A new approach to quantify the mechanical and radiation isocentres of radiotherapy treatment machine gantries.

Piotr Skworcow1, John A Mills, Olivier C L Haas, Keith J Burnham.   

Abstract

In this paper a new method is proposed to quantify and reduce the radiation beam position uncertainty due to the radiotherapy treatment machine gantry deflection. A new tool has been designed and manufactured to provide the means to measure the alignment of the collimator axis and of the beam central axis in space, using the NDI Polaris optical tracking system and Gafchromic films. The tool can be mounted onto the accessory tray of the linacs from different manufacturers. The approach has been demonstrated with measurements of the mechanical isocentre being performed on ten linacs from three major manufacturers at four clinical sites. Measurements of the radiation isocentre were performed on a single linac. The collimator axis trajectory is modelled using a vector-end effector in order to provide more information than standard mechanical assessment methods. The method uses a mathematical optimization technique to calculate the position of the mechanical isocentre and the 'size' of the collimator axes intersection locus. Deviations of the collimator axes from the isocentre are expressed in terms of systematic and random error. The effects of measurement uncertainties are evaluated both via simulations and experimentally. The use of optical tracking and optimization techniques combined with an operator-induced measurement error compensation algorithm leads to a faster measurement of the mechanical isocentre (20 min for 24 angles) and eliminates operator-induced uncertainties. The uncertainty of the measurement of the mechanical isocentre was between 40 microm and 70 microm in terms of standard deviation. For some of the linacs assessed, the mechanical isocentre obtained using a standard approach with an adjustable pointer was displaced by over 1 mm from that found with the proposed method. The radii of the collimator axes intersection locus found with the proposed method were between 0.4 mm and 0.72 mm for the linacs assessed. Film measurement revealed a misalignment of the mechanical isocentre and the radiation isocentre by 0.4 mm. The proposed method potentially enables a reduction of the beam position uncertainty. This may be achieved at the planning stage by compensating for the identified systematic collimator axes deviations which were found to be reproducible. The method also creates a potential for using different setup margins independently for each axis and for each gantry angle, calculated specifically for a given linac.

Entities:  

Mesh:

Year:  2007        PMID: 18029996     DOI: 10.1088/0031-9155/52/23/022

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  7 in total

1.  Quantifying the gantry sag on linear accelerators and introducing an MLC-based compensation strategy.

Authors:  Weiliang Du; Song Gao; Xiaochun Wang; Rajat J Kudchadker
Journal:  Med Phys       Date:  2012-04       Impact factor: 4.071

2.  Evaluation of a magnetic resonance guided linear accelerator for stereotactic radiosurgery treatment.

Authors:  Ning Wen; Joshua Kim; Anthony Doemer; Carri Glide-Hurst; Indrin J Chetty; Chang Liu; Eric Laugeman; Ilma Xhaferllari; Akila Kumarasiri; James Victoria; Maria Bellon; Steve Kalkanis; M Salim Siddiqui; Benjamin Movsas
Journal:  Radiother Oncol       Date:  2018-05-25       Impact factor: 6.280

3.  A novel method to determine linac mechanical isocenter position and size and examples of specific QA applications.

Authors:  Jacek M Chojnowski; Jonathan R Sykes; David I Thwaites
Journal:  J Appl Clin Med Phys       Date:  2021-05-30       Impact factor: 2.102

Review 4.  Isocenter verification for linac-based stereotactic radiation therapy: review of principles and techniques.

Authors:  Pejman Rowshanfarzad; Mahsheed Sabet; Daryl J O'Connor; Peter B Greer
Journal:  J Appl Clin Med Phys       Date:  2011-11-15       Impact factor: 2.102

5.  Improved accuracy for noncoplanar radiotherapy: an EPID-based method for submillimeter alignment of linear accelerator table rotation with MV isocenter.

Authors:  Matthew J Nyflot; Ning Cao; Juergen Meyer; Eric C Ford
Journal:  J Appl Clin Med Phys       Date:  2014-03-06       Impact factor: 2.102

6.  A photographic technique for quick assessment of mechanical isocenter of a linear accelerator.

Authors:  Janne Heikkilä; Jan Seppälä; Jouni Nuutinen; Antti Vanne; Tapani Lahtinen
Journal:  Tech Innov Patient Support Radiat Oncol       Date:  2017-04-28

7.  Dosimetric evaluation of the gantry sag effect in clinical SRS plans.

Authors:  Egor Borzov; Alex Nevelsky; Rachel Bar-Deroma; Itzhak Orion
Journal:  BJR Open       Date:  2019-02-13
  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.