Evaluation of high-energy brachytherapy source electronic disequilibrium and dose from emitted electrons.

Facundo Ballester1, Domingo Granero, José Pérez-Calatayud, Christopher S Melhus, Mark J Rivard.   

Abstract

PURPOSE: The region of electronic disequilibrium near photon-emitting brachytherapy sources of high-energy radionuclides (60Co, 137CS, 192Ir, and 169Yb) and contributions to total dose from emitted electrons were studied using the GEANT4 and PENELOPE Monte Carlo codes.
METHODS: Hypothetical sources with active and capsule materials mimicking those of actual sources but with spherical shape were examined. Dose contributions due to source photons, x rays, and bremsstrahlung; source beta-, Auger electrons, and internal conversion electrons; and water collisional kerma were scored. To determine if conclusions obtained for electronic equilibrium conditions and electron dose contribution to total dose for the representative spherical sources could be applied to actual sources, the 192Ir mHDR-v2 source model (Nucletron B.V., Veenendaal, The Netherlands) was simulated for comparison to spherical source results and to published data.
RESULTS: Electronic equilibrium within 1% is reached for 60Co, 137CS, 192Ir, and 169Yb at distances greater than 7, 3.5, 2, and 1 mm from the source center, respectively, in agreement with other published studies. At 1 mm from the source center, the electron contributions to total dose are 1.9% and 9.4% for 60Co and 192Ir, respectively. Electron emissions become important (i.e., > 0.5%) within 3.3 mm of 60Co and 1.7 mm of 192Ir sources, yet are negligible over all distances for 137Cs and 169Yb. Electronic equilibrium conditions along the transversal source axis for the mHDR-v2 source are comparable to those of the spherical sources while electron dose to total dose contribution are quite different.
CONCLUSIONS: Electronic equilibrium conditions obtained for spherical sources could be generalized to actual sources while electron contribution to total dose depends strongly on source dimensions, material composition, and electron spectra.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 19810499     DOI: 10.1118/1.3194754

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  9 in total

1.  A dosimetric uncertainty analysis for photon-emitting brachytherapy sources: report of AAPM Task Group No. 138 and GEC-ESTRO.

Authors:  Larry A DeWerd; Geoffrey S Ibbott; Ali S Meigooni; Michael G Mitch; Mark J Rivard; Kurt E Stump; Bruce R Thomadsen; Jack L M Venselaar
Journal:  Med Phys       Date:  2011-02       Impact factor: 4.071

2.  Dosimetric characterization of round HDR 192Ir accuboost applicators for breast brachytherapy.

Authors:  Mark J Rivard; Christopher S Melhus; David E Wazer; Raymond J Bricault
Journal:  Med Phys       Date:  2009-11       Impact factor: 4.071

3.  Evaluation of interpolation methods for TG-43 dosimetric parameters based on comparison with Monte Carlo data for high-energy brachytherapy sources.

Authors:  Ma Carmen Pujades-Claumarchirant; Domingo Granero; Jose Perez-Calatayud; Facundo Ballester; Christopher Melhus; Mark Rivard
Journal:  J Contemp Brachytherapy       Date:  2010-04-01

4.  Revision of the dosimetric parameters of the CSM11 LDR Cs-137 source.

Authors:  Antonio Otal; Juan Manuel Martínez-Fernández; Domingo Granero
Journal:  J Contemp Brachytherapy       Date:  2011-03-31

5.  Dosimetric study of GZP6 60 Co high dose rate brachytherapy source.

Authors:  Qin Lei; Anjian Xu; Chengjun Gou; Yumei Wen; Donglin He; Junxiang Wu; Qing Hou; Zhangwen Wu
Journal:  J Appl Clin Med Phys       Date:  2018-05-28       Impact factor: 2.102

6.  Intra-cavitary brachytherapy in endometrial carcinoma: experience with cobalt-60 source.

Authors:  Jose Antonio Domínguez Rullán; María Teresa Muñoz Miguelañez; Rafael Colmenares Fernández; Miguel Cámara Gallego; Mercedes Martín Sánchez; Feliciano García Vicente; Sonsoles Sancho García
Journal:  J Contemp Brachytherapy       Date:  2022-05-10

7.  Monte Carlo dosimetric study of the Flexisource Co-60 high dose rate source.

Authors:  Javier Vijande; Domingo Granero; Jose Perez-Calatayud; Facundo Ballester
Journal:  J Contemp Brachytherapy       Date:  2012-03-30

8.  A comparative assessment of inhomogeneity and finite patient dimension effects in 60Co and 192Ir high-dose-rate brachytherapy.

Authors:  Irina Fotina; Kyveli Zourari; Vasileios Lahanas; Evaggelos Pantelis; Panagiotis Papagiannis
Journal:  J Contemp Brachytherapy       Date:  2018-02-28

9.  The dosimetric impact of replacing the TG-43 algorithm by model based dose calculation for liver brachytherapy.

Authors:  Anna Sophie Duque; Stefanie Corradini; Florian Kamp; Max Seidensticker; Florian Streitparth; Christopher Kurz; Franziska Walter; Katia Parodi; Frank Verhaegen; Jens Ricke; Claus Belka; Gabriel Paiva Fonseca; Guillaume Landry
Journal:  Radiat Oncol       Date:  2020-03-09       Impact factor: 3.481

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.