Effects of the diameter and shape of orthodontic mini-implants on microdamage to the cortical bone.

Nam-Ki Lee1, Seung-Hak Baek.   

Abstract

INTRODUCTION: The purpose of this study was to investigate the effects of the diameter and shape of orthodontic mini-implants (OMIs) on microdamage to the cortical bone during implant placement.
METHODS: Twenty-eight self-drilling OMIs (Biomaterials Korea, Seoul, Korea; length, 6 mm; diameters, 1.5 and 2 mm; cylindrical and tapered shapes; classified as 1.5C, 2C, 1.5T, and 2T) were placed with a surgical device in the tibias of 7 New Zealand white rabbits (mature males; mean age, 6 months; mean weight, 3.1 kg). Four OMIs of each type per rabbit were placed randomly. Maximum insertion torque (MIT) was measured. Immediately after placement of the OMIs, the block of bone with the OMI was harvested. Cortical bone thickness was measured by using microcomputed tomography, and histomorphometric analyses of the number of cracks (NC), accumulated crack length (ACL), maximum radius of the crack (MRC), and longest crack (LC) were performed. Kruskal-Wallis and Mann-Whitney U tests with the Bonferroni adjustment were done for statistical analyses.
RESULTS: Increased diameter (1.5C<2C and 1.5T<2T) and tapering (1.5C<2T) resulted in increased values of MIT, NC, and LC (P <0.01, respectively). Similarly, with increased diameters (1.5C and 1.5T<2C and 2T), there were increases of ACL and MRC (P <0.001, respectively). However, there were no differences in the values of MIT, NC, ACL, MRC, and LC between the cylindrical and tapered OMIs with the same diameters (1.5C and 1.5T, 2C and 2T).
CONCLUSIONS: OMIs with larger diameters and tapered shapes caused greater microdamage to the cortical bone; this might affect bone remodeling and the stability of the OMIs. Copyright (c) 2010 American Association of Orthodontists. Published by Mosby, Inc. All rights reserved.

Entities:  

Mesh:

Year:  2010        PMID: 20620822     DOI: 10.1016/j.ajodo.2010.02.019

Source DB:  PubMed          Journal:  Am J Orthod Dentofacial Orthop        ISSN: 0889-5406            Impact factor:   2.650


  25 in total

1.  Surface characteristics and mechanical behavior of retrieved orthodontic microimplants.

Authors:  Lu Lu; Hyo-Sang Park
Journal:  J Zhejiang Univ Sci B       Date:  2018-05       Impact factor: 3.066

2.  Histomorphometric analysis of microcrack healing after the installation of mini-implants.

Authors:  Soobin Shin; Pan-Soo Park; Seung-Hak Baek; Il-Hyung Yang
Journal:  J Periodontal Implant Sci       Date:  2015-04-29       Impact factor: 2.614

3.  Assessment of cortical bone microdamage following insertion of microimplants using optical coherence tomography: a preliminary study.

Authors:  Hemanth Tumkur Lakshmikantha; Naresh Kumar Ravichandran; Mansik Jeon; Jeehyun Kim; Hyo-Sang Park
Journal:  J Zhejiang Univ Sci B       Date:  2018 Nov.       Impact factor: 3.066

4.  Group Distal Movement of Teeth using Micro-Screw-Implant Anchorage-A Case Report.

Authors:  Biju Kalarickal
Journal:  J Clin Diagn Res       Date:  2014-05-15

5.  Mechanical stability assessment of novel orthodontic mini-implant designs: Part 2.

Authors:  Christine Hong; Peter Truong; Ha Na Song; Benjamin M Wu; Won Moon
Journal:  Angle Orthod       Date:  2011-06-29       Impact factor: 2.079

6.  Production and repair of implant-induced microdamage in the cortical bone of goats after long-term estrogen deficiency.

Authors:  Z Yu; G Wang; T Tang; L Fu; X Yu; L Cao; Z Zhu; K Dai; S Qiu
Journal:  Osteoporos Int       Date:  2013-08-28       Impact factor: 4.507

7.  Effects of predrilling depth and implant shape on the mechanical properties of orthodontic mini-implants during the insertion procedure.

Authors:  Keun-Chul Cho; Seung-Hak Baek
Journal:  Angle Orthod       Date:  2011-11-03       Impact factor: 2.079

8.  Angled-predrilling depth and mini-implant shape effects on the mechanical properties of self-drilling orthodontic mini-implants during the angled insertion procedure.

Authors:  Yoon-Young Heo; Keun-Chul Cho; Seung-Hak Baek
Journal:  Angle Orthod       Date:  2012-03-03       Impact factor: 2.079

9.  Effect of vertical placement angle on the insertion torque of mini-implants in human alveolar bone.

Authors:  Rafael Ribeiro Maya; Célia Regina Maio Pinzan-Vercelino; Julio de Araujo Gurgel
Journal:  Dental Press J Orthod       Date:  2016 Sep-Oct

10.  In vitro and in vivo mechanical stability of orthodontic mini-implants.

Authors:  Il-Sik Cho; Sung-Kyun Kim; Young-Il Chang; Seung-Hak Baek
Journal:  Angle Orthod       Date:  2011-10-19       Impact factor: 2.079

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.