu3, Qing Hou1, Zhangwen Wu1.   

Abstract

The purpose of this study was to obtain dosimetric parameters of GZP6 60 Co brachytherapy source number 3. The Geant4 MC code has been used to obtain the dose rate distribution following the American Association of Physicists in Medicine (AAPM) TG-43U1 dosimetric formalism. In the simulation, the source was centered in a 50 cm radius water phantom. The cylindrical ring voxels were 0.1 mm thick for r ≤ 1 cm, 0.5 mm for 1 cm < r ≤ 5 cm, and 1 mm for r > 5 cm. The kerma-dose approximation was performed for r > 0.75 cm to increase the simulation efficiency. Based on the numerical results, the dosimetric datasets were obtained. These results were compared with the available data of the similar 60 Co high dose rate sources and the detailed dosimetric characterization was discussed.
© 2018 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals, Inc. on behalf of American Association of Physicists in Medicine.

Entities:  

Keywords:  60Co; Monte Carlo simulation; brachytherapy; dosimetry

Mesh:

Year:  2018        PMID: 29808551      PMCID: PMC6036395          DOI: 10.1002/acm2.12362

Source DB:  PubMed          Journal:  J Appl Clin Med Phys        ISSN: 1526-9914            Impact factor:   2.102


INTRODUCTION

Monte Carlo (MC) simulation of electron and photon transport has been widely used in many areas of interest in medical physics, mainly in the development of the brachytherapy field. Calculation of kerma, absorbed dose, fluence, and related quantities at a geometric point is an important application of the Monte Carlo method. Many kinds of Monte Carlo simulation software have been successfully used for dosimetric studies, including Geant4, MCNP, EGSnrc (an extended and improved version of the EGS4), PENELOPE, and FLUKA. The GZP6 60Co afterloading high dose rate (HDR) unit (Nuclear Power Institute of China) is widely used at home and abroad. It is comprised of six different source designs with a stepping source (source number 6) and five nonstepping sources (source number 1–5). It mainly addresses intracavitary and interstitial applications and is considered an integral part of the treatment of cervical, vaginal, rectal, and esophageal cancers. In the literature, several investigations have been performed on the GZP6 60Co HDR unit. In the study of Mesbahi et al.1, air kerma strengths of source numbers 1, 2, and 5 were obtained by in‐air measurements and a Farmer‐type ionization chamber. In a separate investigation, the radial dose functions of the three sources were calculated by Mesbahi et al.2 using the MC method and GZP6 TPS. Toossi et al.3 estimated the air kerma strength of GZP6 60Co source number 3 by Monte Carlo simulation and in‐air measurements. The dose distribution for GZP6 60Co stepping source was also calculated using the matrix shift method by Toossi et al.4 For the purpose of quality assurance, the dose distributions generated by GZP6 TPS were verified in another investigation.5 The dosimetric parameters of radioactive sources are crucial elements in clinical practice as they are important input data in treatment planning systems. Tabrizi et al. (2012) used the MCNP4C Monte Carlo code to obtain the dosimetric parameters of GZP6 60Co sources. However, in the study of Vijande et al.6, it is indicated that the radial dose function from the study of Tabrizi et al. is inconsistent with other 60Co source data and is difficult to understand from a physical point of view. Hence, additional investigation of the dosimetric parameters of GZP6 60Co sources is needed. This work comprises a full and MC‐based dosimetry report for GZP6 60Co source number 3 according to the recommendations of the American Association of Physicists in Medicine (AAPM) and the European Society for Radiotherapy and Oncology (ESTRO) on dose calculations for high‐energy (average energy higher than 50 keV) photon‐emitting brachytherapy sources.7, 8, 9 The calculated dosimetric parameters were compared with the available data of the similar 60Co high dose rate (HDR) sources6, 10, 11 and the detailed dosimetric characterization of GZP6 60Co source number 3 was discussed.

MATERIAL AND METHODS

GZP6 60Co brachytherapy source number 3 is composed of an active cylindrical 60Co pellet with a 3.5 mm length and a 1.5 mm diameter covered with a titanium layer with a thickness of 0.1 mm. The radioactive 60Co is uniformly distributed in the core. Several nonactive steel pellets 1.5 mm in diameter line up with the active cylindrical 60Co pellet. All the pellets described above were packaged in a steel spring cover with a thickness of 0.5 mm. The detailed information of this source is taken from published studies2, 12 and is illustrated in Fig. 1. The mass density and chemical composition of the materials are shown in Table 1.
Figure 1

A schematic view of the GZP6 60Co HDR source number 3. Dimensions are given in mm.

