Comparison of implant component fractures in external and internal type: A 12-year retrospective study.

Yuseung Yi1, Jai-Young Koak1, Seong-Kyun Kim1, Shin-Jae Lee2, Seong-Joo Heo1.   

Abstract

PURPOSE: The aim of this study was to compare the fracture of implant component behavior of external and internal type of implants to suggest directions for successful implant treatment.
MATERIALS AND METHODS: Data were collected from the clinical records of all patients who received WARANTEC implants at Seoul National University Dental Hospital from February 2002 to January 2014 for 12 years. Total number of implants was 1,289 and an average of 3.2 implants was installed per patient. Information about abutment connection type, implant locations, platform sizes was collected with presence of implant component fractures and their managements. SPSS statistics software (version 24.0, IBM) was used for the statistical analysis.
RESULTS: Overall fracture was significantly more frequent in internal type. The most frequently fractured component was abutment in internal type implants, and screw fracture occurred most frequently in external type. Analyzing by fractured components, screw fracture was the most frequent in the maxillary anterior region and the most abutment fracture occurred in the maxillary posterior region and screw fractures occurred more frequently in NP (narrow platform) and abutment fractures occurred more frequently in RP (regular platform).
CONCLUSION: In external type, screw fracture occurred most frequently, especially in the maxillary anterior region, and in internal type, abutment fracture occurred frequently in the posterior region. placement of an external type implant rather than an internal type is recommended for the posterior region where abutment fractures frequently occur.

Entities:  

Keywords:  Abutment fracture; Fixture fracture; Implant fracture; Screw fracture

Year:  2018        PMID: 29713437      PMCID: PMC5917108          DOI: 10.4047/jap.2018.10.2.155

Source DB:  PubMed          Journal:  J Adv Prosthodont        ISSN: 2005-7806            Impact factor:   1.904


INTRODUCTION

Endosseous implants are reliable choice of treatment for the replacement of missing natural teeth. Although the overall success rate of implant is relatively high, between 95 – 98%,1 they often encounter complications such as periimplantitis and other technical problems. According to several previous studies, the most common cause of implant failure is peri-implantitis and technical complications are not uncommon.234 Technical problems of implant-supported restorations can be classified into two groups: those relating to the prosthesis, and those relating to the implant components.56 Technical problems relating to the prosthesis are such as veneering material or framework fractures and technical problems relating to the implants components include screw loosening or screw fractures, abutment fracture and implant fixture fractures. If the problem is caused by the implant prosthesis itself, the problem can be overcome by refabricating the prosthesis. However, in the case of problems associated with implant components, various solutions must be sought, simply to replace the implant components or, in some cases, remove the implants, which makes it difficult to predict the solution. Adell et al.7 reported a 3.5% implant fixture fracture in a 15-year study of Brånemark implant in 1981. Naert et al.8 reported a 0.53% implant fracture, 8.9% abutment screw fracture, and 1.2% occlusal screw fracture in a case study of implant supporting complete fixed prosthesis with Brånemark implant in 1992. As regards the fracture of implant, Rangert et al.9 said it was associated with bruxism or strong occlusal force, and it occurred more frequently in an single or double implant prosthesis of the posterior region. In a retrospective study of implant complications in 1997, Tolman and Laney reported10 that screw fractures occurred in 87 of 1,250 implants (7.0%). Although there have been several studies on fracture of the implant components, most studies report only failure of the osseointegration as failure of the implant and it is overlooked that implant failure can be caused by implant component fracture. Clinically, in the case of implant abutment fracture or screw fracture, it is recommended to remake the prosthesis after removing the fractured remnant. However, it is difficult to remove the fragments and eventually the implants must be removed, leading to implant failure. The purpose of this study was to predict the prognosis of implants and to suggest directions for successful implant treatment by analyzing the factors affecting the fracture of implant components including abutment connection type, implant location and implant platform size.

