INTRODUCTION: The purpose of this study was to investigate the systematic localization accuracy, treatment planning capability, and delivery accuracy of an integrated magnetic resonance imaging guided Linear Accelerator (MR-Linac) platform for stereotactic radiosurgery.
MATERIALS AND METHODS: The phantom for the end-to-end test comprises three different compartments: a rectangular MR/CT target phantom, a Winston-Lutz cube, and a rectangular MR/CT isocenter phantom. Hidden target tests were performed at gantry angles of 0, 90, 180, and 270 degrees to quantify the systematic accuracy. Five patient plans with a total of eleven lesions were used to evaluate the dosimetric accuracy. Single-isocenter IMRT treatment plans using 10-15 coplanar beams were generated to treat the multiple metastases.
RESULTS: The end-to-end localization accuracy of the system was 1.0 ± 0.1 mm. The conformity index, homogeneity index and gradient index of the plans were 1.26 ± 0.22, 1.22 ± 0.10, and 5.38 ± 1.44, respectively. The average absolute point dose difference between measured and calculated dose was 1.64 ± 1.90%, and the mean percentage of points passing the 3%/1 mm gamma criteria was 96.87%.
CONCLUSIONS: Our experience demonstrates that excellent plan quality and delivery accuracy was achievable on the MR-Linac for treating multiple brain metastases with a single isocenter.
Copyright © 2018 Elsevier B.V. All rights reserved.

Entities:  

Keywords:  Accuracy; MR-Linac; Single isocenter multiple targets; Stereotactic radiosurgery

Mesh:

Year:  2018        PMID: 29807837      PMCID: PMC6223121          DOI: 10.1016/j.radonc.2018.04.034

Source DB:  PubMed          Journal:  Radiother Oncol        ISSN: 0167-8140            Impact factor:   6.280


  28 in total

1.  A new approach to quantify the mechanical and radiation isocentres of radiotherapy treatment machine gantries.

Authors:  Piotr Skworcow; John A Mills; Olivier C L Haas; Keith J Burnham
Journal:  Phys Med Biol       Date:  2007-11-16       Impact factor: 3.609

2.  Magnetic-field-induced dose effects in MR-guided radiotherapy systems: dependence on the magnetic field strength.

Authors:  A J E Raaijmakers; B W Raaymakers; J J W Lagendijk
Journal:  Phys Med Biol       Date:  2008-01-18       Impact factor: 3.609

3.  Fast and robust online adaptive planning in stereotactic MR-guided adaptive radiation therapy (SMART) for pancreatic cancer.

Authors:  O Bohoudi; A M E Bruynzeel; S Senan; J P Cuijpers; B J Slotman; F J Lagerwaard; M A Palacios
Journal:  Radiother Oncol       Date:  2017-08-12       Impact factor: 6.280

4.  Stereotactic radiotherapy of intracranial tumors: a comparison of intensity-modulated radiotherapy and dynamic conformal arc.

Authors:  Ruud G J Wiggenraad; Anna L Petoukhova; Lia Versluis; Jan P C van Santvoort
Journal:  Int J Radiat Oncol Biol Phys       Date:  2009-02-11       Impact factor: 7.038

5.  Radiosurgery of multiple brain metastases with single-isocenter dynamic conformal arcs (SIDCA).

Authors:  Yimei Huang; Karen Chin; Jared R Robbins; Jinkoo Kim; Haisen Li; Hanan Amro; Indrin J Chetty; James Gordon; Samuel Ryu
Journal:  Radiother Oncol       Date:  2014-07-02       Impact factor: 6.280

6.  Volumetric intensity-modulated arc therapy vs. conventional IMRT in head-and-neck cancer: a comparative planning and dosimetric study.

Authors:  Wilko F A R Verbakel; Johan P Cuijpers; Daan Hoffmans; Michael Bieker; Ben J Slotman; Suresh Senan
Journal:  Int J Radiat Oncol Biol Phys       Date:  2009-05-01       Impact factor: 7.038

7.  Whole-procedural radiological accuracy for delivering multi-session gamma knife radiosurgery with a relocatable frame system.

Authors:  Lijun Ma; Dilini Pinnaduwage; Michael McDermott; Penny K Sneed
Journal:  Technol Cancer Res Treat       Date:  2013-08-31

8.  Coplanar versus noncoplanar intensity-modulated radiation therapy (IMRT) and volumetric-modulated arc therapy (VMAT) treatment planning for fronto-temporal high-grade glioma.

