Nels C Knutson1, William R Kennedy1, Francisco J Reynoso1, Jiayuan Peng1, Lauren E Henke1, Eric Laugeman1, Michael Watts1, Jessica Hillard1, Ana Heermann1, Geoffrey D Hugo1, Sasa Mutic1, Bin Cai1.
Abstract
PURPOSE: To test the feasibility of a simplified, robust, workflow for intracranial stereotactic radiation therapy (SRT) using a ring gantry linear accelerator (RGLA) equipped with a dual-layer stacked, staggered, and interdigitating multileaf collimator.Entities:
Year: 2020 PMID: 32529144 PMCID: PMC7276691 DOI: 10.1016/j.adro.2020.01.003
Source DB: PubMed Journal: Adv Radiat Oncol ISSN: 2452-1094
Volumetric modulated arc therapy optimization objectives
| Optimization object | Optimization specification | Typical relative weight |
|---|---|---|
| OAR | Upper: Whatever is required to meet constraints | Whatever is required |
| NTO | Start distance = 0 mm, start dose = 101%, end dose = 20%, fall off = 0.25 | 150 |
| GTV | Lower: V32 Gy = 100%; upper: V42.5 Gy = 0% | 120 and 100, respectively |
| PTV | Lower: V30.5 Gy = 100% | 110 |
| PTV-GTV | Upper: V36 Gy = 0% | 100 |
| Ring touching PTV | Upper: V30 Gy = 0% | 80 |
| Ring 5 mm from PTV | Upper: V15 Gy = 0% | 80 |
| MUR | Max MU <1800 | 80 |
Abbreviations: GTV = gross tumor volume; NTO = normal tissue objective; OAR = organ at risk; PTV = planning target volume; MUR = monitor unit ratio.
Plan quality metrics and organ at risk constraints
| Quantity | Definition | Clinical goal (acceptable variation) |
|---|---|---|
| CI | V30Gy/VPTV | 1-1.2 (1-1.5) |
| GI | V15Gy/V30Gy | <3 (<4) |
| R50 | V15Gy/VPTV | <4 (<4.5) |
| HI | Max dose/30 Gy | 1.1-1.55 |
| GM | Distance between 30 Gy and 15 Gy equivalent spheres | <0.8 cm (<1.2 cm) |
| MUR | No. of monitor units/3000 cGy | <4 (<5) |
Abbreviations: CI = conformity index; GI = gradient index; GM = gradient measure; HI = homogeneity index; MUR = monitor unit ratio; R50 = relative ratio of the 50 isodose volume hetero.
Figure 1Visual comparison of stereotactic radiation therapy plans with c-arm and ring gantry linear accelerators.
Figure 2Box and swarm plots of plan quality parameters for Halcyon stereotactic radiation therapy plans and the clinical plans.
Figure 3Planning target volume (PTV) and brain dose-volume histogram data for Halcyon stereotactic radiation therapy plans and the clinical plans.
Figure 4Summary of quality assurance (QA) results for stereotactic radiation therapy plans with the ion chamber results shown on top and portal dosimetry pass rates per field shown on bottom.