MATERIALS AND METHODS

Data were collected from the clinical records of all patients who received one or more WARANTEC implants at Seoul National University Dental Hospital from February 2002 to January 2014 for 12 years and the following cases were excluded: i) implants failed in osseointegration, ii) implant placement after jaw resection and reconstruction, iii) the opposite arch was complete denture, iv) implant assisted over denture, v) insufficient clinical chart recording, vi) patients who have not visited since 2012. Data collection included 406 patients (205 males, 201 females), ranging in age from 21 to 94 years (mean 64.6 years, SD 11.5). Total number of implants was 1,289 and an average of 3.2 implants was installed per patient. Information about abutment connection type (internal or external), implant locations, platform sizes was collected with presence of implant component fractures and their managements (Table 1). The information about implant component fractures was divided into three types: screw fractures, abutment fractures, and fixture fractures. The management of fractures was classified as screw replacement, prosthesis refabrication, and fixture removal.
Table 1

Distribution of implants according to the location and platform size

Number of implants LocationMaxillary AnteriorMaxillary PosteriorMandibular AnteriorMandibular PosteriorTotal (%)
External typePlatform sizeNP42322211107 (13.4)
RP7529518243631 (79.0)
WP23002961 (7.6)
Total (%)119 (14.9)357 (44.7)40 (5.0)283 (35.4)799 (100)
Internal typePlatform sizeNP691622 (4.5)
RP142539152428 (87.3)
WP02201840 (8.2)
Total (%)20 (4.1)284 (58.0)10 (2.0)176 (35.9)490 (100)
TOTALPlatform sizeNP48412317129 (10.0)
RP89548273951059 (82.2)
WP252047101 (7.8)
Total (%)139 (10.8)641 (49.7)50 (3.9)459 (35.6)1289 (100)

NP = Narrow Platform, RP = Regular platform, WP = Wide platform

SPSS statistics software (version 24.0, IBM, New York, NY, USA) was used for the statistical analysis. The Pearson chi-square test and Fisher's exact test (P = .05) were used to evaluate the association between implant characteristics and implant components fracture.

RESULTS

A total of 1,289 implants were placed in 406 patients during investigation period. 799 implants had external type abutment connection (62.0%) and 490 implants were internal type abutment connection (38.0%). 139 implants were placed in the maxillary anterior region (10.8%), 641 implants in the maxillary posterior region (49.7%), 50 implants in the mandibular anterior region (3.9%), and 449 implants in the mandibular posterior region (35.6%). According to the platform size, 129 implants (10.0%) with narrow platform (NP), 1059 implants (82.2%) with regular platform (RP), and 101 implants (7.8%) with wide platform (WP) were placed (Table 1, Fig. 1).
Fig. 1

Implant distribution according to the location and the platform sizes.

Of the total 1,289 implants, component fractures occurred in 72 implants (5.6%). In internal type implants, 53 components fractures occurred (10.8%): 8 screw fractures (1.6%), 40 abutment fractures (8.2%), and 5 fixture fracture (1.0%), and in external type implants, 19 components fractures occurred (2.4%): 16 screw fractures (2.0%), 1 abutment fracture (0.1%), and 2 fixture fractures (0.3%) (Table 2).
Table 2

Implant component fractures

Internal type (%)External type (%)Total (%)
Total implants490 (38.0)799 (62.0)1,289
Screw fracture8 (1.6)16 (2.0)24 (1.9)
Abutment fracture40 (8.2)a1 (0.1)b41 (3.2)
Fixture fracture5 (1.0)2 (0.3)7 (0.5)
Total fracture53 (10.8)b19 (2.4)c72 (5.6)

Different letters mean significant difference (P < .05)

Overall fracture was significantly more frequent in internal type (P < .001). The most frequently fractured component was abutment in internal type implants (8.2%), and screw fracture occurred most frequently (2.0%) in external type (Fig. 2).
Fig. 2

Implant component fracture. Overall fracture was significantly more frequent in internal type. The most frequently fractured component was abutment in internal type implants, and screw fracture occurred most frequently in external type.