Authors:  Valerie Panet-Raymond; Will Ansbacher; Sergei Zavgorodni; Bill Bendorffe; Alan Nichol; Pauline T Truong; Wayne Beckham; Maria Vlachaki
Journal:  J Appl Clin Med Phys       Date:  2012-07-05       Impact factor: 2.102

9.  Evaluation of stability of stereotactic space defined by cone-beam CT for the Leksell Gamma Knife Icon.

Authors:  Ismail AlDahlawi; Dheerendra Prasad; Matthew B Podgorsak
Journal:  J Appl Clin Med Phys       Date:  2017-04-17       Impact factor: 2.102

10.  Characteristics of a novel treatment system for linear accelerator-based stereotactic radiosurgery.

Authors:  Ning Wen; Haisen Li; Kwang Song; Karen Chin-Snyder; Yujiao Qin; Jinkoo Kim; Maria Bellon; Misbah Gulam; Stephen Gardner; Anthony Doemer; Suneetha Devpura; James Gordon; Indrin Chetty; Farzan Siddiqui; Munther Ajlouni; Robert Pompa; Zane Hammoud; Michael Simoff; Steven Kalkanis; Benjamin Movsas; M Salim Siddiqui
Journal:  J Appl Clin Med Phys       Date:  2015-07-08       Impact factor: 2.102

View more
  17 in total

1.  Evaluation of diffusion-weighted MRI and geometric distortion on a 0.35T MR-LINAC at multiple gantry angles.

Authors:  Benjamin Lewis; Anamaria Guta; Stacie Mackey; H Michael Gach; Sasa Mutic; Olga Green; Taeho Kim
Journal:  J Appl Clin Med Phys       Date:  2021-01-15       Impact factor: 2.102

2.  Large field of view distortion assessment in a low-field MR-linac.

Authors:  Siamak P Nejad-Davarani; Joshua P Kim; Dongsu Du; Carri Glide-Hurst
Journal:  Med Phys       Date:  2019-03-23       Impact factor: 4.071

3.  Comparison of Motion-Insensitive T2-Weighted MRI Pulse Sequences for Visualization of the Prostatic Urethra During MR Simulation.

Authors:  Kristen L Zakian; Andreas Wibmer; Hebert A Vargas; Eveline Alberts; Mo Kadbi; Borys Mychalczak; Marisa Kollmeier; Daniel Gorovets; Sean McBride; Margie Hunt; Michael J Zelefsky; Neelam Tyagi
Journal:  Pract Radiat Oncol       Date:  2019-06-25

4.  Association Between Internal Organ/Liver Tumor and External Surface Motion From Cine MR Images on an MRI-Linac.

Authors:  Weihua Mao; Joshua Kim; Indrin J Chetty
Journal:  Front Oncol       Date:  2022-06-30       Impact factor: 5.738

5.  Carbon ion and proton beam irradiation of a normal human TK6 lymphoblastoid cell line within a magnetic field of 1.0 tesla.

Authors:  B Yudhistiara; K J Weber; P E Huber; A Ruehle; S Brons; P Haering; J Debus; H Hauswald
Journal:  Cancer Manag Res       Date:  2019-09-12       Impact factor: 3.989

6.  A Method to Determine the Coincidence of MRI-Guided Linac Radiation and Magnetic Isocenters.

Authors:  Kujtim Latifi; Eduardo G Moros; Geoffrey Zhang; Louis Harrison; Vladimir Feygelman
Journal:  Technol Cancer Res Treat       Date:  2019-01-01

7.  Intracranial Stereotactic Radiation Therapy With a Jawless Ring Gantry Linear Accelerator Equipped With New Dual Layer Multileaf Collimator.

Authors:  Nels C Knutson; William R Kennedy; Francisco J Reynoso; Jiayuan Peng; Lauren E Henke; Eric Laugeman; Michael Watts; Jessica Hillard; Ana Heermann; Geoffrey D Hugo; Sasa Mutic; Bin Cai
Journal:  Adv Radiat Oncol       Date:  2020-01-28

Review 8.  Medical physics challenges in clinical MR-guided radiotherapy.

Authors:  Christopher Kurz; Giulia Buizza; Guillaume Landry; Florian Kamp; Moritz Rabe; Chiara Paganelli; Guido Baroni; Michael Reiner; Paul J Keall; Cornelis A T van den Berg; Marco Riboldi
Journal:  Radiat Oncol       Date:  2020-05-05       Impact factor: 3.481

9.  MRI quality control for low-field MR-IGRT systems: Lessons learned.

Authors:  H Michael Gach; Austen N Curcuru; Erin J Wittland; Borna Maraghechi; Bin Cai; Sasa Mutic; Olga L Green
Journal:  J Appl Clin Med Phys       Date:  2019-09-21       Impact factor: 2.102

10.  Clinical utility of Gafchromic film in an MRI-guided linear accelerator.

Authors:  Ilma Xhaferllari; Joshua P Kim; Ruchira Liyanage; Chang Liu; Dongsu Du; Anthony Doemer; Indrin J Chetty; Ning Wen
Journal:  Radiat Oncol       Date:  2021-06-26       Impact factor: 3.481

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.