Table 1

Mass density and composition of the materials of GZP6 60Co HDR source number 3

Material: descriptionMass density (g/cm3)Composition (element/weight fraction)
Cobalt: source core8.85Co/1
Titanium: source capsule4.54Ti/1
Steel pellets: spacers in the source braid7.9Fe/0.71994, C/0.0005, Si/0.0072, Mn/0.0137, S/0.00011, P/0.00025, Cr/0.17, Ni/0.0822, Mo/0.0013, V/0.0006, Ti/0.0042
Steel: spring cover6.999Fe/0.7416, Ni/0.069, S/0.0001, Cr/0.167, C/0.0006, Mn/0.0062, Cu/0.0026, Al/0.0062, Mo/0.0015, Si/0.0052
A schematic view of the GZP6 60Co HDR source number 3. Dimensions are given in mm. Mass density and composition of the materials of GZP6 60Co HDR source number 3 The 60Co brachytherapy source emits two gamma rays with energies of 1.33 MeV and 1.17 MeV, mixed with β rays whose maximum energy and average energy is 0.318 MeV and 0.096 MeV, respectively.8 In the simulation, the β rays are neglected because of absorption in the source steel cover.10 In this study, the Monte Carlo code Geant413 (Geant4.9.6.P02 development version) is used to simulate transport and interaction of gamma rays emitted from the GZP6 60Co HDR brachytherapy source in water. The Evaluated Photon Data Library, 1997 Version (EPDL97) and Evaluated Electron Data Library (EEDL) cross‐section libraries were used for photons and electrons, respectively.14, 15 In the simulation, a spherical liquid water phantom with a 50‐cm radius was utilized to approximate an infinite water environment. The density used for the liquid water has been 0.998 g/cm3 as recommended in the TG‐43 U1 report.7 The GZP6 60Co HDR source was accommodated in the center of the water phantom. Electronic equilibrium within 1% was reached for 60Co at distances greater than 7 mm from the source center.16 Thus, the kerma‐dose approximation was performed for r > 0.75 cm to speed up calculations.17 Dose and collisional kerma rate distributions were used to derive the final dosimetric parameters as described in the AAPM TG‐43U1 report.7 The dose distribution of the source was calculated within the radial distance of 20 cm. The cylindrical ring voxels were 0.1 mm thick for r ≤ 1 cm, 0.5 mm for 1 cm < r ≤ 5 cm, and 1 mm for r > 5 cm, which can provide high‐resolution dosimetry. The cutoff energy was set to 10 keV for both photons and electrons. Collisional kerma and absorbed dose were obtained in cylindrical (y, z) and spherical (r, θ) coordinates. The coordinate axes used are shown in Fig. 1. The air kerma strength was calculated in a separate simulation with the source surrounded by vacuum, except for a cylindrical air cell of 0.1 cm in diameter and 0.1 cm in height at r = 10 cm. As clarified in the TG‐43U1S1 report,18 dry air (0% humidity) is recommended for air kerma strength in contrast to the TG‐43U1 report which recommended air at 40% relative humidity. 6 × 109 photon histories (r ≤ 0.75 cm) were simulated to score dose. 109 photon histories (r > 0.75 cm) were simulated to score kerma.

RESULTS AND DISCUSSION

The air kerma strength calculated for GZP6 60Co source number 3 is 3.004 × 10−7 cGy cm2 h−1 Bq−1 with a statistical uncertainty of 0.14%. The dose rate constant of GZP6 60Co source number 3 obtained is 1.088 ± 0.002 cGy h−1 U−1 (with k = 1), which is comparable to the available data of the similar 60Co HDR sources (see Table 2). The radial dose function values and the anisotropy function values of the GZP6 60Co source number 3 are provided in Table 3 and Table 4, respectively. In Fig. 3, the anisotropy function results of GZP6 60Co source number 3 are plotted vs polar angle at the selected radial distances. In addition, the along‐away data are shown in Table 5.
Table 2

Comparison of dose rate constant values for the comparable 60Co HDR sources

Source typeΛ (cGy h−1 U−1)Λ/G (r 0, θ0)
The new BEBIG 60Co10 1.087 ± 0.00111.098 ± 0.0011
Ralston Type‐2 60Co11 1.101 ± 0.0051.105 ± 0.005
Flexisource 60Co6 1.085 ± 0.0031.096 ± 0.003
GZP6 60Co source num. 3 (this work)1.088 ± 0.0021.099 ± 0.002
Table 3