The implants were most placed in the maxillary posterior region (49.7%), followed by the mandibular posterior region (35.6%) (Table 1). Table 3 shows the most frequent location which implant component fractures occurred was the maxillary posterior region in internal type (10.8%), and maxillary anterior region in external type (5.9%).
Table 3

Fractures depending on implant location

By abutment connection type
Internal type (%)External type (%)Total (%)
Maxillary anterior0 (0.0)7 (5.9)7 (5.0)
Maxillary posterior31 (10.9)11 (3.1)42 (6.6)
Mandibular anterior0 (0.0)0 (0.0)0 (0.0)
Mandibular posterior22 (12.5)1 (0.4)23 (5.0)
Total fracture53 (10.8)19 (2.4)72 (5.6)
Analyzing by fractured components, screw fracture was the most frequent in the maxillary anterior region (5.0%) and the most abutment fracture occurred in the maxillary posterior region (4.1%) significantly (P = .013 < .050) (Table 3, Fig. 3A).
Fig. 3

Implant component fracture. (A) Screw fracture was the most frequent in the maxillary anterior region and the most abutment fracture occurred in the maxillary posterior region significantly. (B) In external type, the number of fractures of NP was larger than that in other platform sizes, and more fractures occurred in the RP than other platform sizes in internal type. Screw fractures occurred more frequently in NP and abutment fractures occurred more frequently in RP. Mx = Maxillary, Ant = Anterior, Mn = Mandibular, Pos = Posterior, NP = Narrow platform, RP = Regular platform, WP = Wide platform.

There was significant relationship between platform size and fractures. In external type, the number of fractures of NP was larger than that in other platform sizes, and more fractures occurred in the RP than other platform sizes in internal type. Analyzing by fractured components, screw fractures occurred more frequently in NP (narrow platform) and abutment fractures occurred more frequently in RP (regular platform) (Table 4, Fig. 3B).
Table 4

Fractures depending on implant platform size

By abutment connection type
Internal type (%)External type (%)Total (%)
NP1 (4.5)8 (7.5)9 (7.0)
RP47 (11.0)11 (1.7)58 (5.5)
WP5 (12.5)0 (0.0)5 (5.0)
Total fracture53 (10.8)19 (2.4)72 (5.6)

NP = Narrow Platform, RP = Regular Platform, WP = Wide Platform

DISCUSSION

The number of implants installed for 12 years from 2002 to 2014 was greater in the external type (62.0%) than in the internal type (38.0%). Depending on implant locations, implants were mostly placed at the maxillary posterior region (49.7%), followed by 35.6% at the mandibular posterior region, 10.8% at the maxillary anterior region and 3.9% at the mandibular anterior region. Depending on implant Platform sizes, RP (Regular platform) was the most placed at all locations with 82.2%, NP (Narrow Platform) was placed in the anterior region with 10.0%, WP (Wide Platform) was placed in the posterior region mostly with 7.8% (Table 1, Fig. 1). Of the total 1,289 implants, component fractures occurred in 72 implants (5.6%). In internal type implants, 53 components fractures occurred (10.8 %): 8 screw fractures (1.6%), 40 abutment fractures (8.2%), and 5 fixture fracture (1.0%), and in external type implants, 19 components fractures occurred (2.4%): 16 screw fractures (2.0%), 1 abutment fracture (0.1%), and 2 fixture fractures (0.3%). Table 5 shows the relationship between various factors and implant component fractures.
Table 5

Implant component fractures

ScrewAbutmentFixtureTotal
External type2.00%0.10%0.30%2.40%
Internal type1.60%8.20%1.00%10.80%
P value.402< .001*> .999< .001*
Component fractures depending on implant location
External typeMx Ant5.90%0.00%0.00%5.90%
Mx Pos2.50%0.00%0.60%3.10%
Mn Ant0.00%0.00%0.00%0.00%
Mn Pos0.00%0.40%0.00%0.40%
PP value.003*.553.683.007*
Internal typeMx Ant0.00%0.00%0.00%0.00%
Mx Pos0.70%9.20%1.10%10.90%
Mn Ant0.00%0.00%0.00%0.00%
Mn Pos3.40%8.00%1.10%12.50%
P value.186.601> .999.327
Component fractures depending on implant platform size
External typeNP7.50%0.00%0.00%7.50%
RP1.30%0.20%0.30%1.70%
WP0.00%0.00%0.00%0.00%
P value.001*> .999> .999.003*
Internal typeNP4.50%0.00%0.00%4.50%
RP1.20%8.60%1.20%11.00%
WP5.00%7.50%0.00%12.50%
P value.092.407> .999.662

*significant (P < .05).

