Surface characteristics and mechanical behavior of retrieved orthodontic microimplants.

Lu Lu1, Hyo-Sang Park1.   

Abstract

OBJECTIVES: To observe the surface characteristics and mechanical behavior of retrieved microimplants under clinically simulating experimental conditions and to investigate the feasibility of reuse of microimplants.
MATERIALS AND METHODS: The microimplants, inserted at different angles, were retrieved from the patients (RMIP) and the artificial bone (RMIA). Surface characteristics, including morphologic changes of tips and thread edges, length reduction, and surface compositional variation, were evaluated using a field emission scanning electron microscope, a stereoscopic microscope, and energy-dispersive X-ray spectroscopy, respectively. Mechanical behavior comprising maximum insertion torque (MIT) and insertion time was tested with the artificial bone under clinically simulating conditions.
RESULTS: The tips and thread edges were worn out to various degrees in retrieved microimplants and thin deposits were observed on the surface in the RMIP group. Traces of foreign elements, such as iron, sulphur, and calcium, were detected on the surface of RMIP. Both MIT and insertion time of retrieved microimplants were increased compared to their initial use, and were much greater in RMIP. The increases of MIT were seen in all groups inserted at the insertion angle of 45° compared with 90°, although the differences were not statistically significant.
CONCLUSIONS: Retrieved microimplants exhibited different degrees of changes on surface characteristics and mechanical behavior, with more changes in RMIP. The reuse of microimplants for immediate relocation in the same patient may be acceptable; however, postponed relocation and allogeneic reuse of microimplants are not recommended in clinical practice.

Entities:  

Keywords:  Orthodontic microimplant; Surface characteristics; Mechanical behavior; Retrieval analysis

Mesh:

Substances:

Year:  2018        PMID: 29732748      PMCID: PMC5962514          DOI: 10.1631/jzus.B1700324

Source DB:  PubMed          Journal:  J Zhejiang Univ Sci B        ISSN: 1673-1581            Impact factor:   3.066


  26 in total

1.  Micro-implant anchorage for treatment of skeletal Class I bialveolar protrusion.

Authors:  H S Park; S M Bae; H M Kyung; J H Sung
Journal:  J Clin Orthod       Date:  2001-07

2.  Effect of cortical bone thickness and implant placement torque on stability of orthodontic mini-implants.

Authors:  Mitsuru Motoyoshi; Tohru Yoshida; Akiko Ono; Noriyoshi Shimizu
Journal:  Int J Oral Maxillofac Implants       Date:  2007 Sep-Oct       Impact factor: 2.804

3.  Insertion torque of orthodontic miniscrews according to changes in shape, diameter and length.

Authors:  Seon-A Lim; Jung-Yul Cha; Chung-Ju Hwang
Journal:  Angle Orthod       Date:  2008-03       Impact factor: 2.079

4.  Characterization of retrieved orthodontic miniscrew implants.

Authors:  Theodore Eliades; Spiros Zinelis; Moschos A Papadopoulos; George Eliades
Journal:  Am J Orthod Dentofacial Orthop       Date:  2009-01       Impact factor: 2.650

5.  Insertion angle impact on primary stability of orthodontic mini-implants.

Authors:  Benedict Wilmes; Yu-Yu Su; Dieter Drescher
Journal:  Angle Orthod       Date:  2008-11       Impact factor: 2.079

6.  Retrieval analysis of immediately loaded orthodontic mini-implants: material and tissue characterization.

Authors:  Michail Vasoglou; Evanthia Chrysomali; Spiros Zinelis; Ilias Bitsanis; Nikos Haralambakis; Margarita Makou; George Epsilonliades
Journal:  Eur J Orthod       Date:  2013-12-28       Impact factor: 3.075

7.  Effects of the diameter and shape of orthodontic mini-implants on microdamage to the cortical bone.

Authors:  Nam-Ki Lee; Seung-Hak Baek
Journal:  Am J Orthod Dentofacial Orthop       Date:  2010-07       Impact factor: 2.650

8.  Biomechanical characteristics and reinsertion guidelines for retrieved orthodontic miniscrews.

Authors:  Chooryung J Chung; Kil-Young Jung; Yoon Jeong Choi; Kyung-Ho Kim
Journal:  Angle Orthod       Date:  2014-03-03       Impact factor: 2.079

9.  The effect of multiple processing and re-use on orthodontic mini-screw torque values.

Authors:  Saeed Noorollahian; Shiva Alavi; Elahe Rafiei
Journal:  Dent Res J (Isfahan)       Date:  2015 May-Jun

10.  Effect of recycling protocol on mechanical strength of used mini-implants.

Authors:  Sérgio Estelita; Guilherme Janson; Kelly Chiqueto; Eduardo Silveira Ferreira
Journal:  Int J Dent       Date:  2014-07-17
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.