Radial dose function calculated for the GZP6 60Co HDR source number 3

r (cm) g L (r)
0.251.05
0.331.035
0.51.022
0.751.008
11
1.50.991
20.984
30.969
40.953
50.937
60.921
70.904
80.887
90.869
100.853
120.816
150.759
200.664
Table 4

Anisotropy function calculated for the GZP6 60Co HDR source number 3

θ(°) r (cm)
0.250.330.50.7511.52345678910121520
00.8740.8740.8760.8850.8850.8870.8880.8940.8960.9010.9030.9070.9150.9270.924
10.8730.8710.8740.8830.8850.8870.8910.8950.8970.9010.9040.9060.9130.9200.925
20.8720.8690.8730.8810.8860.8880.8930.8960.8980.9010.9050.9040.9110.9160.926
30.8770.8800.8830.8880.8910.8920.8950.8970.9000.9030.9060.9050.9110.9160.919
40.8810.8830.8890.8920.8950.8970.9000.9040.9060.9080.9110.9100.9130.9150.920
50.8840.8910.8930.8950.8970.8980.8990.9000.9020.9060.9070.9060.9120.9160.921
60.9000.9030.9010.8970.8960.8980.9010.9040.9060.9080.9100.9100.9140.9180.924
70.9090.9060.9030.9000.8990.9010.9030.9060.9070.9100.9120.9120.9150.9190.924
80.9210.9090.9040.9020.9020.9040.9060.9080.9090.9120.9140.9140.9180.9200.925
90.9250.9100.9050.9040.9050.9060.9090.9100.9120.9150.9160.9170.9210.9240.928
100.9310.9110.9060.9070.9080.9090.9120.9120.9160.9180.9190.9200.9250.9290.931
150.9340.9300.9300.9300.9310.9310.9320.9330.9340.9360.9380.9380.9390.9410.943
200.9630.9430.9520.9520.9510.9510.9510.9520.9520.9520.9530.9550.9540.9550.9590.963
250.9620.9640.9670.9660.9650.9660.9660.9660.9660.9660.9670.9670.9660.9670.9670.970
300.9720.9730.9760.9770.9750.9760.9750.9750.9750.9750.9760.9770.9750.9770.9760.976
400.9910.9860.9830.9890.9870.9880.9870.9870.9870.9870.9870.9860.9860.9870.9850.9860.9840.984
501.0000.9950.9930.9900.9940.9950.9940.9940.9950.9940.9940.9940.9940.9950.9930.9940.9920.991
601.0000.9970.9950.9890.9970.9970.9960.9970.9960.9960.9950.9950.9960.9960.9930.9940.9970.993
701.0000.9980.9980.9831.0000.9990.9980.9990.9990.9980.9980.9970.9980.9990.9980.9970.9970.997
801.0001.0000.9981.0001.0001.0001.0001.0000.9990.9980.9990.9980.9980.9990.9980.9971.0001.000
90111111111111111111
1001.0001.0000.9991.0001.0001.0001.0001.0000.9990.9990.9990.9980.9980.9990.9980.9991.0001.000
1101.0001.0000.9980.9860.9990.9980.9990.9990.9990.9980.9990.9980.9980.9980.9980.9970.9980.997
1200.9980.9960.9950.9910.9970.9970.9960.9970.9970.9970.9960.9960.9960.9950.9940.9950.9970.995
1301.0000.9960.9940.9910.9950.9960.9950.9950.9940.9950.9950.9950.9950.9950.9940.9950.9930.994
1400.9960.9880.9840.9850.9870.9880.9880.9890.9880.9880.9880.9860.9870.9890.9870.9880.9860.986
1500.9710.9740.9770.9770.9770.9760.9760.9760.9760.9760.9760.9780.9780.9780.9780.979
1550.9610.9640.9670.9680.9680.9670.9670.9680.9680.9680.9690.9700.9700.9700.9710.973
1600.9460.9520.9540.9540.9540.9530.9540.9550.9560.9560.9570.9600.9600.9610.9650.969
1650.9330.9330.9320.9310.9290.9300.9310.9310.9330.9340.9340.9370.9410.944
1700.8830.8850.8840.8850.8870.8870.8900.8940.8950.8960.9010.9050.907
1710.8630.8630.8630.8640.8680.8700.8720.8750.8760.8780.8820.8860.890
1720.8510.8510.8510.8530.8570.8600.8620.8660.8690.8690.8730.8750.880
1730.8200.8200.8210.8240.8290.8340.8350.8380.8400.8410.8470.8530.863
1740.7970.7970.7990.8040.8090.8150.8200.8240.8290.8320.8420.8480.856
1750.7760.7780.7850.7920.7990.8040.8120.8170.8180.8280.8380.849
1760.7740.7740.7800.7870.7940.8000.8060.8120.8150.8240.8330.844
1770.7700.7720.7790.7860.7940.7940.8030.8080.8090.8190.8300.841
1780.7690.7700.7780.7840.7930.7930.8020.8060.8070.8170.8290.839
1790.7660.7680.7760.7810.7910.7920.8010.8050.8070.8180.8280.835
1800.7700.7700.7770.7790.7920.7930.8020.8050.8100.8200.8300.840
Table 5