Mx = Maxillary, Mn = Mandibular, Ant = Anterior, Pos = Posterior, NP = Narrow Platform, RP = Regular Platform, WP = Wide Platform

Comparing the overall implant component fracture of the internal and external types, the fracture rate in internal type was significantly higher than in external type (P < .001). By implant components, there were no significant difference between internal and external types in screw or fixture fractures, only in abutment fracture, internal type (8.2%) was significantly higher than external type (0.1%) (P < .001). Depending on the implant location, in internal types, there was no significant relationship between the implant location and each component fracture rates, however, in external types, there was a significant difference in total fracture rate (P = .007 < .05) and screw fracture (P = .003 < .05) in maxillary anterior region was significantly higher (5.8%). Analyzing according to implant platform size, NP (Narrow platform) showed significantly higher total fracture rate (P = .003 < .05) and higher screw fracture rates (P = .001 < .05) in external type and there was no significant difference in internal type. In most cases, screw loosening precedes screw fracture. Screw loosening is a relatively frequent complication and it occurs more frequently in external type than internal type, in screw-retained type than in cemented type, in lower arch than in upper arch, and in single tooth restoration than multiple restorations.1112 In this retrospective study, screw fracture occurred in internal type 1.6% and external type 2.0%, but there was no significant difference according to connection type. There was a significant difference according to implant position and platform size. The most frequent location was in the maxillary anterior region (5.9%), and most of them occurred in NP (narrow platform) according to the platform size (7.5%). In other words, screw fracture occurs most frequently in maxillary anterior region with NP (narrow platform) in external type (Fig. 4A). According to previous studies, screw fracture is caused by various factors such as excessive bite force, improper placement of implant, bone loss, inappropriate fit or design of prosthesis, accumulation of fatigue, defect in fabrication and type of implant.13141516 In internal type, the retention of the implant superstructure is obtained by the friction between the implant abutment and the fixture inner surface and screw. However, in external type, the contact surface between the implant abutment and the fixture is smaller than in internal type structurally, and the retentive force of the superstructure is achieved only by screws. Therefore, when excessive lateral force or other factors are applied to the implant, the force is transmitted to the implant screw, and if the force exceeds the retention threshold of the screw, screw loosening will occur, or if it exceeds the fracture threshold, screw fracture will occur. During eccentric movement, the implants in maxillary anterior region are affected by the lateral force and the bending force transmitted to the implant screw also increases, eventually causes screw fracture. Since the fixation screw of NP (narrow platform) abutment has also narrow diameter, thus, the fracture resistance threshold is small and therefore susceptible to fracture when the factors are applied.
Fig. 4

Implant component fractures (By implant component). (A) Screw fracture occurred in internal type 1.6% and external type 2.0%, but there was no significant difference according to connection type. There was a significant difference according to implant position and platform size. The most frequent location was in the maxillary anterior region (5.9%), and most of them occurred in NP. (B) Abutment fracture occurred more frequently in internal type (8.2%) than in external type (0.1%). There was no significant difference in fracture rate according to the implant location, but all fractures occurred in posterior region. (C) Fixture fracture occurred 1.0% in internal type, and 0.3% in external type, there was no significant difference according to implant connection type or implant location, however, all fixture fractures occurred at the posterior region in RP. Mx = Maxillary, Mn = Mandibular, Ant = Anterior, Pos = Posterior, NP = Narrow Platform, RP = Regular Platform, WP = Wide Platform.

