Jun Liu1, Linping Huang1.
Abstract
Objective The present study was performed to assess the accuracy and clinical value of image-guided vacuum-assisted breast biopsy in the diagnosis of mammography-detected breast microcalcifications. Methods This prospective study involved 100 patients with suspicious mammography-detected microcalcifications who underwent image-guided vacuum-assisted breast biopsies from January 2013 to October 2016. Stereotactic vacuum-assisted breast biopsy (SVAB) was performed in 64 patients, and ultrasound (US)-guided vacuum-assisted breast biopsy (US-VAB) was performed in 36 patients in whom application of SVAB was difficult. The microcalcifications were detectable by mammography or US. The mean follow-up duration was 32 months (range, 9-57 months). The biopsy results were correlated with the pathological examination results. Results The positive predictive value of microcalcifications detected by mammography and US in the diagnosis of breast cancer was 37.0% (37/100) and 52.8% (19/36), respectively. The negative predictive value of microcalcifications detected by US in the diagnosis of breast cancer was 71.9% (46/64). No recurrence was observed in the 37 patients with breast cancer. No evidence of malignancy was found among the 63 patients with benign lesions. Conclusion Image-guided vacuum-assisted breast biopsy is accurate and minimally invasive. It can be used as a safe approach for diagnosis in patients with breast microcalcifications.Entities:
Keywords: Image-guided breast biopsy; breast cancer; mammography; microcalcifications; ultrasonography; vacuum assistance
Mesh:
Year: 2018 PMID: 29772934 PMCID: PMC6124280 DOI: 10.1177/0300060518770577
Source DB: PubMed Journal: J Int Med Res ISSN: 0300-0605 Impact factor: 1.671
Patient and lesion characteristics
| Patient characteristics (n = 100) | |
|---|---|
| Patient age, years | 45.4 (25–78) |
| Menopausal status | |
| Premenopausal | 58 (58) |
| Postmenopausal | 42 (42) |
| Family history of breast cancer | |
| Yes | 24 (24) |
| No | 76 (76) |
| Microcalcification location | |
| Left breast | 59 (59) |
| Right breast | 41 (41) |
| Microcalcification distribution | |
| Clustered calcification | 59 (59) |
| Regional calcifications | 30 (30) |
| Diffuse calcifications | 7 (7) |
| Linear calcifications | 3 (3) |
| Segmental calcifications | 3 (3) |
| BI-RADS (lesion) | |
| 3 | 8 (8) |
| 4 | 87 (87) |
| 5 | 5 (5) |
Data are presented as mean (range) or n (%).
BI-RADS: Breast Imaging-Reporting and Data System.
Figure 1.Stereotactic vacuum-assisted breast biopsy. (a, b) Stereo image before puncture shows the position of calcifications and puncture target spot. (c, d) The pre- and post-biopsy images show that the calcifications were completely removed. The white arrow indicates the position of the microcalcifications.
Figure 2.Ultrasound-guided vacuum-assisted breast biopsy. (a) Breast microcalcifications detected by ultrasound. (b) Microcalcifications detected by ultrasound (firefly technology). (c) Specimen mammography shows that the targeted microcalcifications were retrieved. The white arrow indicates the position of microcalcifications.
Correlation between type of mammographic abnormality and pathological diagnosis
| Mammography | Ductal carcinoma in situ | Infiltrating ductal carcinoma | Infiltrating lobular carcinoma | Benign lesions |
|---|---|---|---|---|
| BI-RADS category | ||||
| 5 (n = 5) | 4 | 1 | 0 | 0 |
| 4 (n = 87) | 20 | 11 | 1 | 55 |
| 3 (n = 8) | 0 | 0 | 0 | 8 |
| Distribution of calcification | ||||
| Cluster (n = 59) | 15 | 4 | 1 | 39 |
| Segmental (n = 3) | 2 | 0 | 0 | 1 |
| Diffuse (n = 7) | 1 | 0 | 0 | 6 |
| Regional (n = 30) | 5 | 8 | 0 | 17 |
| Linear (n = 1) | 1 | 0 | 0 | 0 |
| Morphology of calcification | ||||
| Coarse heterogeneous (n = 8) | 1 | 0 | 0 | 7 |
| Amorphous (n = 14) | 1 | 0 | 1 | 12 |
| Fine pleomorphic (n = 76) | 22 | 12 | 0 | 42 |
| Round (n = 2) | 0 | 0 | 0 | 2 |
Significance of microcalcification in the diagnosis of breast cancer
| Method | Pathological diagnosis | Positive predictive value | Negative predictive value | ||
|---|---|---|---|---|---|
| Malignant | Benign | ||||
| Microcalcification detected by mammography | + | 37 | 63 | 37.0% (37/100) | |
| Microcalcification detected by ultrasound | + | 19 | 17 | 52.8% (19/36) | 71.9% (46/66) |
| − | 18 | 46 | |||
| BI-RADS category of microcalcification detected by mammography | 4, 5 | 37 | 55 | 40.2% (37/92) | |
| 3 | 0 | 8 | |||