Dose rate results (cGy h−1 U−1) around the GZP6 60Co HDR source number 3

z (cm) y (cm)
00.250.330.50.7511.5234567
−70.0160.0160.0160.0160.0160.0170.0180.0180.0160.0150.0130.0110.009
−60.0220.0230.0220.0220.0230.0240.0240.0240.0220.0190.0160.0130.011
−50.0320.0330.0320.0330.0350.0360.0350.0340.0290.0240.0200.0160.013
−40.0510.0510.0510.0540.0560.0570.0550.0500.0410.0320.0240.0190.015
−30.0910.0910.0940.1010.1030.1010.0920.0800.0570.0410.0300.0220.017
−20.2070.2230.2320.2370.2280.2100.1700.1320.0810.0520.0350.0250.019
−1.50.4190.4280.4170.3790.3300.2380.1710.0940.0570.0380.0260.019
−10.9860.9620.8680.6970.5440.3330.2150.1060.0610.0390.0270.020
−0.53.6883.1762.2341.3550.8750.4340.2540.1150.0640.0410.0280.020
016.0889.6094.3231.9351.0880.4820.2700.1180.0650.0410.0280.020
0.53.6883.1762.2341.3550.8740.4340.2540.1150.0640.0410.0280.020
10.9880.9830.9590.8670.6960.5440.3330.2150.1060.0610.0390.0270.019
1.50.4290.4320.4300.4160.3780.3290.2380.1710.0940.0570.0380.0260.019
20.2380.2410.2400.2370.2270.2100.1690.1320.0810.0520.0350.0250.018
30.1050.1050.1050.1050.1040.1010.0910.0800.0570.0410.0300.0220.017
40.0570.0590.0580.0580.0580.0570.0540.0500.0400.0320.0240.0190.015
50.0360.0370.0370.0370.0360.0360.0350.0340.0290.0240.0200.0160.013
60.0250.0250.0250.0250.0250.0250.0240.0240.0220.0190.0160.0130.011
70.0180.0180.0180.0180.0180.0180.0180.0180.0160.0150.0130.0110.009
Comparison of dose rate constant values for the comparable 60Co HDR sources Radial dose function calculated for the GZP6 60Co HDR source number 3 Anisotropy function calculated for the GZP6 60Co HDR source number 3 Dose rate results (cGy h−1 U−1) around the GZP6 60Co HDR source number 3 The radial dose function values of the GZP6 60Co source number 3 were compared with corresponding data from the relevant literature (see Fig. 2). It is observed that the curves of the radial dose functions of the source models match well for r > 1 cm and small differences exist for r < 1 cm. These differences are caused by varying degrees of photon absorption and scattering in the sources.8 In general, the radial dose functions do not depend significantly on source dimensions and encapsulation designs.11
Figure 2

(a) Radial dose function of 60Co source models. (b) Zoom‐in at short distances from the source.

(a) Radial dose function of 60Co source models. (b) Zoom‐in at short distances from the source. As shown in Table 4 and Fig. 3, the anisotropy function values of GZP6 60Co source number 3 are nearly uniform for polar angles 30° ≤ θ ≤ 90°. For example, the anisotropy function values are around 0.998 for θ = 70° and around 0.994 for θ = 50°. However, a strong dependence on radial distance was observed for θ < 30°. As described in Ref. 8, the anisotropy function values decrease for polar angles close to the long axis (see Fig. 3), which is caused by the oblique filtration within the source structure.
Figure 3

Anisotropy function of the GZP6 60Co HDR source number 3 at selected radial distances.