Although many researches about implant screw or fixture fracture have been conducted, there is little analysis of abutment fracture. In this study, abutment fracture occurred more frequently in internal type (8.2%) than in external type (0.1%). There was no significant difference in fracture rate according to the implant location, but all fractures occurred in posterior region. (Fig. 4B) What is remarkable is all abutment fractures occurred in single implant restoration in posterior region using gold cast UCLA type with internal hex abutment connection. This type of implant prosthesis is a superstructure of a unit, so that when the lateral force applies, the force concentrates on the implant abutment neck area, causing a bending moment, which causes fracture of the abutment neck. Implant fixture fracture is classified as late failure and have been reported to be caused by various factors. When the bone resorption occurs due to peri-implantitis, the bending stress from the masticatory force increases and the stress concentrates at the end of the abutment screw and acts as a starting point of the fracture.1718 As a result of this study, fixture fracture occurred 1.0% in internal type, and 0.3% in external type, there was no significant difference according to implant connection type or implant location, however, all fixture fractures occurred at the posterior region in RP (regular platform) (Fig. 4C). Rangert et al.17 reported that 90% of implant fractures are located in the molar and premolar regions of the mouth, where chewing forces and lateral movements associated with cusp inclination generate undesirable forces. When a implant is subjected to a force, external and internal types exhibit different force distribution structurally. In external type, when an external force is applied, the stress is transferred to the screw, which is the most susceptible part, and the screw fracture occurs before the fixture is stressed beyond the fracture resistance, thereby preventing the fracture of the fixture. In internal type, however, the mechanical interface between the abutment and the fixture is present, so that when the external force is applied, the stress transmitted to the abutment is transmitted to the fixture and if the stress exceeds the fracture resistance, fixture fracture is caused. Clinically, when implant component fractures occur, the most important thing is whether the problem can be solved easily. When fixture fractures occur, it is considered as an implant failure since it can no longer function as an implant restoration and must be removed. However, in the case of screw or abutment fractures, the problem can be solved more easily if the fractured fragment can be removed. In other words, obtaining retrievability is the most important point to solve the problem. How to cope with fractures during this retrospective study was summarized in Table 6.
Table 6

How to cope with fractures

Fractured component
ScrewAbutmentFixture
Internal type
 Fixture removal1 (12.5%)10 (25%)5 (100%)
 Prosthesis refabrication (retrieve)3 (37.5%)29 (72.5%)0 (0.0%)
 Screw replacement (retrieve)4 (50.0%)0 (0.0%)0 (0.0%)
 Maintain fracture status0 (0.0%)1* (2.5%)0 (0.0%)
External type
 Fixture removal0 (0.0%)0 (0.0%)2 (100%)
 Prosthesis refabrication (retrieve)0 (0.0%)1 (100%)0 (0.0%)
 Screw replacement (retrieve)15 (93.8%)0 (0.0%)0 (0.0%)
 Maintain fractured status1** (6.2%)0 (0.0%)0 (0.0%)

* Fractured abutment could not be removed, being used.

** Fractured screw could not be removed, using new screw cut.

Only 12.5% of the fractured screws were not removed in internal type, and 6.2% in external type, in the case of abutment fracture, the abutment fragment removal was impossible in 27.5% and fixture were removed.

CONCLUSION

In external type, screw fracture occurred most frequently, especially in the maxillary anterior region, and in internal type, abutment fracture occurred frequently in the posterior region. The screw fracture seems to be easier to solve than the abutment fracture. Therefore, placement of an external type implant rather than an internal type is recommended for the posterior region where abutment fractures frequently occur. If an internal type implant is inevitably installed in the posterior region, the SCRP type prosthesis or multi-unit abutment should be used instead of the one-unit UCLA type implant prosthesis for external force distribution.
  16 in total

1.  Mechanical behavior and failure analysis of prosthetic retaining screws after long-term use in vivo. Part 2: Metallurgical and microhardness analysis.

Authors:  Youssef Al Jabbari; Raymond Fournelle; Gerald Ziebert; Jeffrey Toth; Anthony Iacopino
Journal:  J Prosthodont       Date:  2007-11-28       Impact factor: 2.752

2.  Tissue-integrated prosthesis complications.

Authors:  D E Tolman; W R Laney
Journal:  Int J Oral Maxillofac Implants       Date:  1992       Impact factor: 2.804

Review 3.  Abutment screw loosening in single-implant restorations: a systematic review.

