| Literature DB >> 29698492 |
Hiroyuki Takemura1, Tomohiko Ai2,3, Konobu Kimura3,4, Kaori Nagasaka1, Toshihiro Takahashi1, Koji Tsuchiya1,5, Haeun Yang2, Aya Konishi4, Kinya Uchihashi4, Takashi Horii1, Yoko Tabe2,3, Akimichi Ohsaka1,3,5.
Abstract
The XN series automated hematology analyzer has been equipped with a body fluid (BF) mode to count and differentiate leukocytes in BF samples including cerebrospinal fluid (CSF). However, its diagnostic accuracy is not reliable for CSF samples with low cell concentration at the border between normal and pathologic level. To overcome this limitation, a new flow cytometry-based technology, termed "high sensitive analysis (hsA) mode," has been developed. In addition, the XN series analyzer has been equipped with the automated digital cell imaging analyzer DI-60 to classify cell morphology including normal leukocytes differential and abnormal malignant cells detection. Using various BF samples, we evaluated the performance of the XN-hsA mode and DI-60 compared to manual microscopic examination. The reproducibility of the XN-hsA mode showed good results in samples with low cell densities (coefficient of variation; % CV: 7.8% for 6 cells/μL). The linearity of the XN-hsA mode was established up to 938 cells/μL. The cell number obtained using the XN-hsA mode correlated highly with the corresponding microscopic examination. Good correlation was also observed between the DI-60 analyses and manual microscopic classification for all leukocyte types, except monocytes. In conclusion, the combined use of cell counting with the XN-hsA mode and automated morphological analyses using the DI-60 mode is potentially useful for the automated analysis of BF cells.Entities:
Mesh:
Year: 2018 PMID: 29698492 PMCID: PMC5919509 DOI: 10.1371/journal.pone.0195923
Source DB: PubMed Journal: PLoS One ISSN: 1932-6203 Impact factor: 3.240
Within-run precision.
The mean cell number and coefficient of variations (% CV) of ten different CSF samples were calculated five consecutive times with the XN-hsA and the XN-BF modes.
| XN-hsA mode | XN-BF mode | ||||
|---|---|---|---|---|---|
| Sample# | Mean (cells/μl) | CV (%) | Sample# | Mean (cells/μl) | CV (%) |
| 1 | 256.7 | 5.6 | 6 | 333 | 2.6 |
| 2 | 19.0 | 5.7 | 7 | 100.2 | 5.3 |
| 3 | 6.2 | 7.5 | 8 | 28.6 | 8.0 |
| 4 | 1.3 | 13.6 | 9 | 6.8 | 12.3 |
| 5 | 0.5 | 24.8 | 10 | 1.2 | 37.3 |