Anisotropy function of the GZP6 60Co HDR source number 3 at selected radial distances. According to AAPM TG‐43U1 recommendation,7 the uncertainty of the final dose rate values has been estimated, including statistical (A) and systemic uncertainty (B). The statistical uncertainty in water phantom calculations is less than 0.6% for all the points, except at the points located near the longitudinal axis, where it is less than 1.1%. For the simulation of the air kerma strength, the statistical uncertainty is 0.14%. This gives us a combined A‐type uncertainty of 0.6% for all the points, except near longitudinal axis points where the combined A‐type uncertainty is about 1.1%. The type B uncertainty is negligible for 60Co sources.10, 19 Thus, the total uncertainty is 0.6% for all the points except for the longitudinal axis points, which is 1.1%.

CONCLUSIONS

In this study, the Geant4 MC code was used to study the dose rate distribution around the GZP6 60Co HDR source number 3. The dosimetric parameters for this source are obtained as required by AAPM and ESTRO. The calculated dose rate constant of the GZP6 60Co source number 3 is 1.088 ± 0.002 cGy h−1 U−1 and the radial dose functions are consistent with the available data of the similar 60Co HDR sources. In addition, a 2D rectangular dose rate table is presented.

CONFLICTS OF INTEREST

None declared.
  14 in total

1.  Monte Carlo dosimetric study of the BEBIG Co-60 HDR source.

Authors:  F Ballester; D Granero; J Pérez-Calatayud; E Casal; S Agramunt; R Cases
Journal:  Phys Med Biol       Date:  2005-10-12       Impact factor: 3.609

2.  Supplement to the 2004 update of the AAPM Task Group No. 43 Report.

Authors:  Mark J Rivard; Wayne M Butler; Larry A DeWerd; M Saiful Huq; Geoffrey S Ibbott; Ali S Meigooni; Christopher S Melhus; Michael G Mitch; Ravinder Nath; Jeffrey F Williamson
Journal:  Med Phys       Date:  2007-06       Impact factor: 4.071

3.  Dose calculation for photon-emitting brachytherapy sources with average energy higher than 50 keV: report of the AAPM and ESTRO.

Authors:  Jose Perez-Calatayud; Facundo Ballester; Rupak K Das; Larry A Dewerd; Geoffrey S Ibbott; Ali S Meigooni; Zoubir Ouhib; Mark J Rivard; Ron S Sloboda; Jeffrey F Williamson
Journal:  Med Phys       Date:  2012-05       Impact factor: 4.071

4.  Monte Carlo dose calculation of GZP6 (60)Co stepping source based on a matrix shift technique.

Authors:  Mohammad Taghi Bahreyni Toossi; Malihe Abdollahi; Mahdi Ghorbani
Journal:  Rep Pract Oncol Radiother       Date:  2010-12-21

5.  Evaluation of high-energy brachytherapy source electronic disequilibrium and dose from emitted electrons.

Authors:  Facundo Ballester; Domingo Granero; José Pérez-Calatayud; Christopher S Melhus; Mark J Rivard
Journal:  Med Phys       Date:  2009-09       Impact factor: 4.071

6.  Technical note: Dosimetric study of a new Co-60 source used in brachytherapy.

Authors:  D Granero; J Pérez-Calatayud; F Ballester
Journal:  Med Phys       Date:  2007-09       Impact factor: 4.071

7.  Monte Carlo derivation of AAPM TG-43 dosimetric parameters for GZP6 Co-60 HDR sources.

Authors:  Sanaz Hariri Tabrizi; Alireza Kamali Asl; Zohreh Azma
Journal:  Phys Med       Date:  2011-05-31       Impact factor: 2.685

8.  Monte Carlo evaluation of kerma at a point for photon transport problems.

Authors:  J F Williamson
Journal:  Med Phys       Date:  1987 Jul-Aug       Impact factor: 4.071

9.  RECORDS: improved Reporting of montE CarlO RaDiation transport Studies: Report of the AAPM Research Committee Task Group 268.

Authors:  Ioannis Sechopoulos; D W O Rogers; Magdalena Bazalova-Carter; Wesley E Bolch; Emily C Heath; Michael F McNitt-Gray; Josep Sempau; Jeffrey F Williamson
Journal:  Med Phys       Date:  2017-12-16       Impact factor: 4.071

10.  Monte Carlo dosimetric study of the Flexisource Co-60 high dose rate source.

Authors:  Javier Vijande; Domingo Granero; Jose Perez-Calatayud; Facundo Ballester
Journal:  J Contemp Brachytherapy       Date:  2012-03-30
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.