Authors:  Anna Theoharidou; Haralampos P Petridis; Konstantinos Tzannas; Pavlos Garefis
Journal:  Int J Oral Maxillofac Implants       Date:  2008 Jul-Aug       Impact factor: 2.804

4.  Fractured prosthetic abutments in osseointegrated implants: a technical complication to cope with.

Authors:  S Luterbacher; I Fourmousis; N P Lang; U Brägger
Journal:  Clin Oral Implants Res       Date:  2000-04       Impact factor: 5.977

5.  Osseointegrated implants for single-tooth replacement: progress report from a multicenter prospective study after 3 years.

Authors:  W R Laney; T Jemt; D Harris; P J Henry; P H Krogh; G Polizzi; G A Zarb; I Herrmann
Journal:  Int J Oral Maxillofac Implants       Date:  1994 Jan-Feb       Impact factor: 2.804

6.  A 15-year study of osseointegrated implants in the treatment of the edentulous jaw.

Authors:  R Adell; U Lekholm; B Rockler; P I Brånemark
Journal:  Int J Oral Surg       Date:  1981-12

7.  Bending overload and implant fracture: a retrospective clinical analysis.

Authors:  B Rangert; P H Krogh; B Langer; N Van Roekel
Journal:  Int J Oral Maxillofac Implants       Date:  1995 May-Jun       Impact factor: 2.804

8.  Mechanical behavior and failure analysis of prosthetic retaining screws after long-term use in vivo. Part 4: Failure analysis of 10 fractured retaining screws retrieved from three patients.

Authors:  Youssef S Al Jabbari; Raymond Fournelle; Gerald Ziebert; Jeffrey Toth; Anthony M Iacopino
Journal:  J Prosthodont       Date:  2008-01-15       Impact factor: 2.752

9.  Biological and technical complications and failures with fixed partial dentures (FPD) on implants and teeth after four to five years of function.

Authors:  U Brägger; S Aeschlimann; W Bürgin; C H Hämmerle; N P Lang
Journal:  Clin Oral Implants Res       Date:  2001-02       Impact factor: 5.977

10.  Mechanical behavior and failure analysis of prosthetic retaining screws after long-term use in vivo. Part 1: Characterization of adhesive wear and structure of retaining screws.

Authors:  Youssef S Al Jabbari; Raymond Fournelle; Gerald Ziebert; Jeffrey Toth; Anthony M Iacopino
Journal:  J Prosthodont       Date:  2007-11-15       Impact factor: 2.752

View more
  5 in total

1.  Microstructural Analysis of Fractured Orthopedic Implants.

Authors:  Mateusz Kopec; Adam Brodecki; Grzegorz Szczęsny; Zbigniew L Kowalewski
Journal:  Materials (Basel)       Date:  2021-04-25       Impact factor: 3.623

2.  Fractures of implant fixtures: a retroscccpective clinical study.

Authors:  Han-Chang Yu; Young-Kyun Kim
Journal:  Maxillofac Plast Reconstr Surg       Date:  2020-04-25

Review 3.  Retrieval of Fractured Implant Abutment Screws: A Narrative Review.

Authors:  Prabhu Nayana; Sunil S Nayak; Abhirup Chatterjee; Karthik Sivaraman; Gadicherla Srikanth; Chhavi Singh
Journal:  J Int Soc Prev Community Dent       Date:  2022-06-29

4.  Impact of lattice versus solid structure of 3D-printed multiroot dental implants using Ti-6Al-4V: a preclinical pilot study.

Authors:  Jungwon Lee; Ling Li; Hyun-Young Song; Min-Jung Son; Yong-Moo Lee; Ki-Tae Koo
Journal:  J Periodontal Implant Sci       Date:  2022-08       Impact factor: 2.086

5.  A retrospective comparison of clinical outcomes of implant restorations for posterior edentulous area: 3-unit bridge supported by 2 implants vs 3 splinted implant-supported crowns.

Authors:  Yuseung Yi; Seong-Joo Heo; Jai-Young Koak; Seong-Kyun Kim
Journal:  J Adv Prosthodont       Date:  2022-08-29       Impact factor: 1.989

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.